FormalPara Take-home message

This is the first randomized multicenter study comparing the use of HFNC with SMO for the preoxygenation of non-severely hypoxemic patients in the ICU. In this setting, compared to SMO, preoxygenation with HFNC did not improve the lowest SpO2 during intubation but led to a reduction in intubation-related adverse events.

Introduction

Endotracheal intubation is a procedure that is performed daily in the intensive care units (ICU). Nevertheless, it remains a life-threatening event and causes more than 25% of severe complications such as oxygen desaturation, hemodynamic instability, and death [1,2,3], driving intensive research efforts in the field [4,5,6].

Preoxygenation is a key component of intubation in order to stock up oxygen in functional residual capacity (FRC) volume, representing the main oxygen reservoir during apnea [7]. After the induction of general anesthesia, it extends the duration of apnea until hypoxemia occurs, i.e., oxygen desaturation below a critical pulse oximetry (SpO2) value [8]. In critically ill patients, several pathological conditions compromise optimal preoxygenation such as high peripheral oxygen uptake, hemodynamic instability, or altered consciousness. In this setting, preoxygenation with the standard bag-valve mask oxygenation (SMO) followed by rapid sequence intubation (RSI) is proposed [9] to reduce the risk of desaturation and aspiration. RSI includes immediate intubation after hypnotic administration without manual face mask ventilation to minimize the time from induction to intubation. SMO is removed during laryngoscopy—meaning that oxygenation of the patient is interrupted. Alternatives such as high-flow therapy by nasal cannulae (HFNC) could prove beneficial. Indeed, considering the encouraging results in patients presenting acute hypoxemic respiratory failure [10], HFNC has been proposed as a preoxygenation device for intubation in the ICU. This device can deliver up to 60 L/min with an inspired fraction of oxygen of up to 100% [11], generate a moderate positive supraglottic end-expiratory pressure, and reduce the patient effort [12]. Contrary to SMO, HFNC makes it possible to hold nasal prongs in place after induction of general anesthesia and during laryngoscopy, to perform apneic oxygenation throughout the intubation procedure [13]. Furthermore, tolerance of HFNC has repeatedly been shown to be excellent whereas SMO can cause discomfort and claustrophobia since the mask is tightly applied to the patient [14]. HFNC is a widespread device in the ICU but there is no large randomized study that has evaluated its benefits as a preoxygenation device for intubation in non-severely hypoxemic patients. We recently reported that this device failed to prevent desaturation during intubation in severely hypoxemic patients [3]. In non-severely hypoxemic patients, a recent before–after study suggested that HFNC could improve preoxygenation and reduce severe hypoxemia during intubation [15].

The main objective was to determine the efficacy of HFNC for preoxygenation compared with SMO as measured by the lowest level of pulse oximetry during RSI recorded through SpO2. Secondary objectives were to compare intubation procedures, related side effects, and patient outcome in the two groups.

Methods

Study design, setting, and ethical considerations

The PROTRACH study was a multicenter, randomized, controlled, open-label clinical trial. The main objective was to determine the efficacy of HFNC to prevent oxygen desaturation during intubation in the ICU compared with standard of care. Patients were enrolled from April 2016 to June 2017 in seven French ICUs (four medical, two medical-surgical, and one surgical) and followed for 28 days. An independent safety committee oversaw the trial.

Three methods of consent were available; patients were included after written informed consent whenever possible. Most patients were included after the written informed consent of their next-of-kin or following an emergency inclusion procedure. The appropriate ethics committee approved this study protocol in September 2015 (15/13-975).

Participants

Eligible patients were adults (18 years or older), requiring intubation in the ICU, without severe hypoxemia defined as a ratio of arterial partial pressure in oxygen to fraction of inspired oxygen (PaO2/FiO2) below 200 within 4 h before inclusion. To determine the PaO2/FiO2 ratio on inclusion, the FiO2 specified by the manufacturer for Venturi mask or corresponding to the oxygen flow for standard nasal prongs [16] was considered.

Exclusion criteria were intubation without rapid sequence induction (i.e., in case of cardiac arrest), fiberoptic intubation, asphyxia requiring immediate intubation, any nasopharyngeal anatomical obstacle, grade 4 glottis exposure on the Cormack–Lehane scale documented in the medical history, adults subject to legal protection, pregnancy, lack of consent, patients who had already undergone intubation in the ICU (i.e., extubation failure) or who were already participating in an interventional study on preoxygenation.

Randomization

Randomization used fixed blocks of six patients (ratio 1:1) and was stratified by center. The study statistician generated the allocation list. Patients were allocated to one of the two preoxygenation strategies using a secure computer-generated online remote system controlled by the independent research promotion unit at the University Hospital of Nantes which had no role in patient recruitment.

Intervention: preoxygenation and intubation procedure

The study design is described in Supplement 1. Preoxygenation and intubation were performed according to international recommendations in all centers [9]. The intubation team was composed of one junior and one senior.

  1. 1.

    Preoxygenation was started according to the randomization group for a 4-min period [17] in a head-up position. In the intervention group, preoxygenation was performed with HFNC (AIRVO 2™; Fisher & Paykel Healthcare, Auckland, NZ), nasal prongs set at 60 L/min flow of heated and humidified pure oxygen (100% fraction of inspired oxygen, 37 °C). Large or medium nasal cannulae were chosen according to the patient’s nostril size to limit air contamination. In the control group, preoxygenation was performed with SMO set at 15 L/min oxygen flow with a disposable self-inflating resuscitator with a reservoir bag manually held in place ensuring air tightness.

  2. 2.

    At the end of preoxygenation, RSI including immediate intubation after administration of the hypnotic drugs was performed in all patients. Drugs and the use of a neuromuscular blocking agent were left to the discretion of the physician.

  3. 3.

    During laryngoscopy and intubation, HFNC was maintained in place throughout the procedure in an attempt to achieve apneic oxygenation. In the control group, SMO was removed in order to perform laryngoscopy.

  4. 4.

    In case of difficult intubation, the stylet was the first alternative device which was proposed [18] if external laryngeal manipulation did not improve glottis vision.

  5. 5.

    At the end of intubation, the correct position of the tracheal tube was confirmed by using capnography [19].

Endpoints

The primary endpoint was the median lowest SpO2 during intubation, from the beginning of the laryngoscopy until the patient was connected to the mechanical ventilator. To ensure exhaustive data collection, a research nurse and an external observer concurrently collected SpO2 fluctuations during the intubation.

Secondary endpoints were SpO2 from the beginning of preoxygenation to the end of intubation focusing on desaturation (below 95%, 90%, 80%); the rate of difficult intubation [20] and the intubation difficulty scale (IDS) score [21]; the need to proceed to face mask ventilation to correct desaturation; adverse events a priori classified as severe (death, cardiac arrest, SpO2 < 80%, severe hypotension defined by systolic blood pressure < 80 mmHg or vasopressor initiation or 30% dose increment) or moderate (ventricular or supraventricular arrhythmia requiring intervention, esophageal intubation, vomiting with aspiration of gastric content, dental injury); organ failure during the first 5 days (SOFA score); time on the ventilator; length of stay in ICU; occurrence of ventilator-associated pneumonia and mortality rate on day 28 (full study protocol appears in Supplement 2).

Sample size

On the basis of previous studies we hypothesized that HFNC would improve the mean lowest SpO2 value by 5% (from 88% in the control group to 93% in the HFNC group) with a standard deviation of SpO2 of 12% [15, 22]. With 80% power, a 5% type I error (two-sided tests), and 5% attrition rate, the required sample size was 192 patients.

Statistical analysis

Baseline characteristics were described as count and percentage for categorical variables and mean and standard deviation or median and interquartiles for quantitative variables when assumption of normality was not met. An intent-to-treat statistical analysis was conducted. Comparisons of primary and secondary outcomes were performed using the Mantel–Haenszel Chi-square tests (or two-tailed Fisher exact tests when appropriate) for qualitative data and the Van Elteren tests for quantitative data to account for stratification factor (center). For qualitative data, relative risk (with adjustment for center effects) and 95% confidence interval were estimated with the Mantel–Haenszel method. Multivariable analysis of SpO2 < 90%, SpO2 < 80%, and the occurrence of adverse events were performed using logistic regression. The most clinically relevant variables (skills of the first operator, difficult intubation, MACOCHA score ≥ 3, and PaO2/FiO2 ratio at baseline) or associated with the randomization group (P values < 0.20) in univariate analysis (age and NYHA class) were introduced simultaneously into the models. Odds ratios and their 95% confidence intervals (CI) were estimated. All tests were two-tailed. P values of less than 0.05 were considered significant. No data was missing for the primary outcome. For secondary outcomes, missing data were clearly stated in the tables (see also Supplement 3), and no method of imputation was used. SAS software version 9.4 (SAS Institute Inc.) was used.

Results

Patients

In the seven participating ICUs, 192 patients were randomized in the study between April 2016 and June 2017. Three patients withdrew their consent, two patients were discovered to be under guardianship protection, and three patients improved before the start of the study and were therefore not intubated. We considered 184 patients (95 in HFNC and 89 in SMO) in the intent-to-treat (ITT) and 183 patients in the per-protocol (PP) analyses (Fig. 1). Per-protocol analysis excluded the only patient with protocol deviation in the HFNC group who did not receive preoxygenation with HFNC because no device was available. This patient did not present any complication during intubation.

Fig. 1
figure 1

Study flowchart. HFNC high-flow nasal cannulae, SMO standard bag-valve mask oxygenation

At baseline, patients were mainly intubation for a medical condition with a median [interquartile range, IQR] simplified acute physiology score 2 (SAPS II) of 45 [35; 60] in HFNC and 41 [32; 57] in SMO; median [IQR] PaO2/FiO2 was 318 mmHg [242; 396] in HFNC and 375 mmHg [276; 446] in SMO. In the two groups, intubation was mainly performed for neurologic reasons in comatose patients: 68% in HFNC and 75% in SMO (Table 1). In order to prevent hypotension, fluid loading was performed before anesthesia induction in 41 (43%) patients in HFNC vs. 37 (42%) in SMO. Airways, operators, and technical aspects of intubation are described in Table 2. Mean (standard deviation, SD) MACOCHA score to predict difficult intubation was 2.0 (1.6) in HFNC and 2.2 (2.0) in SMO [23]. There was no between-group difference in doses or nature of hypnotics (see Supplement 3). All intubations were performed with two operators including a senior and the first operators were mainly (84%) juniors. In the HFNC group, cannula sizes were large for 45% and medium for 55%. No patient was lost to follow-up (see Supplement 3).

Table 1 Baseline characteristics of the intent-to-treat population
Table 2 Airway, preoxygenation, and intubation settings at baseline

Primary outcome: lowest SpO2 during the intubation procedure

In the intent-to-treat (ITT) analysis (Table 3, Fig. 2), there was no significant difference between the median [IQR] lowest SpO2 during intubation in the HFNC group compared with the SMO group, 100% [97; 100] versus 99% [95; 100] respectively, (P = 0.30). In the per-protocol (PP) analysis (attrition rate 1%), the lack of difference persisted (P = 0.26). Although only non-severely hypoxemic patients were included, the lowest SpO2 during intubation ranged from 69% to 100% in HFNC and from 43% to 100% in SMO.

Table 3 Primary and secondary outcomes
Fig. 2
figure 2

Primary outcome: lowest SpO2 during the intubation procedure. HFNC high-flow nasal cannulae, SMO standard bag-valve mask oxygenation, ITT intent-to-treat analysis, PP per-protocol analysis, SD standard deviation, IQR interquartile range, SpO2 arterial oxygen saturation measured by pulse oximetry

Secondary outcomes

All of the patients were successfully intubated and only one operator removed HFNC for laryngoscopic vision obstruction. At the end of the preoxygenation period, median [IQR] SpO2 was 100% [100; 100], (P = 0.66) in both groups (Table 3). During preoxygenation, jaw-thrust maneuver was performed in 13% of the patients in HFNC and in 33% in SMO (relative risk 0.38, 95% CI 0.21–0.71, P = 0.0009). The median [IQR, min–max] length of intubation (namely, apnea duration) was higher in HFNC (1 [0.5; 1.9, 0–23]) vs. in SMO (0.8 [0.5; 1.4, 0–18] min, P = 0.04. Eleven patients (12%) in HFNC vs. 20 (23%) in SMO experienced an SpO2 drop below 95% (relative risk 0.51, 95% CI 0.26–0.99, P = 0.045). Six patients (6%) in the HFNC group vs. 12 (14%) in the SMO group experienced an SpO2 < 90% (P = 0.10). Two patients (2%) in the HFNC group vs. 7 (8%) in the SMO group experienced a severe SpO2 < 80% (P = 0.06). Overall, during intubation, 6 patients (6%) in the HFNC group vs. 17 patients (19%) in the SMO group (relative risk 0.31, 95% CI 0.13–0.76, P = 0.007) presented at least one intubation-related complication. Severe complications including SpO2 < 80%, severe hypotension, and cardiac arrest occurred in 6% of the patients in HFNC vs. 16% in SMO (relative risk 0.38, 95% CI 0.15–0.95, P = 0.03). No patient in the HFNC group presented moderate complications vs. 7% in the SMO group, P = 0.01 (Table 3 and see Supplement 3). Multivariable analysis showed that HFNC was associated with less desaturation < 90%, less intubation-related complications, and a trend to reduce oxygen desaturation < 80% (P = 0.058) (see Table 4 and Supplement 3). When the complications during intubation and the following hour were pooled, the results followed the same trends (see Supplement 3). The description of the drugs used for RSI and the type of adverse events are described in Supplement 3.

Table 4 Multivariable analysis for desaturation < 90% or complications during intubation

Patients in the HFNC and SMO groups had comparable median [IQR] durations of mechanical ventilation, respectively 3 [2; 6] vs. 3 [2; 7] days, P = 0.80. There was also no difference in SOFA score during the first 5 days after randomization and the length of stay in the ICU (P = 0.89). The mortality rate on day 28 was 27% in both groups, P = 0.93.

Discussion

During intubation in non-severely hypoxemic critically ill patients, HFNC as a preoxygenation and apneic oxygenation device did not increase the median lowest SpO2 compared with SMO. However, HFNC significantly reduced intubation-related complications by threefold including severe complications. The latter result is a major finding considering the high rate of complications observed in the present study, i.e., occurring in more than 10% of patients, despite inclusion of non-severely hypoxemic patients. Considering that only one operator removed HFNC during intubation, holding the nasal prongs in place during intubation did not appear to obstruct vision or interfere with intubation.

Our team has already reported that preoxygenation in severely hypoxemic patients with HFNC compared with SMO did not improve median lowest SpO2 during intubation [3]. To our knowledge, the present study is the first large multicenter randomized clinical trial assessing HFNC preoxygenation before intubation in non-severely hypoxemic critically ill patients. The only publication available in this setting was the study by Miguel-Montanes et al. [15]. Their study reported an improvement in the median lowest SpO2 during intubation and a reduction in severe oxygen desaturations when using HFNC compared with face mask for preoxygenation. Although it included patients with baseline characteristics close to ours, the before–after and single-center design may have overestimated the treatment effect, limiting the generalization of the results and their reproducibility in multicenter studies. In the PROTRACH study, these trends were not confirmed, but HFNC appeared as a protective factor for desaturation < 90% in multivariable analysis (see Table 4). In line with previous studies [2, 6], and similar to hypoxemic patients [3], our results highlighted that non-severely hypoxemic patients experience numerous life-threatening adverse events during intubation including deep desaturation with minimum SpO2 of 69% in HFNC vs. 43% in SMO.

Each complication related to preoxygenation or intubation was immediately declared by the investigators and reviewed by the safety board, ensuring high reporting accuracy. Patients undergoing HFNC preoxygenation experienced fewer adverse events, including fewer severe (mainly SpO2 < 80%) and moderate complications. Several theoretical advantages of HFNC could explain these results. HFNC has already been reported to improve FRC, one of the main determinants of the duration of apnea without desaturation. Reducing the risk of desaturation, HFNC could in turn reduce complications such as hypotension related to hypoxemia [24]. Compared with SMO requiring two-handed thenar eminence grip to perform jaw-thrust maneuver and ensure airtightness, HFNC is a “hands-free” preoxygenation device that could help operators to be more confident and avoid scrambled intubation. Nevertheless, this aspect was not specifically assessed in the study. The reduction of desaturation < 95% in the HFNC group despite longer and more difficult intubation could also be recognized as a marker of apneic oxygenation even if this remains controversial [25]. Preoxygenation has to be differentiated from apneic oxygenation [26]. As exhaled fraction of oxygen (EtO2) was not assessed, the reduction of desaturation in HFNC should not be interpreted as increased oxygen reserves at the end of the preoxygenation compared with SMO. However, the administration of oxygen during the apnea period with HFNC could explain the lower rate of respiratory adverse events. These results are concordant with the OPTINIV study [13]. This study showed that in severe hypoxemic patients the preoxygenation with non-invasive ventilation (NIV, namely the gold standard device in this setting [22, 27]) combined with HFNC increased the lowest saturation during intubation compared to NIV alone. Overall, a comprehensive improvement of the intubation process including preoxygenation optimization, better ergonomics for the operator, and possible apneic oxygenation to offset oxygen uptake during intubation could account for these results. The technical aspect of preoxygenation with SMO can probably explain the high rate of jaw-thrust maneuver in this group.

The rate of intubation-related complications was higher in the SMO group. Strikingly, in the HFNC group difficult intubation and intubation for acute respiratory failure were more frequent, and median PaO2/FiO2 was lower (P = 0.009) and should have led to an increase in the number of desaturation and adverse events in this group. Thus, in order to account for potential baseline imbalance, a multivariable analysis was performed and found that HFNC preoxygenation was associated with a fivefold decrease in oxygen desaturation below 90% and a fourfold decrease in intubation-related complications. Nevertheless, given the small number of intubation-related adverse events, we need to be cautious before generalizing these results.

The study includes unselected patients regardless of their admission diagnosis, in medical and surgical settings from academic and non-academic hospitals, reflecting daily practice in the ICUs and enhancing external validity. In this setting, albeit negative in terms of the primary outcome, the study provided several findings suggesting that HFNC could secure intubation by reducing complications including desaturations. Although this study enrolled ICU patients, these results could be applied to all intubations in the emergency department in non-severely hypoxemic patients considering that HFNC is now a widespread oxygenation device.

This trial has several limitations. An unblinded preoxygenation device may have interfered with our findings. Although feasible [13], blinding may have been very difficult to maintain in a large sample and multicenter trial. The choice of the primary outcome can be discussed. Considering the Miguel-Montanes et al. results, we decided to assess minimal SpO2 instead of severe desaturation below 80%, a common criteria which, however, is infrequent in the non-severely hypoxemic patient. In the same way, whether a 5% difference of SpO2 is a clinically relevant endpoint may be questionable. However, such a difference can lead to a decision of bag-valve mask ventilation or can initiate a fast and deep hypoxemia given the sigmoidal relationship between the affinity of hemoglobin for oxygen and the PaO2. In this study, the 5% difference was not met. SpO2 is not the most relevant parameter to assess oxygenation [28], but systematic arterial catheterization before intubation to measure arterial oxygen saturation or partial oxygen pressure could not be ethically defended. Similarly, EtO2 could have been a relevant way to assess the quality of preoxygenation, in particular to check that the preoxygenation was well performed in the SMO group, with a mask correctly applied to avoid leaks. It is an indicator of FRC oxygen saturation and when EtO2 > 90% is achieved, preoxygenation can be claimed to be effective [29]. Nevertheless, in a non-occlusive device such as HFNC, measurement of EtO2 was difficult to achieve [30]. In the same way, the precise level of post-apnea end-tidal carbon dioxide (EtCO2) was not recorded and could have provided relevant information [31]. However, we made the pragmatic choice to evaluate SpO2, the only worldwide, non-invasive, and ready-to-use device informing the attending physician on apnea tolerance during intubation. Given the limitations of SpO2 measurement, a research nurse was specifically dedicated to monitor SpO2 throughout the procedure to improve data collection. Finally, NIV was not selected as the active comparator. Although it was found to improve preoxygenation in obese patients in the operating room [32], and in severe hypoxemic patients in ICU [22, 33], no data is available in non-severely hypoxemic patients in the ICU.

Conclusion

Compared with SMO, preoxygenation with HFNC provided similar lowest SpO2 during intubation in the ICU. However, HFNC was associated with enhanced safety of the procedure. These secondary results need to be confirmed and could be in favor of performing apneic oxygenation following preoxygenation in non-severely hypoxemic ICU patients.