Development and validation of a direct-comparison method for cardiac 123I-metaiodobenzylguanidine washout rates derived from late 3-hour and 4-hour imaging

verfasst von: Koichi Okuda, Kenichi Nakajima, Shuichi Sugino, Yumiko Kirihara, Shinro Matsuo, Junichi Taki, Mitsumasa Hashimoto, Seigo Kinuya

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 2/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The washout rate (WR) has been used in 123I-metaiodobenzylguanidine (MIBG) imaging to evaluate cardiac sympathetic innervation. However, WR varies depending on the time between the early and late MIBG scans. Late scans are performed at either 3 or 4 hours after injection of MIBG. The aim of this study was to directly compare the WR at 3 hours (WR3h) with the WR at 4 hours (WR4h).

Methods

We hypothesized that the cardiac count would reduce linearly between the 3-hour and 4-hour scans. A linear regression model for cardiac counts at two time-points was generated. We enrolled a total of 96 patients who underwent planar 123I-MIBG scintigraphy early (15 min) and during the late phase at both 3 and 4 hours. Patients were randomly divided into two groups: a model-creation group (group 1) and a clinical validation group (group 2). Cardiac counts at 15 minutes (countearly), 3 hours (count3h) and 4 hours (count4h) were measured. Cardiac count4h was mathematically estimated using the linear regression model from countearly and count3h.

Results

In group 1, the actual cardiac count4h/countearly was highly significantly correlated with count3h/countearly (r = 0.979). In group 2, the average estimated count4h was 92.8 ± 31.9, and there was no significant difference between this value and the actual count4h (91.9 ± 31.9). Bland-Altman analysis revealed a small bias of −0.9 with 95 % limits of agreement of −6.2 and +4.3. WR4h calculated using the estimated cardiac count4h was comparable to the actual WR4h (24.3 ± 9.6 % vs. 25.1 ± 9.7 %, p = ns). Bland-Altman analysis and the intraclass correlation coefficient showed that there was excellent agreement between the estimated and actual WR4h.

Conclusion

The linear regression model that we used accurately estimated cardiac count4h using countearly and count3h. Moreover, WR4h that was mathematically calculated using the estimated count4h was comparable to the actual WR4h.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef
2.
Zurück zum Zitat Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of (123)I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.PubMedCrossRef Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of (123)I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.PubMedCrossRef
3.
Zurück zum Zitat Kashihara K, Ohno M, Kawada S, Okumura Y. Reduced cardiac uptake and enhanced washout of 123I-MIBG in pure autonomic failure occurs conjointly with Parkinson’s disease and dementia with Lewy bodies. J Nucl Med. 2006;47:1099–101.PubMed Kashihara K, Ohno M, Kawada S, Okumura Y. Reduced cardiac uptake and enhanced washout of 123I-MIBG in pure autonomic failure occurs conjointly with Parkinson’s disease and dementia with Lewy bodies. J Nucl Med. 2006;47:1099–101.PubMed
4.
Zurück zum Zitat Yoshita M, Arai H, Arai H, Arai T, Asada T, Fujishiro H, et al. Diagnostic accuracy of 123I-meta-iodobenzylguanidine myocardial scintigraphy in dementia with Lewy bodies: a multicenter study. PLoS ONE. 2015;10:e0120540. Yoshita M, Arai H, Arai H, Arai T, Asada T, Fujishiro H, et al. Diagnostic accuracy of 123I-meta-iodobenzylguanidine myocardial scintigraphy in dementia with Lewy bodies: a multicenter study. PLoS ONE. 2015;10:e0120540.
5.
Zurück zum Zitat Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9. Nakajima K, Okuda K, Matsuo S, Yoshita M, Taki J, Yamada M, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging. 2012;39:113–9.
6.
Zurück zum Zitat Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedPubMedCentralCrossRef Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (MIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedPubMedCentralCrossRef
7.
Zurück zum Zitat Nakajima K, Okuda K, Yoshimura M, Matsuo S, Wakabayashi H, Imanishi Y, et al. Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations. J Nucl Cardiol. 2014;21:970–8.PubMedPubMedCentralCrossRef Nakajima K, Okuda K, Yoshimura M, Matsuo S, Wakabayashi H, Imanishi Y, et al. Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations. J Nucl Cardiol. 2014;21:970–8.PubMedPubMedCentralCrossRef
8.
Zurück zum Zitat Hasegawa D, Onishi H, Matsutomo N, Kangai Y. Validation of an optimal analysis method and reproducibility to calculate the heart-to-mediastinum ratio and washout rate in the iodine-123-labeled metaiodobenzylguanidine myocardial scintigraphy. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2014;70:1420–8.PubMedCrossRef Hasegawa D, Onishi H, Matsutomo N, Kangai Y. Validation of an optimal analysis method and reproducibility to calculate the heart-to-mediastinum ratio and washout rate in the iodine-123-labeled metaiodobenzylguanidine myocardial scintigraphy. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2014;70:1420–8.PubMedCrossRef
9.
Zurück zum Zitat Verschure DO, de Wit TC, Bongers V, Hagen PJ, Sonneck-Koenne C, D’Aron J, et al. (123)I-MIBG heart-to-mediastinum ratio is influenced by high-energy photon penetration of collimator septa from liver and lung activity. Nucl Med Commun. 2015;36:279–85.PubMedCrossRef Verschure DO, de Wit TC, Bongers V, Hagen PJ, Sonneck-Koenne C, D’Aron J, et al. (123)I-MIBG heart-to-mediastinum ratio is influenced by high-energy photon penetration of collimator septa from liver and lung activity. Nucl Med Commun. 2015;36:279–85.PubMedCrossRef
10.
Zurück zum Zitat Okuda K, Nakajima K, Hosoya T, Ishikawa T, Konishi T, Matsubara K, et al. Semi-automated algorithm for calculating heart-to-mediastinum ratio in cardiac iodine-123 MIBG imaging. J Nucl Cardiol. 2011;18:82–9.PubMedCrossRef Okuda K, Nakajima K, Hosoya T, Ishikawa T, Konishi T, Matsubara K, et al. Semi-automated algorithm for calculating heart-to-mediastinum ratio in cardiac iodine-123 MIBG imaging. J Nucl Cardiol. 2011;18:82–9.PubMedCrossRef
11.
Zurück zum Zitat van der Veen L, Scholte A, Stokkel M. Mathematical methods to determine quantitative parameters of myocardial 123I-MIBG studies: a review of the literature. Nucl Med Commun. 2010;31:617–28.PubMed van der Veen L, Scholte A, Stokkel M. Mathematical methods to determine quantitative parameters of myocardial 123I-MIBG studies: a review of the literature. Nucl Med Commun. 2010;31:617–28.PubMed
12.
Zurück zum Zitat Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schafers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schafers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef
13.
Zurück zum Zitat Glowniak JV. Cardiac studies with metaiodobenzylguanidine: a critique of methods and interpretation of results. J Nucl Med. 1995;36:2133–7. Glowniak JV. Cardiac studies with metaiodobenzylguanidine: a critique of methods and interpretation of results. J Nucl Med. 1995;36:2133–7.
14.
Zurück zum Zitat Glowniak JV, Turner FE, Gray LL, Palac RT, Lagunas-Solar MC, Woodward WR. Iodine-123 metaiodobenzylguanidine imaging of the heart in idiopathic congestive cardiomyopathy and cardiac transplants. J Nucl Med. 1989;30:1182–91. Glowniak JV, Turner FE, Gray LL, Palac RT, Lagunas-Solar MC, Woodward WR. Iodine-123 metaiodobenzylguanidine imaging of the heart in idiopathic congestive cardiomyopathy and cardiac transplants. J Nucl Med. 1989;30:1182–91.
15.
Zurück zum Zitat Yoshita M, Taki J, Yokoyama K, Noguchi-Shinohara M, Matsumoto Y, Nakajima K, et al. Value of 123I-MIBG radioactivity in the differential diagnosis of DLB from AD. Neurology. 2006;66:1850–4. Yoshita M, Taki J, Yokoyama K, Noguchi-Shinohara M, Matsumoto Y, Nakajima K, et al. Value of 123I-MIBG radioactivity in the differential diagnosis of DLB from AD. Neurology. 2006;66:1850–4.
16.
Zurück zum Zitat Lorberboym M, Lampl Y, Nikolov G, Sadeh M, Gilad R. I-123 MIBG cardiac scintigraphy and autonomic test evaluation in multiple sclerosis patients. J Neurol. 2008;255:211–6. Lorberboym M, Lampl Y, Nikolov G, Sadeh M, Gilad R. I-123 MIBG cardiac scintigraphy and autonomic test evaluation in multiple sclerosis patients. J Neurol. 2008;255:211–6.
17.
Zurück zum Zitat Somsen GA, Verberne HJ, Fleury E, Righetti A. Normal values and within-subject variability of cardiac I-123 MIBG scintigraphy in healthy individuals: implications for clinical studies. J Nucl Cardiol. 2004;11:126–33. Somsen GA, Verberne HJ, Fleury E, Righetti A. Normal values and within-subject variability of cardiac I-123 MIBG scintigraphy in healthy individuals: implications for clinical studies. J Nucl Cardiol. 2004;11:126–33.
Metadaten
Titel
Development and validation of a direct-comparison method for cardiac 123I-metaiodobenzylguanidine washout rates derived from late 3-hour and 4-hour imaging
verfasst von
Koichi Okuda
Kenichi Nakajima
Shuichi Sugino
Yumiko Kirihara
Shinro Matsuo
Junichi Taki
Mitsumasa Hashimoto
Seigo Kinuya
Publikationsdatum
01.02.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 2/2016
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-015-3173-8

Weitere Artikel der Ausgabe 2/2016

European Journal of Nuclear Medicine and Molecular Imaging 2/2016 Zur Ausgabe