Skip to main content
Erschienen in: European Radiology 3/2010

01.03.2010 | Interventional

Local skin and eye lens equivalent doses in interventional neuroradiology

verfasst von: Michael Sandborg, Sandro Rossitti, Håkan Pettersson

Erschienen in: European Radiology | Ausgabe 3/2010

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess patient skin and eye lens doses in interventional neuroradiology and to assess both stochastic and deterministic radiation risks.

Methods

Kerma–area product (P KA) was recorded and skin doses measured using thermoluminescence dosimeters. Estimated dose at interventional reference point (IRP) was compared with measured absorbed doses.

Results

The average and maximum fluoroscopy times were 32 and 189 min for coiling and 40 and 144 min for embolisation. The average and maximum P KA for coiling were 121 and 436 Gy cm2, respectively, and 189 and 677 Gy cm2 for embolisation. The average and maximum values of the measured maximum absorbed skin doses were 0.72 and 3.0 Sv, respectively, for coiling and 0.79 and 2.1 Sv for embolisation. Two out of the 52 patients received skin doses in excess of 2 Sv. The average and maximum doses to the eye lens (left eye) were 51 and 515 mSv (coiling) and 71 and 289 mSv (embolisation).

Conclusion

The ratio between the measured dose and the dose at the IRP was 0.44 ± 0.18 mSv/mGy indicating that the dose displayed by the x-ray unit overestimates the maximum skin dose but is still a valuable indication of the dose. The risk of inducing skin erythema and lens cataract during our hospital procedures is therefore small.
Literatur
1.
Zurück zum Zitat Valentin J, ICRP (2000) ICRP publication 85: avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30(2):7–67 Valentin J, ICRP (2000) ICRP publication 85: avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30(2):7–67
2.
Zurück zum Zitat WHO (2000) Efficacy and radiation safety in interventional radiology. World Health Organization, Geneva. ISBN 13 9789241545297, ISBN 10 9241545291 WHO (2000) Efficacy and radiation safety in interventional radiology. World Health Organization, Geneva. ISBN 13 9789241545297, ISBN 10 9241545291
3.
Zurück zum Zitat Koenig TR, Detlev W, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures. Part 1, characteristics of radiation injury. AJR Am J Roentgenol 177:3–11PubMed Koenig TR, Detlev W, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures. Part 1, characteristics of radiation injury. AJR Am J Roentgenol 177:3–11PubMed
4.
Zurück zum Zitat Padovani R, Bernardi G, Quai E, Signor M, Toh HS, Morocutti G, Spedicato L (2005) Retrospective evaluation of occurrence of skin injuries in interventional cardiac procedures. Radiat Prot Dosim 117:247–250CrossRef Padovani R, Bernardi G, Quai E, Signor M, Toh HS, Morocutti G, Spedicato L (2005) Retrospective evaluation of occurrence of skin injuries in interventional cardiac procedures. Radiat Prot Dosim 117:247–250CrossRef
5.
Zurück zum Zitat Vano E, Faulkner K (2006) ICRP special radiation protection issues in interventional radiology, digital and cardiac imaging. Radiat Prot Dosim 117:13–17CrossRef Vano E, Faulkner K (2006) ICRP special radiation protection issues in interventional radiology, digital and cardiac imaging. Radiat Prot Dosim 117:13–17CrossRef
6.
Zurück zum Zitat Faulkner K, Malone J, Vano E, Padovani R, Busch HP, Zoetelief J, Bosmans H (2008) The SENTINEL project. Radiat Prot Dosim 129:3–5CrossRef Faulkner K, Malone J, Vano E, Padovani R, Busch HP, Zoetelief J, Bosmans H (2008) The SENTINEL project. Radiat Prot Dosim 129:3–5CrossRef
7.
Zurück zum Zitat Kosunen A, Komppa T, Toivonen M (2005) Evaluation of methods to estimate the patient dose in interventional radiology. Radiat Prot Dosim 117:178–184CrossRef Kosunen A, Komppa T, Toivonen M (2005) Evaluation of methods to estimate the patient dose in interventional radiology. Radiat Prot Dosim 117:178–184CrossRef
8.
Zurück zum Zitat Van Dam J, Bosmans H, Marchal G, Wambersie A (2005) Characteristics of dosimeter types for skin dose measurement in practice. Radiat Prot Dosim 177:185–189 Van Dam J, Bosmans H, Marchal G, Wambersie A (2005) Characteristics of dosimeter types for skin dose measurement in practice. Radiat Prot Dosim 177:185–189
9.
Zurück zum Zitat Balter S, Moses J (2007) Managing patient dose in interventional cardiology. Catherterization and cardiovascular interventions. 70:244–249CrossRef Balter S, Moses J (2007) Managing patient dose in interventional cardiology. Catherterization and cardiovascular interventions. 70:244–249CrossRef
10.
Zurück zum Zitat International Commission on Radiation Units and Measurements (ICRU) (2005) Patient dosimetry for x rays used in medical imaging (report 74). J ICRU 5(2):i. ISSN 1473–6691, ISBN 0199203208 International Commission on Radiation Units and Measurements (ICRU) (2005) Patient dosimetry for x rays used in medical imaging (report 74). J ICRU 5(2):i. ISSN 1473–6691, ISBN 0199203208
11.
Zurück zum Zitat IAEA (2007) Dosimetry in diagnostic radiology: an international code of practice. Technical reports series No 457. International Atomic Energy Agency (IAEA), Vienna. ISSN 0074–1914, ISBN 92–0–115406–2 IAEA (2007) Dosimetry in diagnostic radiology: an international code of practice. Technical reports series No 457. International Atomic Energy Agency (IAEA), Vienna. ISSN 0074–1914, ISBN 92–0–115406–2
12.
Zurück zum Zitat Chida K, Saito H, Otani H, Kohzuki M, Takahashi S, Yamada S, Shirato K, Zuguchi M (2006) Relationship between fluoroscopy time, dose-area product, body weight and maximum skin dose in cardiac interventional procedures. AJR Am J Roentgenol 186:774–778CrossRefPubMed Chida K, Saito H, Otani H, Kohzuki M, Takahashi S, Yamada S, Shirato K, Zuguchi M (2006) Relationship between fluoroscopy time, dose-area product, body weight and maximum skin dose in cardiac interventional procedures. AJR Am J Roentgenol 186:774–778CrossRefPubMed
13.
Zurück zum Zitat Vano E, Gonzalez L, Guibelalde E, Aviles P, Fernandez JM, Prieto C, Galvan C (2005) Evaluation of risk of deterministic effects in fluoroscopically guided procedures. Radiat Prot Dosim 117:190–194CrossRef Vano E, Gonzalez L, Guibelalde E, Aviles P, Fernandez JM, Prieto C, Galvan C (2005) Evaluation of risk of deterministic effects in fluoroscopically guided procedures. Radiat Prot Dosim 117:190–194CrossRef
15.
Zurück zum Zitat Neofotistou V, Vano E, Padovani R, Kotre J, Dowling A, Toivonen M, Kottou S, Tsapaki V, Willis S, Bernardi G, Faulkner K (2003) Preliminary reference levels in interventional cardiology. Eur Radiol 13:2259–2263CrossRefPubMed Neofotistou V, Vano E, Padovani R, Kotre J, Dowling A, Toivonen M, Kottou S, Tsapaki V, Willis S, Bernardi G, Faulkner K (2003) Preliminary reference levels in interventional cardiology. Eur Radiol 13:2259–2263CrossRefPubMed
16.
Zurück zum Zitat Vano E, Järvinen H, Kosunen A, Bly R, Malone J, Dowling A, Larkin A, Padovani R, Bosmans H, Dragusin O, Jaschke W, Torbica P, Back C, Schreiner A, Bokou C, Kottou S, Tsapaki V, Jankowski J, Papierz S, Domienik J, Werduch A, Nikodemova D, Salat D, Kepler K, Bor MD, Vassileva J, Borisova R, Pellet S, Corbett RH (2008) Patient dose in interventional radiology: a European survey. Radiat Prot Dosim 129:39–45CrossRef Vano E, Järvinen H, Kosunen A, Bly R, Malone J, Dowling A, Larkin A, Padovani R, Bosmans H, Dragusin O, Jaschke W, Torbica P, Back C, Schreiner A, Bokou C, Kottou S, Tsapaki V, Jankowski J, Papierz S, Domienik J, Werduch A, Nikodemova D, Salat D, Kepler K, Bor MD, Vassileva J, Borisova R, Pellet S, Corbett RH (2008) Patient dose in interventional radiology: a European survey. Radiat Prot Dosim 129:39–45CrossRef
17.
Zurück zum Zitat Balter S, Miller DL, Vano E, Ortis Lopes P, Bernardi G, Cotelo E, Faulkner K, Nowotny R, Padovani R, Ramirez A (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35:673–680CrossRefPubMed Balter S, Miller DL, Vano E, Ortis Lopes P, Bernardi G, Cotelo E, Faulkner K, Nowotny R, Padovani R, Ramirez A (2008) A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys 35:673–680CrossRefPubMed
18.
Zurück zum Zitat Faulkner K, Ortiz-Lopes P, Vano E (2005) Patient dosimetry in diagnostic radiology and interventional radiology: a practical approach using trigger levels. Radiat Prot Dosim 117:166–168 Faulkner K, Ortiz-Lopes P, Vano E (2005) Patient dosimetry in diagnostic radiology and interventional radiology: a practical approach using trigger levels. Radiat Prot Dosim 117:166–168
19.
Zurück zum Zitat International Electrotechnical Commission (IEC) (2000) Standard 60601–2–43. Medical electrical equipment, part 2–43: particular requirements for the safety of X-ray equipment for interventional procedures International Electrotechnical Commission (IEC) (2000) Standard 60601–2–43. Medical electrical equipment, part 2–43: particular requirements for the safety of X-ray equipment for interventional procedures
20.
Zurück zum Zitat Rampada O, Rapolo R (2004) A method for real time estimation of entrance skin dose distribution in interventional neuroradiology. Med Phys 31:2356–2361CrossRef Rampada O, Rapolo R (2004) A method for real time estimation of entrance skin dose distribution in interventional neuroradiology. Med Phys 31:2356–2361CrossRef
21.
Zurück zum Zitat Nishizawa K, Moritake T, Matsumaru Y, Tsuboi K, Iwai K (2003) Dose measurements for patients and physicians using a glass dosimeter during endovascular treatment for brain disease. Rad Prot Dosim 107:247–252 Nishizawa K, Moritake T, Matsumaru Y, Tsuboi K, Iwai K (2003) Dose measurements for patients and physicians using a glass dosimeter during endovascular treatment for brain disease. Rad Prot Dosim 107:247–252
22.
Zurück zum Zitat Padovani R, Quai E (2005) Patient dosimetry approaches in interventional cardiology and literature dose data review. Radiat Prot Dosim 117:217–221CrossRef Padovani R, Quai E (2005) Patient dosimetry approaches in interventional cardiology and literature dose data review. Radiat Prot Dosim 117:217–221CrossRef
23.
Zurück zum Zitat Theodorakou C, Horrocks JA (2003) A study on radiation doses and irradiated areas in cerebral embolisation. Br J Radiol 76:546–552CrossRefPubMed Theodorakou C, Horrocks JA (2003) A study on radiation doses and irradiated areas in cerebral embolisation. Br J Radiol 76:546–552CrossRefPubMed
24.
Zurück zum Zitat Bergeron P, Carrier R, Roy D, Blais N, Raymond J (1994) Radiation doses to patients in neurointerventional procedures. AJNR Am J Neuroradiol 15(10):1809–1812PubMed Bergeron P, Carrier R, Roy D, Blais N, Raymond J (1994) Radiation doses to patients in neurointerventional procedures. AJNR Am J Neuroradiol 15(10):1809–1812PubMed
25.
Zurück zum Zitat Mooney RB, McKinstry CS, Kamel HA (2000) Absorbed doses and deterministic effects to patients from interventional neuroradiology. Br J Radiol 73(871):745–751PubMed Mooney RB, McKinstry CS, Kamel HA (2000) Absorbed doses and deterministic effects to patients from interventional neuroradiology. Br J Radiol 73(871):745–751PubMed
26.
Zurück zum Zitat Kuwayama N, Takaku A, Endo S, Nishijima M, Kamei T (1994) Radiation exposure in endovascular surgery of the head and neck. AJNR Am J Neuroradiol 15(10):1801–1808PubMed Kuwayama N, Takaku A, Endo S, Nishijima M, Kamei T (1994) Radiation exposure in endovascular surgery of the head and neck. AJNR Am J Neuroradiol 15(10):1801–1808PubMed
27.
Zurück zum Zitat Gkanatsios NA, Huda W, Peters KR (2002) Adult patient doses in interventional neuroradiology. Med Phys 29:717–723CrossRefPubMed Gkanatsios NA, Huda W, Peters KR (2002) Adult patient doses in interventional neuroradiology. Med Phys 29:717–723CrossRefPubMed
28.
Zurück zum Zitat Alberts WG, Böhm J, Kramer HM, Iles WJ, Schwartz RB, Thompson IMG (1994) International standardization of reference radiations and calibration procedures for radiation protection instruments. In: Kölzer W, Maushart R (eds) Strahlenschutz: Physik und Meßtechnik, vol 1. Verlag TüV Rheinland, pp 181–188 Alberts WG, Böhm J, Kramer HM, Iles WJ, Schwartz RB, Thompson IMG (1994) International standardization of reference radiations and calibration procedures for radiation protection instruments. In: Kölzer W, Maushart R (eds) Strahlenschutz: Physik und Meßtechnik, vol 1. Verlag TüV Rheinland, pp 181–188
29.
Zurück zum Zitat ICRP (2007) ICRP publication 103: recommendations of the IRCP. Ann ICRP 37:2–4 ICRP (2007) ICRP publication 103: recommendations of the IRCP. Ann ICRP 37:2–4
31.
Zurück zum Zitat Moritake T, Matsumau Y, Takigawa T, Nishizawa K, Matsumura A, Tsuboi K (2008) Dose measurements on both patients and operators during neurointerventional procedures using photoluminescence glass dosimeters. AJNR Am J Neuroradiol 29:1910–1917CrossRefPubMed Moritake T, Matsumau Y, Takigawa T, Nishizawa K, Matsumura A, Tsuboi K (2008) Dose measurements on both patients and operators during neurointerventional procedures using photoluminescence glass dosimeters. AJNR Am J Neuroradiol 29:1910–1917CrossRefPubMed
32.
Zurück zum Zitat Persliden (2005) Patient and staff doses in interventional x-ray procedures in Sweden. Radiat Prot Dosim 114:150–157 Persliden (2005) Patient and staff doses in interventional x-ray procedures in Sweden. Radiat Prot Dosim 114:150–157
Metadaten
Titel
Local skin and eye lens equivalent doses in interventional neuroradiology
verfasst von
Michael Sandborg
Sandro Rossitti
Håkan Pettersson
Publikationsdatum
01.03.2010
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 3/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-009-1598-9

Weitere Artikel der Ausgabe 3/2010

European Radiology 3/2010 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.