01.01.2006 | Vascular-Interventional

Whole-body magnetic resonance angiography of patients using a standard clinical scanner

verfasst von: Tomas Hansen, Johan Wikström, Mats-Ola Eriksson, Anders Lundberg, Lars Johansson, Christer Ljungman, Romhild Hoogeven, Håkan Ahlström

Erschienen in: European Radiology | Ausgabe 1/2006

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this study was to evaluate the technique of whole-body magnetic resonance angiography (MRA) of patients with a standard clinical scanner. Thirty-three patients referred for stenoses, occlusions, aneurysms, assessment of patency of vascular grafts, vasculitis and vascular aplasia were examined in a 1.5-T scanner using its standard body coil. Three-dimensional sequences were acquired in four stations after administration of one intravenous injection of 40 ml conventional gadolinium contrast agent. Different vessel segments were evaluated as either diagnostic or nondiagnostic and regarding the presence of stenoses with more than 50% diameter reduction, occlusions or aneurysms. Of 923 vessel segments, 67 were not evaluable because of poor contrast filling (n=31), motion artefacts (n=20), venous overlap (n=12) and other reasons (n=4). Stenoses of more than 50%, occlusions or aneurysms were observed in 26 patients (129 segments). In nine patients additional unsuspected pathology was found. In 10 out of 14 patients (71/79 segments) there was conformity between MRA and digital subtraction angiography regarding the grade of stenosis. This study shows that whole-body MRA with a standard clinical scanner is feasible. Motion artefacts and the timing of the contrast agent through the different segments are still problems to be solved.
Literatur
1.
Zurück zum Zitat Prince MR (1994) Gadolinium-enhanced MR aortography. Radiology 191:155–164PubMed Prince MR (1994) Gadolinium-enhanced MR aortography. Radiology 191:155–164PubMed
2.
Zurück zum Zitat Catalano C, Fraioli F, Laghi A et al (2004) Infrarenal aortic and lower-extremity arterial disease: diagnostic performance of multi-detector row CT angiography. Radiology 231:555–563PubMed Catalano C, Fraioli F, Laghi A et al (2004) Infrarenal aortic and lower-extremity arterial disease: diagnostic performance of multi-detector row CT angiography. Radiology 231:555–563PubMed
3.
Zurück zum Zitat Patel SG, Collie DA, Wardlaw JM et al (2002) Outcome, observer reliability, and patient preferences if CTA, MRA, or Doppler ultrasound were used, individually or together, instead of digital subtraction angiography before carotid endarterectomy. J Neurol Neurosurg Psychiatry 73:21–28CrossRefPubMed Patel SG, Collie DA, Wardlaw JM et al (2002) Outcome, observer reliability, and patient preferences if CTA, MRA, or Doppler ultrasound were used, individually or together, instead of digital subtraction angiography before carotid endarterectomy. J Neurol Neurosurg Psychiatry 73:21–28CrossRefPubMed
4.
Zurück zum Zitat Ho KY, Leiner T, de Haan MW et al (1998) Peripheral vascular tree stenoses: evaluation with moving-bed infusion-tracking MR angiography. Radiology 206:683–692PubMed Ho KY, Leiner T, de Haan MW et al (1998) Peripheral vascular tree stenoses: evaluation with moving-bed infusion-tracking MR angiography. Radiology 206:683–692PubMed
5.
Zurück zum Zitat Yucel EK, Anderson CM, Edelman RR et al (1999) AHA scientific statement. Magnetic resonance angiography: update on applications for extracranial arteries. Circulation 100:2284–2301PubMed Yucel EK, Anderson CM, Edelman RR et al (1999) AHA scientific statement. Magnetic resonance angiography: update on applications for extracranial arteries. Circulation 100:2284–2301PubMed
6.
Zurück zum Zitat Meany JFMRJ, Chakraverty S, Robertson I, Kessel D, Radjenovic A, Kouwenhoven M, Kassner A, Smith MA (1999) Stepping-table gadolinium-enhanced digital subtraction MR angiography of the aorta and lower extremity arteries: preliminary experience. Radiology 211:59–67PubMed Meany JFMRJ, Chakraverty S, Robertson I, Kessel D, Radjenovic A, Kouwenhoven M, Kassner A, Smith MA (1999) Stepping-table gadolinium-enhanced digital subtraction MR angiography of the aorta and lower extremity arteries: preliminary experience. Radiology 211:59–67PubMed
7.
Zurück zum Zitat Ruehm SG, Goyen M, Barkhausen J et al (2001) Rapid magnetic resonance angiography for detection of atherosclerosis. Lancet 357:1086–1091CrossRefPubMed Ruehm SG, Goyen M, Barkhausen J et al (2001) Rapid magnetic resonance angiography for detection of atherosclerosis. Lancet 357:1086–1091CrossRefPubMed
8.
Zurück zum Zitat NASCET (1998) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. North American Symptomatic Carotid Endarterectomy Trial Collaborators. N Engl J Med 325:445–453 NASCET (1998) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. North American Symptomatic Carotid Endarterectomy Trial Collaborators. N Engl J Med 325:445–453
9.
Zurück zum Zitat Goyen M, Herborn CU, Kroger K et al (2003) Detection of atherosclerosis: systemic imaging for systemic disease with whole-body three-dimensional MR angiography—initial experience. Radiology 227:277–282PubMed Goyen M, Herborn CU, Kroger K et al (2003) Detection of atherosclerosis: systemic imaging for systemic disease with whole-body three-dimensional MR angiography—initial experience. Radiology 227:277–282PubMed
10.
Zurück zum Zitat Goyen M, Quick HH, Debatin JF et al (2002) Whole-body three-dimensional MR angiography with a rolling table platform: initial clinical experience. Radiology 224:270–277PubMed Goyen M, Quick HH, Debatin JF et al (2002) Whole-body three-dimensional MR angiography with a rolling table platform: initial clinical experience. Radiology 224:270–277PubMed
Metadaten
Titel
Whole-body magnetic resonance angiography of patients using a standard clinical scanner
verfasst von
Tomas Hansen
Johan Wikström
Mats-Ola Eriksson
Anders Lundberg
Lars Johansson
Christer Ljungman
Romhild Hoogeven
Håkan Ahlström
Publikationsdatum
01.01.2006
Erschienen in
European Radiology / Ausgabe 1/2006
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-2681-5

Weitere Artikel der Ausgabe 1/2006

European Radiology 1/2006 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.