Skip to main content
Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging 10/2013

01.10.2013 | Original Article

Observer reproducibility of results from a low-dose 123I-metaiodobenzylguanidine cardiac imaging protocol in patients with heart failure

verfasst von: Teresa Pellegrino, Mario Petretta, Serena De Luca, Stefania Paolillo, Antonio Boemio, Raffaella Carotenuto, Maria Piera Petretta, Carmine di Nuzzo, Pasquale Perrone-Filardi, Alberto Cuocolo

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 10/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Imaging of the cardiac autonomic system with 123I-metaiodobenzylguanidine (MIBG) is useful in the evaluation of patients with a variety of cardiac conditions, including heart failure (HF), but few data are available about the reproducibility of this technique. We assessed the observer reproducibility of the results from a low-dose 123I-MIBG cardiac imaging protocol in patients with HF.

Methods

A total of 74 patients (62 men, age 67 ± 10 years) with HF and left ventricular systolic dysfunction (ejection fraction 31 ± 7 %) underwent low-dose (111 MBq) planar and single-photon emission computed tomography (SPECT) 123I-MIBG cardiac sympathetic imaging. The intraclass coefficient of correlation (ICC), Lin’s concordance correlation coefficient, and Bland-Altman analysis were used to evaluate the intraobserver and interobserver reproducibility of early and late heart-to-mediastinum (H/M) ratios and of defect scores on SPECT images. The κ statistic was used to evaluate the concordance rates for the identification of patients with a low H/M ratio (<1.60) on late planar imaging.

Results

The differences between measurements obtained twice by the same examiner and by two examiners were negligible for both early and late H/M ratios and for SPECT defect scores. These findings were confirmed by the results of Bland-Altman analysis, and ICC and Lin’s coefficient values were excellent (>0.90) for all measurements. For the identification of patients with a low H/M ratio, the κ values were 0.90 for intraobserver concordance and 0.83 for interobserver concordance.

Conclusion

The present study showed a high observer reproducibility of planar H/M ratios and SPECT defect scores using a low-dose 123I-MIBG cardiac imaging protocol in patients with HF.
Literatur
1.
Zurück zum Zitat Triposkiadis F, Karayannis G, Giamouzis G, Skoularigis J, Louridas G, Butler J. The sympathetic nervous system in heart failure physiology, pathophysiology, and clinical implications. J Am Coll Cardiol. 2009;54:1747–62.PubMedCrossRef Triposkiadis F, Karayannis G, Giamouzis G, Skoularigis J, Louridas G, Butler J. The sympathetic nervous system in heart failure physiology, pathophysiology, and clinical implications. J Am Coll Cardiol. 2009;54:1747–62.PubMedCrossRef
2.
Zurück zum Zitat Carrió I, Cowie MR, Yamazaki J, Udelson J, Camici PG. Cardiac sympathetic imaging with mIBG in heart failure. JACC Cardiovasc Imaging. 2010;3:92–100.PubMedCrossRef Carrió I, Cowie MR, Yamazaki J, Udelson J, Camici PG. Cardiac sympathetic imaging with mIBG in heart failure. JACC Cardiovasc Imaging. 2010;3:92–100.PubMedCrossRef
3.
Zurück zum Zitat Perrone-Filardi P, Paolillo S, Dellegrottaglie S, Gargiulo P, Savarese G, Marciano C, et al. Assessment of cardiac sympathetic activity by MIBG imaging in patients with heart failure: a clinical appraisal. Heart. 2011;97:1828–33.PubMedCrossRef Perrone-Filardi P, Paolillo S, Dellegrottaglie S, Gargiulo P, Savarese G, Marciano C, et al. Assessment of cardiac sympathetic activity by MIBG imaging in patients with heart failure: a clinical appraisal. Heart. 2011;97:1828–33.PubMedCrossRef
4.
Zurück zum Zitat Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef
5.
Zurück zum Zitat Kuwabara Y, Tamaki N, Nakata T, Yamashina S, Yamazaki J. Determination of the survival rate in patients with congestive heart failure stratified by 123I-MIBG imaging: a meta-analysis from the studies performed in Japan. Ann Nucl Med. 2011;25:101–7.PubMedCrossRef Kuwabara Y, Tamaki N, Nakata T, Yamashina S, Yamazaki J. Determination of the survival rate in patients with congestive heart failure stratified by 123I-MIBG imaging: a meta-analysis from the studies performed in Japan. Ann Nucl Med. 2011;25:101–7.PubMedCrossRef
6.
Zurück zum Zitat Paul M, Wichter T, Kies P, Gerss J, Wollmann C, Rahbar K, et al. Cardiac sympathetic dysfunction in genotyped patients with arrhythmogenic right ventricular cardiomyopathy and risk of recurrent ventricular tachyarrhythmias. J Nucl Med. 2011;52:1559–65.PubMedCrossRef Paul M, Wichter T, Kies P, Gerss J, Wollmann C, Rahbar K, et al. Cardiac sympathetic dysfunction in genotyped patients with arrhythmogenic right ventricular cardiomyopathy and risk of recurrent ventricular tachyarrhythmias. J Nucl Med. 2011;52:1559–65.PubMedCrossRef
7.
Zurück zum Zitat Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedCrossRef Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedCrossRef
8.
Zurück zum Zitat Flotats A, Carrió I, Agostini D, Le Guludec D, Marcassa C, Schäfers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef Flotats A, Carrió I, Agostini D, Le Guludec D, Marcassa C, Schäfers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef
9.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.PubMedCrossRef Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.PubMedCrossRef
10.
Zurück zum Zitat de Vet HC, Terwee CB, Knol DL, Bouter LM. When to use agreement versus reliability measures. J Clin Epidemiol. 2006;59:1033–9.PubMedCrossRef de Vet HC, Terwee CB, Knol DL, Bouter LM. When to use agreement versus reliability measures. J Clin Epidemiol. 2006;59:1033–9.PubMedCrossRef
11.
Zurück zum Zitat Bonett DG. Sample size requirements for estimating intraclass correlations with desired precision. Stat Med. 2002;21:1331–5.PubMedCrossRef Bonett DG. Sample size requirements for estimating intraclass correlations with desired precision. Stat Med. 2002;21:1331–5.PubMedCrossRef
13.
14.
Zurück zum Zitat Radiation dose to patients from radiopharmaceuticals (addendum 2 to ICRP publication 53). Ann ICRP. 1998; 28:1–126. Radiation dose to patients from radiopharmaceuticals (addendum 2 to ICRP publication 53). Ann ICRP. 1998; 28:1–126.
15.
Zurück zum Zitat Kottner J, Audigé L, Brorson S, Donner A, Gajewski BJ, Hróbjartsson A, et al. Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol. 2011;64:96–106.PubMedCrossRef Kottner J, Audigé L, Brorson S, Donner A, Gajewski BJ, Hróbjartsson A, et al. Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epidemiol. 2011;64:96–106.PubMedCrossRef
16.
Zurück zum Zitat Watson PF, Petrie A. Method agreement analysis: a review of correct methodology. Theriogenology. 2010;73:1167–79.PubMedCrossRef Watson PF, Petrie A. Method agreement analysis: a review of correct methodology. Theriogenology. 2010;73:1167–79.PubMedCrossRef
17.
Zurück zum Zitat Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef
18.
19.
Zurück zum Zitat Shoukri MM. Measures of interobserver agreement. Boca Raton: Chapman & Hall/CRC; 2004. Shoukri MM. Measures of interobserver agreement. Boca Raton: Chapman & Hall/CRC; 2004.
20.
Zurück zum Zitat Batterham AM, George KP. Reliability in evidence-based clinical practice: a primer for allied health professionals. Phys Ther Sport. 2000;1:54–62.CrossRef Batterham AM, George KP. Reliability in evidence-based clinical practice: a primer for allied health professionals. Phys Ther Sport. 2000;1:54–62.CrossRef
21.
Zurück zum Zitat Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19:231–40.PubMed Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19:231–40.PubMed
22.
Zurück zum Zitat Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45:255–68.PubMedCrossRef Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45:255–68.PubMedCrossRef
23.
Zurück zum Zitat Germano G, Kavanagh PB, Slomka PJ, Berman DS. Tracking a therapeutic response: how reliable are serial measurements of LV perfusion and function? J Nucl Cardiol. 2012;19:360–3.PubMedCrossRef Germano G, Kavanagh PB, Slomka PJ, Berman DS. Tracking a therapeutic response: how reliable are serial measurements of LV perfusion and function? J Nucl Cardiol. 2012;19:360–3.PubMedCrossRef
24.
Zurück zum Zitat Matsui T, Tsutamoto T, Maeda K, Kusukawa J, Kinoshita M. Prognostic value of repeated 123I-metaiodobenzylguanidine imaging in patients with dilated cardiomyopathy with congestive heart failure before and after optimized treatments – comparison with neurohumoral factors. Circ J. 2002;66:537–43.PubMedCrossRef Matsui T, Tsutamoto T, Maeda K, Kusukawa J, Kinoshita M. Prognostic value of repeated 123I-metaiodobenzylguanidine imaging in patients with dilated cardiomyopathy with congestive heart failure before and after optimized treatments – comparison with neurohumoral factors. Circ J. 2002;66:537–43.PubMedCrossRef
25.
Zurück zum Zitat Kasama S, Toyama T, Sumino H, Nakazawa M, Matsumoto N, Sato Y, et al. Prognostic value of serial cardiac 123I-MIBG imaging in patients with stabilized chronic heart failure and reduced left ventricular ejection fraction. J Nucl Med. 2008;49:907–14.PubMedCrossRef Kasama S, Toyama T, Sumino H, Nakazawa M, Matsumoto N, Sato Y, et al. Prognostic value of serial cardiac 123I-MIBG imaging in patients with stabilized chronic heart failure and reduced left ventricular ejection fraction. J Nucl Med. 2008;49:907–14.PubMedCrossRef
26.
Zurück zum Zitat Drakos SG, Athanasoulis T, Malliaras KG, Terrovitis JV, Diakos N, Koudoumas D, et al. Myocardial sympathetic innervation and long-term left ventricular mechanical unloading. JACC Cardiovasc Imaging. 2010;3:64–70.PubMedCrossRef Drakos SG, Athanasoulis T, Malliaras KG, Terrovitis JV, Diakos N, Koudoumas D, et al. Myocardial sympathetic innervation and long-term left ventricular mechanical unloading. JACC Cardiovasc Imaging. 2010;3:64–70.PubMedCrossRef
27.
Zurück zum Zitat Somsen GA, Verberne HJ, Fleury E, Righetti A. Normal values and within-subject variability of cardiac I-123 MIBG scintigraphy in healthy individuals: implications for clinical studies. J Nucl Cardiol. 2004;11:126–33.PubMedCrossRef Somsen GA, Verberne HJ, Fleury E, Righetti A. Normal values and within-subject variability of cardiac I-123 MIBG scintigraphy in healthy individuals: implications for clinical studies. J Nucl Cardiol. 2004;11:126–33.PubMedCrossRef
28.
Zurück zum Zitat Estorch M, Serra-Grima R, Flotats A, Marí C, Bernà L, Catafau A, et al. Myocardial sympathetic innervation in the athlete’s sinus bradycardia: is there selective inferior myocardial wall denervation? J Nucl Cardiol. 2000;7:354–8.PubMedCrossRef Estorch M, Serra-Grima R, Flotats A, Marí C, Bernà L, Catafau A, et al. Myocardial sympathetic innervation in the athlete’s sinus bradycardia: is there selective inferior myocardial wall denervation? J Nucl Cardiol. 2000;7:354–8.PubMedCrossRef
29.
Zurück zum Zitat Matsuo S, Nakajima K, Okuda K, Kawano M, Ishikawa T, Hosoya T, et al. Standardization of the heart-to-mediastinum ratio of 123I-labelled-metaiodobenzylguanidine uptake using the dual energy window method: feasibility of correction with different camera collimator combinations. Eur J Nucl Med Mol Imaging. 2009;36:560–6.PubMedCrossRef Matsuo S, Nakajima K, Okuda K, Kawano M, Ishikawa T, Hosoya T, et al. Standardization of the heart-to-mediastinum ratio of 123I-labelled-metaiodobenzylguanidine uptake using the dual energy window method: feasibility of correction with different camera collimator combinations. Eur J Nucl Med Mol Imaging. 2009;36:560–6.PubMedCrossRef
30.
Zurück zum Zitat Verberne HJ, Feenstra C, de Jong WM, Somsen GA, van Eck-Smit BL, Busemann SE. Influence of collimator choice and simulated clinical conditions on 123I-MIBG heart/mediastinum ratios: a phantom study. Eur J Nucl Med Mol Imaging. 2005;32:1100–7.PubMedCrossRef Verberne HJ, Feenstra C, de Jong WM, Somsen GA, van Eck-Smit BL, Busemann SE. Influence of collimator choice and simulated clinical conditions on 123I-MIBG heart/mediastinum ratios: a phantom study. Eur J Nucl Med Mol Imaging. 2005;32:1100–7.PubMedCrossRef
31.
Zurück zum Zitat Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9.PubMedCrossRef Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9.PubMedCrossRef
32.
Zurück zum Zitat Boogers MJ, Borleffs CJ, Henneman MM, van Bommel RJ, van Ramshorst J, Boersma E, et al. Cardiac sympathetic denervation assessed with 123-iodine metaiodobenzylguanidine imaging predicts ventricular arrhythmias in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:2769–77.PubMedCrossRef Boogers MJ, Borleffs CJ, Henneman MM, van Bommel RJ, van Ramshorst J, Boersma E, et al. Cardiac sympathetic denervation assessed with 123-iodine metaiodobenzylguanidine imaging predicts ventricular arrhythmias in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:2769–77.PubMedCrossRef
Metadaten
Titel
Observer reproducibility of results from a low-dose 123I-metaiodobenzylguanidine cardiac imaging protocol in patients with heart failure
verfasst von
Teresa Pellegrino
Mario Petretta
Serena De Luca
Stefania Paolillo
Antonio Boemio
Raffaella Carotenuto
Maria Piera Petretta
Carmine di Nuzzo
Pasquale Perrone-Filardi
Alberto Cuocolo
Publikationsdatum
01.10.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 10/2013
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-013-2461-4

Weitere Artikel der Ausgabe 10/2013

European Journal of Nuclear Medicine and Molecular Imaging 10/2013 Zur Ausgabe