Erschienen in:
01.12.2012 | Original
Diaphragm electromyographic activity as a predictor of weaning failure
verfasst von:
Martin Dres, Matthieu Schmidt, Alexis Ferre, Julien Mayaux, Thomas Similowski, Alexandre Demoule
Erschienen in:
Intensive Care Medicine
|
Ausgabe 12/2012
Einloggen, um Zugang zu erhalten
Abstract
Purpose
To compare breathing pattern descriptors and diaphragm electromyographic activity (EAdi)-derived indices obtained from a neurally adjusted ventilatory assist catheter during a spontaneous breathing trial (SBT) in patients successfully and unsuccessfully separated from the ventilator and to assess their performance as a potential marker to discriminate these two categories of patients.
Methods
Fifty-seven ready-to-wean patients were included in a prospective observational study. During a 30-min SBT (pressure support 7 cmH2O, zero end expiratory pressure), tidal volume (V
T) and respiratory rate (RR) were obtained from the flow signal at baseline and at 3, 10, 20 and 30 min during the SBT. EAdi-derived indices were simultaneously computed: maximum of the EAdi (EAdimax), area under the inspiratory curve of EAdi (EAdiAUC), the difference between EAdimax and EAdimin (∆EAdi), EAdimax/V
T, EAdiAUC/V
T and ∆EAdi/V
T. Patients, successfully (success group; n = 35) and unsuccessfully (failure group; n = 22) separated from the ventilator were compared.
Results
At baseline, the breathing pattern was similar in the two groups, whereas EAdimax and EAdiAUC were significantly lower in the success group (p < 0.05). In the failure group, RR and RR/V
T increased significantly during the trial, V
T decreased, whereas EAdimax and EAdiAUC did not change. At 3 min, the areas under the receiver operating characteristic-curve of RR/V
T and the EAdi-derived indices to predict weaning outcome were 0.83 for the rapid shallow breathing index (RSBI), 0.84 for EAdimax/V
T , 0.80 for EAdiAUC/V
T (0.80) and 0.82 for ∆EAdi/V
T. The coefficient of variation for V
T decreased in the failure group while that for EAdimax remained unchanged.
Conclusions
EAdi-derived indices provide reliable and early predictors of weaning outcome. However, the performance of these indices is not better than the RR/V
T.