Prospectively gated axial CT coronary angiography: comparison of image quality and effective radiation dose between 64- and 256-slice CT

verfasst von: Oliver Klass, Matthew Walker, Anne Siebach, Tina Stuber, Sebastian Feuerlein, Markus Juchems, Martin H. K. Hoffmann

Erschienen in: European Radiology | Ausgabe 5/2010

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare image quality and effective radiation dose for prospectively gated axial CT coronary angiography on 64- and 256-slice CT.

Methods

The patient cohort consisted of 80 consecutive patients undergoing imaging with 256-slice CT and 80 patients with 64-slice CT. The two patient groups were demographically matched according to age, gender, body mass index and heart rate. For both groups, two observers independently assessed image quality for all coronary segments on a five-point ordinal scale. The two groups were compared with regard to image quality and effective radiation dose. Image quality scores less than 3 were considered non-diagnostic.

Results

Mean effective radiation dose did not significantly differ between the two groups (3.4 ± 0.8 mSv in both groups; p > 0.1). Vessel-based image quality was significantly better in patients undergoing 256-slice CT compared with those undergoing 64-slice CT (p < 0.001). The proportion of assessable coronary segments significantly increased from 95.6% in the 64-slice group to 98.9% in the 256-slice group (p < 0.05).

Conclusion

Prospectively gated axial CT coronary angiography performed on 256-slice CT provides significantly improved and more stable image quality at an equivalent effective radiation dose compared with 64-slice CT.
Literatur
1.
Zurück zum Zitat Earls JP, Berman EL, Urban BA et al (2008 ) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246:742–753CrossRefPubMed Earls JP, Berman EL, Urban BA et al (2008 ) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246:742–753CrossRefPubMed
2.
Zurück zum Zitat Scheffel H, Alkadhi H, Leschka S et al (2008) Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance. Heart 94:1132–1137CrossRefPubMed Scheffel H, Alkadhi H, Leschka S et al (2008) Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance. Heart 94:1132–1137CrossRefPubMed
3.
Zurück zum Zitat Shuman WP, Branch KR, May JM et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed Shuman WP, Branch KR, May JM et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed
4.
Zurück zum Zitat Klass O, Jeltsch M, Feuerlein S et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol. doi:10.1007/s00330-008-1222-4 Klass O, Jeltsch M, Feuerlein S et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol. doi:10.​1007/​s00330-008-1222-4
5.
Zurück zum Zitat Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed
6.
Zurück zum Zitat Weigold WG, Olszewski ME, Walker MJ (2009) Low-dose prospectively gated 256-slice coronary computed tomographic angiography. Int J Cardiovasc Imaging. doi:10.1007/s10554-009-9439-0 Weigold WG, Olszewski ME, Walker MJ (2009) Low-dose prospectively gated 256-slice coronary computed tomographic angiography. Int J Cardiovasc Imaging. doi:10.​1007/​s10554-009-9439-0
7.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging. doi:10.1007/s10554-008-9308-2 Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging. doi:10.​1007/​s10554-008-9308-2
8.
Zurück zum Zitat Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2007) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol. doi:10.1016/j.ejrad.2007.05.001 Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2007) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol. doi:10.​1016/​j.​ejrad.​2007.​05.​001
9.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546CrossRefPubMed Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546CrossRefPubMed
10.
Zurück zum Zitat Walker MJ, Olszewski ME, Desai MY, Halliburton SS, Flamm SD (2009) New radiation dose saving technologies for 256-slice cardiac computed tomography angiography. Int J Cardiovasc Imaging. doi:10.1007/s10554-009-9444-3 Walker MJ, Olszewski ME, Desai MY, Halliburton SS, Flamm SD (2009) New radiation dose saving technologies for 256-slice cardiac computed tomography angiography. Int J Cardiovasc Imaging. doi:10.​1007/​s10554-009-9444-3
11.
Zurück zum Zitat Austen WG, Edwards JE, Frye RL et al (1975 ) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American heart association. Circulation 51(Suppl):5–40PubMed Austen WG, Edwards JE, Frye RL et al (1975 ) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American heart association. Circulation 51(Suppl):5–40PubMed
12.
Zurück zum Zitat Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRefPubMed Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428CrossRefPubMed
15.
Zurück zum Zitat Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80CrossRefPubMed Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80CrossRefPubMed
16.
Zurück zum Zitat Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385CrossRefPubMed Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385CrossRefPubMed
17.
Zurück zum Zitat Soliman OI, Kirschbaum SW, van Dalen BM et al (2008) Accuracy and reproducibility of quantitation of left ventricular function by real-time three-dimensional echocardiography versus cardiac magnetic resonance. Am J Cardiol 102:778–783CrossRefPubMed Soliman OI, Kirschbaum SW, van Dalen BM et al (2008) Accuracy and reproducibility of quantitation of left ventricular function by real-time three-dimensional echocardiography versus cardiac magnetic resonance. Am J Cardiol 102:778–783CrossRefPubMed
18.
Zurück zum Zitat Matt D, Scheffel H, Leschka S et al (2007) Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR Am J Roentgenol 189:567–573CrossRefPubMed Matt D, Scheffel H, Leschka S et al (2007) Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR Am J Roentgenol 189:567–573CrossRefPubMed
19.
Zurück zum Zitat Katritsis D, Efstathopoulos E, Betsou S et al (2000) Radiation exposure of patients and coronary arteries in the stent era: A prospective study. Catheter Cardiovasc Interv 51:259–264CrossRefPubMed Katritsis D, Efstathopoulos E, Betsou S et al (2000) Radiation exposure of patients and coronary arteries in the stent era: A prospective study. Catheter Cardiovasc Interv 51:259–264CrossRefPubMed
20.
Zurück zum Zitat Kocinaj D, Cioppa A, Ambrosini G et al (2006 ) Radiation dose exposure during cardiac and peripheral arteries catheterisation. Int J Cardiol 113:283–284CrossRefPubMed Kocinaj D, Cioppa A, Ambrosini G et al (2006 ) Radiation dose exposure during cardiac and peripheral arteries catheterisation. Int J Cardiol 113:283–284CrossRefPubMed
21.
Zurück zum Zitat Miller SW, Castronovo FP Jr (1985 ) Radiation exposure and protection in cardiac catheterization laboratories. Am J Cardiol 55:171–176CrossRefPubMed Miller SW, Castronovo FP Jr (1985 ) Radiation exposure and protection in cardiac catheterization laboratories. Am J Cardiol 55:171–176CrossRefPubMed
22.
Zurück zum Zitat Watson LE, Riggs MW, Bourland PD (1997 ) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef Watson LE, Riggs MW, Bourland PD (1997 ) Radiation exposure during cardiology fellowship training. Health Phys 73:690–693PubMedCrossRef
23.
Zurück zum Zitat Huda W, Vance A (2007) Patient radiation doses from adult and pediatric CT. AJR Am J Roentgenol 188:540–546CrossRefPubMed Huda W, Vance A (2007) Patient radiation doses from adult and pediatric CT. AJR Am J Roentgenol 188:540–546CrossRefPubMed
24.
Zurück zum Zitat Poll LW, Cohnen M, Brachten S, Ewen K, Modder U (2002) Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation ("ECG pulsing"): phantom measurements. Rofo 174:1500–1505PubMed Poll LW, Cohnen M, Brachten S, Ewen K, Modder U (2002) Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation ("ECG pulsing"): phantom measurements. Rofo 174:1500–1505PubMed
25.
Zurück zum Zitat Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed
26.
Zurück zum Zitat Jakobs TF, Becker CR, Ohnesorge B et al (2002 ) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086CrossRefPubMed Jakobs TF, Becker CR, Ohnesorge B et al (2002 ) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086CrossRefPubMed
27.
Zurück zum Zitat Johnson TR, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415CrossRefPubMed Johnson TR, Nikolaou K, Wintersperger BJ et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415CrossRefPubMed
28.
Zurück zum Zitat Ertel D, Lell MM, Harig F, Flohr T, Schmidt B, Kalender WA (2009) Cardiac spiral dual-source CT with high pitch: a feasibility study. Eur Radiol. doi:10.1007/s00330-009-1503-6 Ertel D, Lell MM, Harig F, Flohr T, Schmidt B, Kalender WA (2009) Cardiac spiral dual-source CT with high pitch: a feasibility study. Eur Radiol. doi:10.​1007/​s00330-009-1503-6
29.
Zurück zum Zitat Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2008) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65:442–448CrossRefPubMed Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2008) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65:442–448CrossRefPubMed
30.
Zurück zum Zitat Jung B, Mahnken AH, Stargardt A et al (2003) Individually weight-adapted examination protocol in retrospectively ECG-gated MSCT of the heart. Eur Radiol 13:2560–2566CrossRefPubMed Jung B, Mahnken AH, Stargardt A et al (2003) Individually weight-adapted examination protocol in retrospectively ECG-gated MSCT of the heart. Eur Radiol 13:2560–2566CrossRefPubMed
31.
Zurück zum Zitat Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed
Metadaten
Titel
Prospectively gated axial CT coronary angiography: comparison of image quality and effective radiation dose between 64- and 256-slice CT
verfasst von
Oliver Klass
Matthew Walker
Anne Siebach
Tina Stuber
Sebastian Feuerlein
Markus Juchems
Martin H. K. Hoffmann
Publikationsdatum
01.05.2010
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 5/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-009-1652-7

Weitere Artikel der Ausgabe 5/2010

European Radiology 5/2010 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.