Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery

verfasst von: Asger Greval Petersen, Søren Eiskjær, Jon Kaspersen

Erschienen in: Pediatric Radiology | Ausgabe 8/2012

Abstract

Background

During surgery for spinal deformities, accurate placement of pedicle screws may be guided by intraoperative cone-beam flat-detector CT.

Objective

The purpose of this study was to identify appropriate paediatric imaging protocols aiming to reduce the radiation dose in line with the ALARA principle.

Materials and methods

Using O-arm® (Medtronic, Inc.), three paediatric phantoms were employed to measure CTDIw doses with default and lowered exposure settings. Images from 126 scans were evaluated by two spinal surgeons and scores were compared (Kappa statistics). Effective doses were calculated. The recommended new low-dose 3-D spine protocols were then used in 15 children.

Results

The lowest acceptable exposure as judged by image quality for intraoperative use was 70 kVp/40 mAs, 70 kVp/80 mAs and 80 kVp/40 mAs for the 1-, 5- and 12-year-old-equivalent phantoms respectively (kappa = 0,70). Optimised dose settings reduced CTDIw doses 89–93%. The effective dose was 0.5 mSv (91–94,5% reduction). The optimised protocols were used clinically without problems.

Conclusions

Radiation doses for intraoperative 3-D CT using a cone-beam flat-detector scanner could be reduced at least 89% compared to manufacturer settings and still be used to safely navigate pedicle screws.

Introduction

It is well-documented that radiographic examinations for spinal deformity during childhood increase the lifetime risk of cancer, particularly breast cancer [1, 2]. It follows that precautions need to be taken to reduce radiation burden.
New technologies have advanced surgery for spinal deformities. The use of pedicle screws requires accurate placement to avoid damage to the spinal cord and the large vessels in front of the spine. To secure the correct placement of pedicle screws, intraoperative imaging is imperative. For many years, fluoroscopy was the only available intraoperative imaging modality. However, in the last two decades, navigation based on preoperative CT has been developed. However, it is not widely used, possibly due to the time-consuming registration process (coupling of the preoperative CT with patient anatomy) of up to 15–20 min per vertebra making it almost impossible to use in spine deformity surgery with instrumentation of 10–15 vertebrae in a single operation. On the other hand, the need for navigation is most pressing in young patients who have both deformed and small pedicles.
Intraoperative cone-beam flat-detector X-ray systems have changed spinal surgery and are rapidly being implemented worldwide. These provide both 2-D fluoroscopic and 3-D images, which when coupled to a navigation system add significant value to surgical outcomes [36]. With such a system, the registration process is fast and navigation can be performed in four to six vertebrae before a new 3-D scan is needed. The drawbacks are the price (scanner and essential navigation system, approximately €700,000 in 2012) and the increased radiation burden.
According to Zhang et al. [7], the patient dose delivered by the scanner in default mode is equivalent to 0.5–0.6 times the dose delivered by a conventional 64-slice CT-scanner. Usually at least two to three scans are required during surgery for a spinal deformity. Other authors [8, 9] have said the resultant dose is associated with increased cancer risk. On the basis of estimates from the United States [8], it is thought that in children younger than 15 years approximately 80–90 per 100,000 CT examinations attribute to cancer mortality.
Our aim was to identify appropriate intraoperative exposure settings for children for a cone-beam flat-detector system, aiming to reduce the radiation dose.

Materials and methods

Scanner

The O-arm® (Medtronic, Inc.) cone-beam flat-detector system was used. The system has an O-ring type gantry and an X-ray tube equivalent to 32 kW, and a X-ray filter of 4 mm Al. The flat-panel has an amorphous silicon-based detector of 30 cm × 40 cm with a 0.194-mm pixel pitch. The system can be configured in 2-D fluoroscopic mode or 3-D mode. In this study, only 3-D mode was used and only the low-definition mode. In the low-definition mode, 192 single images (compared to 392 images in high-definition mode) of slice thickness 0.833 mm were recorded in a 360-degree rotation of the detector and the radiation source with an image matrix of 512 × 512. The time for image acquisition was 13 s with a beam on-time of 3.91 s. The source-to-isocentre distance of the O-arm was 64.7 cm and the source-to-detector distance 116.8 cm. The collimated X-ray beam was 22.18 × 16.62 cm and the 3-D reconstructed volume was 20 cm × 15 cm.
Besides the 16 default imaging protocols (head, chest, abdomen and extremity applied to small, medium, large and extra large patients, respectively), the O-Arm allows manual adjustment of kVp and mA. With the manual setting, the kVp can be varied between 50 and 120 with 1-kVp intervals, and the mA can be varied between 10 and 120 in predefined steps.
The manufacturer’s advice (personal communication) for abdominal imaging is to use the default values for small (waist circumference 12–26 cm) or the default values for medium (waist circumference 20–34 cm).

Phantoms

The studies were conducted using polymethyl-methacrylate (PMMA) phantoms to estimate patient dose equivalence. Four cylindrical PMMA phantoms were made with diameters of 10 cm, 16 cm, 24 cm and 32 cm (Fig. 1). The 10-cm PMMA would be equivalent to the body of a child <1 year old, and the phantoms of 16 cm, 24 cm and 32 cm would be equivalent to the body of a 5-year old, 12-year-old and an adult, respectively [3, 10]. All phantoms were 15 cm long and had standard holes for CTDI dosimetry at the centre and at the perimeter but were also made with a 3-cm cylindrical hole through the phantom in a radial distance from the perimeter corresponding with the position of the spine in a patient of the same size. In this hole, four samples made from human femoral neck (Fig. 2), three of which also contained pedicle screws, were placed and scanned one at a time in the centre of the phantom with the remaining holes filled with solid PMMA rods. The locations of the screws with respect to the cortical edge of the bone were chosen to match clinically realistic scenarios.

Dose measurements

The dose to the patient was expressed as CTDIw and calculated from the dose measured in the PMMA cylinders in the periphery and centre with a 100-mm pencil ion chamber with all other holes filled with solid PMMA rods. The CTDIw was measured for each phantom size using both the default and the optimised values for the setting of kVp and mA. The absorbed doses of radiosensitive organs and total effective doses were calculated for each phantom size, using both the default and the optimised values for kVp and mAs. The calculations were performed with the PCXMC Monte Carlo dose simulation program version 2.0 (STUK - Radiation and Nuclear Safety Authority, Helsinki, Finland) using the ICRP 103 tissue weighting-factors [11]. To perform this calculation, the PCXMC program requires measurements of the incident air kerma at the surface of each phantom without backscatter factor. This was measured at the isocentre with the 100-mm pencil ion chamber and calculated for the surface of each phantom using the inverse square law. To reduce the computation time with PCXMC, the 360-degree arc was divided into 12 parts of 30 degrees. The absorbed doses to the radiosensitive organs and total effective doses were calculated by summing the contributions of each of the 12 dose calculations.

Image quality assessment

Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured. SNR was calculated as S1/σ1, where S1 represents the mean pixel value within a region of interest (ROI) in the PMMA phantom with all PMMA rods inserted at the position of the bone samples and σ1 represents the standard deviation of pixel values within the same ROI. CNR was estimated as |S2−S1|/σ1, where S2 is the mean pixel value in the pure bone sample. For this purpose, ImageJ version 1.32 (http://​rsb.​info.​nih.​gov/​ij/​) was used.
The subjective image quality of all phantom images with pedicle screws (126 scans) was evaluated independently by two spinal surgeons, both with more than 10 years of experience in spinal surgery and intraoperative imaging. The image quality was deemed adequate if the outlines of the bones were visible and if it was possible to discern whether the screw penetrated the bone. If the outlines of the bone sample and the screw position could be visualised with certainty, the image quality was deemed as adequate. In all other cases, the image quality was deemed inadequate. Interobserver agreement was measured with Kappa statistic [12, 13].

Dose optimisation

The scans started with the factory default settings, followed by a series of scans with decreasing mA until the lowest possible tube current of 10 mA was reached. If image quality was still acceptable, additional scans were acquired at 10 mA while reducing kV until an unacceptable image quality was achieved. The scan parameters with the lowest dose to the phantom where the image quality in all four bone samples were accepted were recorded as the suggested optimum low-dose settings for that specific phantom size. Scans were then acquired with all holes filled with solid PMMA rods to measure image noise and dose.
After institutional review board approval, the recommended new low-dose 3-D spine protocol was tested in clinical practice in 15 children (10 females) with severe deformities in whom it would not have been possible to place pedicle screws without navigation. This would otherwise require conventional preoperative CT with a higher dose. Only the 16-cm and 24-cm protocols were used. The average age was 11.5 years (range, 2–17 years). An average of 9.7 vertebrae were scanned (range, 3–16) using an average of 2.5 acquisitions (range, 1–4), as shown in Table 1.
Table 1
Patient characteristics, number of scans, protocols and radiation doses for 15 children undergoing corrective spinal surgery using a cone-beam scanner with radiation-dose optimised exposure for guiding pedicle screw placement
Patient
Sex; age at surgery (years)
Diagnosis
Number of levels
Patient height (cm)/weight (kg)
Number of intraoperative scans
DLP mGy·cma
Scan protocol exposure settings kVp/mA
Instrumentation
1
M, 17
Idiopathic scoliosis
12
185/76
3
94
80/20
Conventionalb
2
F, 2
Hemivertebra
3
82/11
1
17
80/10
Conventionalb
3
M, 13
Neuromuscular scoliosis
16
159/31
4
80
70/20
Conventionalb
4
F, 15
Idiopathic scoliosis
7
158/45
2
65
80/20
Conventionalb
5
F, 15
Idiopathic scoliosis
9
162/52
2
40
70/20
Conventionalb
6
M, 12
Neuromuscular scoliosis
15
152/32
3
60
70/20
Conventionalb
7
M, 6
Infantile scoliosis
4
99/13
2
65
80/20
Growth rod
8
F, 5
Infantile scoliosis
8
109/16
3
86
80/10
Conventionalb
9
F, 14
Idiopathic scoliosis
6
161/48
2
40
70/20
Conventionalb
10
F, 12
Neuromuscular scoliosis
13
121/28
4
80
70/20
Conventionalb
11
F, 11
Neuromuscular scoliosis
11
150/43
2
60
80/10,70/20
Conventionalb
12
M, 14
Neuromuscular scoliosis
16
165/40
4
80
70/20
Conventionalb
13
F, 14
Idiopathic scoliosis
10
175/75
2
40
70/20
Conventionalb
14
F, 8
Spondylolisthesis gr. 4
4
134/28
1
16
80/10
Conventionalb
15
F, 14
Idiopathic scoliosis
11
171/60
3
97
80/10
Conventionalb
a According to the manufacturer
b Conventional segmental instrumentation with pedicle screws and rods

Results

As expected, SNR and CNR decreased nonlinearly with decreasing radiation dose (Fig. 3). SNR and CNR at lowest acceptable dose decrease with increasing phantom size. Therefore, patient size is a crucial factor when choosing the values of SNR and CNR for lowest acceptable image quality.
Table 2 compares the preset and optimised protocols and dose values. For both the 10-cm and the 16-cm phantoms, the optimised dose was compared with the dose of the standard small chest protocol. For the 24-cm phantom, the optimised dose was compared with dose of the standard medium chest protocol. All optimised settings yielded at least a 89% reduction in dose, with the 10-cm protocol yielding a 93% dose reduction as shown in Tables 2 and 3. This was corroborated when the effective dose was calculated with effective dose values approximately 0.5 mSv for the optimised protocols and effective dose values 12–19 times higher for the default protocols as shown in Table 4.
Table 2
Radiation doses at default exposure settings for different sizes of body/anatomy (L large, M medium, S small) and at exposures optimised for low-dose 3-D paediatric spine imaging with dose reduction factors for all phantom sizes
Phantom diameter (cm)
Protocol
kV
mA
mAs
CTDIw (mGy)
Dose reduction, %
32
L
120
50
200
11.3
 
24
M
120
40
160
14.5
 
24
Low dose
70
20
80
1.6
89.2
16
S
120
32
128
14.4
 
16
Low dose
80
10
40
1.5
89.4
10
S
120
32
128
16.6
 
10
Low dose
70
10
40
1.2
93.0
Table 3
Scan protocols for a 1-year-old-equivalent phantom with dose reduction achieved and scoring of image quality
Phantom diameter (cm)
Scan protocol
kV
mA
mAs
CTDIw (mGy)
Dose reduction, %
Image quality scoring
10
1
120
32
128
16.59
0
OK
10
2
120
20
80
10.41
37.3
OK
10
3
120
10
40
5.47
67.0
OK
10
4
100
10
40
3.39
79.6
OK
10
5
80
10
40
1.81
89.1
OK
10
6
70
10
40
1.15
93.1
OK
10
7
60
10
40
0.69
95.8
Not OK
Table 4
Estimated effective doses at default exposure settings for different sizes of body/anatomy (L large, M medium, S small) and at exposures optimised for 3-D paediatric spine imaging with dose reduction factors for all phantom sizes
Phantom diameter (cm)
Protocol
kV
mA
mAs
Effective dose (mSv)
Dose reduction, %
24
M
120
40
160
8.3
 
24
Low dose
70
20
80
0.4
94.7
16
S
120
32
128
6.3
 
16
Low dose
80
10
40
0.5
91.7
10
S
120
32
128
5.8
 
10
Low dose
70
10
40
0.5
92.2
The observers agreed that the lowest acceptable dose for intraoperative imaging was 70 kVp/40 mAs for the 1-year-old-equivalent phantom (10 cm), 70 kVp/80 mAs for the 5-year-old-equivalent phantom (16 cm) and 80 kVp/40 mAs for the 12-year-old-equivalent phantom (24 cm). Figure 4 shows examples of images of pedicle bone implants with minimal penetration at different dose settings.
The interobserver agreement for all scans had a kappa = 0,70 (substantial agreement [14]). For the 10-cm, 16-cm and 24-cm phantoms, kappa was 0.64, 0.72 and 0.70, respectively, also indicating substantial agreement.
In all 15 operations using the optimised setting, the spinal surgeon achieved adequate intraoperative imaging with the cone-beam flat-detector scanner followed by navigation using the Stealth system (Stealth Station®, Medtronic, Minneapolis, USA). Figure 5 shows examples of intraoperative images with different dose settings with and without pedicle screws. The same protocol was used both for the thoracic and abdominal regions. In none of the cases was control CT done postoperatively because of suspected misplacement of screws or neurological deterioration. Conventional radiographs showed satisfactory placement of all screws and the neurological status was unchanged for all children. Spinal cord monitoring (sensory evoked potentials (SEP) and motor evoked potentials (MEP)) did not indicate damage to the spinal cord or the spinal nerves in any of the children.

Discussion

The phantom model allowed a systematic evaluation of different dose settings from the highest to the lowest possible. It also allowed different bone inserts containing pedicle screws to be evaluated with regard to image quality. This is especially important as the cornerstone of image guidance during pedicle screw insertion is the ability to clearly identify the cortex relative to the pedicle screw. Streak artefacts from metallic implants are a major concern, and any realistic model needs to incorporate this. Our phantom model could only be used to evaluate 90-degree perforations of the cortex due to the small size of the bone samples. However, to compensate for this limitation, for the sample with the perforating screw, the perforation was made as small as possible.
The CTDI100 does not measure the actual absorbed dose by the individual patient but should be considered an index for comparisons. The homogeneous PMMA does not simulate the different tissue types and heterogeneities in vivo. The CTDI100 will underestimate the radiation dose because the 100-mm pencil ion chamber only partly covers the collimated beam width. For X-ray beams wider than 40 mm, a pencil chamber longer than 100 mm is required. The CTDIw will underestimate the ideal CTDIw with approximately 20% at a collimated beam width of 20 mm [15]. Dixon [16] has described an alternative method to directly measure the dose at the central scan plane using a small ion chamber.
The patient dose of the cone-beam system relative to CT has been described by Zhang et al. [7], who showed that with identical techniques (kVp, mAs, etc.) and with the same scan length, the cone-beam system in 3-D mode delivered approximately half the radiation dose of a 64-slice CT scanner. Their use of a Farmer ion chamber gave a more correct estimation of patient dose compared to the 100-mm ion chamber used in our study for dose comparison as it could be used to directly measure the point dose at the centre of the scan length. Differences between cone-beam systems and CT are to be expected due to differences in source-to-isocentre distance (SID). The SID of the cone-beam system was 64.7 cm while the SID of the 64-slice CT scanner in question was 57.1 cm. This accounts for approximately 29% of the dose difference [7]. Second, there is a difference in fan beam angle. The cone-beam system has an angle of approximately 20 degrees, and the CT scanner approximately 45 degrees. This leads to less peripheral radiation dose measurements for the cone-beam system because the ion chamber is outside the beam during parts of the scan [7]. The differences in over beaming between the cone-beam system and the CT scanner will account for a slight dose reduction in favour of the cone-beam system [7].
Would a vertebra have been a better bone sample for the phantom part of the study? This would without doubt have better represented the normal anatomy. The cortical thickness of the femoral head is greater than the cortical thickness of the vertebra, and this would have influenced the accuracy of the imaging representation [17]. A more accurate model would probably improve the surgeons’ ability to discern whether a screw was placed correctly. The bone mineral density of children and adolescents changes significantly with age [18, 19]. The bone samples used here were acquired from the local bone bank. The femoral necks used in this study were obtained solely from patients undergoing hip arthroplasty, and these patients are in general not osteoporotic [18, 21]. Mean BMD for a hip arthroplasty patient is approximately 1 g/cm2 with slight variation in several recent studies [20, 21]. Lumbar spine BMD in children varies from approximately 0.6 g/cm2 (age 5 years) to 1.0 g/cm2 (age 17 years) with much greater variation [18, 19]. A single bone sample would not model the bone density of any childhood vertebra since BMD changes with age. In this series (no BMD measurements), the probable difference between BMD in the femoral neck of elderly patients with hip arthrosis and children’s spines would tend to underestimate the radiation dose needed to satisfactorily visualise the correct placement of pedicle screws in young children because the BMD is considerably lower in children 1–9 years of age. However, this did not cause any problems for the implementation of the optimised protocol. For most children who undergo corrective spinal surgery (10–17 years of age), BMD of the lumbar spine is probably not very different from that of the bone samples used in our study.
A substantial number of pedicle screws are misplaced when navigation is not used. The rate of pedicle perforation varies between 10% and 40% [2224]. Our own data (unpublished) for the first 30 children where the screws were placed under fluoroscopic guidance and then position-controlled with volumetric imaging indicate that approximately 8% of a total of 424 screws were misplaced (according to the Ohlin classification [24]) and 1% of screws had to be repositioned or removed. Since we started using the cone-beam flat-panel detector scanner in 2008, we have not had any nerve or spinal cord damage caused by misplaced screws. This has been corroborated by several other studies [2527]. The cost-benefit of navigation is well-documented [5]. It is also well-documented that 3-D techniques are superior to other techniques used for navigation [22, 23]. Some surgeons use freehand techniques relying only on anatomical landmarks. The anatomy of the scoliotic spine in children is highly variable, the pedicles often very small and the deformity often great. Freehand techniques are only feasible for a few very talented and experienced spinal deformity surgeons. Most surgeons would prefer to use some kind of intraoperative imaging guidance. The number of misplaced screws in degenerative adult lumbar spines using freehand techniques was significantly greater than the number of misplaced screws placed following a 3-D scan with subsequent navigation [27]. According to the manufacturers, the radiation dose of one 3-D scan equals that of 35 s of fluoroscopy [7]. Fluoroscopy time has been calculated to be approximately 7–20 s per screw [2831]. With an average ten screws per child in our case study, this equals 68–193 s of fluoroscopy time per operation, or the equivalent of two to six volumetric scans, representing a higher radiation dose compared to our case study. One also needs to take into account that the pedicle screw misplacement rate is significantly reduced when navigation is based on 3-D imaging [2224]. The misplacement rate is even higher in children, which further strengthens the case for using volumetric guidance [27].
There are currently few alternatives to the system used in our study [32, 33]. Other available systems are slower and only two to three spinal levels can be imaged simultaneously, resulting in an imaging time of 8–9 min (as compared to 1 min of discontinuation time for the described system). Other low-dose systems like EOS [34, 35] or the system described by Abul-Kassim et al. [24] cannot be used intraoperatively. The EOS system in particular can only be used for standing or sitting patients [34, 35].
Abul-Kasim et al. [36] have also tried to optimise the radiation exposure and image quality of the same system used in our study. Using a completely different model (anthropomorphic adult chest model and porcine spine with pedicle screws), they arrive at almost the same dose settings as we recommend. The pedicle perforation rate in the porcine spine model was higher than the perforation rate simulated in our model, which may explain some difference in observer agreement at low dose.
The cylindrical-equivalent diameter of a body is defined as the diameter of the cylinder that the body would form if laterally compressed into a cylinder of equal cross-sectional area. This definition should not be confused with the definition of patient-equivalent cylinder based on the patient’s weight and height as suggested by The Danish National Board of Health [37]. The equivalent diameter may be calculated as d = 2√ab where a and b are the minor and major diameters of the ellipse. Note that the circumference of the patient-equivalent circle and the elliptic cross-section of the patient are not equal, and can therefore only estimate the diameter of the patient equivalent cylinder with an error of 10–15% depending on the eccentricity of the patient. Given that the difference in the diameter between the patient equivalent phantoms is about 50%, it might be sufficient to use the circumference to calculate the diameter of the patient to select the appropriate protocol.
The phantoms in our study simulated the lumbar anatomy. The results may therefore not adequately reflect the dose reduction achievable in the thorax. Based on the level of scattered radiation in the thorax compared with the lumbar level, we would anticipate the dose in the thorax to be even lower.
The lower-contrast images of the cone-beam scanner are not comparable with a standard CT [4], but in the case of spinal surgery, spatial resolution is of greater concern. Dealing with very high-density material (bone and metal) relative to water, we showed that the needed spatial resolution can be obtained at a fraction of the preset exposure values (Table 2).
The parameters of the optimised protocols were to some extent dictated by limitations of the generator. Even at the lowest possible tube current (10 mA), all phantoms at all voltages above 90 kVp showed acceptable image quality. If the mA were lowered further, a lower patient dose might have been achievable since the same dose at a higher kVp would result in a lower absorbed dose to the patient. Dose to the patient may have been reduced further by using a bow-tie filter typically employed in standard CT scanners [38]. A preliminary study with such a filter has shown that dose may be additionally reduced by a factor of 2. Automated current modulation as proposed by Kalender et al. [39] might also further reduce the dose as would adaptive statistical iterative reconstruction combined with conventional filtered back projection (a further estimated reduction of 30–40% [40]).

Conclusion

With optimised exposures at 70 kVp/40 mAs for a 1-year-old-equivalent phantom, 70 kVp/80 mAs for a 5-year-old-equivalent phantom and 80 kVp/40 mAs for a 12-year-old-equivalent phantom, radiation doses for intraoperative 3-D imaging with a cone-beam flat-panel detector scanner were reduced at least 89% and could still be used to safely guide the placement of pedicle screws. The effective doses for optimised scans were estimated at approximately 0.5 mSv and were between 91–94,5% lower than the effective dose estimated for the manufacturers' default exposure values.

Open Access

This article is distributed under the terms of the Creative Commons Attribution License, which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Pädiatrie

Kombi-Abonnement

Mit e.Med Pädiatrie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Pädiatrie, den Premium-Inhalten der pädiatrischen Fachzeitschriften, inklusive einer gedruckten Pädiatrie-Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Doody MM, Lonstein JE, Stovall M et al (2000) Breast cancer mortality after diagnostic radiotherapy. Spine 25:2052–2063PubMedCrossRef Doody MM, Lonstein JE, Stovall M et al (2000) Breast cancer mortality after diagnostic radiotherapy. Spine 25:2052–2063PubMedCrossRef
2.
Zurück zum Zitat Ronckers CM, Land CE, Miller JS et al (2010) Cancer mortality among women frequently exposed to radiographic examinations for spinal disorders. Radiat Res 174:83–90PubMedCrossRef Ronckers CM, Land CE, Miller JS et al (2010) Cancer mortality among women frequently exposed to radiographic examinations for spinal disorders. Radiat Res 174:83–90PubMedCrossRef
3.
Zurück zum Zitat Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779PubMedCrossRef Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779PubMedCrossRef
4.
Zurück zum Zitat Tian NF, Xu HZ (2009) Image guided pedicle screw insertion accuracy: a metaanalysis. Int Orthop 18:341–346 Tian NF, Xu HZ (2009) Image guided pedicle screw insertion accuracy: a metaanalysis. Int Orthop 18:341–346
5.
Zurück zum Zitat Watkins RG IV, Gupta A, Watkins RG III (2010) Cost-effectiveness of image-guided spine surgery. Open Orthop J 4:228–233PubMedCrossRef Watkins RG IV, Gupta A, Watkins RG III (2010) Cost-effectiveness of image-guided spine surgery. Open Orthop J 4:228–233PubMedCrossRef
6.
Zurück zum Zitat Atesok K, Finkelstein J, Khoury A et al (2007) The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Int J Care 38:1163–1169 Atesok K, Finkelstein J, Khoury A et al (2007) The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Int J Care 38:1163–1169
7.
Zurück zum Zitat Zhang J, Weir V, Fajardo L et al (2009) Dosimetric characterization of a cone-beam O-arm™ imaging system. J Xray Sci Technol 17:305–317PubMed Zhang J, Weir V, Fajardo L et al (2009) Dosimetric characterization of a cone-beam O-arm™ imaging system. J Xray Sci Technol 17:305–317PubMed
8.
Zurück zum Zitat Brenner DJ, Elliston CD, Hall EJ et al (2001) Estimated risk of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed Brenner DJ, Elliston CD, Hall EJ et al (2001) Estimated risk of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed
9.
Zurück zum Zitat Mueller DL, Hatab M, Al-Senan R et al (2011) Pediatric radiation exposure during the initial evaluation for blunt trauma. J Trauma 70:724–731PubMedCrossRef Mueller DL, Hatab M, Al-Senan R et al (2011) Pediatric radiation exposure during the initial evaluation for blunt trauma. J Trauma 70:724–731PubMedCrossRef
10.
Zurück zum Zitat National Electrical Manufacturers Association (2004) General guidelines for reducing CT pediatric dose through scan protocol adjustments. NEMA white paper, Rosslyn, VA, USA. www.nema.org National Electrical Manufacturers Association (2004) General guidelines for reducing CT pediatric dose through scan protocol adjustments. NEMA white paper, Rosslyn, VA, USA. www.​nema.​org
11.
Zurück zum Zitat International Commission on Radiological Protection (2007) The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103 International Commission on Radiological Protection (2007) The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103
12.
Zurück zum Zitat Siegel J, Castellan NJ (1988) Nonparametric statistics for the behavioural sciences. McGraw-Hill, New York Siegel J, Castellan NJ (1988) Nonparametric statistics for the behavioural sciences. McGraw-Hill, New York
13.
14.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
15.
Zurück zum Zitat Li X, Zhang D, Liu B (2011) A practical approach to estimate the weighted CT dose index over an infinite integration length. Phys Med Biol 56:5789–5803PubMedCrossRef Li X, Zhang D, Liu B (2011) A practical approach to estimate the weighted CT dose index over an infinite integration length. Phys Med Biol 56:5789–5803PubMedCrossRef
16.
Zurück zum Zitat Dixon RL (2003) A new look at CT dose measurement: Beyond CTDI. Med Phys 30:1227–1280 Dixon RL (2003) A new look at CT dose measurement: Beyond CTDI. Med Phys 30:1227–1280
17.
Zurück zum Zitat Prevrhal S, Fox JC, Shepherd JA et al (2003) Accuracy of CT-based thickness measurement of thin structures: modelling of limited spatial resolution in all three dimensions. Med Phys 30:1–8PubMedCrossRef Prevrhal S, Fox JC, Shepherd JA et al (2003) Accuracy of CT-based thickness measurement of thin structures: modelling of limited spatial resolution in all three dimensions. Med Phys 30:1–8PubMedCrossRef
18.
Zurück zum Zitat Khadilkar AV, Sanwalka NJ, Chiplonkar SA et al (2011) Normative data and percentile curves for dual energy x-ray absorptiometry in healthy Indian girls and boys aged 5–17 years. Bone 48:810–819PubMedCrossRef Khadilkar AV, Sanwalka NJ, Chiplonkar SA et al (2011) Normative data and percentile curves for dual energy x-ray absorptiometry in healthy Indian girls and boys aged 5–17 years. Bone 48:810–819PubMedCrossRef
19.
Zurück zum Zitat Shepherd JA, Wang L, Gilsanz V et al (2011) Optimal monitoring time between DXA measures in children. J Bone Miner Res 26:2745–2752PubMedCrossRef Shepherd JA, Wang L, Gilsanz V et al (2011) Optimal monitoring time between DXA measures in children. J Bone Miner Res 26:2745–2752PubMedCrossRef
20.
Zurück zum Zitat Wolf O, Ström H, Milbrink J et al (2009) Differences in hip bone mineral density may explain the hip fracture pattern in osteoarthritic hips. Acta Orthop 80:308–313PubMedCrossRef Wolf O, Ström H, Milbrink J et al (2009) Differences in hip bone mineral density may explain the hip fracture pattern in osteoarthritic hips. Acta Orthop 80:308–313PubMedCrossRef
21.
Zurück zum Zitat Penny JØ, Ovesen O, Brixen K et al (2010) Bone mineral density of the femoral neck in resurfacing hip arthroplasty. Precision of DXA biased by region of interest and rotation of hip. Acta Orthop 81:318–323PubMedCrossRef Penny JØ, Ovesen O, Brixen K et al (2010) Bone mineral density of the femoral neck in resurfacing hip arthroplasty. Precision of DXA biased by region of interest and rotation of hip. Acta Orthop 81:318–323PubMedCrossRef
22.
Zurück zum Zitat Kosmoupoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine 32:E111–E120CrossRef Kosmoupoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine 32:E111–E120CrossRef
23.
Zurück zum Zitat Ledonio CG, Vitale MG, Richards BS (2010) Meta-analysis of the safety and efficacy of pedicle screw spinal instrumentation in pediatric spinal deformity: Results of SRS and POSNA task force. Spine J 10:S4–S5CrossRef Ledonio CG, Vitale MG, Richards BS (2010) Meta-analysis of the safety and efficacy of pedicle screw spinal instrumentation in pediatric spinal deformity: Results of SRS and POSNA task force. Spine J 10:S4–S5CrossRef
24.
Zurück zum Zitat Abul-Kassim K, Strömbeck A, Ohlin A et al (2009) Reliability of low radiation dose CT in the assessment of screw placement after posterior scoliosis surgery evaluated with a new grading system. Spine 34:941–948. doi:10.1097/BSD.0b013e318211fdea CrossRef Abul-Kassim K, Strömbeck A, Ohlin A et al (2009) Reliability of low radiation dose CT in the assessment of screw placement after posterior scoliosis surgery evaluated with a new grading system. Spine 34:941–948. doi:10.​1097/​BSD.​0b013e318211fdea​ CrossRef
25.
Zurück zum Zitat Silberman J, Riese F, Allam Y et al (2011) Computer tomography assessment of pedicle screw placement in lumbar and sacral spine: comparison between freehand and O-arm based navigation techniques. Eur Spine J 20:875–881CrossRef Silberman J, Riese F, Allam Y et al (2011) Computer tomography assessment of pedicle screw placement in lumbar and sacral spine: comparison between freehand and O-arm based navigation techniques. Eur Spine J 20:875–881CrossRef
26.
Zurück zum Zitat Oertel MF, Hobart J, Stein M et al (2011) Clinical and methodological precision of spinal navigation assisted by 3D intraoperative O-arm radiographic imaging. J Neurosurg Spine 14:532–536PubMedCrossRef Oertel MF, Hobart J, Stein M et al (2011) Clinical and methodological precision of spinal navigation assisted by 3D intraoperative O-arm radiographic imaging. J Neurosurg Spine 14:532–536PubMedCrossRef
27.
28.
Zurück zum Zitat Jones DP, Robertson PA, Lunt B et al (2000) Radiation exposure during fluoroscopically assisted pedicle screw insertion in the lumbar spine. Spine 25:1538–1541PubMedCrossRef Jones DP, Robertson PA, Lunt B et al (2000) Radiation exposure during fluoroscopically assisted pedicle screw insertion in the lumbar spine. Spine 25:1538–1541PubMedCrossRef
29.
Zurück zum Zitat Ul Haque M, Shufflebagger HL, O’Brien M et al (2006) Radiation exposure during pedicle screw placement in adolescent idiopathic scoliosis: is fluoroscopy safe? Spine 31:2516–2520PubMedCrossRef Ul Haque M, Shufflebagger HL, O’Brien M et al (2006) Radiation exposure during pedicle screw placement in adolescent idiopathic scoliosis: is fluoroscopy safe? Spine 31:2516–2520PubMedCrossRef
30.
Zurück zum Zitat Rampersaud YR, Foley KT, Shen AC et al (2000) Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine 25:2637–2645PubMedCrossRef Rampersaud YR, Foley KT, Shen AC et al (2000) Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion. Spine 25:2637–2645PubMedCrossRef
31.
Zurück zum Zitat Slomczykowski M, Roberto M, Schneeberger P et al (1999) Radiation dose for pedicle screw insertion. Fluoroscopic method versus computer-assisted surgery. Spine 24:975–982PubMedCrossRef Slomczykowski M, Roberto M, Schneeberger P et al (1999) Radiation dose for pedicle screw insertion. Fluoroscopic method versus computer-assisted surgery. Spine 24:975–982PubMedCrossRef
32.
Zurück zum Zitat Zausinger S, Scheder B, Uhl E et al (2009) Intraoperative computed tomography with integrated navigation system in spinal stabilization. Spine 34:2919–2926PubMedCrossRef Zausinger S, Scheder B, Uhl E et al (2009) Intraoperative computed tomography with integrated navigation system in spinal stabilization. Spine 34:2919–2926PubMedCrossRef
33.
Zurück zum Zitat Beck M, Mittelmeier T, Gierer P et al (2009) Benefit and accuracy of intraoperative 3D-imaging after pedicle screw placement: a prospective study in stabilizing thoracolumbar fractures. Eur Spine J 18:1469–1477PubMedCrossRef Beck M, Mittelmeier T, Gierer P et al (2009) Benefit and accuracy of intraoperative 3D-imaging after pedicle screw placement: a prospective study in stabilizing thoracolumbar fractures. Eur Spine J 18:1469–1477PubMedCrossRef
34.
Zurück zum Zitat Illés T, Tunyogi-Csapó M, Somoskeöy S (2011) Breakthrough in three-dimensional scoliosis diagnosis: significance of horizontal plane view and vertebra vectors. Eur Spine J 20:135–143PubMedCrossRef Illés T, Tunyogi-Csapó M, Somoskeöy S (2011) Breakthrough in three-dimensional scoliosis diagnosis: significance of horizontal plane view and vertebra vectors. Eur Spine J 20:135–143PubMedCrossRef
35.
Zurück zum Zitat Deschênes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities. Reducing patient radiation dose with a new slot-scanning X-ray imager. Spine 35:989–994PubMedCrossRef Deschênes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities. Reducing patient radiation dose with a new slot-scanning X-ray imager. Spine 35:989–994PubMedCrossRef
36.
Zurück zum Zitat Abul-Kasim K, Söderberg M, Selariu E et al (2012) Optimization of radiation exposure and image quality of the cone-beam O-arm intraoperative imaging system in spinal surgery. J Spinal Disord Tech 25:52–58. doi:10.1097/BSD.0b013e318211fdea Abul-Kasim K, Söderberg M, Selariu E et al (2012) Optimization of radiation exposure and image quality of the cone-beam O-arm intraoperative imaging system in spinal surgery. J Spinal Disord Tech 25:52–58. doi:10.​1097/​BSD.​0b013e318211fdea​
37.
Zurück zum Zitat Statens Institut for Strålehygiejne (2006) Børnedoser ved radiologi. Sundhedsstyrelsen, Copenhagen Statens Institut for Strålehygiejne (2006) Børnedoser ved radiologi. Sundhedsstyrelsen, Copenhagen
38.
Zurück zum Zitat Mail N, Moseley DJ, Siewerdsen JH et al (2009) The influence of bowtie filtration on cone-beam CT image quality. Med Phys 36:22–32 Mail N, Moseley DJ, Siewerdsen JH et al (2009) The influence of bowtie filtration on cone-beam CT image quality. Med Phys 36:22–32
39.
Zurück zum Zitat Kalender WA, Buchenau S, Deak P et al (2008) Technical approaches to the optimization of CT. Phys Med 24:71–79PubMedCrossRef Kalender WA, Buchenau S, Deak P et al (2008) Technical approaches to the optimization of CT. Phys Med 24:71–79PubMedCrossRef
40.
Zurück zum Zitat Miéville FA, Gudinchet F, Rizzo E et al (2011) Paediatric cardiac CT examinations: impact of the iterative reconstruction method ASIR on image quality—preliminary findings. Pediatr Radiol 41:1154–1164PubMedCrossRef Miéville FA, Gudinchet F, Rizzo E et al (2011) Paediatric cardiac CT examinations: impact of the iterative reconstruction method ASIR on image quality—preliminary findings. Pediatr Radiol 41:1154–1164PubMedCrossRef
Metadaten
Titel
Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery
verfasst von
Asger Greval Petersen
Søren Eiskjær
Jon Kaspersen
Publikationsdatum
01.08.2012
Verlag
Springer-Verlag
Erschienen in
Pediatric Radiology / Ausgabe 8/2012
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-012-2396-0

Weitere Artikel der Ausgabe 8/2012

Pediatric Radiology 8/2012 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.