Tube current reduction in pediatric non-ECG-gated heart CT by combined tube current modulation

verfasst von: Hyum Woo Goo, Dong Soo Suh

Erschienen in: Pediatric Radiology | Ausgabe 4/2006

Einloggen, um Zugang zu erhalten

Abstract

Background

Dose reduction by combined tube current modulation has not been evaluated in pediatric low-dose CT.

Objective

This study was performed to evaluate tube current reduction in non-ECG-gated heart CT angiography in children by combined tube current modulation, and to assess the effects of body weight, age, sex, and scan direction on tube current reduction.

Materials and methods

From September 2004 to January 2005, non-ECG-gated heart CT examinations were performed with combined tube current modulation in 50 children (median age 1 year, range 1 month to 16 years; M/F 29/21) with congenital heart disease. Sixteen-slice spiral CT studies were obtained using a weight-based low-dose protocol (80 kVp, 50–160 effective mA). CT scans were obtained in the craniocaudal direction when leg veins (n=36) were used for IV injection of contrast agent and in the caudocranial direction when arm veins (n=13) were used. In one child who underwent a Fontan operation, both arm and leg veins were used. We calculated tube current reduction by combined tube current modulation and evaluated the effects of body weight, age, sex, and scan technique on tube current reduction. The quality of CT angiography images was visually evaluated by an experienced pediatric radiologist.

Results

Overall tube current reduction by combined tube current modulation was 15.8±11.1%. The reduction was variable among five body weight groups (9.3±7.9% for <4.9 kg, 14.3±9.3% for 5.0–9.9 kg, 16.4±12.5% for 10.0–19.9 kg, 25.8±9.8% for 20.0–39.9 kg, 15.9±12.7% for 40.0–59.9 kg) and was significantly different among the five groups (P=0.017). When eight children in the 40.0–59.9-kg group were excluded, age showed a significant positive correlation with tube current reduction (γ=0.4, P=0.003). There was no significant difference in tube current reduction between boys and girls. Tube current reduction was significantly greater for the caudocranial scan (21.2±9.2%) than for the craniocaudal scan (14.2±11.3%) (P=0.049). All CT angiography images were of diagnostic quality.

Conclusions

Combined tube current modulation reduced tube current in pediatric non-ECG-gated heart CT by 15.8% without degradation of image quality. With our CT protocol, for body weights up to 39.9 kg, age showed a significant positive correlation with tube current reduction, but there was a lower dose-saving effect in children heavier than 40 kg. Tube current reduction was not affected by sex, but was affected by scan direction.
Literatur
1.
Zurück zum Zitat Linton OW, Mettler FA Jr (2003) National conference on dose reduction in CT with an emphasis on pediatric patients. AJR 181:321–329PubMed Linton OW, Mettler FA Jr (2003) National conference on dose reduction in CT with an emphasis on pediatric patients. AJR 181:321–329PubMed
2.
Zurück zum Zitat Brenner DJ, Elliston C, Hall E, et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed Brenner DJ, Elliston C, Hall E, et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed
3.
Zurück zum Zitat Goo HW (2005) Pediatric CT: understanding of radiation dose and optimization of imaging techniques. J Korean Radiol Soc 52:1–5 Goo HW (2005) Pediatric CT: understanding of radiation dose and optimization of imaging techniques. J Korean Radiol Soc 52:1–5
4.
Zurück zum Zitat Kalra MK, Maher MM, Toth TL, et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628PubMedCrossRef Kalra MK, Maher MM, Toth TL, et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628PubMedCrossRef
5.
Zurück zum Zitat Boone JM, Geraghty EM, Seibert JA, et al (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228:352–360PubMedCrossRef Boone JM, Geraghty EM, Seibert JA, et al (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228:352–360PubMedCrossRef
6.
Zurück zum Zitat Frush DP, Soden B, Frush KS, et al (2002) Improved pediatric multi-detector body CT using a size-based color-coded format. AJR 178:721–726PubMed Frush DP, Soden B, Frush KS, et al (2002) Improved pediatric multi-detector body CT using a size-based color-coded format. AJR 178:721–726PubMed
7.
Zurück zum Zitat Starck G, Lonn L, Cederblad A, et al (2002) A method to obtain the same level of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol 75:140–150PubMed Starck G, Lonn L, Cederblad A, et al (2002) A method to obtain the same level of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol 75:140–150PubMed
8.
Zurück zum Zitat Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174PubMedCrossRef Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231:169–174PubMedCrossRef
9.
Zurück zum Zitat Kalra MK, Maher MM, Toth TL, et al (2004) Techniques and applications of automatic tube current modulation for CT. Radiology 233:649–657PubMedCrossRef Kalra MK, Maher MM, Toth TL, et al (2004) Techniques and applications of automatic tube current modulation for CT. Radiology 233:649–657PubMedCrossRef
10.
Zurück zum Zitat Flohr TG, Schaller S, Stierstorfer K, et al (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773PubMedCrossRef Flohr TG, Schaller S, Stierstorfer K, et al (2005) Multi-detector row CT systems and image-reconstruction techniques. Radiology 235:756–773PubMedCrossRef
11.
Zurück zum Zitat Goo HW, Park IS, Ko JK, et al (2003) CT of congenital heart disease: normal anatomy and typical pathologic conditions. Radiographics 23:S147–S165PubMedCrossRef Goo HW, Park IS, Ko JK, et al (2003) CT of congenital heart disease: normal anatomy and typical pathologic conditions. Radiographics 23:S147–S165PubMedCrossRef
12.
Zurück zum Zitat Goo HW (2004) Evaluation of the airways in patients with congenital heart disease using multislice CT. J Korean Pediatr Cardiol Soc 8:37–43 Goo HW (2004) Evaluation of the airways in patients with congenital heart disease using multislice CT. J Korean Pediatr Cardiol Soc 8:37–43
13.
Zurück zum Zitat Goo HW, Park IS, Ko JK, et al (2005) Computed tomography for the diagnosis of congenital heart disease in pediatric and adult patients. Int J Cardiovasc Imaging 21:347–365PubMedCrossRef Goo HW, Park IS, Ko JK, et al (2005) Computed tomography for the diagnosis of congenital heart disease in pediatric and adult patients. Int J Cardiovasc Imaging 21:347–365PubMedCrossRef
14.
Zurück zum Zitat Goo HW, Park IS, Ko JK, et al (2005) Visibility of the origin and proximal course of coronary arteries on non-ECG-gated heart CT in patients with congenital heart disease. Pediatr Radiol 35:792–798PubMedCrossRef Goo HW, Park IS, Ko JK, et al (2005) Visibility of the origin and proximal course of coronary arteries on non-ECG-gated heart CT in patients with congenital heart disease. Pediatr Radiol 35:792–798PubMedCrossRef
15.
Zurück zum Zitat Mulkens TH, Bellinck P, Baeyaert M, et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector CT examinations: clinical evaluation. Radiology 237:213–223PubMedCrossRef Mulkens TH, Bellinck P, Baeyaert M, et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector CT examinations: clinical evaluation. Radiology 237:213–223PubMedCrossRef
16.
Zurück zum Zitat Greess H, Lutze J, Nomayr A, et al (2004) Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol 14:995–999PubMedCrossRef Greess H, Lutze J, Nomayr A, et al (2004) Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol 14:995–999PubMedCrossRef
17.
Zurück zum Zitat Greess H, Wolf H, Baum U, et al (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10:391–394PubMedCrossRef Greess H, Wolf H, Baum U, et al (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10:391–394PubMedCrossRef
18.
Zurück zum Zitat Das M, Mahnken AH, Mühlenbruch G, et al (2005) Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations. AJR 184:1437–1443PubMed Das M, Mahnken AH, Mühlenbruch G, et al (2005) Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations. AJR 184:1437–1443PubMed
19.
Zurück zum Zitat Kalra MK, Rizzo S, Maher MM, et al (2005) Chest CT performed with z-axis modulation: scanning protocol and radiation dose. Radiology 237:303–308PubMedCrossRef Kalra MK, Rizzo S, Maher MM, et al (2005) Chest CT performed with z-axis modulation: scanning protocol and radiation dose. Radiology 237:303–308PubMedCrossRef
20.
Zurück zum Zitat Tack D, De Maertelaer V, Gevenois PA (2003) Dose reduction in multidetector CT using attenuation-based online tube current modulation. AJR 181:331–334PubMed Tack D, De Maertelaer V, Gevenois PA (2003) Dose reduction in multidetector CT using attenuation-based online tube current modulation. AJR 181:331–334PubMed
21.
Zurück zum Zitat Mastora I, Remy-Jardin M, Delannoy V, et al (2004) Multi-detector row spiral CT angiography of the thoracic outlet: dose reduction with anatomically adapted online tube current modulation and preset dose savings. Radiology 230:116–124PubMedCrossRef Mastora I, Remy-Jardin M, Delannoy V, et al (2004) Multi-detector row spiral CT angiography of the thoracic outlet: dose reduction with anatomically adapted online tube current modulation and preset dose savings. Radiology 230:116–124PubMedCrossRef
22.
Zurück zum Zitat Greess H, Nomayr A, Wolf H, et al (2002) Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol 12:1571–1576PubMedCrossRef Greess H, Nomayr A, Wolf H, et al (2002) Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol 12:1571–1576PubMedCrossRef
23.
Zurück zum Zitat Rizzo SM, Kalra MK, Schmidt B (2005) Automatic exposure control techniques for individual dose adaptation (letter). Radiology 235:335–336PubMedCrossRef Rizzo SM, Kalra MK, Schmidt B (2005) Automatic exposure control techniques for individual dose adaptation (letter). Radiology 235:335–336PubMedCrossRef
24.
Zurück zum Zitat Menke J (2005) Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology 236:565–571PubMedCrossRef Menke J (2005) Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology 236:565–571PubMedCrossRef
25.
Zurück zum Zitat Rizzo SM, Kalra MK, Maher MM, et al (2005) Do metallic endoprostheses increase radiation dose associated with automatic tube-current modulation in abdominal-pelvic MDCT? A phantom and patient study. AJR 184:491–496PubMed Rizzo SM, Kalra MK, Maher MM, et al (2005) Do metallic endoprostheses increase radiation dose associated with automatic tube-current modulation in abdominal-pelvic MDCT? A phantom and patient study. AJR 184:491–496PubMed
26.
Zurück zum Zitat Dalal T, Kalra MK, Rizzo SM, et al (2005) Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients. Radiology 236:671–675PubMedCrossRef Dalal T, Kalra MK, Rizzo SM, et al (2005) Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients. Radiology 236:671–675PubMedCrossRef
Metadaten
Titel
Tube current reduction in pediatric non-ECG-gated heart CT by combined tube current modulation
verfasst von
Hyum Woo Goo
Dong Soo Suh
Publikationsdatum
01.04.2006
Verlag
Springer-Verlag
Erschienen in
Pediatric Radiology / Ausgabe 4/2006
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-005-0105-y

Weitere Artikel der Ausgabe 4/2006

Pediatric Radiology 4/2006 Zur Ausgabe

Minisymposium

Our story

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.