Skip to main content
Erschienen in: Pediatric Radiology 5/2015

01.05.2015 | Original Article

Radiation dose from multidetector CT studies in children: results from the first Italian nationwide survey

verfasst von: Claudio Granata, Daniela Origgi, Federica Palorini, Domenica Matranga, Sergio Salerno

Erschienen in: Pediatric Radiology | Ausgabe 5/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

Multidetector CT (MDCT) scanners have contributed to the widespread use of CT in paediatric imaging. However, concerns are raised for the associated radiation exposure. Very few surveys on radiation exposure from MDCT studies in children are available.

Objective

The aim of this study was to outline the status of radiation exposure in children from MDCT practice in Italy.

Materials and methods

In this retrospective multicentre study we asked Italian radiology units with an MDCT scanner with at least 16 slices to provide dosimetric and acquisition parameters of CT examinations in three age groups (1–5, 6–10, 11–15 years) for studies of head, chest and abdomen. The dosimetric results were reported in terms of third-quartile volumetric CT dose index (CTDIvol) (mGy), size-specific dose estimate (SSDE) (mGy), dose length product (DLP) (mGy cm), and total DLP for multiphase studies. These results were compared with paediatric European and adult Italian published data. A multivariate analysis assessed the association of CTDIvol with patient characteristics and scanning modalities.

Results

We collected data from 993 MDCT examinations performed at 25 centres. For age groups 1–5 years, 6–10 years and 11–15 years, the CTDIvol, DLP and total DLP values were statistically significantly below the values observed in our analogous national survey in adults, although the difference decreased with increasing age. CTDIvol variability among centres was statistically significant (variance = 0.07; 95% confidence interval = 0.03–0.16; P < 0.001).

Conclusions

This study reviewed practice in Italian centres performing paediatric imaging with MDCT scanners. The variability of doses among centres suggests that the use of standardised CT protocols should be encouraged.
Literatur
1.
Zurück zum Zitat Goske MJ, Applegate KE, Boylan J et al (2008) The ‘Image Gently’ campaign: increasing CT radiation dose awareness through a national education and awareness program. Pediatr Radiol 38:265–269 Goske MJ, Applegate KE, Boylan J et al (2008) The ‘Image Gently’ campaign: increasing CT radiation dose awareness through a national education and awareness program. Pediatr Radiol 38:265–269
2.
Zurück zum Zitat Tsushima Y, Taketomi-Takahashi A, Takei H et al (2010) Radiation exposure from CT examination in Japan. BMC Med Imaging 10:1–8CrossRef Tsushima Y, Taketomi-Takahashi A, Takei H et al (2010) Radiation exposure from CT examination in Japan. BMC Med Imaging 10:1–8CrossRef
3.
Zurück zum Zitat Aroura A, Bouchud FO, Valley J-F et al (2007) Number of x-ray examinations performed on paediatric and geriatric patients compared with adult patients. Radiat Protect Dosim 123:402–408CrossRef Aroura A, Bouchud FO, Valley J-F et al (2007) Number of x-ray examinations performed on paediatric and geriatric patients compared with adult patients. Radiat Protect Dosim 123:402–408CrossRef
6.
Zurück zum Zitat Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505CrossRefPubMedCentralPubMed Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505CrossRefPubMedCentralPubMed
7.
Zurück zum Zitat Mathews JD, Forsythe AV, Brady Z et al (2013) Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. Br Med J 346:f2360–2378CrossRef Mathews JD, Forsythe AV, Brady Z et al (2013) Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. Br Med J 346:f2360–2378CrossRef
9.
Zurück zum Zitat Teunen D (1998) European directive on health protection of individuals against the dangers of ionizing radiation in relation to medical exposure (97/43/EURATOM). J Radiol Prot 18:133–137CrossRefPubMed Teunen D (1998) European directive on health protection of individuals against the dangers of ionizing radiation in relation to medical exposure (97/43/EURATOM). J Radiol Prot 18:133–137CrossRefPubMed
11.
Zurück zum Zitat Decreto Legislativo 26 maggio 2000, n.187. Attuazione della direttiva 97/43/Euratom in materia di protezione sanitaria delle persone contro i pericoli delle radiazioni ionizzanti connesse ad esposizioni mediche (Legislative Decree n.187 of May 26, 2000. Implementation of 97/43/Euratom directive on individual health protection against ionizing radiations due to medical exposure). Gazzetta Ufficiale n.157 del 7 luglio 2000, suppl. ordinario n.105. http://people.unica.it/radioprotezione/files/2009/11/DLgs_187_26052000.pdf. Accessed 7 Oct 2014 Decreto Legislativo 26 maggio 2000, n.187. Attuazione della direttiva 97/43/Euratom in materia di protezione sanitaria delle persone contro i pericoli delle radiazioni ionizzanti connesse ad esposizioni mediche (Legislative Decree n.187 of May 26, 2000. Implementation of 97/43/Euratom directive on individual health protection against ionizing radiations due to medical exposure). Gazzetta Ufficiale n.157 del 7 luglio 2000, suppl. ordinario n.105. http://​people.​unica.​it/​radioprotezione/​files/​2009/​11/​DLgs_​187_​26052000.​pdf. Accessed 7 Oct 2014
12.
Zurück zum Zitat Società Italiana di Radiologia Medica (SIRM) [Italian Society of Medical Radiology], Società di Ricerca per l’Organizzazione Sanitaria (Research Society for Health Care Organization)(2010) Censimento Nazionale delle risorse umane e tecnologiche dell’area radiologica (National Survey of human and technological resources of the radiological area) Il Radiologo s2:3–39 Società Italiana di Radiologia Medica (SIRM) [Italian Society of Medical Radiology], Società di Ricerca per l’Organizzazione Sanitaria (Research Society for Health Care Organization)(2010) Censimento Nazionale delle risorse umane e tecnologiche dell’area radiologica (National Survey of human and technological resources of the radiological area) Il Radiologo s2:3–39
15.
Zurück zum Zitat Palorini F, Origgi D, Granata C et al (2014) Adult exposure from MDCT including multiphase studies: first Italian nationwide survey. Eur Radiol 24:469–483 Palorini F, Origgi D, Granata C et al (2014) Adult exposure from MDCT including multiphase studies: first Italian nationwide survey. Eur Radiol 24:469–483
16.
Zurück zum Zitat Shrimpton PC, Hillier MC, Lewis MA et al (2006) National survey of doses from CT in the UK: 2003. Br J Radiol 79:968–980CrossRefPubMed Shrimpton PC, Hillier MC, Lewis MA et al (2006) National survey of doses from CT in the UK: 2003. Br J Radiol 79:968–980CrossRefPubMed
17.
Zurück zum Zitat Verdun FR, Gutierrez D, Vader JP et al (2008) CT radiation dose in children: a survey to establish age-based diagnostic reference levels in Switzerland. Eur Radiol 18:1980–1986CrossRefPubMed Verdun FR, Gutierrez D, Vader JP et al (2008) CT radiation dose in children: a survey to establish age-based diagnostic reference levels in Switzerland. Eur Radiol 18:1980–1986CrossRefPubMed
18.
Zurück zum Zitat Brisse HJ, Aubert B (2009) CT exposure from pediatric MDCT: results from the 2007–2008 SFIPP/ISRN survey. J Radiol 90:207–215CrossRefPubMed Brisse HJ, Aubert B (2009) CT exposure from pediatric MDCT: results from the 2007–2008 SFIPP/ISRN survey. J Radiol 90:207–215CrossRefPubMed
19.
Zurück zum Zitat Yakoumakis E, Karlatira M, Gialousis G et al (2009) Effective dose variation in pediatric computed tomography: dose reference levels in Greece. Health Phys 97:595–603CrossRefPubMed Yakoumakis E, Karlatira M, Gialousis G et al (2009) Effective dose variation in pediatric computed tomography: dose reference levels in Greece. Health Phys 97:595–603CrossRefPubMed
21.
Zurück zum Zitat Goske MJ, Strauss KJ, Coombs LP et al (2013) Diagnostic reference ranges for pediatric abdominal CT. Radiology 268:208–218CrossRefPubMed Goske MJ, Strauss KJ, Coombs LP et al (2013) Diagnostic reference ranges for pediatric abdominal CT. Radiology 268:208–218CrossRefPubMed
22.
Zurück zum Zitat Kleinman PL, Strauss KJ, Zurakowski D et al (2010) Patient size measured on CT images as a function of age at a tertiary care children’s hospital. AJR Am J Roentgenol 194:1611–1619CrossRefPubMed Kleinman PL, Strauss KJ, Zurakowski D et al (2010) Patient size measured on CT images as a function of age at a tertiary care children’s hospital. AJR Am J Roentgenol 194:1611–1619CrossRefPubMed
23.
Zurück zum Zitat Rizzo S, Kalra M, Schmidt B et al (2006) Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis. AJR Am J Roentgenol 186:673–679CrossRefPubMed Rizzo S, Kalra M, Schmidt B et al (2006) Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis. AJR Am J Roentgenol 186:673–679CrossRefPubMed
24.
Zurück zum Zitat Papadakis AE, Perisinakis K, Damilakis J (2008) Automatic exposure control in paediatric and adult multidetector CT examinations: a phantom study on dose reduction and image quality. Med Phys 35:4567–4576CrossRefPubMed Papadakis AE, Perisinakis K, Damilakis J (2008) Automatic exposure control in paediatric and adult multidetector CT examinations: a phantom study on dose reduction and image quality. Med Phys 35:4567–4576CrossRefPubMed
25.
Zurück zum Zitat Mulkens TH, Bellinck P, Baeyaert M et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237:213–223CrossRefPubMed Mulkens TH, Bellinck P, Baeyaert M et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237:213–223CrossRefPubMed
26.
Zurück zum Zitat Sutton DG, McVey S, Gentle D et al (2014) CT chest abdomen pelvis doses in Scotland: has the DRL had its day? Br J Radiol 87:20140157CrossRefPubMed Sutton DG, McVey S, Gentle D et al (2014) CT chest abdomen pelvis doses in Scotland: has the DRL had its day? Br J Radiol 87:20140157CrossRefPubMed
Metadaten
Titel
Radiation dose from multidetector CT studies in children: results from the first Italian nationwide survey
verfasst von
Claudio Granata
Daniela Origgi
Federica Palorini
Domenica Matranga
Sergio Salerno
Publikationsdatum
01.05.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 5/2015
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-014-3201-z

Weitere Artikel der Ausgabe 5/2015

Pediatric Radiology 5/2015 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.