Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

verfasst von: Sebastian Leschka, Lars Husmann, Lotus M. Desbiolles, Oliver Gaemperli, Tiziano Schepis, Pascal Koepfli, Thomas Boehm, Borut Marincek, Philipp A. Kaufmann, Hatem Alkadhi

Erschienen in: European Radiology | Ausgabe 9/2006

Einloggen, um Zugang zu erhalten

Abstract

The reconstruction intervals providing best image quality for non-invasive coronary angiography with 64-slice computed tomography (CT) were evaluated. Contrast-enhanced, retrospectively electrocardiography (ECG)-gated 64-slice CT coronary angiography was performed in 80 patients (47 male, 33 female; mean age 62.1±10.6 years). Thirteen data sets were reconstructed in 5% increments from 20 to 80% of the R-R interval. Depending on the average heart rate during scanning, patients were grouped as <65 bpm (n=49) and ≥65 bpm (n=31). Two blinded and independent readers assessed the image quality of each coronary segment with a diameter ≥1.5 mm using the following scores: 1, no motion artifacts; 2, minor artifacts; 3, moderate artifacts; 4, severe artifacts; and 5, not evaluative. The average heart rate was 63.3±13.1 bpm (range 38–102). Acceptable image quality (scores 1–3) was achieved in 99.1% of all coronary segments (1,162/1,172; mean image quality score 1.55±0.77) in the best reconstruction interval. Best image quality was found at 60% and 65% of the R-R interval for all patients and for each heart rate subgroup, whereas motion artifacts occurred significantly more often (P<0.01) at other reconstruction intervals. At heart rates <65 bpm, acceptable image quality was found in all coronary segments at 60%. At heart rates ≥65 bpm, the whole coronary artery tree could be visualized with acceptable image quality in 87% (27/31) of the patients at 60%, while ten segments in four patients were rated as non-diagnostic (scores 4–5) at any reconstruction interval. In conclusion, 64-slice CT coronary angiography provides best overall image quality in mid-diastole. At heart rates <65 bpm, diagnostic image quality of all coronary segments can be obtained at a single reconstruction interval of 60%.
Literatur
1.
Zurück zum Zitat Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef
2.
Zurück zum Zitat Raff GL, Gallagher MJ, O'Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef Raff GL, Gallagher MJ, O'Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef
3.
Zurück zum Zitat Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef
4.
Zurück zum Zitat Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef
5.
Zurück zum Zitat Pugliese F et al (2005) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol:1–8 Pugliese F et al (2005) Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris. Eur Radiol:1–8
6.
Zurück zum Zitat Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32:2536–2547PubMedCrossRef Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32:2536–2547PubMedCrossRef
7.
Zurück zum Zitat Giesler T et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol 179:911–916PubMed Giesler T et al (2002) Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol 179:911–916PubMed
8.
Zurück zum Zitat Hong C et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMedCrossRef Hong C et al (2001) ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology 220:712–717PubMedCrossRef
9.
Zurück zum Zitat Kopp AF et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMedCrossRef Kopp AF et al (2001) Coronary arteries: retrospectively ECG-gated multi-detector row CT angiography with selective optimization of the image reconstruction window. Radiology 221:683–688PubMedCrossRef
10.
Zurück zum Zitat Hamoir XL et al (2005) Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol 15:296–304PubMedCrossRef Hamoir XL et al (2005) Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time. Eur Radiol 15:296–304PubMedCrossRef
11.
Zurück zum Zitat Bley TA et al (2005) Computed tomography coronary angiography with 370-millisecond gantry rotation time: evaluation of the best image reconstruction interval. J Comput Assist Tomogr 29:1–5PubMedCrossRef Bley TA et al (2005) Computed tomography coronary angiography with 370-millisecond gantry rotation time: evaluation of the best image reconstruction interval. J Comput Assist Tomogr 29:1–5PubMedCrossRef
12.
Zurück zum Zitat Sanz J et al (2005) The importance of end-systole for optimal reconstruction protocol of coronary angiography with 16-slice multidetector computed tomography. Invest Radiol 40:155–163PubMedCrossRef Sanz J et al (2005) The importance of end-systole for optimal reconstruction protocol of coronary angiography with 16-slice multidetector computed tomography. Invest Radiol 40:155–163PubMedCrossRef
13.
Zurück zum Zitat Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef
14.
Zurück zum Zitat Kachelriess M, Ulzheimer S, Kalender WA (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef Kachelriess M, Ulzheimer S, Kalender WA (2000) ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys 27:1881–1902PubMedCrossRef
15.
Zurück zum Zitat Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25:907–923PubMedCrossRef Flohr T, Ohnesorge B (2001) Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart. J Comput Assist Tomogr 25:907–923PubMedCrossRef
16.
Zurück zum Zitat Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc committee for grading of coronary artery disease, council on cardiovascular surgery, American heart association. Circulation 51:5–40PubMed
17.
Zurück zum Zitat Shim SS, Kim Y, Lim SM (2005) Improvement of image quality with beta-blocker premedication on ECG-gated 16-MDCT coronary angiography. AJR Am J Roentgenol 184:649–654PubMed Shim SS, Kim Y, Lim SM (2005) Improvement of image quality with beta-blocker premedication on ECG-gated 16-MDCT coronary angiography. AJR Am J Roentgenol 184:649–654PubMed
18.
Zurück zum Zitat Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef
19.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
20.
Zurück zum Zitat Choi HS et al (2004) Pitfalls, artifacts, and remedies in multi- detector row CT coronary angiography. Radiographics 24:787–800PubMedCrossRef Choi HS et al (2004) Pitfalls, artifacts, and remedies in multi- detector row CT coronary angiography. Radiographics 24:787–800PubMedCrossRef
21.
Zurück zum Zitat Hoffmann MH et al (2005) Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology 234:86–97PubMedCrossRef Hoffmann MH et al (2005) Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology 234:86–97PubMedCrossRef
22.
Zurück zum Zitat Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054PubMedCrossRef
23.
Zurück zum Zitat Cademartiri F et al (2005) Diagnostic accuracy of multislice computed tomography coronary angiography is improved at low heart rates. Int J Cardiovasc Imaging:1–5 Cademartiri F et al (2005) Diagnostic accuracy of multislice computed tomography coronary angiography is improved at low heart rates. Int J Cardiovasc Imaging:1–5
Metadaten
Titel
Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT
verfasst von
Sebastian Leschka
Lars Husmann
Lotus M. Desbiolles
Oliver Gaemperli
Tiziano Schepis
Pascal Koepfli
Thomas Boehm
Borut Marincek
Philipp A. Kaufmann
Hatem Alkadhi
Publikationsdatum
01.09.2006
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 9/2006
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-006-0262-x

Weitere Artikel der Ausgabe 9/2006

European Radiology 9/2006 Zur Ausgabe

Calendar of Events

Calendar of Events

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.