01.03.2009 | Vascular-Interventional

Ascending aorta measurements as assessed by ECG-gated multi-detector computed tomography: a pilot study to establish normative values for transcatheter therapies

verfasst von: Tri-Linh C. Lu, Christoph H. Huber, Elena Rizzo, Jashmid Dehmeshki, Ludwig K. von Segesser, Salah D. Qanadli

Erschienen in: European Radiology | Ausgabe 3/2009

Einloggen, um Zugang zu erhalten

Abstract

The aim of this study was to provide an insight into normative values of the ascending aorta in regards to novel endovascular procedures using ECG-gated multi-detector CT angiography. Seventy-seven adult patients without ascending aortic abnormalities were evaluated. Measurements at relevant levels of the aortic root and ascending aorta were obtained. Diameter variations of the ascending aorta during cardiac cycle were also considered. Mean diameters (mm) were as follows: LV outflow tract 20.3 ± 3.4, coronary sinus 34.2 ± 4.1, sino-tubular junction 29.7 ± 3.4 and mid ascending aorta 32.7 ± 3.8 with coefficients of variation (CV) ranging from 12 to 17%. Mean distances (mm) were: from the plane passing through the proximal insertions of the aortic valve cusps to the right brachio-cephalic artery (BCA) 92.6 ± 11.8, from the plane passing through the proximal insertions of the aortic valve cusps to the proximal coronary ostium 12.1 ± 3.7, and between both coronary ostia 7.2 ± 3.1, minimal arc of the ascending aorta from left coronary ostium to right BCA 52.9 ± 9.5, and the fibrous continuity between the aortic valve and the anterior leaflet of the mitral valve 14.6 ± 3.3, CV 13–43%. Mean aortic valve area was 582.0 ± 131.9 mm2. The variation of the antero-posterior and transverse diameters of the ascending aorta during the cardiac cycle were 8.4% and 7.3%, respectively. Results showed large inter-individual variations in diameters and distances but with limited intra-individual variations during the cardiac cycle. A personalized approach for planning endovascular devices must be considered.
Literatur
1.
Zurück zum Zitat Dubost C, Allary M, Oeconomos N (1951) Aneurysm of the abdominal aorta treated by resection and graft. Arch Mal Coeur Vaiss 44:848–851PubMed Dubost C, Allary M, Oeconomos N (1951) Aneurysm of the abdominal aorta treated by resection and graft. Arch Mal Coeur Vaiss 44:848–851PubMed
2.
Zurück zum Zitat Dubost C, Allary M, Oeconomos N (1952) Resection of an aneurysm of the abdominal aorta: reestablishment of the continuity by a preserved human arterial graft, with result after five months. AMA Arch Surg 64:405–408PubMed Dubost C, Allary M, Oeconomos N (1952) Resection of an aneurysm of the abdominal aorta: reestablishment of the continuity by a preserved human arterial graft, with result after five months. AMA Arch Surg 64:405–408PubMed
3.
Zurück zum Zitat Creech O Jr (1966) Endo-aneurysmorrhaphy and treatment of aortic aneurysm. Ann Surg 164:935–946CrossRefPubMed Creech O Jr (1966) Endo-aneurysmorrhaphy and treatment of aortic aneurysm. Ann Surg 164:935–946CrossRefPubMed
4.
Zurück zum Zitat Parodi JC, Palmaz JC, Barone HD (1991) Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 5:491–499CrossRefPubMed Parodi JC, Palmaz JC, Barone HD (1991) Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 5:491–499CrossRefPubMed
5.
Zurück zum Zitat Parker MV, O’Donnell SD, Chang AS, Johnson CA, Gillespie DL, Goff JM, Rasmussen TE, Rich NM (2005) What imaging studies are necessary for abdominal aortic endograft sizing? A prospective blinded study using conventional computed tomography, aortography, and three-dimensional computed tomography. J Vasc surg 41:199–205CrossRefPubMed Parker MV, O’Donnell SD, Chang AS, Johnson CA, Gillespie DL, Goff JM, Rasmussen TE, Rich NM (2005) What imaging studies are necessary for abdominal aortic endograft sizing? A prospective blinded study using conventional computed tomography, aortography, and three-dimensional computed tomography. J Vasc surg 41:199–205CrossRefPubMed
6.
Zurück zum Zitat Moon MC, Morales JP, Greenberg RK (2007) The aortic arch and ascending aorta: are they within the endovascular realm? Semin Vasc Surg 20:97–107CrossRefPubMed Moon MC, Morales JP, Greenberg RK (2007) The aortic arch and ascending aorta: are they within the endovascular realm? Semin Vasc Surg 20:97–107CrossRefPubMed
7.
Zurück zum Zitat Huber CH, von Segesser LK (2006) Direct access valve replacement (DAVR) – are we entering a new era in cardiac surgery? Eur J Cardiothorac Surg 29:380–385CrossRefPubMed Huber CH, von Segesser LK (2006) Direct access valve replacement (DAVR) – are we entering a new era in cardiac surgery? Eur J Cardiothorac Surg 29:380–385CrossRefPubMed
8.
Zurück zum Zitat Ishimaru S (2004) Endografting of the aortic arch. J Endovasc Ther 11 Suppl 2:II62–71 Ishimaru S (2004) Endografting of the aortic arch. J Endovasc Ther 11 Suppl 2:II62–71
9.
Zurück zum Zitat Kasirajan K (2006) Thoracic endografts: procedural steps, technical pitfalls and how to avoid them. Semin Vasc Surg 19:3–10CrossRefPubMed Kasirajan K (2006) Thoracic endografts: procedural steps, technical pitfalls and how to avoid them. Semin Vasc Surg 19:3–10CrossRefPubMed
10.
Zurück zum Zitat Aronberg DJ, Glazer HS, Madsen K, Sagel SS (1984) Normal thoracic aortic diameters by computed tomography. J Comput Assist Tomogr 8:247–250PubMed Aronberg DJ, Glazer HS, Madsen K, Sagel SS (1984) Normal thoracic aortic diameters by computed tomography. J Comput Assist Tomogr 8:247–250PubMed
11.
Zurück zum Zitat Pearce WH, Slaughter MS, LeMaire S, Salyapongse AN, Feinglass J, McCarthy WJ, Yao JS (1993) Aortic diameter as a function of age, gender, and body surface area. Surgery 114:691–697PubMed Pearce WH, Slaughter MS, LeMaire S, Salyapongse AN, Feinglass J, McCarthy WJ, Yao JS (1993) Aortic diameter as a function of age, gender, and body surface area. Surgery 114:691–697PubMed
12.
Zurück zum Zitat Fitzgerald SW, Donaldson JS, Poznanski AK (1987) Pediatric thoracic aorta: normal measurements determined with CT. Radiology 165:667–669PubMed Fitzgerald SW, Donaldson JS, Poznanski AK (1987) Pediatric thoracic aorta: normal measurements determined with CT. Radiology 165:667–669PubMed
13.
Zurück zum Zitat Hager A, Kaemmerer H, Rapp-Berhhardt U, Blücher S, Rapp K, Bernhardt TM, Galanski M, Hess J (2002) Diameters of the thoracic aorta throughout life as measured with helical computed tomography. J Thorac Cardiovasc Surg 123:1060–1066CrossRefPubMed Hager A, Kaemmerer H, Rapp-Berhhardt U, Blücher S, Rapp K, Bernhardt TM, Galanski M, Hess J (2002) Diameters of the thoracic aorta throughout life as measured with helical computed tomography. J Thorac Cardiovasc Surg 123:1060–1066CrossRefPubMed
14.
Zurück zum Zitat Rubin GD, Leung AN, Robertson VJ, Stark P (1998) Thoracic spiral CT: influence of subsecond gantry rotation on image quality. Radiology 208:771–776PubMed Rubin GD, Leung AN, Robertson VJ, Stark P (1998) Thoracic spiral CT: influence of subsecond gantry rotation on image quality. Radiology 208:771–776PubMed
15.
Zurück zum Zitat Qanadli SD, El Hajiam M, Mesurolle B, Lavisse L, Jourdan O, Randoux B, Chagnon S, Lacombe P (1999) Motion artifacts of the aorta simulating aortic dissection on spiral CT. J Comput Assist Tomogr 23:1–6CrossRefPubMed Qanadli SD, El Hajiam M, Mesurolle B, Lavisse L, Jourdan O, Randoux B, Chagnon S, Lacombe P (1999) Motion artifacts of the aorta simulating aortic dissection on spiral CT. J Comput Assist Tomogr 23:1–6CrossRefPubMed
16.
Zurück zum Zitat Duvernoy O, Coulden R, Ytterberg C (1995) Aortic motion: a potential pitfall in CT imaging of dissection in the ascending aorta. J Comput Assist Tomogr 19:569–572CrossRefPubMed Duvernoy O, Coulden R, Ytterberg C (1995) Aortic motion: a potential pitfall in CT imaging of dissection in the ascending aorta. J Comput Assist Tomogr 19:569–572CrossRefPubMed
17.
Zurück zum Zitat Marten K, Funke M, Rummeny EJ, Engelke C (2005) Electrocardiographic assistance in multidetector CT of thoracic disorders. Clin Radiol 60:8–21CrossRefPubMed Marten K, Funke M, Rummeny EJ, Engelke C (2005) Electrocardiographic assistance in multidetector CT of thoracic disorders. Clin Radiol 60:8–21CrossRefPubMed
18.
Zurück zum Zitat Roos JE, William JK, Weishaupt D, Lachat M, Marincek B, Hilfiker PR (2002) Thoracic aorta: motion artifact reduction with retrospective and prospective electrocardiography-assisted multi-detector row CT. Radiology 222:271–277CrossRefPubMed Roos JE, William JK, Weishaupt D, Lachat M, Marincek B, Hilfiker PR (2002) Thoracic aorta: motion artifact reduction with retrospective and prospective electrocardiography-assisted multi-detector row CT. Radiology 222:271–277CrossRefPubMed
19.
Zurück zum Zitat Schertler T, Glücker T, Wildermuth S, Jungius KP, Marincek B, Boehm T (2005) Comparison of retrospectively ECG-gated and nongated MDCT of the chest in an emergency setting regarding workflow, image quality, and diagnostic certainty. Emerg Radiol 12:19–29CrossRefPubMed Schertler T, Glücker T, Wildermuth S, Jungius KP, Marincek B, Boehm T (2005) Comparison of retrospectively ECG-gated and nongated MDCT of the chest in an emergency setting regarding workflow, image quality, and diagnostic certainty. Emerg Radiol 12:19–29CrossRefPubMed
20.
Zurück zum Zitat O’Brien JP, Srichai MB, Hecht EM, Kim DC, Jacobs JE (2007) Anatomy of the heart at multidetector CT: what the radiologist needs to know. Radiographics 27:1569–1582CrossRefPubMed O’Brien JP, Srichai MB, Hecht EM, Kim DC, Jacobs JE (2007) Anatomy of the heart at multidetector CT: what the radiologist needs to know. Radiographics 27:1569–1582CrossRefPubMed
21.
Zurück zum Zitat Horiguchi J, Kiguchi M, Fujioka C, Shen Y, Arie R, Sunasaka K, Ito K (2008) Radiation dose, image quality, stenosis mesurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol 190:315–320CrossRefPubMed Horiguchi J, Kiguchi M, Fujioka C, Shen Y, Arie R, Sunasaka K, Ito K (2008) Radiation dose, image quality, stenosis mesurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol 190:315–320CrossRefPubMed
22.
Zurück zum Zitat van Prehn J, Vincken KL, Muhs BE, Barwegen GK, Bartels LW, Prokop M, Moll FL, Verhagen HJ (2007) Toward endografting of the ascending aorta: insight into dynamics using dynamic cine-CTA. J Endovasc Ther 14:551–60CrossRefPubMed van Prehn J, Vincken KL, Muhs BE, Barwegen GK, Bartels LW, Prokop M, Moll FL, Verhagen HJ (2007) Toward endografting of the ascending aorta: insight into dynamics using dynamic cine-CTA. J Endovasc Ther 14:551–60CrossRefPubMed
23.
Zurück zum Zitat Ganten M, Krautter U, Hosch W, Hansmann J, von Tengg-Kobligk H, Delorme S, Kauczor HU, Kauffmann GW, Bock M (2007) Age related change of human aortic distensibility: evaluation with ECG-gated CT. Eur Radiol 17:701–708CrossRefPubMed Ganten M, Krautter U, Hosch W, Hansmann J, von Tengg-Kobligk H, Delorme S, Kauczor HU, Kauffmann GW, Bock M (2007) Age related change of human aortic distensibility: evaluation with ECG-gated CT. Eur Radiol 17:701–708CrossRefPubMed
24.
Zurück zum Zitat Poutanen T, Tikanoja T, Sairanen H, Jokinen E (2003) Normal aortic dimensions and flow in 168 children and young adults. Clin Physiol Funct Imaging 23:224–229CrossRefPubMed Poutanen T, Tikanoja T, Sairanen H, Jokinen E (2003) Normal aortic dimensions and flow in 168 children and young adults. Clin Physiol Funct Imaging 23:224–229CrossRefPubMed
25.
Zurück zum Zitat Smiseth OA, Stoylen A, Ihlen H (2004) Tissue Doppler imaging for the diagnosis of coronary artery disease. Curr Opin Cardiol 19:421–429CrossRefPubMed Smiseth OA, Stoylen A, Ihlen H (2004) Tissue Doppler imaging for the diagnosis of coronary artery disease. Curr Opin Cardiol 19:421–429CrossRefPubMed
26.
Zurück zum Zitat Van Dantzig JM (2006) Echocardiography in the emergency department. Semin Cardiothorac Vasc Anesth 10:79–81CrossRefPubMed Van Dantzig JM (2006) Echocardiography in the emergency department. Semin Cardiothorac Vasc Anesth 10:79–81CrossRefPubMed
27.
Zurück zum Zitat Erbel R, Mohr-Kahaly S, Oelert H, Iversen S, Jakob H, Thelen M, Just M, Meyer J (1992) Diagnostic goals in aortic dissection. Value of transthoracic and transesophageal echocardiography. Herz 17:321–37PubMed Erbel R, Mohr-Kahaly S, Oelert H, Iversen S, Jakob H, Thelen M, Just M, Meyer J (1992) Diagnostic goals in aortic dissection. Value of transthoracic and transesophageal echocardiography. Herz 17:321–37PubMed
28.
Zurück zum Zitat Appelbe AF, Walker PG, Yeoh JK, Bonitatibus A, Yoganathan AP, Martin RP (1993) Clinical significance and origin of artifacts in transesophageal echocardiography of the thoracic aorta. J Am Coll Cardiol 21:754–760PubMedCrossRef Appelbe AF, Walker PG, Yeoh JK, Bonitatibus A, Yoganathan AP, Martin RP (1993) Clinical significance and origin of artifacts in transesophageal echocardiography of the thoracic aorta. J Am Coll Cardiol 21:754–760PubMedCrossRef
29.
Zurück zum Zitat Clarkson PM, Brandt PW (1985) Aortic diameters in infants and young children: normative angiographic data. Pediatr Cardiol 6:3–6CrossRefPubMed Clarkson PM, Brandt PW (1985) Aortic diameters in infants and young children: normative angiographic data. Pediatr Cardiol 6:3–6CrossRefPubMed
30.
Zurück zum Zitat Trivedi KR, Pinzon JL, McCrindle BW, Burrows PE, Freedom RM, Benson LN (2002) Cineangiographic aortic dimensions in normal children. Cardiol Young 12:339–344CrossRefPubMed Trivedi KR, Pinzon JL, McCrindle BW, Burrows PE, Freedom RM, Benson LN (2002) Cineangiographic aortic dimensions in normal children. Cardiol Young 12:339–344CrossRefPubMed
31.
Zurück zum Zitat Finet G, Masquet C, Eifferman A, Funck F, Lefèvre T, Marco J, Amiel M, Beaune J, Liénard J (1996) Can we optimize our angiographic views every time? Qualitative and quantitative evaluation of a new functionality. Invest Radiol 31:523–531CrossRefPubMed Finet G, Masquet C, Eifferman A, Funck F, Lefèvre T, Marco J, Amiel M, Beaune J, Liénard J (1996) Can we optimize our angiographic views every time? Qualitative and quantitative evaluation of a new functionality. Invest Radiol 31:523–531CrossRefPubMed
32.
Zurück zum Zitat Lorenz CH (2000) The range of normal values of cardiovascular structures in infants, children, and adolescents measured by magnetic resonance imaging. Pediatr Cardiol 21:37–46CrossRefPubMed Lorenz CH (2000) The range of normal values of cardiovascular structures in infants, children, and adolescents measured by magnetic resonance imaging. Pediatr Cardiol 21:37–46CrossRefPubMed
33.
Zurück zum Zitat Kim YC, Nielsen JF, Nayak KS (2008) Automatic correction of echo-planar imaging (EPI) ghosting artifacts in real-time interactive cardiac MRI using sensitivity encoding. J Magn Reson Imaging 27:239–245CrossRefPubMed Kim YC, Nielsen JF, Nayak KS (2008) Automatic correction of echo-planar imaging (EPI) ghosting artifacts in real-time interactive cardiac MRI using sensitivity encoding. J Magn Reson Imaging 27:239–245CrossRefPubMed
34.
Zurück zum Zitat Tenforde TS (2005) Magnetically induced electric fields and currents in the circulatory system. Prog Biophys Mol Biol 87:279–288CrossRefPubMed Tenforde TS (2005) Magnetically induced electric fields and currents in the circulatory system. Prog Biophys Mol Biol 87:279–288CrossRefPubMed
35.
Zurück zum Zitat Wyers MC, Fillinger MF, Schermerhorn ML, Powell RJ, Rzucidio EM, Walsh DB, Zwolak RM, Cronenwett JL (2003) Endovascular repair of abdominal aortic aneurysm without preoperative arteriography. J Vasc Surg 38:730–738CrossRefPubMed Wyers MC, Fillinger MF, Schermerhorn ML, Powell RJ, Rzucidio EM, Walsh DB, Zwolak RM, Cronenwett JL (2003) Endovascular repair of abdominal aortic aneurysm without preoperative arteriography. J Vasc Surg 38:730–738CrossRefPubMed
36.
Zurück zum Zitat von Ludinghausen M (2003) The clinical anatomy of coronary arteries. Adv Anat Embryol Cell Biol 167:III–VIII, 1–111 von Ludinghausen M (2003) The clinical anatomy of coronary arteries. Adv Anat Embryol Cell Biol 167:III–VIII, 1–111
37.
Zurück zum Zitat Huber CH, Tozzi P, Corno AF, Marty B, Ruchat P, Gersbach P, Nasratulla M, von Segesser LK (2004) Do valved stents compromise coronary flow? Eur J Cardiothorac Surg 25:754–759CrossRefPubMed Huber CH, Tozzi P, Corno AF, Marty B, Ruchat P, Gersbach P, Nasratulla M, von Segesser LK (2004) Do valved stents compromise coronary flow? Eur J Cardiothorac Surg 25:754–759CrossRefPubMed
Metadaten
Titel
Ascending aorta measurements as assessed by ECG-gated multi-detector computed tomography: a pilot study to establish normative values for transcatheter therapies
verfasst von
Tri-Linh C. Lu
Christoph H. Huber
Elena Rizzo
Jashmid Dehmeshki
Ludwig K. von Segesser
Salah D. Qanadli
Publikationsdatum
01.03.2009
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 3/2009
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-008-1182-8

Weitere Artikel der Ausgabe 3/2009

European Radiology 3/2009 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.