Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T

Erschienen in: European Radiology | Ausgabe 6/2010

Einloggen, um Zugang zu erhalten

Abstract

Objective

As high-field cardiac MRI (CMR) becomes more widespread the propensity of ECG to interference from electromagnetic fields (EMF) and to magneto-hydrodynamic (MHD) effects increases and with it the motivation for a CMR triggering alternative. This study explores the suitability of acoustic cardiac triggering (ACT) for left ventricular (LV) function assessment in healthy subjects (n = 14).

Methods

Quantitative analysis of 2D CINE steady-state free precession (SSFP) images was conducted to compare ACT’s performance with vector ECG (VCG). Endocardial border sharpness (EBS) was examined paralleled by quantitative LV function assessment.

Results

Unlike VCG, ACT provided signal traces free of interference from EMF or MHD effects. In the case of correct R-wave recognition, VCG-triggered 2D CINE SSFP was immune to cardiac motion effects—even at 3.0 T. However, VCG-triggered 2D SSFP CINE imaging was prone to cardiac motion and EBS degradation if R-wave misregistration occurred. ACT-triggered acquisitions yielded LV parameters (end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and left ventricular mass (LVM)) comparable with those derived from VCG-triggered acquisitions (1.5 T: ESVVCG = (56 ± 17) ml, EDVVCG = (151 ± 32) ml, LVMVCG = (97 ± 27) g, SVVCG = (94 ± 19) ml, EFVCG = (63 ± 5)% cf. ESVACT = (56 ± 18) ml, EDVACT = (147 ± 36) ml, LVMACT = (102 ± 29) g, SVACT = (91 ± 22) ml, EFACT = (62 ± 6)%; 3.0 T: ESVVCG = (55 ± 21) ml, EDVVCG = (151 ± 32) ml, LVMVCG = (101 ± 27) g, SVVCG = (96 ± 15) ml, EFVCG = (65 ± 7)% cf. ESVACT = (54 ± 20) ml, EDVACT = (146 ± 35) ml, LVMACT = (101 ± 30) g, SVACT = (92 ± 17) ml, EFACT = (64 ± 6)%).

Conclusions

ACT’s intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical CMR.
Literatur
1.
Zurück zum Zitat Hudsmith LE, Petersen SE, Francis JM, Robson MD, Neubauer S (2005) Normal human left and right ventricular and left atrial dimensions using steady state free precession magnetic resonance imaging. J Cardiovasc Magn Reson 7:775–782CrossRefPubMed Hudsmith LE, Petersen SE, Francis JM, Robson MD, Neubauer S (2005) Normal human left and right ventricular and left atrial dimensions using steady state free precession magnetic resonance imaging. J Cardiovasc Magn Reson 7:775–782CrossRefPubMed
2.
Zurück zum Zitat Maceira A, Prasad S, Khan M, Pennell D (2006) Normalized left ventricular systolic and diastolic function by steady state free precession cardiovascular magnetic resonance. J Magn Reson Imaging 8:417–426 Maceira A, Prasad S, Khan M, Pennell D (2006) Normalized left ventricular systolic and diastolic function by steady state free precession cardiovascular magnetic resonance. J Magn Reson Imaging 8:417–426
3.
Zurück zum Zitat Gutberlet M, Schwinge K, Freyhardt P, Spors B, Grothoff M, Denecke T, Ludemann L, Noeske R, Niendorf T, Felix R (2005) Influence of high magnetic field strengths and parallel acquisition strategies on image quality in cardiac 2D CINE magnetic resonance imaging: comparison of 1.5 T vs 3.0 T. Eur Radiol 15:1586–1597CrossRef Gutberlet M, Schwinge K, Freyhardt P, Spors B, Grothoff M, Denecke T, Ludemann L, Noeske R, Niendorf T, Felix R (2005) Influence of high magnetic field strengths and parallel acquisition strategies on image quality in cardiac 2D CINE magnetic resonance imaging: comparison of 1.5 T vs 3.0 T. Eur Radiol 15:1586–1597CrossRef
4.
Zurück zum Zitat Lima JA, Desai MY (2004) Cardiovascular magnetic resonance imaging: current and emerging applications. J Am Coll Cardiol 44:1164–1171CrossRefPubMed Lima JA, Desai MY (2004) Cardiovascular magnetic resonance imaging: current and emerging applications. J Am Coll Cardiol 44:1164–1171CrossRefPubMed
5.
Zurück zum Zitat Niendorf T, Saranathan M, Lingamneni A, Pedrosa I, Spencer M, Cline H, Foo TK, Rofsky NM (2005) Short breath-hold, volumetric coronary MR angiography employing steady-state free precession in conjunction with parallel imaging. Magn Reson Med 53:885–894CrossRefPubMed Niendorf T, Saranathan M, Lingamneni A, Pedrosa I, Spencer M, Cline H, Foo TK, Rofsky NM (2005) Short breath-hold, volumetric coronary MR angiography employing steady-state free precession in conjunction with parallel imaging. Magn Reson Med 53:885–894CrossRefPubMed
6.
Zurück zum Zitat Pennell DJ, Sechtem UP, Higgins CB, Manning WJ, Pohost GM, Rademakers FE, van Rossum AC, Shaw LJ, Yucel EK (2004) Clinical indications for cardiovascular magnetic resonance (CMR): Consensus panel report. Eur Heart J 25:1940–1965CrossRefPubMed Pennell DJ, Sechtem UP, Higgins CB, Manning WJ, Pohost GM, Rademakers FE, van Rossum AC, Shaw LJ, Yucel EK (2004) Clinical indications for cardiovascular magnetic resonance (CMR): Consensus panel report. Eur Heart J 25:1940–1965CrossRefPubMed
7.
Zurück zum Zitat Roes SD, Korosoglou G, Schar M, Westenberg JJ, van Osch MJP, de Roos A, Stuber M (2008) Correction for heart rate variability during 3D whole heart MR coronary angiography. J Magn Reson Imaging 27:1046–1053CrossRefPubMed Roes SD, Korosoglou G, Schar M, Westenberg JJ, van Osch MJP, de Roos A, Stuber M (2008) Correction for heart rate variability during 3D whole heart MR coronary angiography. J Magn Reson Imaging 27:1046–1053CrossRefPubMed
8.
Zurück zum Zitat Sodickson DK, Hardy CJ, Zhu Y, Giaquinto RO, Gross P, Kenwood G, Niendorf T, Lejay H, McKenzie CA, Ohliger MA, Grant AK, Rofsky NM (2005) Rapid volumetric MRI using parallel imaging with order-of-magnitude accelerations and a 32-element RF coil array: feasibility and implications. Acad Radiol 12:626–635CrossRefPubMed Sodickson DK, Hardy CJ, Zhu Y, Giaquinto RO, Gross P, Kenwood G, Niendorf T, Lejay H, McKenzie CA, Ohliger MA, Grant AK, Rofsky NM (2005) Rapid volumetric MRI using parallel imaging with order-of-magnitude accelerations and a 32-element RF coil array: feasibility and implications. Acad Radiol 12:626–635CrossRefPubMed
9.
Zurück zum Zitat Spuentrup E, Botnar RM (2006) Coronary magnetic resonance imaging: visualization of the vessel lumen and the vessel wall and molecular imaging of arteriothrombosis. Eur Radiol 16:1–14CrossRefPubMed Spuentrup E, Botnar RM (2006) Coronary magnetic resonance imaging: visualization of the vessel lumen and the vessel wall and molecular imaging of arteriothrombosis. Eur Radiol 16:1–14CrossRefPubMed
10.
Zurück zum Zitat Lanzer P, Barta C, Botvinick EH, Wiesendanger HU, Modin G, Higgins CB (1985) ECG-synchronized cardiac MR imaging: method and evaluation. Radiology 155:681–686PubMed Lanzer P, Barta C, Botvinick EH, Wiesendanger HU, Modin G, Higgins CB (1985) ECG-synchronized cardiac MR imaging: method and evaluation. Radiology 155:681–686PubMed
11.
Zurück zum Zitat Fischer SE, Wickline SA, Lorenz CH (1999) Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med 42:361–370CrossRefPubMed Fischer SE, Wickline SA, Lorenz CH (1999) Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med 42:361–370CrossRefPubMed
12.
Zurück zum Zitat Chia JM, Fischer SE, Wickline SA, Lorenz CH (2000) Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method. J Magn Reson Imaging 12:678–688CrossRefPubMed Chia JM, Fischer SE, Wickline SA, Lorenz CH (2000) Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method. J Magn Reson Imaging 12:678–688CrossRefPubMed
13.
Zurück zum Zitat Kugel H, Bremer C, Puschel M, Fischbach R, Lenzen H, Tombach B, Van Aken H, Heindel W (2003) Hazardous situation in the MR bore: induction in ECG leads causes fire. Eur Radiol 13:690–694PubMed Kugel H, Bremer C, Puschel M, Fischbach R, Lenzen H, Tombach B, Van Aken H, Heindel W (2003) Hazardous situation in the MR bore: induction in ECG leads causes fire. Eur Radiol 13:690–694PubMed
14.
Zurück zum Zitat Shellock FG, Kanal E (1996) Burns associated with the use of monitoring equipment during MR procedures. J Magn Reson Imaging 6:271–272CrossRefPubMed Shellock FG, Kanal E (1996) Burns associated with the use of monitoring equipment during MR procedures. J Magn Reson Imaging 6:271–272CrossRefPubMed
15.
Zurück zum Zitat Shellock FG, Crues JV (2004) MR procedures: biologic effects, safety, and patient care. Radiology 232:635–652CrossRefPubMed Shellock FG, Crues JV (2004) MR procedures: biologic effects, safety, and patient care. Radiology 232:635–652CrossRefPubMed
16.
Zurück zum Zitat Stralka JP, Bottomley PA (2007) A prototype RF dosimeter for independent measurement of the average specific absorption rate (SAR) during MRI. J Magn Reson Imaging 26:1296–1302CrossRefPubMed Stralka JP, Bottomley PA (2007) A prototype RF dosimeter for independent measurement of the average specific absorption rate (SAR) during MRI. J Magn Reson Imaging 26:1296–1302CrossRefPubMed
17.
Zurück zum Zitat Stecco A, Saponaro A, Carriero A (2007) Patient safety issues in magnetic resonance imaging: state of the art. Radiol Med 112:491–508CrossRefPubMed Stecco A, Saponaro A, Carriero A (2007) Patient safety issues in magnetic resonance imaging: state of the art. Radiol Med 112:491–508CrossRefPubMed
18.
Zurück zum Zitat Stuber M, Botnar RM, Fischer SE, Lamerichs R, Smink J, Harvey P, Manning WJ (2002) Preliminary report on in vivo coronary MRA at 3 Tesla in humans. Magn Reson Med 48:425–429CrossRefPubMed Stuber M, Botnar RM, Fischer SE, Lamerichs R, Smink J, Harvey P, Manning WJ (2002) Preliminary report on in vivo coronary MRA at 3 Tesla in humans. Magn Reson Med 48:425–429CrossRefPubMed
19.
Zurück zum Zitat Frauenrath T, Niendorf T, Kob M (2008) Acoustic method for synchronization of magnetic resonance imaging (MRI). Acta Acust United Acust 94:148–155CrossRef Frauenrath T, Niendorf T, Kob M (2008) Acoustic method for synchronization of magnetic resonance imaging (MRI). Acta Acust United Acust 94:148–155CrossRef
20.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical method for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical method for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
21.
Zurück zum Zitat Rangayyan RM, Lehner RJ (1987) Phonocardiogram signal analysis: a review. Crit Rev Biomed Eng 15:211–236PubMed Rangayyan RM, Lehner RJ (1987) Phonocardiogram signal analysis: a review. Crit Rev Biomed Eng 15:211–236PubMed
22.
Zurück zum Zitat Larson AC, Kellman P, Arai A, Hirsch GA, McVeigh E, Li D, Simonetti OP (2005) Preliminary investigation of respiratory self-gating for free-breathing segmented cine MRI. Magn Reson Med 53:159–168CrossRefPubMed Larson AC, Kellman P, Arai A, Hirsch GA, McVeigh E, Li D, Simonetti OP (2005) Preliminary investigation of respiratory self-gating for free-breathing segmented cine MRI. Magn Reson Med 53:159–168CrossRefPubMed
23.
Zurück zum Zitat Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP (2004) Self-gated cardiac cine MRI. Magn Reson Med 51:93–102CrossRefPubMed Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP (2004) Self-gated cardiac cine MRI. Magn Reson Med 51:93–102CrossRefPubMed
24.
Zurück zum Zitat Nijm GM, Sahakian AV, Swiryn S, Carr JC, Sheehan JJ, Larson AC (2008) Comparison of self–gated cine MRI retrospective cardiac synchronization algorithms. J Magn Reson Imaging 28:767–772CrossRefPubMed Nijm GM, Sahakian AV, Swiryn S, Carr JC, Sheehan JJ, Larson AC (2008) Comparison of self–gated cine MRI retrospective cardiac synchronization algorithms. J Magn Reson Imaging 28:767–772CrossRefPubMed
25.
Zurück zum Zitat Saleem SN (2008) Feasibility of MRI of the fetal heart with balanced steady-state free precession sequence along fetal body and cardiac planes. AJR Am J Roentgenol 191:1208–1215CrossRefPubMed Saleem SN (2008) Feasibility of MRI of the fetal heart with balanced steady-state free precession sequence along fetal body and cardiac planes. AJR Am J Roentgenol 191:1208–1215CrossRefPubMed
26.
Zurück zum Zitat Snyder CJ, DelaBarre L, Metzger GJ, van de Moortele PF, Akgun C, Ugurbil K, Vaughan JT (2009) Initial results of cardiac imaging at 7 Tesla. Magn Reson Med 61:517–524CrossRefPubMed Snyder CJ, DelaBarre L, Metzger GJ, van de Moortele PF, Akgun C, Ugurbil K, Vaughan JT (2009) Initial results of cardiac imaging at 7 Tesla. Magn Reson Med 61:517–524CrossRefPubMed
27.
Zurück zum Zitat Vaughan JT, Snyder CJ, DelaBarre LJ, Bolan PJ, Tian J, Bolinger L, Adriany G, Andersen P, Strupp J, Ugurbil K (2009) Whole-body imaging at 7 T: preliminary results. Magn Reson Med 61:244–248CrossRefPubMed Vaughan JT, Snyder CJ, DelaBarre LJ, Bolan PJ, Tian J, Bolinger L, Adriany G, Andersen P, Strupp J, Ugurbil K (2009) Whole-body imaging at 7 T: preliminary results. Magn Reson Med 61:244–248CrossRefPubMed
Metadaten
Titel
Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T
Publikationsdatum
01.06.2010
Erschienen in
European Radiology / Ausgabe 6/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-009-1676-z

Weitere Artikel der Ausgabe 6/2010

European Radiology 6/2010 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.