Open Access 04.05.2022 | Original Article

Deep learning–based time-of-flight (ToF) image enhancement of non-ToF PET scans

verfasst von: Abolfazl Mehranian, Scott D. Wollenweber, Matthew D. Walker, Kevin M. Bradley, Patrick A. Fielding, Martin Huellner, Fotis Kotasidis, Kuan-Hao Su, Robert Johnsen, Floris P. Jansen, Daniel R. McGowan

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 11/2022

Abstract

Purpose

To improve the quantitative accuracy and diagnostic confidence of PET images reconstructed without time-of-flight (ToF) using deep learning models trained for ToF image enhancement (DL-ToF).

Methods

A total of 273 [18F]-FDG PET scans were used, including data from 6 centres equipped with GE Discovery MI ToF scanners. PET data were reconstructed using the block-sequential-regularised-expectation–maximisation (BSREM) algorithm with and without ToF. The images were then split into training (n = 208), validation (n = 15), and testing (n = 50) sets. Three DL-ToF models were trained to transform non-ToF BSREM images to their target ToF images with different levels of DL-ToF strength (low, medium, high). The models were objectively evaluated using the testing set based on standardised uptake value (SUV) in 139 identified lesions, and in normal regions of liver and lungs. Three radiologists subjectively rated the models using testing sets based on lesion detectability, diagnostic confidence, and image noise/quality.

Results

The non-ToF, DL-ToF low, medium, and high methods resulted in − 28 ± 18, − 28 ± 19, − 8 ± 22, and 1.7 ± 24% differences (mean; SD) in the SUVmax for the lesions in testing set, compared to ToF-BSREM image. In background lung VOIs, the SUVmean differences were 7 ± 15, 0.6 ± 12, 1 ± 13, and 1 ± 11% respectively. In normal liver, SUVmean differences were 4 ± 5, 0.7 ± 4, 0.8 ± 4, and 0.1 ± 4%. Visual inspection showed that our DL-ToF improved feature sharpness and convergence towards ToF reconstruction. Blinded clinical readings of testing sets for diagnostic confidence (scale 0–5) showed that non-ToF, DL-ToF low, medium, and high, and ToF images scored 3.0, 3.0, 4.1, 3.8, and 3.5 respectively. For this set of images, DL-ToF medium therefore scored highest for diagnostic confidence.

Conclusion

Deep learning–based image enhancement models may provide converged ToF-equivalent image quality without ToF reconstruction. In clinical scoring DL-ToF-enhanced non-ToF images (medium and high) on average scored as high as, or higher than, ToF images. The model is generalisable and hence, could be applied to non-ToF images from BGO-based PET/CT scanners.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s00259-022-05824-7.
This article is part of the Topical Collection on Advanced Image Analyses (Radiomics and Artificial Intelligence)

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Time-of-flight (ToF) positron emission tomography (PET) is a detector technology that measures the arrival times of the annihilation photons with an uncertainty governed by the coincidence timing resolution (CTR) of the scanner [1]. The first generation of ToF PET scanners were equipped with either caesium or barium fluoride scintillators coupled with photomultiplier tubes, providing CTR of 400–600 ps. However, neither their sensitivity nor spatial resolution could compete with that of non-ToF bismuth germanate (BGO)–based scanners. With the advent of lutetium (Lu)-based scintillators with better sensitivity and spatial resolution, conventional photomultiplier tube–based ToF PET scanners became commercially available around 2006 with a CTR in the range of 450–600 ps. Since then, advancements in silicon photomultiplier detectors have led to the next generation of ToF scanners with CTR of 214–380 ps [2]. Compared to Lu-based scintillators, BGO has a higher stopping power, therefore higher sensitivity for a given crystal size, at the expense of poor timing resolution.
Using ToF capability, the location of emission points along each line of response (LOR) is estimated and utilised during PET image reconstruction to update image voxels only along each segment of response, defined by ToF resolution, instead of the whole LOR. Consequently, the cross-dependencies between image voxels are reduced, which results in (i) reduced noise propagation with fast and space-invariant convergence, which in turn improves the detectability of lesions [3, 4], and (ii) reduced sensitivity to errors in normalization, attenuation correction, and scatter correction [5, 6]. As the CTR is improved, cross-dependencies between image voxels are reduced leading to further ToF benefits.
ToF technology together with advanced image reconstruction algorithms (such as ordered subsets expectation maximisation — OSEM, or block sequential regularised expectation maximisation — BSREM [7]) have led to improved diagnostic confidence and lesion detectability [8]. With the recent advancements in artificial intelligence, deep learning (DL) techniques have found promising applications in PET imaging from photon detection to image reconstruction [911]. Recently, deep convolutional neural networks have been extensively used to reduce PET acquisition time or radiotracer dose by reducing image noise [1214] or reconstruction time with improved image convergence [15]. For ToF technology, DL has been used for data-driven time-of-flight estimation which can lead to about 20% improvement in CTR [16].
Given that the benefits of ToF technology are directly translated into image space and a number of current clinical PET scanners are non-ToF (i.e. BGO-based), there is a desire to improve the diagnostic value of non-ToF scanners compared to ToF scanners [17, 18]. In this study, we aim to leverage deep learning to enable ToF benefits for PET images reconstructed without ToF information. To the best of our knowledge, the proposed deep learning for ToF image enhancement (DL-ToF) is the first-ever attempt to transform non-ToF PET images to ToF-like images. Although related, this challenge differs substantially from the aforementioned uses of DL for noise-reduction due to the variety of ways in which ToF information influences the image appearance. The neural network is not required to add time of flight information to the PET coincidence data, but it is required to learn how ToF information alters many image characteristics, and then to replicate these changes when supplied with non-TOF input images. Three DL-ToF models with different levels of contrast-enhancement-to-noise trade off (low: L, medium: M and high: H) were trained in supervised learning sessions for transforming non-ToF BSREM images, each reconstructed with a range of regularization parameters (beta), to ToF BSREM images, reconstructed with specific regularization values to reflect the intended level of contrast-to-noise. The performance of the three models was quantitatively and qualitatively evaluated using ToF and non-ToF PET scans for [18F]-FDG oncology exams.

Materials and methods

Data acquisition and processing

The PET list-mode data and CT-based attenuation correction (CTAC) images of a total of 273 whole-body oncology [18F]-FDG PET exams were retrospectively collected, as summarised in Supp. Materials Table 1, from six clinical sites equipped with Discovery MI (DMI) and D710 ToF PET/CT scanners. Using training datasets from various clinical sites improves the generalisability of DL-ToF models to account for the fact that each site uses different acquisition protocols and reconstruction parameters. The DMI’s PET subsystem has ToF resolution of 385 ps and different sensitivity depending on the number of detector rings. The DMI system can be configured with 3, 4, or 5 rings of detectors, for axial FOV of 15, 20 or 25 cm. The D710 has a ToF resolution of 550 ps and an axial FOV of 15.7 cm. The use of different scanners and imaging protocols at different sites led to a range of injected [18F]-FDG activity (mean ± SD: 348 ± 118 MBq) and scan duration (161 ± 46 s/bed). Moreover, there were variations in patient size (body mass index, BMI, 27.3 ± 6.0 kg/m2). The [18F]-FDG uptake time varied between sites: 82 ± 26 min. For each subject, a whole-body CT scan protocol was performed for PET attenuation correction using 100–120 kVp.
The 273 DMI/D710 exams were divided into training (n = 208), validation (n = 15), and testing (n = 50) sets. The validation and testing DMI exams were chosen by two nuclear medicine experts for pathologically interesting cases with small lesions. Each dataset was reconstructed using the BSREM algorithm into 4 image series, one ToF (target), and three non-ToF images (input) with different beta values. Supp. Materials Table 1 summarises the beta values chosen for each DL-ToF model, clinical site, and target-input pair. The beta value of ToF BSREM images were experimentally adjusted per site in order to achieve the same low noise level, based on visual inspection, across data from all sites. Each image was reconstructed with a matrix size of 256 × 256 and field-of-view of 700 mm (xy pixel size: 2.73 mm, slice thickness: 2.79 mm). The whole-body image volumes used for training and validation were axially divided into equally spaced contiguous 3D sub-volumes, each of 100 slices (28 cm).

Model training

A 3D residual U-Net network [19] was developed and implemented in PyTorch 1.6 (www.​pytorch.​org) shown in Supp. Figure 1. DL-ToF networks were trained in a supervised session in which their predicted ToF BSREM images were compared to target ToF BSREM ones based on a mean squared error (MSE) loss function. Supp. Materials Table 2 summarises the network and training hyperparameters that were optimised experimentally. The ADAM algorithm [20] was used to update the networks’ trainable parameters for a maximum of 100 epochs on a workstation with two RTX6000 GPUs. The validation set was used to monitor the network’s generalisation error to avoid over-fitting. The epoch at which the model had the lowest validation loss was chosen as a stopping point.

Evaluation

The performance of our trained DL models was quantitatively evaluated using the testing sets based on standardised uptake values (SUVs) including SUVmax (maximum voxel intensity) in lesions, SUVmean (mean intensity of voxels) in normal liver and lungs and the noise in the liver using volumes of interest (VOIs) selected per subject. For each subject, 5 VOIs of size 7 × 7 × 7 voxels (~ 7 mL) were defined in the lungs, and 5 similar VOIs in liver. Noise in liver was calculated as standard deviation of the five VOI mean values. For each subject, up to 5 small lesions were visually identified and segmented using an adaptive thresholding method (42% of maximum minus minimum SUV in a 7 × 7 × 7 bounding box). For the evaluation of DL-ToF models, the beta value of both non-ToF and ToF BSREM images were set to 350. The difference in SUV values (compared to the target ToF BSREM SUVs), scatter plots, and Bland–Altman plots were generated. The statistical significance of differences in SUVs was evaluated using the Wilcoxon signed rank test. Additionally, root-mean-square error (RMSE) between reference ToF images (x) and other images (y) over the whole-body (WB) was calculated by \(RMSE=\sqrt{\frac{1}{N}{\sum }_{i}^{N}{\left({x}_{i}-{y}_{i}\right)}^{2}}\), where N is the total number of voxels in the body. Also, WB SUVmean was calculated by averaging the SUV values in the body.
Three radiologists, (KMB, PAF, and MH), blinded to image reconstruction, independently rated all 50 testing sets. Each patient had 5 image series (Non-ToF and ToF BSREM, DL-ToF low (L), medium (M) and high (H)); these were assessed, with corresponding CT, based on Likert scores considering three image features (lesion detectability, diagnostic confidence, and image noise/quality). The Likert scale used was 0 (non-diagnostic), 1 (poor,), 2 (satisfactory), 3 (good), 4 (very good), and 5 (excellent) with image noise metrics scored on the same 0–5 scale as described previously [21]. In addition, the 5 series were ranked in order of preference from 1 (best) to 5 (worst) for each imaging feature category. Using SPSS v27, the interclass correlation coefficient (ICC) was calculated between the radiologists’ image scores to assess the level to which the average reader scores are generalisable to a wider population of readers [22]. Post hoc pairwise testing against both non-ToF and ToF BSREM images was then performed using Dunn’s method, with Bonferroni correction applied to the reported p-values.

Application to a BGO-PET camera

Ten exams acquired on a GE Discovery IQ (DIQ) scanner were used to illustrate the generalisability of the trained models for a non-ToF BGO-based scanner. This final step of running DL-ToF models on data from a non-TOF, BGO PET scanner had the goal of demonstrating the potential of the DL approach. However, with no ground truth or target solution to compare against, we limited this part of the study to a small number of subjects, and the image analysis to a visual verification that the images were free of obvious artefacts and showed visual changes in accordance with expectations for the three DL models.

Results

Figures 1 and 2 compare the performance of our three DL-ToF models in comparison with the input non-ToF BSREM (beta = 350) and target ToF BSREM (beta = 350) images for two representative, example patients with different BMI scanned on a Discovery MI PET/CT scanner (further examples are shown in Supp. Figures 2–5). As shown by the arrows, the patients have multiple small lesions in different areas (neck, mediastinum, breast, and vertebrae) which have a lower contrast in the non-ToF image. The DL-ToF models improve the detectability and contrast of the lesions towards their target ToF images, with DL-ToF(H) providing the closest match visually. Since the models were trained to provide different levels of smoothness, the liver noise as well as lesion contrast is different among these three models. As shown in Fig. 1, the DL-ToF models improve the overall image quality and feature sharpness of the non-ToF PET images.
Table 1
Quantitative performance of the DL-ToF models evaluated on 50 test exams, expressed as a percentage difference with ToF BSREM (taken as ground truth), for lesion SUVmax, lung SUVmean, liver SUVmean, and noise in liver (the standard deviation of noise averaged over all exams) for each type of reconstruction. P-values (parentheses) show significance of difference from ToF BSREM
 
Lesion SUVmax (%)
Lung SUVmean (%)
Liver SUVmean (%)
Liver noise (SUV)
Non-ToF BSREM
 − 28.6 ± 18.3 (< 0.0001)
7.7 ± 15.0 (< 0.0001)
4.3 ± 5.6 (< 0.0001)
0.16
DL-ToF(L)
 − 28.7 ± 19.0 (< 0.0001)
0.6 ± 12.1 (0.179)
0.7 ± 4.6 (0.067)
0.10
DL-ToF(M)
 − 8.0 ± 22.5 (< 0.0001)
1.3 ± 13.0 (0.083)
0.8 ± 4.4 (0.016)
0.13
DL-ToF(H)
1.7 ± 23.9 (0.57)
1.4 ± 11.5 (0.50)
0.1 ± 4.5 (0.86)
0.19
ToF-BSREM
0.19
Table 1 shows the quantitative performance of non-ToF BSREM and DL-ToF methods on the DMI’s testing set (n = 50) for SUVmax of the 139 identified lesions, and SUVmean in normal lungs and liver. The percentage difference from the target ToF BSREM method is provided (mean ± standard deviation), along with the result of the tests of statistical significance. As seen, DL-ToF methods reduce the lesion’s SUVmax difference from − 28.6 to 1.7%, depending on their level of smoothness. In this test dataset, the difference of 1.7% between DL-ToF(H) and ToF BSREM was not statistically significant. In the lungs, DL-ToF models reduce the SUVmean differences from 7.7% to less than 2%. These results demonstrate that DL-ToF models make lesions hotter and lungs colder; in other words, they improve the accuracy of the non-ToF BSREM images. Assessment of the noise in the liver, via the average of the liver standard deviations from 50 testing datasets, shows that the DL-ToF models provide different level of smoothness, and all provide some level of noise reduction. Supp. Materials Table 3 shows the RMSE and SUVmean over whole-body and lesion/lung/liver ROIs for the test cases.
Figure 3 shows scatter plots of lesion SUVmax for non-ToF BSREM and DL-ToF images compared to reference ToF BSREM images including slope and R-square of regression lines. As could be expected, the non-ToF BSREM method shows a lower lesion SUVmax and hence a less steep slope of the fitted line compared to ToF BSREM method. As the strength of DL-ToF is increased, the slope of fitted line for DL-ToF methods gets closer to identity: this indicates contrast convergence enhancement of the input non-ToF images. As shown, DL-ToF(H) increases the slope from 0.84 to 1.02 and increases the coefficient of determination (0.96 to 0.97).
Figure 4 shows Bland–Altman plots comparing the concordance of lesion SUVmax between target ToF BSREM and other methods. Consistent with the other quantification measures, the plots show a systematic difference in SUVmax which is reduced by DL-ToF methods.
Table 2 and Supp. Materials Table 4 show the scores and ranking results for different reconstruction methods for 50 testing exams from three independent readers. Table 2 also provides p-values for the scores, using pairwise comparisons with respect to ToF BSREM methods. Supp. Materials Table 5 shows p-values with respect to non-ToF BSREM. The lesion detectability results show that DL-ToF(H) significantly improves lesion detectability in the images so much that their p-values become lower than 0.001. In terms of diagnostic confidence, DL-ToF(M) achieves the best score whereas for image noise/quality DL-ToF(L) scores the best. These results highlight that the strength of DL-ToF can be chosen to provide a balance between lesion detection and noise reduction, according to the preference of the image reader. Results in Supp. Materials Table 4 also show DL-ToF(H) achieves the best rank for lesion detectability, whereas DL-ToF (M) has the best rank for diagnostic confidence (i.e., better than ToF BSREM) and DL-ToF(L) achieves the best rank in terms of noise and image quality.
Table 2
Clinical image quality scoring from three readers of 50 test whole-body scans based on different criteria, mean ± standard deviation. 0 is non-diagnostic; 5 is excellent. Bold indicates the best (highest) score for each metric. The intraclass correlation coefficient (ICC) is also provided for each metric (95% confidence interval) to show reader agreement. P-values (in parentheses) are given with respect to ToF BSREM (with p-values with respect to non-ToF BSREM shown in Suppl. Table 5)
Scores
Diagnostic confidence
Lesion detectability
Image noise/quality
Non-ToF BSREM
3.03 ± 0.40 (< 0.001)
3.03 ± 0.43 (< 0.001)
3.36 ± 0.40 (1.000)
DL-ToF(L)
2.98 ± 0.34 (< 0.001)
2.88 ± 0.35 (< 0.001)
4.52 ± 0.27 (< 0.001)
DL-ToF(M)
4.07 ± 0.47 (< 0.001)
3.99 ± 0.48 (1.000)
4.09 ± 0.34 (< 0.001)
DL-ToF(H)
3.83 ± 0.38 (0.11)
4.18 ± 0.39 (1.000)
3.39 ± 0.40 (0.96)
ToF BSREM
3.53 ± 0.53
4.08 ± 0.54
3.08 ± 0.55
ICC
0.67 (0.60, 0.74)
0.68 (0.61, 0.74)
0.58 (0.48, 0.66)
The application of the technique to data from a BGO non-ToF DIQ PET scanner (on which the algorithm had not been trained) provided images which, visually, met our expectations. This is shown in Figs. 5 and 6, which although from different patients, can be compared to Figs. 1 and 2. Eight further examples are presented in Supplementary Figs. 6–9. In these cases, the models showed similar image enhancement as was achieved with non-ToF DMI data. In Fig. 6, the patient presents attenuation correction artefacts near to the diaphragm, which are often reduced by ToF reconstruction [6]; in this instance, all DL-ToF models show reduction of the artefacts.

Discussion

In this study, the feasibility of utilising deep learning for enhancing the ToF features in PET images reconstructed without ToF capability was explored with the intended application of improving lesion detectability and diagnostic confidence for [18F]-FDG PET scans acquired in non-ToF PET scanners. Three DL-ToF models were trained with different levels of smoothness or ToF strength to demonstrate the flexibility of the proposed DL solution to meet radiologists’ preferences.
Our models were trained to transform non-ToF PET images reconstructed by the BSREM algorithm to their corresponding ToF BSREM images as faithfully as possible. The BSREM algorithm was chosen over OSEM as it provides a higher convergence and lower noise, therefore giving more accurate quantification [7, 21]. The DL-ToF(H) algorithm that was developed, when applied on non-ToF BSREM (beta = 350) input data, achieves quantification errors less than 5% compared to the target ToF BSREM. On the other hand, the results show that all DL-ToF models provide some level of noise reduction, which translates to improved contrast-to-noise ratio (CNR), indicating improved feature sharpness and lesion detectability.
As shown in Fig. 3, in our primary testing set of 50 exams, the identified lesions are mostly clustered around an SUVmax of 5 in the ToF BSREM images (median = 5.6). These lesions, as well as those of lower SUVmax, are often diagnostically important and affected by ToF reconstruction method. The scatter plots show that as one moves from DL-ToF(L) to DL-ToF(H) the set of lesion SUVs are increased toward their target ToF SUVs.
The clinical reading results showed that DL-ToF models present favourable performance. For instance, in Table 2, DL-ToF(H), which has the least smooth DL-ToF model or highest ToF strength, achieves on average 3.83, 4.18, and 3.39 scores for the key metrics of diagnostic confidence, lesion detectability, and image noise/quality compared to reference ToF images with the corresponding scores of 3.53, 4.08, and 3.08, respectively. In our test set, the best score for lesion detectability was from DLT(H) with the best score for image noise/quality from DLT(L) and diagnostic confidence from DLT(M). Overall, in terms of diagnostic confidence, the DL-ToF(M) model provides a better trade-off in our test set as a lower noise and improved detectability are desirable features for an image reconstruction or enhancement technique. It is the balance of good performance regarding lesion detectability and image noise that leads to best diagnostic confidence as shown by the blinded clinician scores and ranking. In some implementations of PET image reconstruction, including that used here, the use of ToF information can lead to faster convergence and hence a noisier image as compared to non-TOF for the same number of iterations. It is possible to smooth ToF images with a Gaussian filter or in the case of BSREM to use a larger beta value but this may come at the cost of a reduction in lesion detectability [7].
This study utilised a U-Net model, as an encoder-decoder CNN, that was trained using a diverse set of DMI datasets. The quality and diversity of the training set is one of the key factors in the performance and generalisability of a CNN model. To exemplify this generalisability, the models were also applied to ten patients scanned on a BGO non-ToF scanner (GE Healthcare Discovery IQ), as shown in Figs. 5 and 6 and Supplementary Figs. 6–9. These examples suggest that the models may work on data from scanners that were not part of the training dataset; examining this in more detail will be the subject of a follow-on study.
This study has a number of limitations. Our testing sets do not include randomly selected exams (i.e. combination of normal/abnormal) but rather patients with lesions that showed low activity or which were completely missed in non-ToF BSREM images. Therefore, our results might be biased to highlight the gap between ToF and non-ToF reconstructions. However, our results with another testing set used during model validation (not shown in this study) demonstrated that our DL-ToF models show the greatest enhancement for patients with the highest BMI; this result is in line with the expected behaviour of ToF reconstruction. Another limitation could be that the readers were shown all 5 sets of images (blinded) of a subject at the same time. This might bias the scoring of the images, although was considered advantageous, in order to facilitate the detection of false positive or missing lesions by comparing images all at once and furthermore is required in order to produce a rank ordering. Further, DL-ToF models were not compared to any other ToF image enhancement technique given the novelty of our methodology and they were not tested for non-FDG tracers. Therefore, this work opens new research topics for future studies. Future work should include further clinical evaluation using a cohort of FDG exams with the possibility of getting the clinical feedback into the training cycle of our models.

Conclusion

This study developed three deep convolutional neural networks for ToF-like enhancement of PET images acquired in non-ToF PET/CT scanners. Our results demonstrate that the proposed networks improve the feature quantification (lesions, liver and lungs), overall image sharpness (as seen with ToF, e.g. organ delineation, ribs, vertebrae), and overall diagnostic value (particularly in terms of lesion detectability and diagnostic confidence). Depending on the model ToF strength, DL-ToF(L) showed more noise reduction, whereas DL-ToF(H) had the greatest improvement in lesion detection. DL-ToF(M) presented a balanced performance and best diagnostic confidence. We conclude that deep learning–enhanced image reconstruction can markedly improve non-ToF PET images towards their corresponding ToF images.

Acknowledgements

The authors would like to thank Prof. Luca Guerra, University of Milano-Bicocca, Milan, Italy, for providing the non-ToF raw data for the DIQ images used in this study.

Declarations

Ethics approval

Patient data was used in this project in the form of anonymised retrospectively reconstructed image datasets. This practice does not require informed consent nor was specific ethics approval required.
Not applicable.
Not applicable.

Conflict of interest

AM, SDW, FK, KHS, RJ, and FPJ are all employees of GE Healthcare. No other conflicts of interest to declare.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Schaart DR. Physics and technology of time-of-flight PET detectors. Phys Med Biol. 2021;66(9):09TR01.CrossRef Schaart DR. Physics and technology of time-of-flight PET detectors. Phys Med Biol. 2021;66(9):09TR01.CrossRef
2.
Zurück zum Zitat Surti S, Karp JS. Update on latest advances in time-of-flight PET. Phys Med. 2020;80:251–8.CrossRef Surti S, Karp JS. Update on latest advances in time-of-flight PET. Phys Med. 2020;80:251–8.CrossRef
3.
Zurück zum Zitat El Fakhri G, et al. Improvement in lesion detection with whole-body oncologic time-of-flight PET. J Nucl Med. 2011;52(3):347–53.CrossRef El Fakhri G, et al. Improvement in lesion detection with whole-body oncologic time-of-flight PET. J Nucl Med. 2011;52(3):347–53.CrossRef
4.
Zurück zum Zitat Schaefferkoetter J, et al. Clinical impact of time-of-flight and point response modeling in PET reconstructions: a lesion detection study. Phys Med Biol. 2013;58(5):1465–78.CrossRef Schaefferkoetter J, et al. Clinical impact of time-of-flight and point response modeling in PET reconstructions: a lesion detection study. Phys Med Biol. 2013;58(5):1465–78.CrossRef
5.
Zurück zum Zitat Conti M. Why is TOF PET reconstruction a more robust method in the presence of inconsistent data? Phys Med Biol. 2011;56(1):155–68.CrossRef Conti M. Why is TOF PET reconstruction a more robust method in the presence of inconsistent data? Phys Med Biol. 2011;56(1):155–68.CrossRef
6.
Zurück zum Zitat Mehranian A, Zaidi H. Impact of time-of-flight PET on quantification errors in MR imaging-based attenuation correction. J Nucl Med. 2015;56(4):635–41.CrossRef Mehranian A, Zaidi H. Impact of time-of-flight PET on quantification errors in MR imaging-based attenuation correction. J Nucl Med. 2015;56(4):635–41.CrossRef
7.
Zurück zum Zitat Teoh EJ, et al. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q. Clear on an LYSO PET/CT System. J Nucl Med. 2015;56(9):1447–52.CrossRef Teoh EJ, et al. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q. Clear on an LYSO PET/CT System. J Nucl Med. 2015;56(9):1447–52.CrossRef
8.
Zurück zum Zitat Teoh EJ, et al. 18F-FDG PET/CT assessment of histopathologically confirmed mediastinal lymph nodes in non-small cell lung cancer using a penalised likelihood reconstruction. Eur Radiol. 2016;26(11):4098–106.CrossRef Teoh EJ, et al. 18F-FDG PET/CT assessment of histopathologically confirmed mediastinal lymph nodes in non-small cell lung cancer using a penalised likelihood reconstruction. Eur Radiol. 2016;26(11):4098–106.CrossRef
9.
Zurück zum Zitat Arabi H, et al. The promise of artificial intelligence and deep learning in PET and SPECT imaging. Physica Med. 2021;83:122–37.CrossRef Arabi H, et al. The promise of artificial intelligence and deep learning in PET and SPECT imaging. Physica Med. 2021;83:122–37.CrossRef
10.
Zurück zum Zitat Gong K, et al. Machine learning in PET: from photon detection to quantitative image reconstruction. Proc IEEE. 2020;108(1):51–68.CrossRef Gong K, et al. Machine learning in PET: from photon detection to quantitative image reconstruction. Proc IEEE. 2020;108(1):51–68.CrossRef
11.
Zurück zum Zitat Reader AJ, et al. Deep learning for PET image reconstruction. IEEE Trans Radiat Plasma Med Sci. 2021;5(1):1–25.CrossRef Reader AJ, et al. Deep learning for PET image reconstruction. IEEE Trans Radiat Plasma Med Sci. 2021;5(1):1–25.CrossRef
12.
Zurück zum Zitat Lu W, et al. An investigation of quantitative accuracy for deep learning based denoising in oncological PET. Phys Med Biol. 2019;64(16):165019.CrossRef Lu W, et al. An investigation of quantitative accuracy for deep learning based denoising in oncological PET. Phys Med Biol. 2019;64(16):165019.CrossRef
14.
Zurück zum Zitat Mehranian A, et al. Image enhancement of whole-body oncology [18F]-FDG PET scans using deep neural networks to reduce noise. Eur J Nucl Med Mol Imaging. 2022;49(2):539–49.CrossRef Mehranian A, et al. Image enhancement of whole-body oncology [18F]-FDG PET scans using deep neural networks to reduce noise. Eur J Nucl Med Mol Imaging. 2022;49(2):539–49.CrossRef
15.
Zurück zum Zitat Cheng L et al. Accelerated iterative image reconstruction using a deep learning based leapfrogging strategy. Int Conf Fully Three-Dimensional Image Reconstr Radiol Nucl Med. 2017; 715–720. Cheng L et al. Accelerated iterative image reconstruction using a deep learning based leapfrogging strategy. Int Conf Fully Three-Dimensional Image Reconstr Radiol Nucl Med. 2017; 715–720.
16.
Zurück zum Zitat Berg E, Cherry SR. Using convolutional neural networks to estimate time-of-flight from PET detector waveforms. Phys Med Biol. 2018;63(2):02LT01.CrossRef Berg E, Cherry SR. Using convolutional neural networks to estimate time-of-flight from PET detector waveforms. Phys Med Biol. 2018;63(2):02LT01.CrossRef
17.
Zurück zum Zitat Karp JS, et al. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med. 2008;49(3):462–70.CrossRef Karp JS, et al. Benefit of time-of-flight in PET: experimental and clinical results. J Nucl Med. 2008;49(3):462–70.CrossRef
18.
Zurück zum Zitat Conti M. Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging. 2011;38(6):1147–57.CrossRef Conti M. Focus on time-of-flight PET: the benefits of improved time resolution. Eur J Nucl Med Mol Imaging. 2011;38(6):1147–57.CrossRef
21.
Zurück zum Zitat Chilcott AK, Bradley KM, McGowan DR. Effect of a Bayesian penalized likelihood PET reconstruction compared with ordered subset expectation maximization on clinical image quality over a wide range of patient weights. AJR Am J Roentgenol. 2018;210(1):153–7.CrossRef Chilcott AK, Bradley KM, McGowan DR. Effect of a Bayesian penalized likelihood PET reconstruction compared with ordered subset expectation maximization on clinical image quality over a wide range of patient weights. AJR Am J Roentgenol. 2018;210(1):153–7.CrossRef
22.
Zurück zum Zitat McGraw K, Wong SP. Forming inferences about some intraclass correlation coefficients. Psychol Methods. 1996;1(1):30–46.CrossRef McGraw K, Wong SP. Forming inferences about some intraclass correlation coefficients. Psychol Methods. 1996;1(1):30–46.CrossRef
Metadaten
Titel
Deep learning–based time-of-flight (ToF) image enhancement of non-ToF PET scans
verfasst von
Abolfazl Mehranian
Scott D. Wollenweber
Matthew D. Walker
Kevin M. Bradley
Patrick A. Fielding
Martin Huellner
Fotis Kotasidis
Kuan-Hao Su
Robert Johnsen
Floris P. Jansen
Daniel R. McGowan
Publikationsdatum
04.05.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 11/2022
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-022-05824-7

Weitere Artikel der Ausgabe 11/2022

European Journal of Nuclear Medicine and Molecular Imaging 11/2022 Zur Ausgabe