Double intramedullary cortical button versus suture anchors for distal biceps tendon repair: a biomechanical comparison

verfasst von: Sebastian Siebenlist, Arne Buchholz, Julian Zapf, Gunther H. Sandmann, Karl F. Braun, Frank Martetschläger, Alexander Hapfelmeier, Tobias M. Kraus, Andreas Lenich, Peter Biberthaler, Florian Elser

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 3/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this biomechanical in vitro study was to compare the novel technique of double intramedullary cortical button (DICB) fixation with the well-established method of suture anchor (SA) fixation for distal biceps tendon repair.

Methods

A matched-pair analysis (24 human cadaveric radii) was performed with respect to cyclic loadings and failure strengths. Twelve specimens per group were cyclically loaded for 1,000 cycles at 1.5 Hz from 5 to 50 N and from 5 to 100 N, respectively. The tendon–bone displacement was optically analysed using the Image J Software (National Institute of Health). Afterwards, all specimens were pulled to failure. Maximum load to failure and mode of failure were recorded.

Results

All DICB constructs passed the cyclic loading test, whereas 4 of the 12 specimens within the SA group failed by anchor pull-out. Cyclic loading showed a mean tendon–bone displacement of 0.6 ± 1.4 mm for the DICB group and 1.4 ± 1.4 mm for the SA group (n.s.) after 1,000 cycles with 50 N, and a mean displacement of 2.1 ± 2.4 mm for the DICB group and 3.5 ± 3.7 mm for the SA group (n.s.) after 1,000 cycles with 100 N. Load to failure testing showed a mean failure load of 312 ± 76 N and a stiffness of 67.1 ± 11.7 N/mm for the DICB technique. The mean load to failure for the SA repair was 200 ± 120 N (n.s.) and the stiffness was 55.9 ± 21.3 N/mm (n.s.).

Conclusions

The novel technique of DICB fixation showed small tendon–bone displacement during cyclic testing and reliable fixation strength to the bone in load to failure. Moreover, all DICB constructs passed cyclic loadings without failure. Based on the current findings, a more aggressive postoperative rehabilitation may be allowed for the DICB repair in clinical use.
Literatur
1.
Zurück zum Zitat An KN, Hui FC, Morrey BF, Linscheid RL, Chao EY (1981) Muscles across the elbow joint: a biomechanical analysis. J Biomech 14:659–669CrossRefPubMed An KN, Hui FC, Morrey BF, Linscheid RL, Chao EY (1981) Muscles across the elbow joint: a biomechanical analysis. J Biomech 14:659–669CrossRefPubMed
2.
Zurück zum Zitat Athwal GS, Steinmann SP, Rispoli DM (2007) The distal biceps tendon: footprint and relevant clinical anatomy. J Hand Surg Am 32:1225–1229CrossRefPubMed Athwal GS, Steinmann SP, Rispoli DM (2007) The distal biceps tendon: footprint and relevant clinical anatomy. J Hand Surg Am 32:1225–1229CrossRefPubMed
3.
Zurück zum Zitat Bain GI, Prem H, Heptinstall RJ, Verhellen R, Paix D (2000) Repair of distal biceps tendon rupture: a new technique using the Endobutton. J Should Elb Surg 9:120–126CrossRef Bain GI, Prem H, Heptinstall RJ, Verhellen R, Paix D (2000) Repair of distal biceps tendon rupture: a new technique using the Endobutton. J Should Elb Surg 9:120–126CrossRef
4.
Zurück zum Zitat Baker BE, Bierwagen D (1985) Rupture of the distal biceps brachii. Operative versus non-operative treatment. J Bone Joint Surg Am 67:414–417PubMed Baker BE, Bierwagen D (1985) Rupture of the distal biceps brachii. Operative versus non-operative treatment. J Bone Joint Surg Am 67:414–417PubMed
5.
Zurück zum Zitat Berlet GC, Johnson JA, Milne AD, Patterson SD, King GJ (1998) Distal biceps brachii tendon repair: an in vitro biomechanical study of tendon reattachment. Am J Sports Med 26:428–432PubMed Berlet GC, Johnson JA, Milne AD, Patterson SD, King GJ (1998) Distal biceps brachii tendon repair: an in vitro biomechanical study of tendon reattachment. Am J Sports Med 26:428–432PubMed
6.
Zurück zum Zitat Chavan PR, Duquin TR, Bisson LJ (2008) Repair of the ruptured distal biceps tendon: a systematic review. Am J Sports Med 36:1618–1624CrossRefPubMed Chavan PR, Duquin TR, Bisson LJ (2008) Repair of the ruptured distal biceps tendon: a systematic review. Am J Sports Med 36:1618–1624CrossRefPubMed
7.
Zurück zum Zitat Cho CH, Song KS, Choi IJ et al (2011) Insertional anatomy and clinical relevance of the distal biceps tendon. Knee Surg Sports Traumatol Arthrosc 19:1930–1935CrossRefPubMed Cho CH, Song KS, Choi IJ et al (2011) Insertional anatomy and clinical relevance of the distal biceps tendon. Knee Surg Sports Traumatol Arthrosc 19:1930–1935CrossRefPubMed
8.
Zurück zum Zitat Dirim B, Brouha SS, Pretterklieber ML et al (2008) Terminal bifurcation of the biceps brachii muscle and tendon: anatomic considerations and clinical implications. AJR Am J Roentgenol 191:245–255CrossRef Dirim B, Brouha SS, Pretterklieber ML et al (2008) Terminal bifurcation of the biceps brachii muscle and tendon: anatomic considerations and clinical implications. AJR Am J Roentgenol 191:245–255CrossRef
9.
Zurück zum Zitat Eames MH, Bain GI, Fogg QA, van Riet RP (2007) Distal biceps tendon anatomy: a cadaveric study. J Bone Joint Surg Am 89:1044–1049CrossRefPubMed Eames MH, Bain GI, Fogg QA, van Riet RP (2007) Distal biceps tendon anatomy: a cadaveric study. J Bone Joint Surg Am 89:1044–1049CrossRefPubMed
10.
Zurück zum Zitat Forthman CL, Zimmerman RM, Sullivan MJ, Gabel GT (2008) Cross-sectional anatomy of the bicipital tuberosity and biceps brachii tendon insertion: relevance to anatomic tendon repair. J Should Elb Surg 17:522–526CrossRef Forthman CL, Zimmerman RM, Sullivan MJ, Gabel GT (2008) Cross-sectional anatomy of the bicipital tuberosity and biceps brachii tendon insertion: relevance to anatomic tendon repair. J Should Elb Surg 17:522–526CrossRef
11.
Zurück zum Zitat Greenberg JA, Fernandez JJ, Wang T, Turner C (2003) EndoButton-assisted repair of distal biceps tendon ruptures. J Should Elb Surg 12:484–490CrossRef Greenberg JA, Fernandez JJ, Wang T, Turner C (2003) EndoButton-assisted repair of distal biceps tendon ruptures. J Should Elb Surg 12:484–490CrossRef
12.
Zurück zum Zitat Grewal R, Athwal GS, Macdermid JC, Faber KJ et al (2012) Single versus double-incision technique for the repair of acute distal biceps tendon ruptures: a randomized clinical trial. J Bone Joint Surg 94:1166–1174CrossRefPubMed Grewal R, Athwal GS, Macdermid JC, Faber KJ et al (2012) Single versus double-incision technique for the repair of acute distal biceps tendon ruptures: a randomized clinical trial. J Bone Joint Surg 94:1166–1174CrossRefPubMed
13.
Zurück zum Zitat Heinzelmann AD, Savoie FH 3rd, Ramsey JR, Field LD, Mazzocca AD (2009) A combined technique for distal biceps repair using a soft tissue button and biotenodesis interference screw. Am J Sports Med 37:989–994CrossRefPubMed Heinzelmann AD, Savoie FH 3rd, Ramsey JR, Field LD, Mazzocca AD (2009) A combined technique for distal biceps repair using a soft tissue button and biotenodesis interference screw. Am J Sports Med 37:989–994CrossRefPubMed
14.
Zurück zum Zitat Hetsroni I, Pilz-Burstein R, Nyska M et al (2008) Avulsion of the distal biceps brachii tendon in middle-aged population: is surgical repair advisable? A comparative study of 22 patients treated with either nonoperative management or early anatomical repair. Injury 39:753–760CrossRefPubMed Hetsroni I, Pilz-Burstein R, Nyska M et al (2008) Avulsion of the distal biceps brachii tendon in middle-aged population: is surgical repair advisable? A comparative study of 22 patients treated with either nonoperative management or early anatomical repair. Injury 39:753–760CrossRefPubMed
15.
Zurück zum Zitat Hutchinson HL, Gloystein D, Gillespie M (2008) Distal biceps tendon insertion: an anatomic study. J Should Elb Surg 17:342–346CrossRef Hutchinson HL, Gloystein D, Gillespie M (2008) Distal biceps tendon insertion: an anatomic study. J Should Elb Surg 17:342–346CrossRef
16.
Zurück zum Zitat Iriuchishima T, Shirakura K, Yorifuji H, Aizawa S, Murakami T, Fu FH (2013) ACL footprint size is correlated with the height and area of the lateral wall of femoral intercondylar notch. Knee Surg Sports Traumatol Arthrosc 21:789–796CrossRefPubMed Iriuchishima T, Shirakura K, Yorifuji H, Aizawa S, Murakami T, Fu FH (2013) ACL footprint size is correlated with the height and area of the lateral wall of femoral intercondylar notch. Knee Surg Sports Traumatol Arthrosc 21:789–796CrossRefPubMed
17.
Zurück zum Zitat Iriuchishima T, Yorifuji H, Aizawa S, Tajika Y, Murakami T, Fu FH (2012) Evaluation of ACL mid-substance cross-sectional area for reconstructed autograft selection. Knee Surg Sports Traumatol Arthrosc, doi:10.1007/s00167-012-2356-0 Iriuchishima T, Yorifuji H, Aizawa S, Tajika Y, Murakami T, Fu FH (2012) Evaluation of ACL mid-substance cross-sectional area for reconstructed autograft selection. Knee Surg Sports Traumatol Arthrosc, doi:10.​1007/​s00167-012-2356-0
18.
Zurück zum Zitat Jobin CM, Kippe MA, Gardner TR, Levine WN, Ahmad CS (2009) Distal biceps tendon repair: a cadaveric analysis of suture anchor and interference screw restoration of the anatomic footprint. Am J Sports Med 37:2214–2221CrossRefPubMed Jobin CM, Kippe MA, Gardner TR, Levine WN, Ahmad CS (2009) Distal biceps tendon repair: a cadaveric analysis of suture anchor and interference screw restoration of the anatomic footprint. Am J Sports Med 37:2214–2221CrossRefPubMed
19.
Zurück zum Zitat John CK, Field LD, Weiss KS, Savoie FH 3rd (2007) Single incision repair of acute distal biceps ruptures by use of suture anchors. J Should Elb Surg 16:78–83CrossRef John CK, Field LD, Weiss KS, Savoie FH 3rd (2007) Single incision repair of acute distal biceps ruptures by use of suture anchors. J Should Elb Surg 16:78–83CrossRef
20.
Zurück zum Zitat Khan AD, Penna S, Yin Q, Sinopidis C, Brownson P, Frostick SP (2008) Repair of distal biceps tendon ruptures using suture anchors through a single anterior incision. Arthroscopy 24:39–45CrossRefPubMed Khan AD, Penna S, Yin Q, Sinopidis C, Brownson P, Frostick SP (2008) Repair of distal biceps tendon ruptures using suture anchors through a single anterior incision. Arthroscopy 24:39–45CrossRefPubMed
21.
Zurück zum Zitat Kettler M, Lunger J, Kuhn V, Mutschler W, Tingart MJ (2007) Failure strengths in distal biceps tendon repair. Am J Sports Med 35:1544–1548CrossRefPubMed Kettler M, Lunger J, Kuhn V, Mutschler W, Tingart MJ (2007) Failure strengths in distal biceps tendon repair. Am J Sports Med 35:1544–1548CrossRefPubMed
22.
Zurück zum Zitat Kettler M, Tingart MJ, Lunger J, Kuhn V (2008) Reattachment of the distal tendon of biceps: factors affecting the failure strength of the repair. J Bone Joint Surg Br 90:103–106CrossRefPubMed Kettler M, Tingart MJ, Lunger J, Kuhn V (2008) Reattachment of the distal tendon of biceps: factors affecting the failure strength of the repair. J Bone Joint Surg Br 90:103–106CrossRefPubMed
23.
Zurück zum Zitat King J, Bollier M (2008) Repair of distal biceps tendon ruptures using the Endobutton. J Am Acad Orthop Surg 16:490–494PubMed King J, Bollier M (2008) Repair of distal biceps tendon ruptures using the Endobutton. J Am Acad Orthop Surg 16:490–494PubMed
24.
Zurück zum Zitat Krushinski EM, Brown JA, Murthi AM (2007) Distal biceps tendon rupture: biomechanical analysis of repair strength of the Bio-Tenodesis screw versus suture anchors. J Should Elb Surg 16:218–223CrossRef Krushinski EM, Brown JA, Murthi AM (2007) Distal biceps tendon rupture: biomechanical analysis of repair strength of the Bio-Tenodesis screw versus suture anchors. J Should Elb Surg 16:218–223CrossRef
25.
Zurück zum Zitat Lemos SE, Ebramzedeh E, Kvitne RS (2004) A new technique: in vitro suture anchor fixation has superior yield strength to bone tunnel fixation for distal biceps tendon repair. Am J Sports Med 32:406–410CrossRefPubMed Lemos SE, Ebramzedeh E, Kvitne RS (2004) A new technique: in vitro suture anchor fixation has superior yield strength to bone tunnel fixation for distal biceps tendon repair. Am J Sports Med 32:406–410CrossRefPubMed
26.
Zurück zum Zitat Lo EY, Li CS, Van den Bogaerde JM (2011) The effect of drill trajectory on proximity to the posterior interosseous nerve during cortical button distal biceps repair. Arthroscopy 27:1048–1054CrossRefPubMedCentralPubMed Lo EY, Li CS, Van den Bogaerde JM (2011) The effect of drill trajectory on proximity to the posterior interosseous nerve during cortical button distal biceps repair. Arthroscopy 27:1048–1054CrossRefPubMedCentralPubMed
27.
Zurück zum Zitat Mazzocca A, Burton KJ, Romeo AA, Santangelo S, Adams DA, Arciero RA (2007) Biomechanical evaluation of 4 techniques of distal biceps brachii tendon repair. Am J Sports Med 35:252–258CrossRefPubMed Mazzocca A, Burton KJ, Romeo AA, Santangelo S, Adams DA, Arciero RA (2007) Biomechanical evaluation of 4 techniques of distal biceps brachii tendon repair. Am J Sports Med 35:252–258CrossRefPubMed
28.
Zurück zum Zitat Mazzocca AD, Cohen M, Berkson E et al (2007) The anatomy of the bicipital tuberosity and distal biceps tendon. J Should Elb Surg 16:122–127CrossRef Mazzocca AD, Cohen M, Berkson E et al (2007) The anatomy of the bicipital tuberosity and distal biceps tendon. J Should Elb Surg 16:122–127CrossRef
29.
Zurück zum Zitat Mokhtee DB, Brown JM, Mackinnon SE, Tung TH (2009) Reconstruction of posterior interosseous nerve injury following biceps tendon repair: case report and cadaveric study. Hand 4:134–139CrossRefPubMedCentralPubMed Mokhtee DB, Brown JM, Mackinnon SE, Tung TH (2009) Reconstruction of posterior interosseous nerve injury following biceps tendon repair: case report and cadaveric study. Hand 4:134–139CrossRefPubMedCentralPubMed
30.
Zurück zum Zitat Peeters T, Soon C, Jansen N et al (2009) Functional outcome after repair of distal biceps tendon ruptures using the Endobutton technique. J Should Elb Surg 18:283–287CrossRef Peeters T, Soon C, Jansen N et al (2009) Functional outcome after repair of distal biceps tendon ruptures using the Endobutton technique. J Should Elb Surg 18:283–287CrossRef
31.
Zurück zum Zitat Pereira DS, Kvitne RS, Liang M, Giacobetti FB, Ebramzadeh E (2002) Surgical repair of distal biceps tendon ruptures: a biomechanical comparison of two techniques. Am J Sports Med 30:432–436PubMed Pereira DS, Kvitne RS, Liang M, Giacobetti FB, Ebramzadeh E (2002) Surgical repair of distal biceps tendon ruptures: a biomechanical comparison of two techniques. Am J Sports Med 30:432–436PubMed
32.
Zurück zum Zitat Rantanen J, Orava S (1999) Rupture of the distal biceps tendon: a report of 19 patients treated with anatomic reinsertion, and a meta-analysis of 147 cases found in the literature. Am J Sports Med 27:128–132PubMed Rantanen J, Orava S (1999) Rupture of the distal biceps tendon: a report of 19 patients treated with anatomic reinsertion, and a meta-analysis of 147 cases found in the literature. Am J Sports Med 27:128–132PubMed
33.
Zurück zum Zitat Rose DM, Archibald JD, Sutter EG, Belkoff SM, Wilckens JH (2011) Biomechanical analysis suggests early rehabilitation is possible after single-incision EndoButton distal biceps repair with FiberWire. Knee Surg Sports Traumatol Arthrosc 19:1019–1022CrossRefPubMed Rose DM, Archibald JD, Sutter EG, Belkoff SM, Wilckens JH (2011) Biomechanical analysis suggests early rehabilitation is possible after single-incision EndoButton distal biceps repair with FiberWire. Knee Surg Sports Traumatol Arthrosc 19:1019–1022CrossRefPubMed
34.
Zurück zum Zitat Safran MR, Graham SM (2004) Distal biceps tendon ruptures: incidence, demographics, and the effect of smoking. Clin Orthop Relat Res 404:275–283CrossRef Safran MR, Graham SM (2004) Distal biceps tendon ruptures: incidence, demographics, and the effect of smoking. Clin Orthop Relat Res 404:275–283CrossRef
35.
Zurück zum Zitat Saldua N, Carney J, Dewing C, Thompson M (2008) The effect of drilling angle on posterior interosseous nerve safety during open and endoscopic anterior single-incision repair of the distal biceps tendon. Arthroscopy 24:305–310CrossRefPubMed Saldua N, Carney J, Dewing C, Thompson M (2008) The effect of drilling angle on posterior interosseous nerve safety during open and endoscopic anterior single-incision repair of the distal biceps tendon. Arthroscopy 24:305–310CrossRefPubMed
36.
Zurück zum Zitat Shin SH, Jeon IH, Kim HJ et al (2010) Articular surface area of the coronoid process and radial head in elbow extension: surface ration in cadavers and a computed tomography in vivo. J Hand Surg Am 35:1120–1125CrossRefPubMed Shin SH, Jeon IH, Kim HJ et al (2010) Articular surface area of the coronoid process and radial head in elbow extension: surface ration in cadavers and a computed tomography in vivo. J Hand Surg Am 35:1120–1125CrossRefPubMed
37.
Zurück zum Zitat Siebenlist S, Elser F, Sandmann GH et al (2011) The double intramedullary cortical button fixation for distal biceps tendon repair. Knee Surg Sports Traumatol Arthrosc 19:1925–1929CrossRefPubMed Siebenlist S, Elser F, Sandmann GH et al (2011) The double intramedullary cortical button fixation for distal biceps tendon repair. Knee Surg Sports Traumatol Arthrosc 19:1925–1929CrossRefPubMed
38.
Zurück zum Zitat Siebenlist S, Lenich A, Buchholz A et al (2011) Biomechanical in vitro validation of intramedullary cortical button fixation for distal biceps tendon repair–a new technique. Am J Sports Med 39:1762–1768CrossRefPubMed Siebenlist S, Lenich A, Buchholz A et al (2011) Biomechanical in vitro validation of intramedullary cortical button fixation for distal biceps tendon repair–a new technique. Am J Sports Med 39:1762–1768CrossRefPubMed
39.
Zurück zum Zitat Spang JT, Weinhold PS, Karas SG (2006) A biomechanical comparison of Endobutton versus suture anchor repair of distal biceps tendon injuries. J Should Elb Surg 15:509–514CrossRef Spang JT, Weinhold PS, Karas SG (2006) A biomechanical comparison of Endobutton versus suture anchor repair of distal biceps tendon injuries. J Should Elb Surg 15:509–514CrossRef
40.
Zurück zum Zitat Thiel W (1992) The preservation of the whole corpse with natural color. Ann Anat 174:185–195CrossRefPubMed Thiel W (1992) The preservation of the whole corpse with natural color. Ann Anat 174:185–195CrossRefPubMed
Metadaten
Titel
Double intramedullary cortical button versus suture anchors for distal biceps tendon repair: a biomechanical comparison
verfasst von
Sebastian Siebenlist
Arne Buchholz
Julian Zapf
Gunther H. Sandmann
Karl F. Braun
Frank Martetschläger
Alexander Hapfelmeier
Tobias M. Kraus
Andreas Lenich
Peter Biberthaler
Florian Elser
Publikationsdatum
01.03.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 3/2015
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2590-0

Weitere Artikel der Ausgabe 3/2015

Knee Surgery, Sports Traumatology, Arthroscopy 3/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.