Objective

The relationship between metal wear debris, pseudotumor formation and metal hypersensitivity is complex and not completely understood. The purpose of this study was to assess the prevalence of pseudotumor formation in a consecutive series of metal-on-metal (MoM) total hip arthroplasty (THA) and to investigate its relationship to serum metal-ion levels and hypersensitivity to metal.

Methods

Forty-one patients (31 males), mean age 52 (28–68) years, with a total of 49 large-head MoM THA participated in a 5–7-year follow-up study. Patients underwent ultrasonography, serum metal-ion concentrations were measured, metal allergy and atopic dermatitis were evaluated, and the questionnaires of the Oxford Hip Score (OHS), Harris Hip Score (HHS) and the Short-Form Health Survey (SF-36) were completed.

Results

Pseudotumors were found in eight patients, but they were asymptomatic and their serum metal-ion levels were similar to those observed in patients with no pseudotumors (p > 0.36). The capsule-stem distance of mean 8.6 mm (SD 3.82, 95 % CI: 5.40–11.79) was wider (p = 0.02) in patients with pseudotumours than in patients without pseudotumors of mean 5.6 mm (SD 2.89, 95 % CI: 4.68–6.58). Positive patch test reactions were seen in three patients. Higher serum metal-ion levels of chromium and cobalt were significantly correlated with steeper cup inclination and smaller femoral head sizes, and were associated with female gender (p < 0.04).

Conclusion

We found no association between pseudotumor formation, serum metal-ion levels, metal patch test reactivity, and atopic dermatitis. However, clinicians should be aware of asymptomatic pseudotumors, and we advise further exploration into the mechanisms involved in the pathogenesis of pseudotumors.
Literatur
1.
Zurück zum Zitat Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93(5):608–15.CrossRefPubMed Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93(5):608–15.CrossRefPubMed
2.
Zurück zum Zitat Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94(7):895–900.CrossRefPubMed Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94(7):895–900.CrossRefPubMed
3.
Zurück zum Zitat Malviya A, Ramaskandhan JR, Bowman R, Hashmi M, Holland JP, Kometa S, et al. What advantage is there to be gained using large modular metal-on-metal bearings in routine primary hip replacement? A preliminary report of a prospective randomised controlled trial. J Bone Joint Surg Br. 2011;93(12):1602–9.CrossRefPubMed Malviya A, Ramaskandhan JR, Bowman R, Hashmi M, Holland JP, Kometa S, et al. What advantage is there to be gained using large modular metal-on-metal bearings in routine primary hip replacement? A preliminary report of a prospective randomised controlled trial. J Bone Joint Surg Br. 2011;93(12):1602–9.CrossRefPubMed
4.
Zurück zum Zitat Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90(7):847–51.CrossRefPubMed Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90(7):847–51.CrossRefPubMed
6.
Zurück zum Zitat Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, et al. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87(1):28–36.CrossRefPubMed Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, et al. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87(1):28–36.CrossRefPubMed
7.
Zurück zum Zitat Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92(1):38–46.CrossRefPubMed Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92(1):38–46.CrossRefPubMed
9.
Zurück zum Zitat Lombardi Jr AV, Barrack RL, Berend KR, Cuckler JM, Jacobs JJ, Mont MA, et al. The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br. 2012;94(11 Suppl A):14–8.CrossRefPubMed Lombardi Jr AV, Barrack RL, Berend KR, Cuckler JM, Jacobs JJ, Mont MA, et al. The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br. 2012;94(11 Suppl A):14–8.CrossRefPubMed
10.
Zurück zum Zitat Sofka CM, Potter HG, Adler RS, Pavlov H. Musculoskeletal imaging update: current applications of advanced imaging techniques to evaluate the early and long-term complications of patients with orthopedic implants. HSS J. 2006;2(1):73–7.PubMedCentralCrossRefPubMed Sofka CM, Potter HG, Adler RS, Pavlov H. Musculoskeletal imaging update: current applications of advanced imaging techniques to evaluate the early and long-term complications of patients with orthopedic implants. HSS J. 2006;2(1):73–7.PubMedCentralCrossRefPubMed
11.
Zurück zum Zitat Bosker BH, Ettema HB, Boomsma MF, Kollen BJ, Maas M, Verheyen CC. High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br. 2012;94(6):755–61.CrossRefPubMed Bosker BH, Ettema HB, Boomsma MF, Kollen BJ, Maas M, Verheyen CC. High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br. 2012;94(6):755–61.CrossRefPubMed
12.
Zurück zum Zitat Hart AJ, Satchithananda K, Liddle AD, Sabah SA, McRobbie D, Henckel J, et al. Pseudotumors in association with well-functioning metal-on-metal hip prostheses: a case-control study using three-dimensional computed tomography and magnetic resonance imaging. J Bone Joint Surg Am. 2012;94(4):317–25.PubMed Hart AJ, Satchithananda K, Liddle AD, Sabah SA, McRobbie D, Henckel J, et al. Pseudotumors in association with well-functioning metal-on-metal hip prostheses: a case-control study using three-dimensional computed tomography and magnetic resonance imaging. J Bone Joint Surg Am. 2012;94(4):317–25.PubMed
13.
Zurück zum Zitat Chang EY, McAnally JL, Van Horne JR, Statum S, Wolfson T, Gamst A, et al. Metal-on-metal total hip arthroplasty: do symptoms correlate with MR imaging findings? Radiology. 2012;265(3):848–57.CrossRefPubMed Chang EY, McAnally JL, Van Horne JR, Statum S, Wolfson T, Gamst A, et al. Metal-on-metal total hip arthroplasty: do symptoms correlate with MR imaging findings? Radiology. 2012;265(3):848–57.CrossRefPubMed
14.
Zurück zum Zitat Donell ST, Darrah C, Nolan JF, Wimhurst J, Toms A, Barker TH, et al. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. 2010;92(11):1501–8.CrossRefPubMed Donell ST, Darrah C, Nolan JF, Wimhurst J, Toms A, Barker TH, et al. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. 2010;92(11):1501–8.CrossRefPubMed
15.
Zurück zum Zitat Matthies AK, Skinner JA, Osmani H, Henckel J, Hart AJ. Pseudotumors are common in well-positioned low-wearing metal-on-metal hips. Clin Orthop Relat Res. 2012;470(7):1895–906.PubMedCentralCrossRefPubMed Matthies AK, Skinner JA, Osmani H, Henckel J, Hart AJ. Pseudotumors are common in well-positioned low-wearing metal-on-metal hips. Clin Orthop Relat Res. 2012;470(7):1895–906.PubMedCentralCrossRefPubMed
16.
Zurück zum Zitat Thomas P, Braathen LR, Dorig M, Aubock J, Nestle F, Werfel T, et al. Increased metal allergy in patients with failed metal-on-metal hip arthroplasty and peri-implant T-lymphocytic inflammation. Allergy. 2009;64(8):1157–65.CrossRefPubMed Thomas P, Braathen LR, Dorig M, Aubock J, Nestle F, Werfel T, et al. Increased metal allergy in patients with failed metal-on-metal hip arthroplasty and peri-implant T-lymphocytic inflammation. Allergy. 2009;64(8):1157–65.CrossRefPubMed
17.
Zurück zum Zitat Malajian D, Belsito DV. Cutaneous delayed-type hypersensitivity in patients with atopic dermatitis. J Am Acad Dermatol. 2013;69(2):232–7.CrossRefPubMed Malajian D, Belsito DV. Cutaneous delayed-type hypersensitivity in patients with atopic dermatitis. J Am Acad Dermatol. 2013;69(2):232–7.CrossRefPubMed
19.
Zurück zum Zitat Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, et al. The use of the Oxford hip and knee scores. J Bone Joint Surg Br. 2007;89(8):1010–4.CrossRefPubMed Murray DW, Fitzpatrick R, Rogers K, Pandit H, Beard DJ, Carr AJ, et al. The use of the Oxford hip and knee scores. J Bone Joint Surg Br. 2007;89(8):1010–4.CrossRefPubMed
20.
Zurück zum Zitat Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty—an end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51(4):737–55.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty—an end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51(4):737–55.PubMed
21.
Zurück zum Zitat Bjorner JB, Thunedborg K, Kristensen TS, Modvig J, Bech P. The Danish SF-36 Health Survey: translation and preliminary validity studies. J Clin Epidemiol. 1998;51(11):991–9.CrossRefPubMed Bjorner JB, Thunedborg K, Kristensen TS, Modvig J, Bech P. The Danish SF-36 Health Survey: translation and preliminary validity studies. J Clin Epidemiol. 1998;51(11):991–9.CrossRefPubMed
22.
Zurück zum Zitat Craig JG. Ultrasound of the postoperative hip. Semin Musculoskelet Radiol. 2013;17(1):49–55.CrossRefPubMed Craig JG. Ultrasound of the postoperative hip. Semin Musculoskelet Radiol. 2013;17(1):49–55.CrossRefPubMed
23.
Zurück zum Zitat Long SS, Surrey D, Nazarian LN. Common sonographic findings in the painful hip after hip arthroplasty. J Ultrasound Med. 2012;31(2):301–12.PubMed Long SS, Surrey D, Nazarian LN. Common sonographic findings in the painful hip after hip arthroplasty. J Ultrasound Med. 2012;31(2):301–12.PubMed
24.
Zurück zum Zitat Hailer NP, Bengtsson M, Lundberg C, Milbrink J. High metal ion levels after use of the ASR device correlate with development of pseudotumors and T cell activation. Clin Orthop Relat Res. 2014;472(3):953–61. Hailer NP, Bengtsson M, Lundberg C, Milbrink J. High metal ion levels after use of the ASR device correlate with development of pseudotumors and T cell activation. Clin Orthop Relat Res. 2014;472(3):953–61.
25.
Zurück zum Zitat Cervini P, Hesley GK, Thompson RL, Sampathkumar P, Knudsen JM. Incidence of infectious complications after an ultrasound-guided intervention. AJR Am J Roentgenol. 2010;195(4):846–50.CrossRefPubMed Cervini P, Hesley GK, Thompson RL, Sampathkumar P, Knudsen JM. Incidence of infectious complications after an ultrasound-guided intervention. AJR Am J Roentgenol. 2010;195(4):846–50.CrossRefPubMed
26.
Zurück zum Zitat Dramis A, Aldlyami E, Grimer RJ, Dunlop DJ, O’Connell N, Elliott T. What is the significance of a positive Propionibacterium acnes culture around a joint replacement? Int Orthop. 2009;33(3):829–33.PubMedCentralCrossRefPubMed Dramis A, Aldlyami E, Grimer RJ, Dunlop DJ, O’Connell N, Elliott T. What is the significance of a positive Propionibacterium acnes culture around a joint replacement? Int Orthop. 2009;33(3):829–33.PubMedCentralCrossRefPubMed
28.
Zurück zum Zitat Thyssen JP, Linneberg A, Menne T, Nielsen NH, Johansen JD. The association between hand eczema and nickel allergy has weakened among young women in the general population following the Danish nickel regulation: results from two cross-sectional studies. Contact Dermatitis. 2009;61(6):342–8.CrossRefPubMed Thyssen JP, Linneberg A, Menne T, Nielsen NH, Johansen JD. The association between hand eczema and nickel allergy has weakened among young women in the general population following the Danish nickel regulation: results from two cross-sectional studies. Contact Dermatitis. 2009;61(6):342–8.CrossRefPubMed
29.
Zurück zum Zitat Hallab N, Merritt K, Jacobs JJ. Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am. 2001;83-A(3):428–36.PubMed Hallab N, Merritt K, Jacobs JJ. Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am. 2001;83-A(3):428–36.PubMed
30.
Zurück zum Zitat MacDonald SJ, Brodner W, Jacobs JJ. A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty. 2004;19(8 Suppl 3):12–6.CrossRefPubMed MacDonald SJ, Brodner W, Jacobs JJ. A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty. 2004;19(8 Suppl 3):12–6.CrossRefPubMed
31.
Zurück zum Zitat Penny JO, Overgaard S. Serum chromium levels sampled with steel needle versus plastic IV cannula: does method matter? J Biomed Mater Res B Appl Biomater. 2010;92(1):1–4.CrossRefPubMed Penny JO, Overgaard S. Serum chromium levels sampled with steel needle versus plastic IV cannula: does method matter? J Biomed Mater Res B Appl Biomater. 2010;92(1):1–4.CrossRefPubMed
32.
Zurück zum Zitat Papageorgiou I, Yin Z, Ladon D, Baird D, Lewis AC, Sood A, et al. Genotoxic effects of particles of surgical cobalt chrome alloy on human cells of different age in vitro. Mutat Res. 2007;619(1–2):45–58.CrossRefPubMed Papageorgiou I, Yin Z, Ladon D, Baird D, Lewis AC, Sood A, et al. Genotoxic effects of particles of surgical cobalt chrome alloy on human cells of different age in vitro. Mutat Res. 2007;619(1–2):45–58.CrossRefPubMed
33.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
34.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
35.
Zurück zum Zitat Brooker AF, Bowerman JW, Robinson RA, Riley Jr LH. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg Am. 1973;55(8):1629–32.PubMed Brooker AF, Bowerman JW, Robinson RA, Riley Jr LH. Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg Am. 1973;55(8):1629–32.PubMed
36.
Zurück zum Zitat Altman DG. Practical statistics for medical research. London: Chapman and Hall, 2009. ISBN 9780412276309 Altman DG. Practical statistics for medical research. London: Chapman and Hall, 2009. ISBN 9780412276309
37.
Zurück zum Zitat Nishii T, Sakai T, Takao M, Yoshikawa H, Sugano N. Ultrasound screening of periarticular soft tissue abnormality around metal-on-metal bearings. J Arthroplasty. 2012;27(6):895–900.CrossRefPubMed Nishii T, Sakai T, Takao M, Yoshikawa H, Sugano N. Ultrasound screening of periarticular soft tissue abnormality around metal-on-metal bearings. J Arthroplasty. 2012;27(6):895–900.CrossRefPubMed
38.
Zurück zum Zitat Williams DH, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. Prevalence of pseudotumor in asymptomatic patients after metal-on-metal hip arthroplasty. J Bone Joint Surg Am. 2011;93(23):2164–71.CrossRefPubMed Williams DH, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. Prevalence of pseudotumor in asymptomatic patients after metal-on-metal hip arthroplasty. J Bone Joint Surg Am. 2011;93(23):2164–71.CrossRefPubMed
39.
Zurück zum Zitat Wynn-Jones H, Macnair R, Wimhurst J, Chirodian N, Derbyshire B, Toms A, et al. Silent soft tissue pathology is common with a modern metal-on-metal hip arthroplasty. Acta Orthop. 2011;82(3):301–7.PubMedCentralCrossRefPubMed Wynn-Jones H, Macnair R, Wimhurst J, Chirodian N, Derbyshire B, Toms A, et al. Silent soft tissue pathology is common with a modern metal-on-metal hip arthroplasty. Acta Orthop. 2011;82(3):301–7.PubMedCentralCrossRefPubMed
40.
Zurück zum Zitat Siddiqui IA, Sabah SA, Satchithananda K, Lim AK, Henckel J, Skinner JA, et al. Cross-sectional imaging of the metal-on-metal hip prosthesis: the London ultrasound protocol. Clin Radiol. 2013;68(8):e472–8.CrossRefPubMed Siddiqui IA, Sabah SA, Satchithananda K, Lim AK, Henckel J, Skinner JA, et al. Cross-sectional imaging of the metal-on-metal hip prosthesis: the London ultrasound protocol. Clin Radiol. 2013;68(8):e472–8.CrossRefPubMed
41.
Zurück zum Zitat van Holsbeeck MT et al. Detection of infection in loosened hip prostheses: efficacy of sonography. AJR Am J Roentgenol. 1994;163:381–4.CrossRefPubMed van Holsbeeck MT et al. Detection of infection in loosened hip prostheses: efficacy of sonography. AJR Am J Roentgenol. 1994;163:381–4.CrossRefPubMed
42.
Zurück zum Zitat Glyn-Jones S, Pandit H, Kwon YM, Doll H, Gill HS, Murray DW. Risk factors for inflammatory pseudotumour formation following hip resurfacing. J Bone Joint Surg Br. 2009;91(12):1566–74.CrossRefPubMed Glyn-Jones S, Pandit H, Kwon YM, Doll H, Gill HS, Murray DW. Risk factors for inflammatory pseudotumour formation following hip resurfacing. J Bone Joint Surg Br. 2009;91(12):1566–74.CrossRefPubMed
43.
Zurück zum Zitat Thyssen JP, Linneberg A, Menne T, Johansen JD. The epidemiology of contact allergy in the general population: prevalence and main findings. Contact Dermatitis. 2007;57(5):287–99.CrossRefPubMed Thyssen JP, Linneberg A, Menne T, Johansen JD. The epidemiology of contact allergy in the general population: prevalence and main findings. Contact Dermatitis. 2007;57(5):287–99.CrossRefPubMed
44.
Zurück zum Zitat Thyssen JP, Linneberg A, Menne T, Nielsen NH, Johansen JD. Contact allergy to allergens of the TRUE-test (panels 1 and 2) has decreased modestly in the general population. Br J Dermatol. 2009;161(5):1124–9.CrossRefPubMed Thyssen JP, Linneberg A, Menne T, Nielsen NH, Johansen JD. Contact allergy to allergens of the TRUE-test (panels 1 and 2) has decreased modestly in the general population. Br J Dermatol. 2009;161(5):1124–9.CrossRefPubMed
45.
Zurück zum Zitat Thomas P, Stauner K, Schraml A, Mahler V, Banke IJ, Gollwitzer H, et al. Characteristics of 200 patients with suspected implant allergy compared to 100 symptom-free arthroplasty patients. Orthopade. 2013;42(8):607–13.CrossRefPubMed Thomas P, Stauner K, Schraml A, Mahler V, Banke IJ, Gollwitzer H, et al. Characteristics of 200 patients with suspected implant allergy compared to 100 symptom-free arthroplasty patients. Orthopade. 2013;42(8):607–13.CrossRefPubMed
46.
Zurück zum Zitat Jacobi-Gresser E, Huesker K, Schutt S. Genetic and immunological markers predict titanium implant failure: a retrospective study. Int J Oral Maxillofac Surg. 2013;42(4):537–43.CrossRefPubMed Jacobi-Gresser E, Huesker K, Schutt S. Genetic and immunological markers predict titanium implant failure: a retrospective study. Int J Oral Maxillofac Surg. 2013;42(4):537–43.CrossRefPubMed
47.
Zurück zum Zitat Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90(9):1143–51.CrossRefPubMed Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90(9):1143–51.CrossRefPubMed
48.
Zurück zum Zitat de Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De SK. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90(10):1291–7.CrossRefPubMed de Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De SK. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90(10):1291–7.CrossRefPubMed
49.
Zurück zum Zitat Hart AJ, Buddhdev P, Winship P, Faria N, Powell JJ, Skinner JA. Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int. 2008;18(3):212–9.PubMed Hart AJ, Buddhdev P, Winship P, Faria N, Powell JJ, Skinner JA. Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int. 2008;18(3):212–9.PubMed
50.
Zurück zum Zitat De HR, Campbell P, Reid S, Skipor AK, De SK. Metal ion levels in a triathlete with a metal-on-metal resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2007;89(4):538–41. De HR, Campbell P, Reid S, Skipor AK, De SK. Metal ion levels in a triathlete with a metal-on-metal resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2007;89(4):538–41.
51.
Zurück zum Zitat Pattyn CA, Lauwagie SN, Verdonk RC. Whole blood metal ion concentrations in correlation with activity level in three different metal-on-metal bearings. J Arthroplasty. 2011;26(1):58–64.CrossRefPubMed Pattyn CA, Lauwagie SN, Verdonk RC. Whole blood metal ion concentrations in correlation with activity level in three different metal-on-metal bearings. J Arthroplasty. 2011;26(1):58–64.CrossRefPubMed
Metadaten
Titel
No association between pseudotumors, high serum metal-ion levels and metal hypersensitivity in large-head metal-on-metal total hip arthroplasty at 5–7-year follow-up
verfasst von
Mette Holm Hjorth
Maiken Stilling
Kjeld Soballe
Lars Hans Bolvig
Jacob Pontoppidan Thyssen
Inger Mechlenburg
Stig Storgaard Jakobsen
Publikationsdatum
01.01.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 1/2016
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-015-2264-8

Weitere Artikel der Ausgabe 1/2016

Skeletal Radiology 1/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.