Computer-assisted surgery improves rotational positioning of the femoral component but not the tibial component in total knee arthroplasty

verfasst von: Daniel Hernandez-Vaquero, Alfonso Noriega-Fernandez, Jose Manuel Fernandez-Carreira, Jose Manuel Fernandez-Simon, Jimena Llorens de los Rios

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 12/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Computer-assisted surgery (CAS) may facilitate better positioning of total knee arthroplasty (TKA) along the coronal and lateral axes; however, there are doubts as to its usefulness in the rotational plane.

Methods

This is a prospective study of 95 TKAs comparing two groups: the CAS group and the standard equipment group. The series comprises 95 cases. A radiography of the lower limb and computer tomographies (CTs) of the femoral condylar region, the proximal end of the tibia and the ankle were performed to measure rotational angulation. A month after TKA surgery, the radiography and the CTs were repeated to analyze the position of the prosthetic components in the rotational plane.

Results

In the coronal axis, both CAS and mechanical technique improved femoro-tibial alignment, but when there are preexisting deformities ≥4°, CAS obtains better results. A strong correlation (R = 0.94, p = 0.001) was observed between the mean rotational axis measured with CT in the tibial plateau and that measured from the axis of the ankle. The mean initial femoral rotation of the complete series was 6.7° and 2.7° at 1-month follow-up (p < 0.001). In the standard instrumentation group, the femoral rotation went from 6.8° to 2.3°, whereas in the CAS group the femoral rotation went from 6.5° to 3.1° (p = 0.039), which is very close to the ideal 3° angle of external rotation. Tibial rotation changed by 5.28° for the entire patient population, but no differences were found when comparing CAS and standard instrumentation.

Conclusion

CAS improves frontal alignment in TKA, especially in the presence of preoperative deformities. In the femoral component, navigation most closely replicated the ideal 3° external rotation of the femoral component, but tibial rotation did not differ when comparing CAS to standard instrumentation.

Level of evidence

II.
Literatur
1.
Zurück zum Zitat Aglietti P, Sensi L, Cuomo P, Ciardullo A (2008) Rotational position of femoral and tibial components in TKA using the femoral transepicondylar axis. Clin Orthop Relat Res 466:2751–2755PubMedCentralPubMedCrossRef Aglietti P, Sensi L, Cuomo P, Ciardullo A (2008) Rotational position of femoral and tibial components in TKA using the femoral transepicondylar axis. Clin Orthop Relat Res 466:2751–2755PubMedCentralPubMedCrossRef
2.
Zurück zum Zitat Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219PubMedCrossRef Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219PubMedCrossRef
3.
Zurück zum Zitat Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after knee arthroplasty. Clin Orthop Rel Res 392:46–55CrossRef Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L (2001) Component rotation and anterior knee pain after knee arthroplasty. Clin Orthop Rel Res 392:46–55CrossRef
4.
Zurück zum Zitat Bedard M, Vince KJ, Redfern J, Collen SR (2011) Internal rotation of the tibial component is frequent in stiff total knee arthroplasty. Clin Orthop Rel Res 469:2346–2355CrossRef Bedard M, Vince KJ, Redfern J, Collen SR (2011) Internal rotation of the tibial component is frequent in stiff total knee arthroplasty. Clin Orthop Rel Res 469:2346–2355CrossRef
6.
Zurück zum Zitat Bonnin MP, Saffarini M, Mercier PE, Laurent JR, Carrillon Y (2011) Is the anterior tibial tuberosity a reliable rotational landmark for the tibial component in total knee arthroplasty? J Arthroplasty 26:260–267PubMedCrossRef Bonnin MP, Saffarini M, Mercier PE, Laurent JR, Carrillon Y (2011) Is the anterior tibial tuberosity a reliable rotational landmark for the tibial component in total knee arthroplasty? J Arthroplasty 26:260–267PubMedCrossRef
7.
Zurück zum Zitat Chauhan SK, Clark GW, Lloyd S, Scott G, Breidahl W, Sikorski JM (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multiparameter quantitative CT assessment of alignment (The Perth CT protocol). J Bone Joint Surg 86-B:818–823CrossRef Chauhan SK, Clark GW, Lloyd S, Scott G, Breidahl W, Sikorski JM (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multiparameter quantitative CT assessment of alignment (The Perth CT protocol). J Bone Joint Surg 86-B:818–823CrossRef
8.
Zurück zum Zitat Cheng T, Zhang G, Zhang X (2011) Imageless navigation system does not improve component rotational alignment in total knee arthroplasty. J Surg Res 171:590–600PubMedCrossRef Cheng T, Zhang G, Zhang X (2011) Imageless navigation system does not improve component rotational alignment in total knee arthroplasty. J Surg Res 171:590–600PubMedCrossRef
9.
Zurück zum Zitat Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef
10.
Zurück zum Zitat Cobb JP, Dixon H, Dandachli W, Iranpour F (2008) The anatomical tibial axis. Reliable rotational orientation in knee replacement. J Bone Joint Surg 90B:1032–1038CrossRef Cobb JP, Dixon H, Dandachli W, Iranpour F (2008) The anatomical tibial axis. Reliable rotational orientation in knee replacement. J Bone Joint Surg 90B:1032–1038CrossRef
11.
Zurück zum Zitat Czurda T, Fennema P, Baumgartner M, Ritschl P (2010) The association between component malalignment and post-operative pain following navigation-assisted total knee arthroplasty: results of a cohort/nested case-control study. Knee Surg Sports Traumatol Arthrosc 18:863–869PubMedCrossRef Czurda T, Fennema P, Baumgartner M, Ritschl P (2010) The association between component malalignment and post-operative pain following navigation-assisted total knee arthroplasty: results of a cohort/nested case-control study. Knee Surg Sports Traumatol Arthrosc 18:863–869PubMedCrossRef
12.
Zurück zum Zitat Hernandez-Vaquero D, Suarez-Vazquez A, Sandoval-Garcia MA, Noriega-Fernandez A (2010) Computer assistance increases precision of component placement in total knee arthroplasty with articular deformity. Clin Orthop Relat Res 468:1237–1241PubMedCentralPubMedCrossRef Hernandez-Vaquero D, Suarez-Vazquez A, Sandoval-Garcia MA, Noriega-Fernandez A (2010) Computer assistance increases precision of component placement in total knee arthroplasty with articular deformity. Clin Orthop Relat Res 468:1237–1241PubMedCentralPubMedCrossRef
13.
Zurück zum Zitat Hernandez-Vaquero D, Suarez-Vazquez A, Iglesias-Fernandez S (2011) Can computer assistance improve the clinical and functional scores in total knee arthroplasty? Clin Orthop Relat Res 469:3436–3442PubMedCentralPubMedCrossRef Hernandez-Vaquero D, Suarez-Vazquez A, Iglesias-Fernandez S (2011) Can computer assistance improve the clinical and functional scores in total knee arthroplasty? Clin Orthop Relat Res 469:3436–3442PubMedCentralPubMedCrossRef
14.
Zurück zum Zitat Hiscox CM, Bohm ER, Turgeon TR, Hedden DR, Burnell CD (2011) Randomized trial of computer-assisted knee arthroplasty: impact on clinical and radiographic outcomes. J Arthroplasty 26:1259–1264PubMedCrossRef Hiscox CM, Bohm ER, Turgeon TR, Hedden DR, Burnell CD (2011) Randomized trial of computer-assisted knee arthroplasty: impact on clinical and radiographic outcomes. J Arthroplasty 26:1259–1264PubMedCrossRef
15.
Zurück zum Zitat Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21:2271–2280PubMedCrossRef Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21:2271–2280PubMedCrossRef
16.
Zurück zum Zitat Hutter EE, Granger JF, Beal MD, Siston RA (2013) Is There a gold standard for TKA tibial component rotational alignment? Clin Orthop Relat Res 471:1646–1653PubMedCentralPubMedCrossRef Hutter EE, Granger JF, Beal MD, Siston RA (2013) Is There a gold standard for TKA tibial component rotational alignment? Clin Orthop Relat Res 471:1646–1653PubMedCentralPubMedCrossRef
17.
Zurück zum Zitat Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T (2007) Determining the rotational alignment of the tibial component at total knee replacement. A comparison of two techniques. J Bone Joint Surg Br 89-B:45–49CrossRef Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T (2007) Determining the rotational alignment of the tibial component at total knee replacement. A comparison of two techniques. J Bone Joint Surg Br 89-B:45–49CrossRef
18.
Zurück zum Zitat Incavo SJ, Coughlin KM, Pappas Ch, Beynnon BD (2003) Anatomic rotational relationships of the proximal tibia, distal femur, and patella. Implications for rotational alignment in total knee arthroplasty. J Arthroplasty 18:643–648PubMedCrossRef Incavo SJ, Coughlin KM, Pappas Ch, Beynnon BD (2003) Anatomic rotational relationships of the proximal tibia, distal femur, and patella. Implications for rotational alignment in total knee arthroplasty. J Arthroplasty 18:643–648PubMedCrossRef
19.
Zurück zum Zitat Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE (2000) The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthroplasty 15:761–766PubMedCrossRef Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE (2000) The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthroplasty 15:761–766PubMedCrossRef
20.
Zurück zum Zitat Jenny JY, Boeri C (2004) Low reproducibility of the intra–operative measurement of the transepicondylar axis during total knee replacement. Acta Orthop Scand 75:74–77PubMedCrossRef Jenny JY, Boeri C (2004) Low reproducibility of the intra–operative measurement of the transepicondylar axis during total knee replacement. Acta Orthop Scand 75:74–77PubMedCrossRef
21.
Zurück zum Zitat Kaipel M, Gergely I, Sinz K, Neumann C, Sinz G (2013) Femoral rotation in ligament balanced knee arthroplasty. A prospective clinical study. J Arthroplasty 28:1103–1106CrossRef Kaipel M, Gergely I, Sinz K, Neumann C, Sinz G (2013) Femoral rotation in ligament balanced knee arthroplasty. A prospective clinical study. J Arthroplasty 28:1103–1106CrossRef
22.
Zurück zum Zitat Kim H-O, Park J-W, Kim J-S, Park S-D (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop 38:379–385PubMedCrossRef Kim H-O, Park J-W, Kim J-S, Park S-D (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop 38:379–385PubMedCrossRef
23.
Zurück zum Zitat Kim SW, Lee H-J, Jung H-J, Lee J-S, Kim KS (2013) Less femoral lift-off and better femoral alignment in TKA using computer-assisted surgery. Knee Surg Sports Traumatol Arthrosc 21:2255–2262PubMedCrossRef Kim SW, Lee H-J, Jung H-J, Lee J-S, Kim KS (2013) Less femoral lift-off and better femoral alignment in TKA using computer-assisted surgery. Knee Surg Sports Traumatol Arthrosc 21:2255–2262PubMedCrossRef
24.
Zurück zum Zitat Koch PP, Müller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21:2200–2205PubMedCrossRef Koch PP, Müller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21:2200–2205PubMedCrossRef
25.
Zurück zum Zitat de Ladoucette A (2012) CT measurements prior to computer-assisted total knee arthroplasty do not improve rotational placement of the femoral component. Knee Surg Sports Traumatol Arthrosc 20:2471–2475CrossRef de Ladoucette A (2012) CT measurements prior to computer-assisted total knee arthroplasty do not improve rotational placement of the femoral component. Knee Surg Sports Traumatol Arthrosc 20:2471–2475CrossRef
26.
Zurück zum Zitat Michaut M, Beaufils P, Galaud B, Abadie P, Boisrenoult P, Fallet L (2008) Rotational alignment of femoral component with computed-assisted surgery (CAS) during total knee arthroplasty. Rev Chir Orthop Reparatrice Appar Mot 94:580–584PubMedCrossRef Michaut M, Beaufils P, Galaud B, Abadie P, Boisrenoult P, Fallet L (2008) Rotational alignment of femoral component with computed-assisted surgery (CAS) during total knee arthroplasty. Rev Chir Orthop Reparatrice Appar Mot 94:580–584PubMedCrossRef
27.
Zurück zum Zitat Moon YW, Ha CW, Do KH, Kim CY, Han JH, Na SE, Lee CH, Kim JG, Park YS (2012) Comparison of robot-assisted and conventional total knee arthroplasty: a controlled cadaver study using multiparameter quantitative three-dimensional CT assessment of alignment. Comput Aided Surg 17:86–95PubMedCrossRef Moon YW, Ha CW, Do KH, Kim CY, Han JH, Na SE, Lee CH, Kim JG, Park YS (2012) Comparison of robot-assisted and conventional total knee arthroplasty: a controlled cadaver study using multiparameter quantitative three-dimensional CT assessment of alignment. Comput Aided Surg 17:86–95PubMedCrossRef
28.
Zurück zum Zitat Nicoll D, Rowley DI (2010). Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg 92-B:1238–1244 Nicoll D, Rowley DI (2010). Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg 92-B:1238–1244
29.
Zurück zum Zitat Parratte S, Blanc G, Boussemart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21:2213–2219PubMedCrossRef Parratte S, Blanc G, Boussemart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21:2213–2219PubMedCrossRef
30.
Zurück zum Zitat Scout RD (2013) Femoral and tibial component rotation in total knee arthroplasty. Methods and consequences. Bone Joint J 95B:140–143CrossRef Scout RD (2013) Femoral and tibial component rotation in total knee arthroplasty. Methods and consequences. Bone Joint J 95B:140–143CrossRef
31.
Zurück zum Zitat Siston RA, Patel JJ, Goodman SB, Delp SL, Giori NJ (2005) The variability of femoral rotational alignment in total knee arthroplasty. J Bone Joint Surg 87A:2276–2280CrossRef Siston RA, Patel JJ, Goodman SB, Delp SL, Giori NJ (2005) The variability of femoral rotational alignment in total knee arthroplasty. J Bone Joint Surg 87A:2276–2280CrossRef
32.
Zurück zum Zitat Siston RA, Cromie MJ, Gold GE, Goodman SB, Delp SL, Maloney WJ, Giori NJ (2008) Averaging different alignment axes improves femoral rotational alignment in computer-navigated total knee arthroplasty. J Bone Joint Surg 90A:2098–2104CrossRef Siston RA, Cromie MJ, Gold GE, Goodman SB, Delp SL, Maloney WJ, Giori NJ (2008) Averaging different alignment axes improves femoral rotational alignment in computer-navigated total knee arthroplasty. J Bone Joint Surg 90A:2098–2104CrossRef
33.
Zurück zum Zitat Stöckl B, Nogler M, Rosiek R, Fischer M, Krismer M, Kessler O (2004) Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res 426:180–186PubMedCrossRef Stöckl B, Nogler M, Rosiek R, Fischer M, Krismer M, Kessler O (2004) Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res 426:180–186PubMedCrossRef
34.
Zurück zum Zitat van der Linden-van der Zwaag HM, Bos J, van der Heide HJ, Nelissen RG (2011). A computed tomography based study on rotational alignment accuracy of the femoral component in total knee arthroplasty using computer-assisted orthopaedic surgery. Int Orthop 35:845–850 van der Linden-van der Zwaag HM, Bos J, van der Heide HJ, Nelissen RG (2011). A computed tomography based study on rotational alignment accuracy of the femoral component in total knee arthroplasty using computer-assisted orthopaedic surgery. Int Orthop 35:845–850
35.
Zurück zum Zitat Victor J (2009) Rotational alignment of the distal femur: a literature review. Orthop Traumat Surg Res 95:365–372CrossRef Victor J (2009) Rotational alignment of the distal femur: a literature review. Orthop Traumat Surg Res 95:365–372CrossRef
36.
Zurück zum Zitat Walmsley D, Backstein D (2013) Rotational alignment in total knee arthroplasty. Sem Arthroplasty 24:156–159CrossRef Walmsley D, Backstein D (2013) Rotational alignment in total knee arthroplasty. Sem Arthroplasty 24:156–159CrossRef
37.
Zurück zum Zitat Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non–image-based navigation-assisted in total knee arthroplasty. J Arthroplasty 22:1150–1161PubMedCrossRef Yau WP, Leung A, Liu KG, Yan CH, Wong LLS, Chiu KY (2007) Interobserver and intra-observer errors in obtaining visually selected anatomical landmarks during registration process in non–image-based navigation-assisted in total knee arthroplasty. J Arthroplasty 22:1150–1161PubMedCrossRef
Metadaten
Titel
Computer-assisted surgery improves rotational positioning of the femoral component but not the tibial component in total knee arthroplasty
verfasst von
Daniel Hernandez-Vaquero
Alfonso Noriega-Fernandez
Jose Manuel Fernandez-Carreira
Jose Manuel Fernandez-Simon
Jimena Llorens de los Rios
Publikationsdatum
01.12.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 12/2014
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-014-3233-9

Weitere Artikel der Ausgabe 12/2014

Knee Surgery, Sports Traumatology, Arthroscopy 12/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.