Retrospective comparison of gradient recalled echo R2* and spin-echo R2 magnetic resonance analysis methods for estimating liver iron content in children and adolescents
verfasst von:
Suraj D. Serai, Robert J. Fleck, Charles T. Quinn, Bin Zhang, Daniel J. Podberesky
Erschienen in:
Pediatric Radiology
|
Ausgabe 11/2015
Einloggen, um Zugang zu erhalten
Abstract
Background
Serial surveillance of liver iron concentration (LIC) provides guidance for chelation therapy in patients with iron overload. The diagnosis of iron overload traditionally relies on core liver biopsy, which is limited by invasiveness, sampling error, cost and general poor acceptance by pediatric patients and parents. Thus noninvasive diagnostic methods such as MRI are highly attractive for quantification of liver iron concentration.
Objective
To compare two MRI-based methods for liver iron quantification in children.
Materials and methods
64 studies on 48 children and young adults (age range 4–21 years) were examined by gradient recalled echo (GRE) R2* and spin-echo R2 MRI at 1.5T to evaluate liver iron concentration. Scatter plots and Bland–Altman difference plots were generated to display and assess the relationship between the methods.
Results
With the protocols used in this investigation, Bland–Altman agreement between the methods is best when LIC is <20 mg/g dry tissue. Scatter plots show that all values with LIC <20 mg/g dry tissue fall within the 95% prediction limits.
Conclusion
Liver iron concentration as determined by the R2* and R2 MR methods is statistically comparable, with no statistical difference between these methods for LIC <20 mg/g.