Management of patient dose and image noise in routine pediatric CT abdominal examinations

verfasst von: Francis R. Verdun, D. Lepori, Pascal Monnin, Jean-François Valley, Pierre Schnyder, F. Gudinchet

Erschienen in: European Radiology | Ausgabe 5/2004

Einloggen, um Zugang zu erhalten

Abstract

The aim was to propose a strategy for finding reasonable compromises between image noise and dose as a function of patient weight. Weighted CT dose index (CTDIw) was measured on a multidetector-row CT unit using CTDI test objects of 16, 24 and 32 cm in diameter at 80, 100, 120 and 140 kV. These test objects were then scanned in helical mode using a wide range of tube currents and voltages with a reconstructed slice thickness of 5 mm. For each set of acquisition parameter image noise was measured and the Rose model observer was used to test two strategies for proposing a reasonable compromise between dose and low-contrast detection performance: (1) the use of a unique noise level for all test object diameters, and (2) the use of a unique dose efficacy level defined as the noise reduction per unit dose. Published data were used to define four weight classes and an acquisition protocol was proposed for each class. The protocols have been applied in clinical routine for more than one year. CTDIvol values of 6.7, 9.4, 15.9 and 24.5 mGy were proposed for the following weight classes: 2.5–5, 5–15, 15–30 and 30–50 kg with image noise levels in the range of 10–15 HU. The proposed method allows patient dose and image noise to be controlled in such a way that dose reduction does not impair the detection of low-contrast lesions. The proposed values correspond to high- quality images and can be reduced if only high-contrast organs are assessed.
Literatur
1.
Zurück zum Zitat Shrimpton PC, Jones DG, Hillier MC, Wall BF, Le Heron JC, Faulkner K (1991) Survey of CT practice in the UK. Part 2: dosimetric aspects, NRPB R249. NRPB, Chilton, UK Shrimpton PC, Jones DG, Hillier MC, Wall BF, Le Heron JC, Faulkner K (1991) Survey of CT practice in the UK. Part 2: dosimetric aspects, NRPB R249. NRPB, Chilton, UK
2.
Zurück zum Zitat Brix G, Nagel HD, Stamm G, Veit R, Lechel U, Griebel J, Galanski M (2003) Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey. Eur Radiol 13(8):1979–1991CrossRefPubMed Brix G, Nagel HD, Stamm G, Veit R, Lechel U, Griebel J, Galanski M (2003) Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey. Eur Radiol 13(8):1979–1991CrossRefPubMed
3.
Zurück zum Zitat Golding SJ, Shrimpton PC (2002) Radiation dose in CT: are we meeting the challenge? Br J Radiol 75:1–4PubMed Golding SJ, Shrimpton PC (2002) Radiation dose in CT: are we meeting the challenge? Br J Radiol 75:1–4PubMed
4.
Zurück zum Zitat Kaul A, Bauer B, Bernhardt J, Nosske D, Veit R (1998) Effective doses to members of the public from the diagnostic application of ionizing radiation in Germany. Eur Radiol 7(7):1127–1132CrossRef Kaul A, Bauer B, Bernhardt J, Nosske D, Veit R (1998) Effective doses to members of the public from the diagnostic application of ionizing radiation in Germany. Eur Radiol 7(7):1127–1132CrossRef
5.
Zurück zum Zitat Giacomuzzi SM, Erckert B, Schopf T, Freund MC, Springer P, Dessl A, Jaschke W (1996) The smart-scan procedure of spiral computed tomography. A new method for dose reduction. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 165(1):10–16PubMed Giacomuzzi SM, Erckert B, Schopf T, Freund MC, Springer P, Dessl A, Jaschke W (1996) The smart-scan procedure of spiral computed tomography. A new method for dose reduction. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 165(1):10–16PubMed
6.
Zurück zum Zitat Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9(2):323–328CrossRefPubMed Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9(2):323–328CrossRefPubMed
7.
Zurück zum Zitat Greess H, Wolf H, Baum U, Lell M, Pirkl M, Kalender W, Bautz WA (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMed Greess H, Wolf H, Baum U, Lell M, Pirkl M, Kalender W, Bautz WA (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMed
8.
Zurück zum Zitat Toth TL, Bromberg NB, Pan TS, Rabe J, Woloschek SJ, Li J, Seidenschnur GE (2000) A dose reduction X-ray beam positioning system for high-speed multislice CT scanners. Med Phys 27(12):2659–2668 Toth TL, Bromberg NB, Pan TS, Rabe J, Woloschek SJ, Li J, Seidenschnur GE (2000) A dose reduction X-ray beam positioning system for high-speed multislice CT scanners. Med Phys 27(12):2659–2668
9.
Zurück zum Zitat Kearney SE, Jones P, Meakin K, Garvey CJ (1997) CT scanning of the paranasal sinuses: the effect of reducing mAs. Br J Radiol 70:1071–1074PubMed Kearney SE, Jones P, Meakin K, Garvey CJ (1997) CT scanning of the paranasal sinuses: the effect of reducing mAs. Br J Radiol 70:1071–1074PubMed
10.
Zurück zum Zitat Cohnen M, Fischer H, Hamacher J, Lins E, Kotter R, Modder U (2000) CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. Am J Neuroradiol 21(9):1654–1660PubMed Cohnen M, Fischer H, Hamacher J, Lins E, Kotter R, Modder U (2000) CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. Am J Neuroradiol 21(9):1654–1660PubMed
11.
Zurück zum Zitat Ravenel JG, Scalzetti EM, Huda W, Garrisi W (2001) Radiation exposure and image quality in chest CT examinations. Am J Roentgenol 177:279–284PubMed Ravenel JG, Scalzetti EM, Huda W, Garrisi W (2001) Radiation exposure and image quality in chest CT examinations. Am J Roentgenol 177:279–284PubMed
12.
Zurück zum Zitat Kalra MK, Prasad S, Saini S, Blake MA, Varghese J, Halpern EF, Rhea JT, Thrall JH (2002) Clinical comparison of standard-dose and 50% reduced-dose abdominal CT: effect on image quality. Am J Roentgenol 179:1101–1106PubMed Kalra MK, Prasad S, Saini S, Blake MA, Varghese J, Halpern EF, Rhea JT, Thrall JH (2002) Clinical comparison of standard-dose and 50% reduced-dose abdominal CT: effect on image quality. Am J Roentgenol 179:1101–1106PubMed
13.
Zurück zum Zitat Huda W, Atherton JV, Ware DE, Cumming WA (1997) An approach for the estimation of effective radiation dose at CT in pediatric patients. Radiology 203:417–422PubMed Huda W, Atherton JV, Ware DE, Cumming WA (1997) An approach for the estimation of effective radiation dose at CT in pediatric patients. Radiology 203:417–422PubMed
14.
Zurück zum Zitat Chapple CL, Willis S, Frame J (2002) Effective dose in pediatric computed tomography. Phys Med Biol 47:107–115PubMed Chapple CL, Willis S, Frame J (2002) Effective dose in pediatric computed tomography. Phys Med Biol 47:107–115PubMed
15.
Zurück zum Zitat Paterson A, Frush DP, Donnelly LF (2001) Helical CT of the body: are settings adjusted for pediatric patients? AJR Am J Roentgenol 176:297–301PubMed Paterson A, Frush DP, Donnelly LF (2001) Helical CT of the body: are settings adjusted for pediatric patients? AJR Am J Roentgenol 176:297–301PubMed
16.
Zurück zum Zitat Donnelly LF, Emery KH, Brody AS, Laor T, Gylys-Morin VM, Anton CG, Thomas SR, Frush DP (2001) Minimizing radiation dose for pediatric body, application of single-detector helical CT: strategies at a large children’s hospital. AJR Am J Roentgenol 176:303–306PubMed Donnelly LF, Emery KH, Brody AS, Laor T, Gylys-Morin VM, Anton CG, Thomas SR, Frush DP (2001) Minimizing radiation dose for pediatric body, application of single-detector helical CT: strategies at a large children’s hospital. AJR Am J Roentgenol 176:303–306PubMed
17.
Zurück zum Zitat Wormanns D, Diederich S, Lenzen H, Lange P, Link TM, Ludwig K, Papke K, Hagedorn C, Heindel W (2001) Abdominal spiral CT in children: which radiation exposure is required? Eur Radiol 11(11):2262–2266PubMed Wormanns D, Diederich S, Lenzen H, Lange P, Link TM, Ludwig K, Papke K, Hagedorn C, Heindel W (2001) Abdominal spiral CT in children: which radiation exposure is required? Eur Radiol 11(11):2262–2266PubMed
18.
Zurück zum Zitat Christoph Suess C, Chen X (2002) Dose optimization in pediatric CT: current technology and future innovations. Pediatr Radiol 32:729–734CrossRefPubMed Christoph Suess C, Chen X (2002) Dose optimization in pediatric CT: current technology and future innovations. Pediatr Radiol 32:729–734CrossRefPubMed
19.
Zurück zum Zitat Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as function of patient weight at abdominal CT. Radiology 217:430–435PubMed Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as function of patient weight at abdominal CT. Radiology 217:430–435PubMed
21.
Zurück zum Zitat Boone JM, Geraghty EM, Seibert JA, Wootton-Gorges SL (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228(2):352–360PubMed Boone JM, Geraghty EM, Seibert JA, Wootton-Gorges SL (2003) Dose reduction in pediatric CT: a rational approach. Radiology 228(2):352–360PubMed
22.
Zurück zum Zitat International Electrotechnical Committee (1999) Medical diagnostic X-ray equipment—particular requirements for the safety of X-ray equipment for CT; Standard IEC # 60601-2-44 International Electrotechnical Committee (1999) Medical diagnostic X-ray equipment—particular requirements for the safety of X-ray equipment for CT; Standard IEC # 60601-2-44
23.
Zurück zum Zitat International Electrotechnical Committee (2001) Medical diagnostic X-ray equipment—particular requirements for the safety of X-ray equipment for CT; Standard IEC # 60601-2-44 amendment 1 International Electrotechnical Committee (2001) Medical diagnostic X-ray equipment—particular requirements for the safety of X-ray equipment for CT; Standard IEC # 60601-2-44 amendment 1
24.
Zurück zum Zitat Boone JM, Nelson TR, Lindfors KK, Seibert JA (2001) Dedicated breast CT: radiation dose and image quality evaluation. Radiology 221:657–667PubMed Boone JM, Nelson TR, Lindfors KK, Seibert JA (2001) Dedicated breast CT: radiation dose and image quality evaluation. Radiology 221:657–667PubMed
25.
Zurück zum Zitat Shrimpton PC, Wall BF (2000) Reference doses for pediatric computed tomography. Radiat Protect Dosim 90(1–2):249–252 Shrimpton PC, Wall BF (2000) Reference doses for pediatric computed tomography. Radiat Protect Dosim 90(1–2):249–252
26.
Zurück zum Zitat Frush DP, Soden B, Frush KS, Lowry C (2002) Improved pediatric multidetector body CT using a size-based color-coded format. Am J Roentgenol 178:721–726PubMed Frush DP, Soden B, Frush KS, Lowry C (2002) Improved pediatric multidetector body CT using a size-based color-coded format. Am J Roentgenol 178:721–726PubMed
27.
Zurück zum Zitat Greess H, Nömayr A, Wolf H, Baum U, Lell M, Böwing B, Kalender W, Bautz WA (2002) Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol 12:1571–1576PubMed Greess H, Nömayr A, Wolf H, Baum U, Lell M, Böwing B, Kalender W, Bautz WA (2002) Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose). Eur Radiol 12:1571–1576PubMed
28.
Zurück zum Zitat Cohnen M, Poll LJ, Puettmann C, Ewen K, Saleh A, Modder U (2003) Effective doses in standard protocoles for multi-slice CT scanning. Eur Radiol 13(5):1148–1153PubMed Cohnen M, Poll LJ, Puettmann C, Ewen K, Saleh A, Modder U (2003) Effective doses in standard protocoles for multi-slice CT scanning. Eur Radiol 13(5):1148–1153PubMed
29.
Zurück zum Zitat Commission of the European Communities (2000) European guidelines on quality criteria for computed tomography, Luxembourg EUR 16262 EN Commission of the European Communities (2000) European guidelines on quality criteria for computed tomography, Luxembourg EUR 16262 EN
30.
Zurück zum Zitat Starck G, Lönn L, Cederblad Å, Forsell-Aronsson E, Sjöström L, Alpsten M (2002) A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol 75:140–150PubMed Starck G, Lönn L, Cederblad Å, Forsell-Aronsson E, Sjöström L, Alpsten M (2002) A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol 75:140–150PubMed
Metadaten
Titel
Management of patient dose and image noise in routine pediatric CT abdominal examinations
verfasst von
Francis R. Verdun
D. Lepori
Pascal Monnin
Jean-François Valley
Pierre Schnyder
F. Gudinchet
Publikationsdatum
01.05.2004
Erschienen in
European Radiology / Ausgabe 5/2004
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-003-2206-z

Weitere Artikel der Ausgabe 5/2004

European Radiology 5/2004 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.