Erschienen in: CardioVascular and Interventional Radiology 6/2012

01.12.2012 | Laboratory Investigation

Efficacy of Lower-Body Shielding in Computed Tomography Fluoroscopy-Guided Interventions

verfasst von: Andreas H. Mahnken, Martin Sedlmair, Christine Ritter, Rosemarie Banckwitz, Thomas Flohr

Erschienen in: CardioVascular and Interventional Radiology | Ausgabe 6/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Computed tomography (CT) fluoroscopy-guided interventions pose relevant radiation exposure to the interventionalist. The goal of this study was to analyze the efficacy of lower-body shielding as a simple structural method for decreasing radiation dose to the interventionalist without limiting access to the patient.

Material and Methods

All examinations were performed with a 128-slice dual source CT scanner (12 × 1.2-mm collimation; 120 kV; and 20, 40, 60, and 80 mAs) and an Alderson-Rando phantom. Scatter radiation was measured with an ionization chamber and a digital dosimeter at standardized positions and heights with and without a lower-body lead shield (0.5-mm lead equivalent; Kenex, Harlow, UK). Dose decreases were computed for the different points of measurement.

Results

On average, lower-body shielding decreased scatter radiation by 38.2% within a 150-cm radius around the shielding. This decrease is most significant close to the gantry opening and at low heights of 50 and 100 cm above the floor with a maximum decrease of scatter radiation of 95.9% close to the scanner’s isocentre. With increasing distance to the gantry opening, the effect decreased. There is almost no dose decrease effect at ≥150 above the floor. Scatter radiation and its decrease were linearly correlated with the tube current-time product (r 2 = 0.99), whereas percent scatter radiation decrease was independent of the tube current-time product.

Conclusion

Lower-body shielding is an effective way to decrease radiation exposure to the interventionalist and should routinely be used in CT fluoroscopy-guided interventions.
Literatur
1.
Zurück zum Zitat Katada K, Anno H, Takeshita G et al (1994) Development of CT fluoroscopy using fast CT [in Japanese]. Nihon Igaku Hoshasen Gakkai Zasshi 54:1172–1174PubMed Katada K, Anno H, Takeshita G et al (1994) Development of CT fluoroscopy using fast CT [in Japanese]. Nihon Igaku Hoshasen Gakkai Zasshi 54:1172–1174PubMed
2.
Zurück zum Zitat Katada K, Kato R, Anno A et al (1996) Guidance with real-time CT fluoroscopy: early clinical experience. Radiology 200:851–856PubMed Katada K, Kato R, Anno A et al (1996) Guidance with real-time CT fluoroscopy: early clinical experience. Radiology 200:851–856PubMed
3.
Zurück zum Zitat Kim GR, Hur J, Lee SM et al (2011) CT fluoroscopy-guided lung biopsy versus conventional CT-guided lung biopsy: a prospective controlled study to assess radiation doses and diagnostic performance. Eur Radiol 21:232–239PubMedCrossRef Kim GR, Hur J, Lee SM et al (2011) CT fluoroscopy-guided lung biopsy versus conventional CT-guided lung biopsy: a prospective controlled study to assess radiation doses and diagnostic performance. Eur Radiol 21:232–239PubMedCrossRef
4.
Zurück zum Zitat Silverman SG, Tuncali K, Adams DF et al (1999) CT-fluoroscopy guided abdominal interventions: techniques, results, and radiation exposure. Radiology 212:673–681PubMed Silverman SG, Tuncali K, Adams DF et al (1999) CT-fluoroscopy guided abdominal interventions: techniques, results, and radiation exposure. Radiology 212:673–681PubMed
5.
Zurück zum Zitat de Mey J, Op de Beeck B, Meysman M et al (2000) Real-time CT fluoroscopy: diagnostic and therapeutic applications. Eur J Radiol 32:32–40CrossRef de Mey J, Op de Beeck B, Meysman M et al (2000) Real-time CT fluoroscopy: diagnostic and therapeutic applications. Eur J Radiol 32:32–40CrossRef
6.
Zurück zum Zitat Mahnken AH, Penzkofer T, Grommes J et al (2010) Carbon dioxide-enhanced CT-guided placement of aortic stent-grafts: feasibility in an animal model. J Endovasc Ther 17:332–339PubMedCrossRef Mahnken AH, Penzkofer T, Grommes J et al (2010) Carbon dioxide-enhanced CT-guided placement of aortic stent-grafts: feasibility in an animal model. J Endovasc Ther 17:332–339PubMedCrossRef
7.
Zurück zum Zitat Paulson EK, Sheafor DH, Enterline DS et al (2001) CT fluoroscopy-guided interventional procedures: techniques and radiation dose to radiologists. Radiology 220:161–167PubMed Paulson EK, Sheafor DH, Enterline DS et al (2001) CT fluoroscopy-guided interventional procedures: techniques and radiation dose to radiologists. Radiology 220:161–167PubMed
8.
Zurück zum Zitat Neeman Z, Dromi SA, Sarin S et al (2006) CT fluoroscopy shielding: decreases in scattered radiation for the patient and operator. J Vasc Interv Radiol 17:1999–2004PubMedCrossRef Neeman Z, Dromi SA, Sarin S et al (2006) CT fluoroscopy shielding: decreases in scattered radiation for the patient and operator. J Vasc Interv Radiol 17:1999–2004PubMedCrossRef
9.
Zurück zum Zitat Iida H, Chabatake M, Shimizu M et al (2002) The radiation protective devices for interventional procedures using computed tomography [in Japanese]. Nihon Hoshasen Gijutsu Gakkai Zasshi 58:101–108PubMed Iida H, Chabatake M, Shimizu M et al (2002) The radiation protective devices for interventional procedures using computed tomography [in Japanese]. Nihon Hoshasen Gijutsu Gakkai Zasshi 58:101–108PubMed
10.
Zurück zum Zitat Stoeckelhuber BM, Leibecke T, Schulz E et al (2005) Radiation dose to the radiologist’s hand during continuous CT fluoroscopy-guided interventions. Cardiovasc Intervent Radiol 28:589–594PubMedCrossRef Stoeckelhuber BM, Leibecke T, Schulz E et al (2005) Radiation dose to the radiologist’s hand during continuous CT fluoroscopy-guided interventions. Cardiovasc Intervent Radiol 28:589–594PubMedCrossRef
11.
Zurück zum Zitat Haipt F, Kirsch M, Hosten N (2010) Evaluation of a leaden radiation protection barrier for dose reduction for the physician during CT fluoroscopy-guided interventions [in German]. Rofo 182:512–517PubMedCrossRef Haipt F, Kirsch M, Hosten N (2010) Evaluation of a leaden radiation protection barrier for dose reduction for the physician during CT fluoroscopy-guided interventions [in German]. Rofo 182:512–517PubMedCrossRef
12.
Zurück zum Zitat Sanchez R, Vano E, Fernandez JM et al (2010) Staff radiation doses in a real-time display inside the angiography room. Cardiovasc Intervent Radiol 33:1210–1214PubMedCrossRef Sanchez R, Vano E, Fernandez JM et al (2010) Staff radiation doses in a real-time display inside the angiography room. Cardiovasc Intervent Radiol 33:1210–1214PubMedCrossRef
13.
Zurück zum Zitat Shortt CP, Al-Hashimi H, Malone L et al (2007) Staff radiation doses to the lower extremities in interventional radiology. Cardiovasc Intervent Radiol 30:1206–1209PubMedCrossRef Shortt CP, Al-Hashimi H, Malone L et al (2007) Staff radiation doses to the lower extremities in interventional radiology. Cardiovasc Intervent Radiol 30:1206–1209PubMedCrossRef
14.
Zurück zum Zitat Solomon S, Patricu A, Bohlman M et al (2002) Robotically driven interventions: a method of using CT fluoroscopy without radiation exposure to the physician. Radiology 225:277–282PubMedCrossRef Solomon S, Patricu A, Bohlman M et al (2002) Robotically driven interventions: a method of using CT fluoroscopy without radiation exposure to the physician. Radiology 225:277–282PubMedCrossRef
15.
Zurück zum Zitat Hohl C, Suess C, Wildberger JE et al (2008) Dose reduction during CT fluoroscopy: phantom study of angular beam modulation. Radiology 246:519–525PubMedCrossRef Hohl C, Suess C, Wildberger JE et al (2008) Dose reduction during CT fluoroscopy: phantom study of angular beam modulation. Radiology 246:519–525PubMedCrossRef
16.
Zurück zum Zitat Nickoloff EL, Khandji A, Dutta A (2000) Radiation doses during CT fluoroscopy. Health Phys 79:675–681PubMedCrossRef Nickoloff EL, Khandji A, Dutta A (2000) Radiation doses during CT fluoroscopy. Health Phys 79:675–681PubMedCrossRef
17.
Zurück zum Zitat Parker DL (1982) Optimal short scan convolution reconstruction for fan beam CT. Med Phys 9:254–257PubMedCrossRef Parker DL (1982) Optimal short scan convolution reconstruction for fan beam CT. Med Phys 9:254–257PubMedCrossRef
Metadaten
Titel
Efficacy of Lower-Body Shielding in Computed Tomography Fluoroscopy-Guided Interventions
verfasst von
Andreas H. Mahnken
Martin Sedlmair
Christine Ritter
Rosemarie Banckwitz
Thomas Flohr
Publikationsdatum
01.12.2012
Verlag
Springer-Verlag
Erschienen in
CardioVascular and Interventional Radiology / Ausgabe 6/2012
Print ISSN: 0174-1551
Elektronische ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-011-0338-0

Weitere Artikel der Ausgabe 6/2012

CardioVascular and Interventional Radiology 6/2012 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.