01.05.2012 | Original Paper
The primary stability of angle-stable versus conventional locked intramedullary nails
verfasst von:
Dirk Wähnert, Yves Stolarczyk, Konrad L. Hoffmeier, Michael J. Raschke, Gunther O. Hofmann, Thomas Mückley
Erschienen in:
International Orthopaedics
|
Ausgabe 5/2012
Einloggen, um Zugang zu erhalten
Abstract
Purpose
The aim of this study was to compare the initial biomechanical characteristics of the angle-stable locking system for intramedullary nails using the new biodegradable sleeve with conventional locking in the treatment of unstable distal tibial fractures.
Methods
Eight pairs of fresh, frozen porcine tibiae were used for this study. The expert tibial nail (Synthes) was equipped with either conventional locking screws (CL) or the angle-stable locking system (AS). This system consists of a special ASLS screw with a biodegradable sleeve. For this investigation distal tibias (5.5 cm) were used and the nails were locked with three screws in both groups. Biomechanical testing included non-destructive torsional and axial loading.
Results
The AS group showed a significantly higher torsional stiffness (70%) compared to the CL group. The range of motion was 0.5 times smaller for the AS constructs. The neutral zone was eight times higher in the CL group (p < 0.001). In axial loading the AS group also showed a 10% higher axial stiffness and a 12% lower range of motion (p < 0.001).
Conclusion
The angle-stable locking system (ASLS) using a special screw and sleeve locking for intramedullary nails provides a significantly higher primary stability. The differences determined in this study may have clinical relevance particularly for torsional loads. For the new biodegradable angle-stable sleeve we found a comparable stability to the PEEK-based sleeve system. This system has the potential to decrease complications such as secondary loss of reduction and mal-/non-union.