Functional biomechanical performance of a novel anatomically shaped polycarbonate urethane total meniscus replacement

verfasst von: A. C. T. Vrancken, F. Eggermont, T. G. van Tienen, G. Hannink, P. Buma, D. Janssen, N. Verdonschot

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 5/2016

Abstract

Purpose

To evaluate the functional biomechanical performance of a novel anatomically shaped, polycarbonate urethane total meniscus implant.

Methods

Five human cadaveric knees were flexed between 0° and 90° under compressive loads mimicking a squat movement. Anteroposterior (AP) laxity tests were performed in 30° and 90° flexion. Meniscal kinematics and knee laxity were quantified using roentgen stereophotogrammetric analysis. Tibial cartilage contact mechanics were determined in 90° flexion. Measurements were repeated for the native medial meniscus, the implant, after total medial meniscectomy and allograft transplantation.

Results

The implant and allograft displayed increased posterior and medial displacements compared to the native meniscus, yet no differences were found between the implant and allograft. Meniscal condition did not affect rotational laxity. Compared to the native joint, AP laxity for the implant was increased in 30° flexion, but not in 90°. The implant reduced the mean contact pressure compared to meniscectomy but could not restore contact pressures to native meniscus levels. Compared to the native meniscus, the implant significantly increased the peak pressure, while the contact area was reduced. Contact mechanics of the implant and allograft were never statistically different.

Conclusions

Biomechanical performance was similar for the implant and allograft. However, both meniscal replacements could not restore outcomes to native meniscus levels or sufficiently improve outcomes after meniscectomy. This was presumably caused by the mobility allowed by the suture-only horn fixation. The similarity of implant and allograft performance suggests that the novel implant has the biomechanical potential to serve as an alternative to meniscal allograft transplantation.

Introduction

Being involved in 15 % of the knee injuries in an active population, the menisci are amongst the most vulnerable tissues in the knee joint [19]. In the USA, this results in 650,000 meniscus-related surgeries performed annually, most of which encompass (partial) meniscectomy [3]. However, in the long term, 50 % of the meniscectomized patients develop symptomatic osteoarthritis [12]. Currently, meniscal allograft transplantation is the only treatment option for symptomatic total meniscectomy patients. Although allograft transplantation generally relieves pain and improves knee function [11, 18, 23], several studies suggest structural remodelling and shrinkage of the graft tissue, which may compromise its function in the long term [14, 16, 22]. Also, the availability of allografts is restricted by a limited supply and size-matching requirements.
A synthetic, non-resorbable total meniscus replacement could overcome these shortcomings related to the use of meniscal allografts. Since both the fit and fixation of a meniscal replacement have been shown to be important determinants of its functioning [4, 10, 17, 20], we have developed an anatomically shaped, polycarbonate urethane (PCU), total meniscus replacement with extensions for horn fixation to the tibia plateau (Fig. 1b). If this implant is capable of mimicking the native meniscus functions as a load redistributor and a knee stabilizer [5, 20, 27, 32], it may have the potential to put a hold on post-meniscectomy knee pain and functional limitations and ideally also to stop the development of osteoarthritic changes. Therefore, the objective of the current study was to evaluate the functional biomechanical performance of this novel implant and to compare its performance to the clinically relevant control cases. Specifically, whether the implant influenced meniscal kinematics, knee stability and tibial contact mechanics in comparison with the native medial meniscus, after total meniscectomy and meniscal allograft transplantation was assessed.

Materials and methods

In a knee loading rig, cadaveric knee joints were subjected to a quasi-static loading cycle. Meniscal kinematics and knee stability were evaluated using roentgen stereophotogrammetric analysis (RSA), and tibial contact mechanics were recorded using pressure sensors. The measurements were repeated for the native meniscus, the PCU implant, after total meniscectomy and for a meniscal allograft.

Implant

Implant geometry was established by segmenting the medial menisci from eight routine knee MR scans of healthy male subjects (aged 21–67) using Mimics (v14.0, Materialise, Leuven, Belgium), and 3D models were created. Meniscal length and width were determined for each 3D model and the model that approached mean dimensions closest was chosen to serve as the implant geometry. The meniscal horns were extended 5 mm to allow for suture fixation to the tibia plateau. Bionate® PCU (grade II 80A, DSM Biomedical, Geleen, Netherlands) implants were subsequently produced through injection moulding. The implants were hydrated in 37 °C saline for 2 weeks prior to testing.

Specimens

Fresh frozen, left human cadaveric knees were obtained from our institutional Department of Anatomy. Calibrated anteroposterior and lateral knee radiographs were taken to exclude specimens that showed joint space narrowing, osteophytes or chondrocalcinosis. In addition, the images were used to estimate medial meniscus length and width from tibial plateau size measurements [21]. As the prototype implant was available in one size only, knees were excluded from this study when their estimated medial meniscus length or width deviated more than ten per cent from the implant dimensions. In total, ten knees were excluded based on radiographic signs of joint degeneration and seven additional knee joints because of size mismatching. Five male knees (aged 70–88) were included for testing.

Knee loading rig and specimen preparation

To fit the specimens in the loading rig, the femur and the tibia were transected approximately 15 cm above and below the joint line, while taking care to leave the knee capsule intact. All soft tissues surrounding the knee capsule were left intact, including the full quadriceps tendon. The specimens were then transferred to the knee loading rig, which allowed the joints to move within their natural six degrees of freedom (Fig. 1a) [28]. To align the knee joint, the femur was positioned with its intercondylar notch in the intersection of the flexion/extension axis and the varus/valgus axis, and the tibial shaft was aligned with the internal/external rotation axis of the apparatus. Upon alignment, the specimens were permanently fixed to the holders of the loading rig with polymethyl methacrylate. These holders allowed repositioning of the knee joint without affecting its alignment.
Subsequently, the joint capsule was opened through a medial parapatellar incision. To access the full medial knee compartment, the femoral attachment of the medial collateral ligament (MCL) was released by isolation of a bone cap that could be repositioned with a screw (Fig. 1b). The bone cap technique was chosen to allow easy and reproducible repositioning of the MCL, without changing its kinematics [34]. Three tantalum beads (0.8 mm diameter) were inserted in the periphery of the anterior and posterior horn, and the mid-region of the medial meniscus and the implant (Fig. 1c). One bead (1 mm diameter) was inserted in the tibial origin of the anterior cruciate ligament, to define a reproducible origin of the coordinate system. Four additional tantalum markers (1 mm diameter) were inserted into both the proximal tibia and the distal femur. A metal ring was sutured to the distal end of the rectus femoris tendon, to allow for patellofemoral and tibial load transfer from the loading apparatus.

Loading protocol

The loading protocol (Table 1) was designed such that we could analyse meniscal displacement and rotational laxity during a simulated squat movement (between 0° and 90° flexion), knee AP laxity in an unloaded knee joint (in 30° and 90° flexion) and cartilage contact pressures under the maximum compressive load (in 90° flexion). Compressive loads in each flexion angle were derived by solving the equilibrium equations belonging to a free body diagram of a squat [31]. Pilot tests demonstrated that the cadaveric tissue could not withstand physiological loads. To prevent tissue damage, the compressive loads, as derived by solving the equilibrium equations, were linearly down scaled to a maximum of 1000 N. For the same reason, loading of the rectus femoris tendon was restricted to 250 N, which was kept constant during all tests. Since the quadriceps loads resulting from the equilibrium equations were larger than 190 N, the quadriceps loads applied in this study should mainly be considered as joint stabilizers.
Table 1
Loading protocol describing all loads applied during the testing of one meniscal condition
Flexion angle (°)
Axial compressive load (N)
Quadriceps load (N)
Tibial torque (Nm)
Tibial drawer load (N)
0
357
250
357
250
3.4 internal
357
250
3.4 external
30
572
250
572
250
3.4 internal
572
250
3.4 external
67
60
824
250
824
250
3.4 internal
824
250
3.4 external
90
1000
250
1000
250
3.4 internal
1000
250
3.4 external
67
0
357
250
357
250
3.4 internal
357
250
3.4 external
90
1000
250
In each flexion angle, the knee was tested without tibial torque (neutral), and with an internal and external tibial torque of 3.4 Nm [15]. In addition, in 30° and 90° flexion, AP stability of the knee joint was tested through anterior drawer tests. Anterior loads of 67 N were applied to the proximal tibia, to mimic the load that is applied in clinical knee laxity testers [7]. To assess whether repetitive loading of the cadaveric tissue influenced the outcomes of our tests due to tissue degeneration, a second set of measurements in 0° flexion was performed (Table 1).

Meniscal conditions

Testing of the native meniscus (condition 1, Fig. 2a) was followed by the implant (condition 2, Fig. 2b), a total meniscectomy (condition 3, Fig. 2c) and a meniscal allograft (condition 4, Fig. 2d). To start condition 2, a complete medial meniscectomy was performed. The excised meniscus was wrapped in saline-soaked gauze and frozen at −80 °C until its use as allograft meniscus. Tibial bone tunnels (2.5 mm diameter) were drilled from the anteromedial side of the tibia to the anterior and posterior attachment site of the meniscus on the tibial plateau. Two suture threads (FiberWire #2, Arthrex, Naples, FL, USA) were fixed to each horn of the implant and guided through the tunnels. The implant was positioned and fixed by knotting the sutures together. For testing of condition 3, the implant was removed, and the knee was retested for the meniscectomy condition. Lastly, for condition 4, the tests were repeated for an allograft meniscus, which was obtained from the preceding experiment. As all knees were size matched to the implant, their menisci were regarded to be properly sized allografts as well. The horns of the allograft were fixed in a similar fashion as those of the implant. To comply with the surgical standards that are clinically applied, in addition, the allograft was circumferentially fixed to the joint capsule using four vertical mattress sutures.

Biomechanical outcome measures

RSA was used to determine meniscal kinematics, knee AP laxity and tibial rotational laxity [26]. For each separate loading step from Table 1, a bi-planar roentgen image was taken. The 3D locations of the tantalum beads inside the tibia, native meniscus, implant and allograft were obtained from each image by relating their 3D positions to the distances known from a calibration image. All tibial beads were superimposed, allowing the determination of the different locations of the meniscus on the tibia for each condition. The meniscal locations were converted to meniscal translations along the AP and ML axes for the flexion intervals 30°–0°, 60°–0° and 90°–0° for the anterior, mid and posterior regions of the meniscus or implant. To determine AP laxity (in 30° and 90° flexion), the centroid of the tantalum beads inside the tibia was determined in unloaded and loaded state. AP laxity was defined as the translation of the centroid along the AP axis. Rotational laxity was defined as the angular rotation of the tibia around its longitudinal axis, as a result of the application of the internal or external torque. The set-up used in our laboratory allowed to determine bead location with an accuracy of 50 μm and repeated measurements can be performed with an accuracy of 200 µm [6].
In 90° flexion, tibial contact mechanics were evaluated using piezo-electric pressure sensors (Fig. 1d, e, K-scan 4010 N, Tekscan, South Boston, MA, USA). To prevent fluid penetration into the sensor, the sensor was placed between two layers of Tegaderm film dressing (3M, St. Paul, MN, USA). After sealing, all sensors were conditioned, equilibrated and calibrated using a materials testing system (MMED, Materials Technology Corporation, La Canada, CA, USA) and a custom calibration tool consisting of two Teflon plates. To optimize the accuracy of the pressure sensor, the calibration curve was derived from a ten point polynomial fit through the calibration points that were equally distributed between 0.9 and 9.0 MPa [8]. Consequently, cartilage contact pressure could be determined with a precision of 0.035 MPa. To allow sensor insertion, the knee was distracted by the weight of the tibia and a 2-cm incision was made to the anterior meniscotibial ligament (Fig. 1b). A suture with a 10″ straight needle (PDS II 2-0, Ethicon, Somerville, NJ, USA) was knotted to the most posteromedial corner of the sensor and guided along the cruciate ligaments. This suture allowed to pull the sensor exactly underneath the meniscus and was re-used to place the sensors for the implant, meniscectomy and allograft conditions in the same position. Additionally, the anteromedial edge of the pressure sensor was fixed by a suture through the knee capsule. A new sensor tab was used for each meniscal condition. Peak pressure, mean pressure and contact area were determined from the pressure maps for each meniscal condition.

Statistical analysis

A sample size calculation was based on a previous study that evaluated the difference in knee contact mechanics after total medial meniscectomy and subsequent meniscal transplantation [25]. Five knees were required to detect a difference of 1.1 MPa (standard deviation 0.5 MPa) in mean pressure, using a significance level of 0.05 and a power of 80 %.
Linear mixed models were used to study the effect of meniscal condition on contact mechanics, knee laxity and meniscal kinematics. Each outcome measure (peak pressure, mean pressure, contact area, AP laxity in 30° and 90° flexion, internal and external tibial rotation, AP and ML meniscal translation) was analysed separately. The models included specimen as a random factor and all other variables that applied to the outcome measure under analysis (meniscal condition, flexion angle or flexion interval, meniscal region) as fixed factors. A random intercept and slope were included to account for the specimen’s individual response to each experimental condition. For rotational laxity and meniscal displacement, models including the interaction terms between the fixed factors were also evaluated. As the addition of the interaction terms did not significantly increase the model fits based on likelihood ratio tests, the interaction terms were omitted from the models used for the final analysis. Pairwise comparisons between the different levels of the fixed variables were performed by Tukey’s tests that were Bonferroni corrected to account for multiple comparisons. p values below 0.05 were considered statistically significant.

Results

The dimensions of the implant corresponded well to those of the native meniscus. Deviations (mean ± standard deviation) were 4.0 ± 4.1 % for the length and 5.1 ± 4.5 % for the width. Mean size deviations between the allograft and native meniscus were similar; 5.3 ± 4.1 % in length and 4.5 ± 4.6 % in width.

Meniscal kinematics

For all three meniscal conditions, the translation patterns along the AP axis were similar; translation occurred in posterior direction and increased with flexion angle, and the anterior horn was more mobile than the mid-region (p = 0.002) and posterior horn (p < 0.001) (Fig. 3a). A significant increase in the posterior translation was observed when replacing the native meniscus with the implant (p < 0.001) or an allograft (p < 0.001). When comparing the outcomes for each location–flexion interval combination, the differences in posterior translation between the implant and allograft were never statistically significant, whereas those between the meniscal replacements and the native meniscus were always significant (Fig. 3a).
Translations along the ML axis were less regular, particularly for the anterior horn. With increasing flexion angle, the implant and allograft anterior horn both moved medially, whereas the native meniscus moved laterally (Fig. 3b). Also along the ML axis, meniscal mobility increased when the native meniscus was replaced by the implant (p < 0.001) or an allograft (p = 0.001). ML displacement of the implant and allograft was not significantly different. When comparing the outcomes for each location–flexion interval combination, the differences in ML translation between the implant and allograft were always non-significant, whereas the differences between the implant and native meniscus were always significant (Fig. 3b).

Knee stability

The effect of changing the meniscal condition on AP laxity was larger in 30° flexion than in 90° flexion (Fig. 4). In 30° flexion, AP laxity was significantly increased for the implant, meniscectomy and allograft conditions (all p < 0.001). However, replacing the native meniscus by the implant did not significantly increase AP laxity in 90° flexion, whereas the increases in AP laxity found after meniscectomy and allograft transplantation were significant (p = 0.012 and p = 0.016). The changes in internal and external rotational laxity observed between the four meniscal conditions at each separate flexion angle were minor. The differences in rotational laxity between the first and second evaluation in 0° flexion for each meniscal condition were small and not significant (Fig. 5).

Contact mechanics

A representative example of the pressure distribution on the tibia articular cartilage for the four meniscal conditions is shown in Fig. 6. The mean pressure for the native meniscus condition was significantly increased for the implanted knee (p < 0.001), after meniscectomy (p < 0.001) and after allograft transplantation (p = 0.019, Fig. 7). Both the implant condition and allograft condition did significantly reduce the mean contact pressure compared to the meniscectomy condition (p = 0.018 and p < 0.001, Fig. 7). Also the peak pressure was significantly elevated for the implant, meniscectomy and allograft conditions with respect to the native meniscus (all p < 0.001). The peak pressures for the implant and allograft were highly similar. The contact area on the tibia articular cartilage was largest for the native meniscus and was significantly reduced for the implant, allograft and meniscectomy conditions (all p < 0.001). Whereas the implant only slightly increased the contact area with respect to the meniscectomized condition, meniscal allograft transplantation resulted in a significant increase (p < 0.001).

Discussion

The most important finding of the present study was that our anatomically shaped PCU total meniscus implant showed a similar biomechanical performance as a meniscal allograft, but was unable to restore the functionality of the native meniscus. The implant was more mobile than the native meniscus and not capable of restoring native contact mechanics. The implant significantly reduced the mean contact pressure relative to the meniscectomized knee, but was not capable of improving peak pressure, contact area or AP stability. On the other hand, the differences between the implant and allograft were small and non-significant, indicating that the implant could match the biomechanical performance of a meniscal allograft. Previously, it has been shown that meniscal allografts cannot fully restore the biomechanical functionality of the native meniscus [4, 20, 33]. Nevertheless, allografts have shown to be an effective treatment to relieve pain and resume activity for symptomatic meniscectomy patients [11, 18, 23]. This illustrates that it may not be necessary that a meniscal replacement fully restores the biomechanical outcomes to native meniscus levels.
The translations of the implant and allograft were always larger than those of the native meniscus, which may be due to the suture-only fixation technique. The sutures were knotted together to fix the meniscal replacements; however, the threads had the freedom to move within the tibial bone tunnels when the knee was loaded. Whereas the allograft had an additional peripheral fixation to the joint capsule, we did not observe significant differences in mobility between the allograft and implant. This corresponds to a previous observation that the integration between the medial meniscus and the MCL-joint capsule complex does not influence meniscal mobility [31].
The results on AP knee laxity of this study are in agreement with previous findings indicating that complete removal of the medial meniscus significantly increases anterior knee laxity, which suggests that the medial meniscus plays a primary role in AP knee stability [5, 27]. Although the maximum increase in anterior knee laxity was 89 % (meniscectomy versus native meniscus in 30° flexion), the corresponding absolute increase in anterior displacement was only 1.8 mm. In the scope of the classification by Daniel et al. [9] who considered a left to right AP laxity difference of smaller than 2 mm as normal, our changes in knee laxity are marginal.
Whereas the implant and allograft significantly decreased the mean pressure with respect to the meniscectomy condition, the peak pressure and contact area were not improved. These observations correspond with previously reported findings [4, 33]. The mean differences between the implant and native meniscus are close to the values reported by Alhalki et al. [4]. Recently, Wang et al. [33] reported inferior contact mechanics for allografts fixed with the suture-only technique compared to grafts fixed with bone plugs, which was associated with increased extrusion in the former condition. This suggests that our choice for the suture-only fixation technique may have been responsible for the deficient contact mechanics of the meniscus replacements. Clinical evaluations of allografts with suture-only and bone plug fixation have shown similar Lysholm, Tegner and VAS scores [1, 2]. However, contrary to meniscal allografts that over time will be reinforced by tissue ingrowth [18, 30], our implant will permanently depend on its initial fixation. This imposes extra demands on the fixation of the implant and suggests that future implant designs could benefit from a more rigid fixation strategy.
Some limitations should be addressed when interpreting the results of this study. Firstly, all tests were performed at room temperature, while PCU is known to be softer at body temperature and in a fully humidified environment [13]. To minimize the influence of temperature differences, the implant was kept in a 37 °C saline solution up to the moment of implantation. Secondly, the repeated measures design chosen for this study may have biased our results. However, similar values were recorded for tibial rotation at the beginning and end of a loading cycle. Therefore, we are confident that the outcomes of this study were not influenced by the repeated measures design. Lastly, the compressive loads used in this study were lower than physiological loads. Based on previous evaluations of meniscal motion and tibial contact mechanics at multiple load levels, it is expected that the absolute translations and pressures would be increased when higher loads would have been applied [24, 29]. However, it is not anticipated that an increased load would affect the relative differences between the four conditions in this study.

Conclusions

The biomechanical performance of the novel PCU meniscal implant matched that of an allograft meniscus. Nevertheless, neither of the meniscal replacements could reproduce native meniscus performance or improve all aspects of contact mechanics compared to total meniscectomy. Based on the similarity between implant and allograft performance in this study, this implant may become an alternative to meniscal allograft transplantation.

Acknowledgments

This research forms part of the Project P2.03 TRAMMPOLIN of the research program of the BioMedical Materials institute, co-funded by the Dutch Ministry of Economic Affairs.

Conflict of interest

The authors declare that they have no conflict of interest.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Abat F, Gelber PE, Erquicia JI, Pelfort X, Gonzalez-Lucena G, Monllau JC (2012) Suture-only fixation technique leads to a higher degree of extrusion than bony fixation in meniscal allograft transplantation. Am J Sports Med 40(7):1591–1596CrossRefPubMed Abat F, Gelber PE, Erquicia JI, Pelfort X, Gonzalez-Lucena G, Monllau JC (2012) Suture-only fixation technique leads to a higher degree of extrusion than bony fixation in meniscal allograft transplantation. Am J Sports Med 40(7):1591–1596CrossRefPubMed
2.
Zurück zum Zitat Abat F, Gelber PE, Erquicia JI, Tey M, Gonzalez-Lucena G, Monllau JC (2013) Prospective comparative study between two different fixation techniques in meniscal allograft transplantation. Knee Surg Sports Traumatol Arthrosc 21(7):1516–1522CrossRefPubMed Abat F, Gelber PE, Erquicia JI, Tey M, Gonzalez-Lucena G, Monllau JC (2013) Prospective comparative study between two different fixation techniques in meniscal allograft transplantation. Knee Surg Sports Traumatol Arthrosc 21(7):1516–1522CrossRefPubMed
3.
Zurück zum Zitat Abrams GD, Frank RM, Gupta AK, Harris JD, McCormick FM, Cole BJ (2013) Trends in meniscus repair and meniscectomy in the United States, 2005–2011. Am J Sports Med 41(10):2333–2339CrossRefPubMed Abrams GD, Frank RM, Gupta AK, Harris JD, McCormick FM, Cole BJ (2013) Trends in meniscus repair and meniscectomy in the United States, 2005–2011. Am J Sports Med 41(10):2333–2339CrossRefPubMed
4.
Zurück zum Zitat Alhalki MM, Howell SM, Hull ML (1999) How three methods for fixing a medial meniscal autograft affect tibial contact mechanics. Am J Sports Med 27(3):320–328PubMed Alhalki MM, Howell SM, Hull ML (1999) How three methods for fixing a medial meniscal autograft affect tibial contact mechanics. Am J Sports Med 27(3):320–328PubMed
5.
Zurück zum Zitat Arno S, Hadley S, Campbell KA, Bell CP, Hall M, Beltran LS, Recht MP, Sherman OH, Walker PS (2013) The effect of arthroscopic partial medial meniscectomy on tibiofemoral stability. Am J Sports Med 41(1):73–79CrossRefPubMed Arno S, Hadley S, Campbell KA, Bell CP, Hall M, Beltran LS, Recht MP, Sherman OH, Walker PS (2013) The effect of arthroscopic partial medial meniscectomy on tibiofemoral stability. Am J Sports Med 41(1):73–79CrossRefPubMed
6.
Zurück zum Zitat Blankevoort L, Huiskes R, de Lange A (1988) The envelope of passive knee joint motion. J Biomech 21(9):705–720CrossRefPubMed Blankevoort L, Huiskes R, de Lange A (1988) The envelope of passive knee joint motion. J Biomech 21(9):705–720CrossRefPubMed
7.
Zurück zum Zitat Boyer P, Djian P, Christel P, Paoletti X, Degeorges R (2004) Reliability of the KT-1000 arthrometer (Medmetric) for measuring anterior knee laxity: comparison with Telos in 147 knees. Rev Chir Orthop Reparatrice Appar Mot 90(8):757–764CrossRefPubMed Boyer P, Djian P, Christel P, Paoletti X, Degeorges R (2004) Reliability of the KT-1000 arthrometer (Medmetric) for measuring anterior knee laxity: comparison with Telos in 147 knees. Rev Chir Orthop Reparatrice Appar Mot 90(8):757–764CrossRefPubMed
8.
Zurück zum Zitat Brimacombe JM, Wilson DR, Hodgson AJ, Ho KC, Anglin C (2009) Effect of calibration method on Tekscan sensor accuracy. J Biomech Eng 131(3):034503CrossRefPubMed Brimacombe JM, Wilson DR, Hodgson AJ, Ho KC, Anglin C (2009) Effect of calibration method on Tekscan sensor accuracy. J Biomech Eng 131(3):034503CrossRefPubMed
9.
Zurück zum Zitat Daniel DM, Stone ML, Sachs R, Malcom L (1985) Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 13(6):401–407CrossRefPubMed Daniel DM, Stone ML, Sachs R, Malcom L (1985) Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 13(6):401–407CrossRefPubMed
10.
Zurück zum Zitat Dienst M, Greis PE, Ellis BJ, Bachus KN, Burks RT (2007) Effect of lateral meniscal allograft sizing on contact mechanics of the lateral tibial plateau: an experimental study in human cadaveric knee joints. Am J Sports Med 35(1):34–42CrossRefPubMed Dienst M, Greis PE, Ellis BJ, Bachus KN, Burks RT (2007) Effect of lateral meniscal allograft sizing on contact mechanics of the lateral tibial plateau: an experimental study in human cadaveric knee joints. Am J Sports Med 35(1):34–42CrossRefPubMed
11.
Zurück zum Zitat Elattar M, Dhollander A, Verdonk R, Almqvist KF, Verdonk P (2010) Twenty-six years of meniscal allograft transplantation: is it still experimental? A meta-analysis of 44 trials. Knee Surg Sports Traumatol Arthrosc 19:147–157CrossRefPubMed Elattar M, Dhollander A, Verdonk R, Almqvist KF, Verdonk P (2010) Twenty-six years of meniscal allograft transplantation: is it still experimental? A meta-analysis of 44 trials. Knee Surg Sports Traumatol Arthrosc 19:147–157CrossRefPubMed
12.
Zurück zum Zitat Englund M, Roos EM, Lohmander LS (2003) Impact of type of meniscal tear on radiographic and symptomatic knee osteoarthritis: a sixteen-year followup of meniscectomy with matched controls. Arthritis Rheum 48(8):2178–2187CrossRefPubMed Englund M, Roos EM, Lohmander LS (2003) Impact of type of meniscal tear on radiographic and symptomatic knee osteoarthritis: a sixteen-year followup of meniscectomy with matched controls. Arthritis Rheum 48(8):2178–2187CrossRefPubMed
13.
Zurück zum Zitat Flanagan S (2010) Friction, lubrication and wear of total joint replacements. University of Limerick, Limerick Flanagan S (2010) Friction, lubrication and wear of total joint replacements. University of Limerick, Limerick
14.
Zurück zum Zitat Gelber PE, Gonzalez G, Lloreta JL, Reina F, Caceres E, Monllau JC (2008) Freezing causes changes in the meniscus collagen net: a new ultrastructural meniscus disarray scale. Knee Surg Sports Traumatol Arthrosc 16(4):353–359CrossRefPubMed Gelber PE, Gonzalez G, Lloreta JL, Reina F, Caceres E, Monllau JC (2008) Freezing causes changes in the meniscus collagen net: a new ultrastructural meniscus disarray scale. Knee Surg Sports Traumatol Arthrosc 16(4):353–359CrossRefPubMed
15.
Zurück zum Zitat Kutzner I, Heinlein B, Graichen F, Bender A, Rohlmann A, Halder A, Beier A, Bergmann G (2010) Loading of the knee joint during activities of daily living measured in vivo in five subjects. J Biomech 43(11):2164–2173CrossRefPubMed Kutzner I, Heinlein B, Graichen F, Bender A, Rohlmann A, Halder A, Beier A, Bergmann G (2010) Loading of the knee joint during activities of daily living measured in vivo in five subjects. J Biomech 43(11):2164–2173CrossRefPubMed
16.
Zurück zum Zitat Lee BS, Chung JW, Kim JM, Cho WJ, Kim KA, Bin SI (2012) Morphologic changes in fresh-frozen meniscus allografts over 1 year: a prospective magnetic resonance imaging study on the width and thickness of transplants. Am J Sports Med 40(6):1384–1391CrossRefPubMed Lee BS, Chung JW, Kim JM, Cho WJ, Kim KA, Bin SI (2012) Morphologic changes in fresh-frozen meniscus allografts over 1 year: a prospective magnetic resonance imaging study on the width and thickness of transplants. Am J Sports Med 40(6):1384–1391CrossRefPubMed
17.
Zurück zum Zitat Linder-Ganz E, Elsner JJ, Danino A, Guilak F, Shterling A (2010) A novel quantitative approach for evaluating contact mechanics of meniscal replacements. J Biomech Eng 132(2):024501CrossRefPubMed Linder-Ganz E, Elsner JJ, Danino A, Guilak F, Shterling A (2010) A novel quantitative approach for evaluating contact mechanics of meniscal replacements. J Biomech Eng 132(2):024501CrossRefPubMed
18.
Zurück zum Zitat Lubowitz JH, Verdonk PC, Reid JB 3rd, Verdonk R (2007) Meniscus allograft transplantation: a current concepts review. Knee Surg Sports Traumatol Arthrosc 15(5):476–492CrossRefPubMed Lubowitz JH, Verdonk PC, Reid JB 3rd, Verdonk R (2007) Meniscus allograft transplantation: a current concepts review. Knee Surg Sports Traumatol Arthrosc 15(5):476–492CrossRefPubMed
19.
Zurück zum Zitat Majewski M, Susanne H, Klaus S (2006) Epidemiology of athletic knee injuries: a 10-year study. Knee 13(3):184–188CrossRefPubMed Majewski M, Susanne H, Klaus S (2006) Epidemiology of athletic knee injuries: a 10-year study. Knee 13(3):184–188CrossRefPubMed
20.
Zurück zum Zitat Paletta GA, Manning T, Snell E, Parker R, Bergfeld J (1997) The effect of allograft meniscal replacement on intraarticular contact area and pressures in the human knee. Am J Sports Med 25(5):692–698CrossRefPubMed Paletta GA, Manning T, Snell E, Parker R, Bergfeld J (1997) The effect of allograft meniscal replacement on intraarticular contact area and pressures in the human knee. Am J Sports Med 25(5):692–698CrossRefPubMed
21.
Zurück zum Zitat Pollard ME, Kang Q, Berg EE (1995) Radiographic sizing for meniscal transplantation. Arthroscopy 11(6):684–687CrossRefPubMed Pollard ME, Kang Q, Berg EE (1995) Radiographic sizing for meniscal transplantation. Arthroscopy 11(6):684–687CrossRefPubMed
22.
Zurück zum Zitat Rodeo SA, Seneviratne A, Suzuki K, Felker K, Wickiewicz TL, Warren RF (2000) Histological analysis of human meniscal allografts. A preliminary report. J Bone Joint Surg Am 82-A(8):1071–1082PubMed Rodeo SA, Seneviratne A, Suzuki K, Felker K, Wickiewicz TL, Warren RF (2000) Histological analysis of human meniscal allografts. A preliminary report. J Bone Joint Surg Am 82-A(8):1071–1082PubMed
23.
Zurück zum Zitat Rosso F, Bisicchia S, Bonasia DE, Amendola A (2015) Meniscal allograft transplantation: a systematic review. Am J Sports Med 43(4):998–1007CrossRefPubMed Rosso F, Bisicchia S, Bonasia DE, Amendola A (2015) Meniscal allograft transplantation: a systematic review. Am J Sports Med 43(4):998–1007CrossRefPubMed
24.
Zurück zum Zitat Seitz AM, Lubomierski A, Friemert B, Ignatius A, Durselen L (2012) Effect of partial meniscectomy at the medial posterior horn on tibiofemoral contact mechanics and meniscal hoop strains in human knees. J Orthop Res 30(6):934–942CrossRefPubMed Seitz AM, Lubomierski A, Friemert B, Ignatius A, Durselen L (2012) Effect of partial meniscectomy at the medial posterior horn on tibiofemoral contact mechanics and meniscal hoop strains in human knees. J Orthop Res 30(6):934–942CrossRefPubMed
25.
Zurück zum Zitat Sekaran SV, Hull ML, Howell SM (2002) Nonanatomic location of the posterior horn of a medial meniscal autograft implanted in a cadaveric knee adversely affects the pressure distribution on the tibial plateau. Am J Sports Med 30(1):74–82PubMed Sekaran SV, Hull ML, Howell SM (2002) Nonanatomic location of the posterior horn of a medial meniscal autograft implanted in a cadaveric knee adversely affects the pressure distribution on the tibial plateau. Am J Sports Med 30(1):74–82PubMed
26.
Zurück zum Zitat Selvik G (1989) Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl 232:1–51CrossRefPubMed Selvik G (1989) Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl 232:1–51CrossRefPubMed
27.
Zurück zum Zitat Spang JT, Dang AB, Mazzocca A, Rincon L, Obopilwe E, Beynnon B, Arciero RA (2010) The effect of medial meniscectomy and meniscal allograft transplantation on knee and anterior cruciate ligament biomechanics. Arthroscopy 26(2):192–201CrossRefPubMed Spang JT, Dang AB, Mazzocca A, Rincon L, Obopilwe E, Beynnon B, Arciero RA (2010) The effect of medial meniscectomy and meniscal allograft transplantation on knee and anterior cruciate ligament biomechanics. Arthroscopy 26(2):192–201CrossRefPubMed
28.
Zurück zum Zitat van Tienen T, Verdonschot N, Heijkants R, Buma P, Scholten J, Van Kampen A, Veth R (2004) Prosthetic replacement of the medial meniscus in cadaveric knees. Am J Sports Med 32(5):1182–1188CrossRefPubMed van Tienen T, Verdonschot N, Heijkants R, Buma P, Scholten J, Van Kampen A, Veth R (2004) Prosthetic replacement of the medial meniscus in cadaveric knees. Am J Sports Med 32(5):1182–1188CrossRefPubMed
29.
Zurück zum Zitat Vedi V, Spouse E, Williams A, Tennant SJ, Hunt DM, Gedroyc WMW (1999) Meniscal movement: an in vivo study using dynamic MRI. J Bone Joint Surg Br 81(1):37–41CrossRefPubMed Vedi V, Spouse E, Williams A, Tennant SJ, Hunt DM, Gedroyc WMW (1999) Meniscal movement: an in vivo study using dynamic MRI. J Bone Joint Surg Br 81(1):37–41CrossRefPubMed
30.
Zurück zum Zitat von Lewinski G, Hurschler C, Allmann C, Wirth CJ (2006) The influence of pre-tensioning of meniscal transplants on the tibiofemoral contact area. Knee Surg Sports Traumatol Arthrosc 14(5):425–436CrossRef von Lewinski G, Hurschler C, Allmann C, Wirth CJ (2006) The influence of pre-tensioning of meniscal transplants on the tibiofemoral contact area. Knee Surg Sports Traumatol Arthrosc 14(5):425–436CrossRef
31.
Zurück zum Zitat Vrancken ACT, van Tienen TG, Hannink G, Janssen D, Verdonschot N, Buma P (2014) Releasing the circumferential fixation of the medial meniscus does not affect its kinematics. Knee 21(6):1033–1038CrossRefPubMed Vrancken ACT, van Tienen TG, Hannink G, Janssen D, Verdonschot N, Buma P (2014) Releasing the circumferential fixation of the medial meniscus does not affect its kinematics. Knee 21(6):1033–1038CrossRefPubMed
32.
Zurück zum Zitat Walker PS, Erkiuan MJ (1975) The role of the menisci in force transmission across the knee. Clin Orthop Relat Res 109:184–192CrossRefPubMed Walker PS, Erkiuan MJ (1975) The role of the menisci in force transmission across the knee. Clin Orthop Relat Res 109:184–192CrossRefPubMed
33.
Zurück zum Zitat Wang H, Gee AO, Hutchinson ID, Stoner K, Warren RF, Chen TO, Maher SA (2014) Bone plug versus suture-only fixation of meniscal grafts: effect on joint contact mechanics during simulated gait. Am J Sports Med 42(7):1682–1689CrossRefPubMedPubMedCentral Wang H, Gee AO, Hutchinson ID, Stoner K, Warren RF, Chen TO, Maher SA (2014) Bone plug versus suture-only fixation of meniscal grafts: effect on joint contact mechanics during simulated gait. Am J Sports Med 42(7):1682–1689CrossRefPubMedPubMedCentral
34.
Zurück zum Zitat Watanabe Y, Scyoc AV, Tsuda E, Debski RE, Woo SL (2004) Biomechanical function of the posterior horn of the medial meniscus: a human cadaveric study. J Orthop Sci 9(3):280–284CrossRefPubMed Watanabe Y, Scyoc AV, Tsuda E, Debski RE, Woo SL (2004) Biomechanical function of the posterior horn of the medial meniscus: a human cadaveric study. J Orthop Sci 9(3):280–284CrossRefPubMed
Metadaten
Titel
Functional biomechanical performance of a novel anatomically shaped polycarbonate urethane total meniscus replacement
verfasst von
A. C. T. Vrancken
F. Eggermont
T. G. van Tienen
G. Hannink
P. Buma
D. Janssen
N. Verdonschot
Publikationsdatum
14.05.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 5/2016
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-015-3632-6

Weitere Artikel der Ausgabe 5/2016

Knee Surgery, Sports Traumatology, Arthroscopy 5/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.