01.02.2011 | Original Article

Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease

verfasst von: Jean-François Paul, Adela Rohnean, Eric Elfassy, Anne Sigal-Cinqualbre

Erschienen in: Pediatric Radiology | Ausgabe 2/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

For coronary artery visualization, retrospective ECG-gated acquisition by dual-source computed tomography (DSCT) was superior to spiral non-ECG-gated acquisition in a paediatric population of congenital heart disease (CHD) patients. However, retrospective cardiac CT is associated with substantial radiation doses to the patient. Recently, DSCT with end-systolic reconstruction was found to be robust for imaging the coronary arteries in patients with high heart rates.

Objective

To evaluate step-and-shoot DSCT with end-systolic reconstruction for evaluating the heart, coronary arteries and other thoracic structures in young children with CHD.

Materials and methods

All neonates and children younger than 6 years of age who were referred to our institution for CHD evaluation between September and October 2009 were included in the study. ECG-gated DSCT was performed in sequential prospective mode centred on the systolic phase identified by ECG analysis. To assess the radiation dose, we recorded the dose-length product (DLP) in mGy·cm and the effective dose in mSv estimated from the DLP. Overall image quality was evaluated using a 5-grade scoring system and was assessed by looking at cardiac and vascular structures. The image quality for the proximal and middle segments of the right and left coronary arteries was also evaluated using a 5-grade scale.

Results

Images of diagnostic quality (grade ≥ 3) were obtained in all 30 children with a mean image quality grade of 4.7 ± 0.6 (range, 3–5). Mean DLP was 5.7 ± 4.8 mGy*cm (range, 1–22 mGy cm) and mean effective radiation dose was 0.26 ± 0.16 mSv (range, 0.05–0.8 mSv).

Conclusion

Prospective ECG-gated thoracic DSCT at end-systole usually provides adequate thoracic and coronary artery image quality in neonates, infants and young children with CHD, independent of heart rate. This new method is associated with lower radiation doses compared to previous literature (mean effective dose, 0.26 mSv).
Literatur
1.
Zurück zum Zitat Westra SJ, Hill JA, Alejos JC et al (1999) Three-dimensional helical CT of pulmonary arteries in infants and children with congenital heart disease. AJR 173:109–115PubMed Westra SJ, Hill JA, Alejos JC et al (1999) Three-dimensional helical CT of pulmonary arteries in infants and children with congenital heart disease. AJR 173:109–115PubMed
2.
Zurück zum Zitat Lambert V, Sigal-Cinqualbre A, Belli E et al (2005) Preoperative and postoperative evaluation of airways compression in pediatric patients with 3-dimensional multislice computed tomographic scanning: effect on surgical management. J Thorac Cardiovasc Surg 129:1111–1118CrossRefPubMed Lambert V, Sigal-Cinqualbre A, Belli E et al (2005) Preoperative and postoperative evaluation of airways compression in pediatric patients with 3-dimensional multislice computed tomographic scanning: effect on surgical management. J Thorac Cardiovasc Surg 129:1111–1118CrossRefPubMed
3.
Zurück zum Zitat Goo HW, Park IS, Ko JK et al (2005) Computed tomography for the diagnosis of congenital heart disease in pediatric and adult patients. Int J Cardiovasc Imaging 21:347–365CrossRefPubMed Goo HW, Park IS, Ko JK et al (2005) Computed tomography for the diagnosis of congenital heart disease in pediatric and adult patients. Int J Cardiovasc Imaging 21:347–365CrossRefPubMed
4.
Zurück zum Zitat Gilkeson RC, Ciancibello L, Zahka K (2003) Pictorial essay.Multidetector CT evaluation of congenital heart disease in pediatric and adult patients. AJR 180:973–980PubMed Gilkeson RC, Ciancibello L, Zahka K (2003) Pictorial essay.Multidetector CT evaluation of congenital heart disease in pediatric and adult patients. AJR 180:973–980PubMed
5.
Zurück zum Zitat Bean MJ, Pannu H, Fishman EK (2005) Three-dimensional computed tomographic imaging of complex congenital cardiovascular abnormalities. J Comput Assist Tomogr 29:721–724CrossRefPubMed Bean MJ, Pannu H, Fishman EK (2005) Three-dimensional computed tomographic imaging of complex congenital cardiovascular abnormalities. J Comput Assist Tomogr 29:721–724CrossRefPubMed
6.
Zurück zum Zitat Ou P, Celermajer DS, Calcagni G et al (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93:908–913CrossRefPubMed Ou P, Celermajer DS, Calcagni G et al (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93:908–913CrossRefPubMed
7.
Zurück zum Zitat Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed
8.
Zurück zum Zitat Willis CE, Slovis TL (2005) The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams–a white paper conference. AJR 184:373–374PubMed Willis CE, Slovis TL (2005) The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams–a white paper conference. AJR 184:373–374PubMed
10.
Zurück zum Zitat Ben Saad M, Rohnean A, Sigal-Cinqualbre A et al (2009) Evaluation of image quality and radiation dose of thoracic and coronary dual-source CT in 110 infants with congenital heart disease. Pediatr Radiol 39:668–676CrossRefPubMed Ben Saad M, Rohnean A, Sigal-Cinqualbre A et al (2009) Evaluation of image quality and radiation dose of thoracic and coronary dual-source CT in 110 infants with congenital heart disease. Pediatr Radiol 39:668–676CrossRefPubMed
11.
Zurück zum Zitat Stolzmann P, Goetti R, Baumueller S et al (2010) Prospective and retrospective ECG-gating for CT coronary angiography perform similarly accurate at low heart rates. Eur J Radiol Jan 14 [Epub ahead of print] Stolzmann P, Goetti R, Baumueller S et al (2010) Prospective and retrospective ECG-gating for CT coronary angiography perform similarly accurate at low heart rates. Eur J Radiol Jan 14 [Epub ahead of print]
12.
Zurück zum Zitat Herzog BA, Husmann L, Burkhard N et al (2008) Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience. Eur Heart J 29:3037–3042CrossRefPubMed Herzog BA, Husmann L, Burkhard N et al (2008) Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience. Eur Heart J 29:3037–3042CrossRefPubMed
13.
Zurück zum Zitat Araoz PA, Kirsch J, Primak AN et al (2009) Optimal image reconstruction phase at low and high heart rates in dual-source CT coronary angiography. Int J Cardiovasc Imaging 25:837–845CrossRefPubMed Araoz PA, Kirsch J, Primak AN et al (2009) Optimal image reconstruction phase at low and high heart rates in dual-source CT coronary angiography. Int J Cardiovasc Imaging 25:837–845CrossRefPubMed
14.
Zurück zum Zitat Bastarrika G, De Cecco CN, Arraiza M et al (2008) Dual-source CT for visualization of the coronary arteries in heart transplant patients with high heart rates. AJR 191:448–454CrossRefPubMed Bastarrika G, De Cecco CN, Arraiza M et al (2008) Dual-source CT for visualization of the coronary arteries in heart transplant patients with high heart rates. AJR 191:448–454CrossRefPubMed
15.
Zurück zum Zitat Adler G, Meille L, Rohnean A et al (2009) Robustness of end-systolic reconstructions in coronary dual-source CT angiography for high heart rate patients. Eur Radiol Nov 5 [Epub ahead of print] Adler G, Meille L, Rohnean A et al (2009) Robustness of end-systolic reconstructions in coronary dual-source CT angiography for high heart rate patients. Eur Radiol Nov 5 [Epub ahead of print]
16.
Zurück zum Zitat Thomas KE, Wang B (2008) Age-specific effective doses for pediatric MSCT examinations at a large children’s hospital using DLP conversion coefficients: a simple estimation method. Pediatr Radiol 38:645–656CrossRefPubMed Thomas KE, Wang B (2008) Age-specific effective doses for pediatric MSCT examinations at a large children’s hospital using DLP conversion coefficients: a simple estimation method. Pediatr Radiol 38:645–656CrossRefPubMed
17.
Zurück zum Zitat Westra SJ, Hurteau J, Galindo A et al (1999) Cardiac electron-beam CT in children undergoing surgical repair for pulmonary atresia. Radiology 213:502–512PubMed Westra SJ, Hurteau J, Galindo A et al (1999) Cardiac electron-beam CT in children undergoing surgical repair for pulmonary atresia. Radiology 213:502–512PubMed
18.
Zurück zum Zitat Paul JF, Lambert V, Losay J et al (2002) Three-dimensional multislice CT scanner: value in patients with pulmonary atresia with septal defect. Arch Mal Coeur Vaiss 95:427–432PubMed Paul JF, Lambert V, Losay J et al (2002) Three-dimensional multislice CT scanner: value in patients with pulmonary atresia with septal defect. Arch Mal Coeur Vaiss 95:427–432PubMed
19.
Zurück zum Zitat Khatri S, Varma SK, Khatri P et al (2008) 64-slice multidetector-row computed tomographic angiography for evaluating congenital heart disease. Pediatr Cardiol 29:755–762CrossRefPubMed Khatri S, Varma SK, Khatri P et al (2008) 64-slice multidetector-row computed tomographic angiography for evaluating congenital heart disease. Pediatr Cardiol 29:755–762CrossRefPubMed
20.
Zurück zum Zitat Goo HW, Park IS, Ko JK et al (2005) Visibility of the origin and proximal course of coronary arteries on non-ECG-gated heart CT in patients with congenital heart disease. Pediatr Radiol 35:792–798CrossRefPubMed Goo HW, Park IS, Ko JK et al (2005) Visibility of the origin and proximal course of coronary arteries on non-ECG-gated heart CT in patients with congenital heart disease. Pediatr Radiol 35:792–798CrossRefPubMed
21.
Zurück zum Zitat Tsai IC, Lee T, Chen MC et al (2007) Visualization of neonatal coronary arteries on multidetector row CT: ECG-gated versus non-ECG-gated technique. Pediatr Radiol 37:818–825CrossRefPubMed Tsai IC, Lee T, Chen MC et al (2007) Visualization of neonatal coronary arteries on multidetector row CT: ECG-gated versus non-ECG-gated technique. Pediatr Radiol 37:818–825CrossRefPubMed
22.
Zurück zum Zitat Paul JF, Abada HT, Sigal-Cinqualbre A (2004) Should low-kilovoltage chest CT protocols be the rule for pediatric patients? AJR 183:1172, author reply 1172PubMed Paul JF, Abada HT, Sigal-Cinqualbre A (2004) Should low-kilovoltage chest CT protocols be the rule for pediatric patients? AJR 183:1172, author reply 1172PubMed
23.
Zurück zum Zitat Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17:2028–2037CrossRefPubMed Paul JF, Abada HT (2007) Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol 17:2028–2037CrossRefPubMed
24.
Zurück zum Zitat Goo HW, Suh DS (2006) Tube current reduction in pediatric non-ECG-gated heart CT by combined tube current modulation. Pediatr Radiol 36:344–351CrossRefPubMed Goo HW, Suh DS (2006) Tube current reduction in pediatric non-ECG-gated heart CT by combined tube current modulation. Pediatr Radiol 36:344–351CrossRefPubMed
25.
Zurück zum Zitat Frush DP, Yoshizumi T (2006) Conventional and CT angiography in children: dosimetry and dose comparisons. Pediatr Radiol 36:154–158CrossRefPubMed Frush DP, Yoshizumi T (2006) Conventional and CT angiography in children: dosimetry and dose comparisons. Pediatr Radiol 36:154–158CrossRefPubMed
26.
Zurück zum Zitat Lee T, Tsai IC, Fu YC et al (2006) Using multidetector-row CT in neonates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation: initial experiences in technical and clinical feasibility. Pediatr Radiol 36:1273–1282CrossRefPubMed Lee T, Tsai IC, Fu YC et al (2006) Using multidetector-row CT in neonates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation: initial experiences in technical and clinical feasibility. Pediatr Radiol 36:1273–1282CrossRefPubMed
27.
Zurück zum Zitat Kroft LJ, Roelofs JJ, Geleijns J (2010) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300CrossRefPubMed Kroft LJ, Roelofs JJ, Geleijns J (2010) Scan time and patient dose for thoracic imaging in neonates and small children using axial volumetric 320-detector row CT compared to helical 64-, 32-, and 16- detector row CT acquisitions. Pediatr Radiol 40:294–300CrossRefPubMed
Metadaten
Titel
Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease
verfasst von
Jean-François Paul
Adela Rohnean
Eric Elfassy
Anne Sigal-Cinqualbre
Publikationsdatum
01.02.2011
Verlag
Springer-Verlag
Erschienen in
Pediatric Radiology / Ausgabe 2/2011
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-010-1804-6

Weitere Artikel der Ausgabe 2/2011

Pediatric Radiology 2/2011 Zur Ausgabe

Hermes

Hermes

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.