First performance evaluation of a dual-source CT (DSCT) system

verfasst von: Thomas G. Flohr, Cynthia H. McCollough, Herbert Bruder, Martin Petersilka, Klaus Gruber, Christoph Süβ, Michael Grasruck, Karl Stierstorfer, Bernhard Krauss, Rainer Raupach, Andrew N. Primak, Axel Küttner, Stefan Achenbach, Christoph Becker, Andreas Kopp, Bernd M. Ohnesorge

Erschienen in: European Radiology | Ausgabe 2/2006

Einloggen, um Zugang zu erhalten

Abstract

We present a performance evaluation of a recently introduced dual-source computed tomography (DSCT) system equipped with two X-ray tubes and two corresponding detectors, mounted onto the rotating gantry with an angular offset of 90°. We introduce the system concept and derive its consequences and potential benefits for echocardiograph (ECG)-controlled cardiac CT and for general radiology applications. We evaluate both temporal and spatial resolution by means of phantom scans. We present first patient scans to illustrate the performance of DSCT for ECG-gated cardiac imaging, and we demonstrate first results using a dual-energy acquisition mode. Using ECG-gated single-segment reconstruction, the DSCT system provides 83 ms temporal resolution independent of the patient’s heart rate for coronary CT angiography (CTA) and evaluation of basic functional parameters. With dual-segment reconstruction, the mean temporal resolution is 60 ms (minimum temporal resolution 42 ms) for advanced functional evaluation. The z-flying focal spot technique implemented in the evaluated DSCT system allows 0.4 mm cylinders to be resolved at all heart rates. First clinical experience shows a considerably increased robustness for the imaging of patients with high heart rates. As a potential application of the dual-energy acquisition mode, the automatic separation of bones and iodine-filled vessels is demonstrated.
Literatur
1.
Zurück zum Zitat Ohnesorge B, Flohr T, Becker C, Kopp A, Schoepf U, Baum U, Knez A, Klingenbeck Regn K, Reiser M (2000) Cardiac imaging by means of electro- cardiographically gated multisection spiral CT-initial experience. Radiology 217:564–571PubMed Ohnesorge B, Flohr T, Becker C, Kopp A, Schoepf U, Baum U, Knez A, Klingenbeck Regn K, Reiser M (2000) Cardiac imaging by means of electro- cardiographically gated multisection spiral CT-initial experience. Radiology 217:564–571PubMed
2.
Zurück zum Zitat Kachelriess M, Ulzheimer S, Kalender W (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef Kachelriess M, Ulzheimer S, Kalender W (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef
3.
Zurück zum Zitat Taguchi K, Anno H (2000) High temporal resolution for multi-slice helical computed tomography. Med Phys 27(5):861–872PubMedCrossRef Taguchi K, Anno H (2000) High temporal resolution for multi-slice helical computed tomography. Med Phys 27(5):861–872PubMedCrossRef
4.
Zurück zum Zitat Hong C, Becker CR, Huber A, Schoepf UJ, Ohnesorge B, Knez A, Brüning R, Reiser MF (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMed Hong C, Becker CR, Huber A, Schoepf UJ, Ohnesorge B, Knez A, Brüning R, Reiser MF (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMed
5.
Zurück zum Zitat Achenbach S, Ulzheimer S, Baum U et al (2000) Noninvasive coronary angiography by retrospectively ECG-gated multi-slice spiral CT. Circulation 102:2823–2828PubMed Achenbach S, Ulzheimer S, Baum U et al (2000) Noninvasive coronary angiography by retrospectively ECG-gated multi-slice spiral CT. Circulation 102:2823–2828PubMed
6.
Zurück zum Zitat Becker C, Knez A, Ohnesorge B, Schöpf U, Reiser M (2000) Imaging of non calcified coronary plaques using helical CT with retrospective EKG gating. AJR Am J Roentgenol175:423–424 Becker C, Knez A, Ohnesorge B, Schöpf U, Reiser M (2000) Imaging of non calcified coronary plaques using helical CT with retrospective EKG gating. AJR Am J Roentgenol175:423–424
7.
Zurück zum Zitat Knez A, Becker C, Leber A, Ohnesorge B, Reiser M, Haberl R (2000) Non-invasive assessment of coronary artery stenoses with multidetector helical computed tomography. Circulation 101:e221–e222PubMed Knez A, Becker C, Leber A, Ohnesorge B, Reiser M, Haberl R (2000) Non-invasive assessment of coronary artery stenoses with multidetector helical computed tomography. Circulation 101:e221–e222PubMed
8.
Zurück zum Zitat Nieman K, Oudkerk M, Rensing B, van Oijen P, Munne A, van Geuns R, de Feyter P (2001) Coronary angiography with multi-slice computed tomography. Lancet 357:599–603PubMedCrossRef Nieman K, Oudkerk M, Rensing B, van Oijen P, Munne A, van Geuns R, de Feyter P (2001) Coronary angiography with multi-slice computed tomography. Lancet 357:599–603PubMedCrossRef
9.
Zurück zum Zitat Schroeder S, Kopp A, Baumbach A, Meisner C, Kuettner A, Georg C, Ohnesorge B, Herdeg C, Claussen C, Karsch K (2001) Noninvasive detection and evaluation of atherosclerotic coronary plaques with multi-slice computed tomography. J Am Coll Cardiol 37(5):1430–1435PubMedCrossRef Schroeder S, Kopp A, Baumbach A, Meisner C, Kuettner A, Georg C, Ohnesorge B, Herdeg C, Claussen C, Karsch K (2001) Noninvasive detection and evaluation of atherosclerotic coronary plaques with multi-slice computed tomography. J Am Coll Cardiol 37(5):1430–1435PubMedCrossRef
10.
Zurück zum Zitat Schroeder S, Flohr T, Kopp A F, Meisner C, Kuettner A, Herdeg C, Baumbach A, Ohnesorge B (2001) Accuracy of density measurements within plaques located in artificial coronary arteries by X-ray multislice CT: results of a phantom study. J Comput Assist Tomogr 25(6):900–906PubMedCrossRef Schroeder S, Flohr T, Kopp A F, Meisner C, Kuettner A, Herdeg C, Baumbach A, Ohnesorge B (2001) Accuracy of density measurements within plaques located in artificial coronary arteries by X-ray multislice CT: results of a phantom study. J Comput Assist Tomogr 25(6):900–906PubMedCrossRef
11.
Zurück zum Zitat Kopp AF, Ohnesorge B, Becker C, Schröder S, Heuschmid M, Küttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurement with multidetector-row versus electron beam CT. Radiology 225:113–119PubMed Kopp AF, Ohnesorge B, Becker C, Schröder S, Heuschmid M, Küttner A, Kuzo R, Claussen CD (2002) Reproducibility and accuracy of coronary calcium measurement with multidetector-row versus electron beam CT. Radiology 225:113–119PubMed
12.
Zurück zum Zitat Becker CR, Kleffel T, Crispin A, Knez A, Young Y, Schöpf UJ, Haberl R, Reiser MF (2001) Coronary artery calcium measurement: agreement of multirow detector and electron beam CT. AJR Am J Roentgenol 176:1295–1298 Becker CR, Kleffel T, Crispin A, Knez A, Young Y, Schöpf UJ, Haberl R, Reiser MF (2001) Coronary artery calcium measurement: agreement of multirow detector and electron beam CT. AJR Am J Roentgenol 176:1295–1298
13.
Zurück zum Zitat Juergens KU, Grude M, Fallenberg EM, Heindel W, Fischbach R (2002) Using ECG-gated multidetector CT to evaluate global left ventricular myocardial function in patients with coronary artery disease. AJR Am J Roentgenol 179:1545–1550 Juergens KU, Grude M, Fallenberg EM, Heindel W, Fischbach R (2002) Using ECG-gated multidetector CT to evaluate global left ventricular myocardial function in patients with coronary artery disease. AJR Am J Roentgenol 179:1545–1550
14.
Zurück zum Zitat Kopp A, Schröder S, Küttner A et al (2001) Coronary arteries: retrospectively ECG-gated multidetector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMed Kopp A, Schröder S, Küttner A et al (2001) Coronary arteries: retrospectively ECG-gated multidetector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMed
15.
Zurück zum Zitat Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B (2002) New technical developments in multislice CT, part 2: sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac imaging. Röfo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 174:1022–1027PubMedCrossRef Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B (2002) New technical developments in multislice CT, part 2: sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac imaging. Röfo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 174:1022–1027PubMedCrossRef
16.
Zurück zum Zitat Flohr T, Schoepf UJ, Kuettner A, Halliburton S, Bruder H, Suess C, Schmidt B, Hofmann L, Yucel E K, Schaller S, Ohnesorge B (2003) Advances in cardiac imaging with 16-section CT-systems. Acad Radiol 10:386–401PubMedCrossRef Flohr T, Schoepf UJ, Kuettner A, Halliburton S, Bruder H, Suess C, Schmidt B, Hofmann L, Yucel E K, Schaller S, Ohnesorge B (2003) Advances in cardiac imaging with 16-section CT-systems. Acad Radiol 10:386–401PubMedCrossRef
17.
Zurück zum Zitat Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PMT, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PMT, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef
18.
Zurück zum Zitat Ropers D, Baum U, Pohle K et al (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef Ropers D, Baum U, Pohle K et al (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666PubMedCrossRef
19.
Zurück zum Zitat Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Image quality and diagnostic accuracy of non-invasive coronary imaging with 16 detector slice spiral computed tomography with 188 ms temporal resolution. Heart 91(7):938–941PubMedCrossRef Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Image quality and diagnostic accuracy of non-invasive coronary imaging with 16 detector slice spiral computed tomography with 188 ms temporal resolution. Heart 91(7):938–941PubMedCrossRef
20.
Zurück zum Zitat Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms temporal resolution. J Am Coll Cardiol 45(1):123–127PubMedCrossRef Kuettner A, Beck T, Drosch T, Kettering K, Heuschmid M, Burgstahler C, Claussen CD, Kopp AF, Schroeder S (2005) Diagnostic accuracy of noninvasive coronary imaging using 16-detector slice spiral computed tomography with 188 ms temporal resolution. J Am Coll Cardiol 45(1):123–127PubMedCrossRef
21.
Zurück zum Zitat Flohr T, Stierstorfer K, Raupach R, Ulzheimer S, Bruder H (2004) Performance evaluation of a 64-slice CT-system with z-flying focal spot. Röfo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:1803–1810PubMedCrossRef Flohr T, Stierstorfer K, Raupach R, Ulzheimer S, Bruder H (2004) Performance evaluation of a 64-slice CT-system with z-flying focal spot. Röfo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 176:1803–1810PubMedCrossRef
22.
Zurück zum Zitat Leber A W, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography. J Am Coll Cardiol 46(1):147–154PubMedCrossRef Leber A W, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography. J Am Coll Cardiol 46(1):147–154PubMedCrossRef
23.
Zurück zum Zitat Raff G L, Gallagher M J, O’Neill W W, Goldstein J A (2005) Diagnostic accuracy of non-invasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef Raff G L, Gallagher M J, O’Neill W W, Goldstein J A (2005) Diagnostic accuracy of non-invasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef
24.
Zurück zum Zitat Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, de Feyter PJ (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–2323PubMedCrossRef Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, de Feyter PJ (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–2323PubMedCrossRef
25.
Zurück zum Zitat Wintersperger BJ, Nikolaou K, von Ziegler F et al (2005) Image quality and reconstruction timing of 64-slice coronary CT augiography with 0.33s/360° rotation speed (in press) Wintersperger BJ, Nikolaou K, von Ziegler F et al (2005) Image quality and reconstruction timing of 64-slice coronary CT augiography with 0.33s/360° rotation speed (in press)
26.
Zurück zum Zitat Halliburton SS, Stillman AE, Flohr T, Ohnesorge B, Obuchowski N, Lieber M, Karim W, Kuzmiak S, Kasper JM, White RD (2003) Do segmented reconstruction algorithms for cardiac multi-slice computed tomography improve image quality? Herz 28(1):20–31PubMedCrossRef Halliburton SS, Stillman AE, Flohr T, Ohnesorge B, Obuchowski N, Lieber M, Karim W, Kuzmiak S, Kasper JM, White RD (2003) Do segmented reconstruction algorithms for cardiac multi-slice computed tomography improve image quality? Herz 28(1):20–31PubMedCrossRef
27.
Zurück zum Zitat Budoff M, Georgiou D, Brody A et al (1996) Ultrafast computed tomography as a diagnostic modality in the detection of coronary artery disease: a multicenter study. Circulation 93:898–904PubMed Budoff M, Georgiou D, Brody A et al (1996) Ultrafast computed tomography as a diagnostic modality in the detection of coronary artery disease: a multicenter study. Circulation 93:898–904PubMed
28.
Zurück zum Zitat Wielopolski P, van Geuns R, de Feyter P, Oudkerk M (1998) Coronary arteries. Eur Radiol 8:873–885PubMedCrossRef Wielopolski P, van Geuns R, de Feyter P, Oudkerk M (1998) Coronary arteries. Eur Radiol 8:873–885PubMedCrossRef
29.
Zurück zum Zitat Achenbach S, Moshage W, Ropers D, Bachmann K (1998) Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron-beam CT of the coronary arteries. AJR Am J Roentgenol 170:895–899 Achenbach S, Moshage W, Ropers D, Bachmann K (1998) Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron-beam CT of the coronary arteries. AJR Am J Roentgenol 170:895–899
30.
Zurück zum Zitat Becker C, Knez A, Jakobs T et al (1999) Detection and quantification of coronary artery calcification with electron-beam and conventional CT. Eur Radiol 9:620–624PubMedCrossRef Becker C, Knez A, Jakobs T et al (1999) Detection and quantification of coronary artery calcification with electron-beam and conventional CT. Eur Radiol 9:620–624PubMedCrossRef
31.
Zurück zum Zitat McCollough CH, Zink FE (1994) The technical design and performance of ultrafast computed tomography. Radiol Clin North Am 32(3):521–536PubMed McCollough CH, Zink FE (1994) The technical design and performance of ultrafast computed tomography. Radiol Clin North Am 32(3):521–536PubMed
32.
Zurück zum Zitat McCollough CH, Zink FE, Morin R (1994) Radiation dosimetry for electron beam CT. Radiology 192(3):637–643PubMed McCollough CH, Zink FE, Morin R (1994) Radiation dosimetry for electron beam CT. Radiology 192(3):637–643PubMed
33.
Zurück zum Zitat McCollough CH, Kanal KM, Lanutti N, Ryan KJ (1999) Experimental determination of section sensitivity profiles and image noise in electron beam computed tomography. Med Phys 26(2):287–295PubMedCrossRef McCollough CH, Kanal KM, Lanutti N, Ryan KJ (1999) Experimental determination of section sensitivity profiles and image noise in electron beam computed tomography. Med Phys 26(2):287–295PubMedCrossRef
34.
Zurück zum Zitat Robb R, Ritman E (1979) High speed synchronous volume computed tomography of the heart. Radiology 133:655–661PubMed Robb R, Ritman E (1979) High speed synchronous volume computed tomography of the heart. Radiology 133:655–661PubMed
35.
Zurück zum Zitat Ritman E, Kinsey J, Robb R, Gilbert B, Harris L, Wood E (1980) Three-dimensional imaging of heart, lungs, and circulation. Science 210:273–280PubMed Ritman E, Kinsey J, Robb R, Gilbert B, Harris L, Wood E (1980) Three-dimensional imaging of heart, lungs, and circulation. Science 210:273–280PubMed
36.
Zurück zum Zitat Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough C H (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32(8):2536–2547PubMedCrossRef Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough C H (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32(8):2536–2547PubMedCrossRef
37.
Zurück zum Zitat Schardt P, Deuringer J, Freudenberger J, Hell E, Knuepfer W, Mattern D, Schild M (2004) New X-ray tube performance in computed tomography by introducing the rotating envelope tube technology. Med Phys 31(9):2699–2706PubMedCrossRef Schardt P, Deuringer J, Freudenberger J, Hell E, Knuepfer W, Mattern D, Schild M (2004) New X-ray tube performance in computed tomography by introducing the rotating envelope tube technology. Med Phys 31(9):2699–2706PubMedCrossRef
38.
Zurück zum Zitat Parker D (1982) Optimal short scan convolution reconstruction for fanbeam CT. Med Phys 9(2):254–257PubMedCrossRef Parker D (1982) Optimal short scan convolution reconstruction for fanbeam CT. Med Phys 9(2):254–257PubMedCrossRef
39.
Zurück zum Zitat Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for ECG-gated multi-slice spiral CT of the heart. J Comput Assist Tomogr 25(6):907–923PubMedCrossRef Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for ECG-gated multi-slice spiral CT of the heart. J Comput Assist Tomogr 25(6):907–923PubMedCrossRef
40.
Zurück zum Zitat Kalender WA, Perman WH, Vetter JR, Klotz E (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13(3):334–339PubMedCrossRef Kalender WA, Perman WH, Vetter JR, Klotz E (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. I. Phantom studies. Med Phys 13(3):334–339PubMedCrossRef
41.
Zurück zum Zitat Vetter JR, Perman WH, Kalender WA, Mazess RB, Holden JE (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. II. Determination of vertebral bone mineral content. Med Phys 13(3):340–343PubMedCrossRef Vetter JR, Perman WH, Kalender WA, Mazess RB, Holden JE (1986) Evaluation of a prototype dual-energy computed tomographic apparatus. II. Determination of vertebral bone mineral content. Med Phys 13(3):340–343PubMedCrossRef
42.
Zurück zum Zitat Achenbach S, Ropers D, Holle J, et al (2000) In-plane coronary arterial motion velocity: measurement with electron beam CT. Radiology 216:457–463PubMed Achenbach S, Ropers D, Holle J, et al (2000) In-plane coronary arterial motion velocity: measurement with electron beam CT. Radiology 216:457–463PubMed
Metadaten
Titel
First performance evaluation of a dual-source CT (DSCT) system
verfasst von
Thomas G. Flohr
Cynthia H. McCollough
Herbert Bruder
Martin Petersilka
Klaus Gruber
Christoph Süβ
Michael Grasruck
Karl Stierstorfer
Bernhard Krauss
Rainer Raupach
Andrew N. Primak
Axel Küttner
Stefan Achenbach
Christoph Becker
Andreas Kopp
Bernd M. Ohnesorge
Publikationsdatum
01.02.2006
Erschienen in
European Radiology / Ausgabe 2/2006
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-005-2919-2

Weitere Artikel der Ausgabe 2/2006

European Radiology 2/2006 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.