Internal fixation of femoral neck fractures with posterior comminution

A biomechanical comparison of DHS® and Intertan nail®

verfasst von: Martin Rupprecht, Lars Grossterlinden, Kai Sellenschloh, Michael Hoffmann, Klaus Püschel, Michael Morlock, Johannes M. Rueger, Wolfgang Lehmann

Erschienen in: International Orthopaedics | Ausgabe 11/2011

Einloggen, um Zugang zu erhalten

Abstract

Background and purpose

Internal fixation is a therapeutic mainstay for treatment of undisplaced femoral neck fractures and fractures without posterior comminution. The best treatment for unstable and comminuted fractures, however, remains controversial, especially in older patients. The present study was designed to assess the utility of the Intertan Nail® (IT) for stabilization of comminuted Pauwels type III fractures compared to dynamic hips screw (DHS).

Methods

Randomized on the basis of bone mineral density, 32 human femurs were assigned to four groups. Pauwels type III fractures were osteomized with a custom-made saw guide. In 16 specimens the posteromedial support was removed and all femurs were instrumented with an IT or a DHS. All constructs were tested with nondestructive axial loading to 700N, cyclical compression to 1,400N (10,000 cycles), and loading to failure. Outcome measures included number of survived cycles, mechanical stiffness, head displacement and load to failure.

Results

Postoperative mechanical stiffness and stiffness after cyclical loading were significantly reduced in all constructs regardless of the presence of a comminution defect (p = 0.02). Specimens stabilized with the IT had a lower construct displacement (IT, 8.5 ± 0.5 mm vs. DHS, 14.5 ± 2.2 mm; p = 0.007) and sustained higher failure loads (IT, 4929 ± 419 N vs. DHS, 3505 ± 453 N; p = 0.036) than the DHS constructs.

Interpretation

In comminuted Pauwels type III fractures, the fixation with the IT provided sufficient postoperative mechanical strength, comparable rate of femoral head displacement, and a similar tolerance of physiological loads compared to fractures without comminution. The absence of the posteromedial support in comminuted fractures tended to reduce the failure load regardless of the fixation method.
Literatur
1.
Zurück zum Zitat Johansson T, Jacobsson SA, Iverson I et al (2000) Internal fixation versus total hip arthroplasty in the treatment of displaced femoral neck fractures: a prospective randomized study of 100 hips. Acta Orthop Scand 71:597–602PubMedCrossRef Johansson T, Jacobsson SA, Iverson I et al (2000) Internal fixation versus total hip arthroplasty in the treatment of displaced femoral neck fractures: a prospective randomized study of 100 hips. Acta Orthop Scand 71:597–602PubMedCrossRef
2.
Zurück zum Zitat Eisler J, Cornwall R, Strauss E et al (2002) Outcomes of elderly patients with nondisplaced femoral neck fractures. Clin Orthop 399:52–58PubMedCrossRef Eisler J, Cornwall R, Strauss E et al (2002) Outcomes of elderly patients with nondisplaced femoral neck fractures. Clin Orthop 399:52–58PubMedCrossRef
3.
Zurück zum Zitat Lu-Yao GL, Keller RB, Littenberg B et al (1994) Outcomes after displaced fractures of the femoral neck. A meta-analysis of one hundred and six published reports. J Bone Joint Surg Am 76(1):15–25PubMed Lu-Yao GL, Keller RB, Littenberg B et al (1994) Outcomes after displaced fractures of the femoral neck. A meta-analysis of one hundred and six published reports. J Bone Joint Surg Am 76(1):15–25PubMed
4.
Zurück zum Zitat Bonnaire FA, Buitrago-Tellez C, Schmal H et al (2002) Correlation of bone density and geometric parameter to mechanical strength of the femoral neck. Injury 33(Suppl 3):C47–C53PubMedCrossRef Bonnaire FA, Buitrago-Tellez C, Schmal H et al (2002) Correlation of bone density and geometric parameter to mechanical strength of the femoral neck. Injury 33(Suppl 3):C47–C53PubMedCrossRef
5.
Zurück zum Zitat Sjostedt A, Zetterberg C, Hansson T et al (1994) Bone mineral content and fixation strength of femoral neck fractures. A cadaver study. Acta Orthop Scand 65(2):161–165PubMedCrossRef Sjostedt A, Zetterberg C, Hansson T et al (1994) Bone mineral content and fixation strength of femoral neck fractures. A cadaver study. Acta Orthop Scand 65(2):161–165PubMedCrossRef
6.
Zurück zum Zitat Audige L, Hanson B, Swiontkowski MF (2003) Implant-related complications in the treatment of unstable intertrochanteric fractures: meta-analysis of dynamic screw versus dynamic screw-intramedullary nail devices. Int Orthop 27(4):197–203PubMedCrossRef Audige L, Hanson B, Swiontkowski MF (2003) Implant-related complications in the treatment of unstable intertrochanteric fractures: meta-analysis of dynamic screw versus dynamic screw-intramedullary nail devices. Int Orthop 27(4):197–203PubMedCrossRef
7.
Zurück zum Zitat Baumgaertner MR, Solberg BD (1997) Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br 79(6):969–971PubMedCrossRef Baumgaertner MR, Solberg BD (1997) Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg Br 79(6):969–971PubMedCrossRef
8.
Zurück zum Zitat Bonnaire F, Zenker H, Lill Ch et al (2005) Treatment strategies for proximal femur fractures in osteoporotic patients. Osteoporos Int 16:93–102CrossRef Bonnaire F, Zenker H, Lill Ch et al (2005) Treatment strategies for proximal femur fractures in osteoporotic patients. Osteoporos Int 16:93–102CrossRef
9.
Zurück zum Zitat Parker MJ (2002) Evidence-based results depending on the implant used for stabilizing femoral neck fractures. Injury 33(Suppl 3):C15–C18PubMedCrossRef Parker MJ (2002) Evidence-based results depending on the implant used for stabilizing femoral neck fractures. Injury 33(Suppl 3):C15–C18PubMedCrossRef
10.
Zurück zum Zitat Ruecker AH, Rupprecht M, Gruber M et al (2009) The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma 23(1):22–30PubMedCrossRef Ruecker AH, Rupprecht M, Gruber M et al (2009) The treatment of intertrochanteric fractures: results using an intramedullary nail with integrated cephalocervical screws and linear compression. J Orthop Trauma 23(1):22–30PubMedCrossRef
11.
Zurück zum Zitat Sievänen H, Koskue V, Rauhio A et al (1998) Peripheral quantitative computed tomography in human long bones: evaluation of in vitro and in vivo precision. J Bone Miner Res 13(5):871–882PubMedCrossRef Sievänen H, Koskue V, Rauhio A et al (1998) Peripheral quantitative computed tomography in human long bones: evaluation of in vitro and in vivo precision. J Bone Miner Res 13(5):871–882PubMedCrossRef
12.
Zurück zum Zitat Bergmann G, Deuretzbacher G, Heller M et al (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34(7):859–871PubMedCrossRef Bergmann G, Deuretzbacher G, Heller M et al (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34(7):859–871PubMedCrossRef
13.
Zurück zum Zitat Windolf M, Braunstein V, Dutoit C et al (2009) Is a helical shaped implant a superior alternative to the dynamic hip screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech 24:59–64CrossRef Windolf M, Braunstein V, Dutoit C et al (2009) Is a helical shaped implant a superior alternative to the dynamic hip screw for unstable femoral neck fractures? A biomechanical investigation. Clin Biomech 24:59–64CrossRef
14.
Zurück zum Zitat Denham RA (1959) Hip mechanics. J Bone Joint Surg Br 41:550–557PubMed Denham RA (1959) Hip mechanics. J Bone Joint Surg Br 41:550–557PubMed
15.
Zurück zum Zitat Kubiak EN, Bong M, Park SS et al (2004) Intramedullary fixation of unstable intertrochanteric hip fractures: one or two lag screws. J Orthop Trauma 18(1):12–17PubMedCrossRef Kubiak EN, Bong M, Park SS et al (2004) Intramedullary fixation of unstable intertrochanteric hip fractures: one or two lag screws. J Orthop Trauma 18(1):12–17PubMedCrossRef
16.
Zurück zum Zitat Conn KS, Parker MJ (2004) Undisplaced intracapsular hip fracture: results of internal fixation in 375 patients. Clin Orthop 421:249–254PubMedCrossRef Conn KS, Parker MJ (2004) Undisplaced intracapsular hip fracture: results of internal fixation in 375 patients. Clin Orthop 421:249–254PubMedCrossRef
17.
Zurück zum Zitat Heetveld MJ, Raaymakers ELFB, van Walsum ADP et al (2005) Observer assessment of femoral neck radiographs after reduction and dynamic hip screw fixation. Arch Orthop Trauma Surg 125:160–165PubMedCrossRef Heetveld MJ, Raaymakers ELFB, van Walsum ADP et al (2005) Observer assessment of femoral neck radiographs after reduction and dynamic hip screw fixation. Arch Orthop Trauma Surg 125:160–165PubMedCrossRef
18.
Zurück zum Zitat Selvan VT, Oakley MJ, Rangan A et al (2004) Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study. Injury 35:136–141PubMedCrossRef Selvan VT, Oakley MJ, Rangan A et al (2004) Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study. Injury 35:136–141PubMedCrossRef
19.
Zurück zum Zitat Bhandari M, Devereaux PJ, Tornetta P 3rd et al (2005) Operative management of displaced femoral neck fractures in elderly patients. An international survey. J Bone Joint Surg Am 87(9):2122–2130PubMedCrossRef Bhandari M, Devereaux PJ, Tornetta P 3rd et al (2005) Operative management of displaced femoral neck fractures in elderly patients. An international survey. J Bone Joint Surg Am 87(9):2122–2130PubMedCrossRef
20.
Zurück zum Zitat Meyers MH, Harvey JP Jr, Moore TM (1974) The muscle pedicle bone graft in the treatment in displaced fractures of the femoral neck: Indications, operative technique, and results. Orthop Clin North Am 779–792 Meyers MH, Harvey JP Jr, Moore TM (1974) The muscle pedicle bone graft in the treatment in displaced fractures of the femoral neck: Indications, operative technique, and results. Orthop Clin North Am 779–792
21.
Zurück zum Zitat Rau FD, Manoli A 2nd, Morawa LG (1982) Treatment of femoral neck fractures with the sliding compression screw. Clin Orthop Relat Res 163:137–140PubMed Rau FD, Manoli A 2nd, Morawa LG (1982) Treatment of femoral neck fractures with the sliding compression screw. Clin Orthop Relat Res 163:137–140PubMed
22.
Zurück zum Zitat Scheck M (1980) The significance of posterior comminution in femoral neck fractures. Clin Orthop Relat Res 138–142 Scheck M (1980) The significance of posterior comminution in femoral neck fractures. Clin Orthop Relat Res 138–142
23.
Zurück zum Zitat Duda GN, Schneider E, Chao EYS (1997) Internal forces and moments in the femur during walking. J Biomech 30:933–941PubMedCrossRef Duda GN, Schneider E, Chao EYS (1997) Internal forces and moments in the femur during walking. J Biomech 30:933–941PubMedCrossRef
Metadaten
Titel
Internal fixation of femoral neck fractures with posterior comminution
A biomechanical comparison of DHS® and Intertan nail®
verfasst von
Martin Rupprecht
Lars Grossterlinden
Kai Sellenschloh
Michael Hoffmann
Klaus Püschel
Michael Morlock
Johannes M. Rueger
Wolfgang Lehmann
Publikationsdatum
01.11.2011
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 11/2011
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-010-1199-x

Weitere Artikel der Ausgabe 11/2011

International Orthopaedics 11/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.