01.07.2011 | Computed Tomography

High diagnostic accuracy of prospective ECG-gating 64-slice computed tomography coronary angiography for the detection of in-stent restenosis

In-stent restenosis assessment by low-dose MDCT

verfasst von: Daniele Andreini, Gianluca Pontone, Antonio L. Bartorelli, Saima Mushtaq, Daniela Trabattoni, Erika Bertella, Sarah Cortinovis, Andrea Annoni, Alberto Formenti, Giovanni Ballerini, Piergiuseppe Agostoni, Cesare Fiorentini, Mauro Pepi

Erschienen in: European Radiology | Ausgabe 7/2011

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The accuracy of computed tomography (CT) for assessment of coronary stents is as yet unproven and radiation exposure has been a concern. The aim of our study is to compare radiation dose and diagnostic performance of CT with prospective ECG-triggering versus retrospective ECG-triggering for the detection of in-stent restenosis (ISR).

Methods

We enrolled 168 consecutive patients with suspected ISR, 83 studied using CT with prospective ECG-triggering (group 1) and 85 using retrospective ECG-triggering (group 2).

Results

Prevalence of ISR according to catheter angiography was 24% in both groups. The overall evaluability was similar (93% in group 1 vs 95% in group 2). Artefact sub-analysis showed a significantly lower number of blooming and higher number of slice misalignment in group 1 vs group 2. In the stent-based analysis using only evaluable stents, specificity, positive predictive value and accuracy were significantly higher in group 1 (100%, 100% and 99%, respectively) than in group 2 (97%, 91% and 95%, respectively, p < 0.05). Group 1 was exposed to a lower radiation dose compared with group 2 (4.3 ± 1.4 mSv vs 18.5 ± 5.5 mSv, p < 00.1).

Conclusions

CT with prospective ECG-triggering can improve diagnostic accuracy of non-invasive imaging of coronary stents with a significant reduction in radiation exposure
Literatur
1.
Zurück zum Zitat Zellweger MJ, Weinbacher M, Zutter AW et al (2003) Long-term outcome of patients with silent versus symptomatic ischemia six months after percutaneous coronary intervention and stenting. J Am Coll Cardiol 42:33–40PubMedCrossRef Zellweger MJ, Weinbacher M, Zutter AW et al (2003) Long-term outcome of patients with silent versus symptomatic ischemia six months after percutaneous coronary intervention and stenting. J Am Coll Cardiol 42:33–40PubMedCrossRef
2.
Zurück zum Zitat Chabbert V, Carrie D, Bennaceur M et al (2007) Evaluation of in-stent restenosis in proximal coronary arteries with multidetector computed tomography (MDCT). Eur Radiol 17:1452–1463PubMedCrossRef Chabbert V, Carrie D, Bennaceur M et al (2007) Evaluation of in-stent restenosis in proximal coronary arteries with multidetector computed tomography (MDCT). Eur Radiol 17:1452–1463PubMedCrossRef
3.
Zurück zum Zitat Achenbach S (2006) Computed tomography coronary angiography. J Am Coll Cardiol 48:1919–1928PubMedCrossRef Achenbach S (2006) Computed tomography coronary angiography. J Am Coll Cardiol 48:1919–1928PubMedCrossRef
4.
Zurück zum Zitat Rixe J, Achenbach S, Ropers D et al (2006) Assessment of coronary artery stent restenosis by 64-slice multi-detector computed tomography. Eur Heart J 27:2567–2572PubMedCrossRef Rixe J, Achenbach S, Ropers D et al (2006) Assessment of coronary artery stent restenosis by 64-slice multi-detector computed tomography. Eur Heart J 27:2567–2572PubMedCrossRef
5.
Zurück zum Zitat Ehara M, Kawai M, Surmely JF et al (2007) Diagnostic accuracy of coronary in-stent restenosis using 64-slice computed tomography: comparison with invasive coronary angiography. J Am Coll Cardiol 49:951–959PubMedCrossRef Ehara M, Kawai M, Surmely JF et al (2007) Diagnostic accuracy of coronary in-stent restenosis using 64-slice computed tomography: comparison with invasive coronary angiography. J Am Coll Cardiol 49:951–959PubMedCrossRef
6.
Zurück zum Zitat Andreini D, Pontone G, Bartorelli AL et al (2009) Comparison of feasibility and diagnostic accuracy of 64-slice multidetector computed tomography coronary angiography versus invasive coronary angiography versus intravascular ultrasound for evaluation of in-stent restenosis. Am J Cardiol 103:1349–1358PubMedCrossRef Andreini D, Pontone G, Bartorelli AL et al (2009) Comparison of feasibility and diagnostic accuracy of 64-slice multidetector computed tomography coronary angiography versus invasive coronary angiography versus intravascular ultrasound for evaluation of in-stent restenosis. Am J Cardiol 103:1349–1358PubMedCrossRef
7.
Zurück zum Zitat Fei X, Du X, Li P, Liao J, Shen Y, Li K (2008) Effect of dose-reduced scan protocols on cardiac coronary image quality with 64-row MDCT: a cardiac phantom study. Eur J Radiol 67:85–91PubMedCrossRef Fei X, Du X, Li P, Liao J, Shen Y, Li K (2008) Effect of dose-reduced scan protocols on cardiac coronary image quality with 64-row MDCT: a cardiac phantom study. Eur J Radiol 67:85–91PubMedCrossRef
8.
Zurück zum Zitat Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2009) Dose performance and image quality: dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400PubMedCrossRef Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2009) Dose performance and image quality: dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400PubMedCrossRef
9.
Zurück zum Zitat Maintz D, Burg MC, Seifarth H et al (2009) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49PubMedCrossRef Maintz D, Burg MC, Seifarth H et al (2009) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49PubMedCrossRef
10.
Zurück zum Zitat Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2008) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65:442–448PubMedCrossRef Mori S, Nishizawa K, Kondo C, Ohno M, Akahane K, Endo M (2008) Effective doses in subjects undergoing computed tomography cardiac imaging with the 256-multislice CT scanner. Eur J Radiol 65:442–448PubMedCrossRef
11.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546PubMedCrossRef Rybicki FJ, Otero HJ, Steigner ML et al (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 24:535–546PubMedCrossRef
12.
Zurück zum Zitat Leschka S, Stolzmann P, Desbiolles L et al (2009) Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience. Eur Radiol 19:2896–2903PubMedCrossRef Leschka S, Stolzmann P, Desbiolles L et al (2009) Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience. Eur Radiol 19:2896–2903PubMedCrossRef
13.
Zurück zum Zitat Stolzmann P, Scheffel H, Leschka S (2008) Influence of calcifications on diagnostic accuracy of coronary CT angiography using prospective ECG triggering. AJR Am J Roentgenol 191:1684–1689PubMedCrossRef Stolzmann P, Scheffel H, Leschka S (2008) Influence of calcifications on diagnostic accuracy of coronary CT angiography using prospective ECG triggering. AJR Am J Roentgenol 191:1684–1689PubMedCrossRef
14.
Zurück zum Zitat Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385PubMedCrossRef Leschka S, Wildermuth S, Boehm T et al (2006) Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality. Radiology 241:378–385PubMedCrossRef
15.
Zurück zum Zitat Pontone G, Andreini D, Bartorelli AL et al (2009) Diagnostic accuracy of coronary computed tomography angiography: a comparison between prospective and retrospective electrocardiogram triggering. J Am Coll Cardiol 54:346–355PubMedCrossRef Pontone G, Andreini D, Bartorelli AL et al (2009) Diagnostic accuracy of coronary computed tomography angiography: a comparison between prospective and retrospective electrocardiogram triggering. J Am Coll Cardiol 54:346–355PubMedCrossRef
16.
Zurück zum Zitat Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197PubMedCrossRef Husmann L, Valenta I, Gaemperli O et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197PubMedCrossRef
17.
Zurück zum Zitat Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248PubMedCrossRef Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248PubMedCrossRef
18.
Zurück zum Zitat Poll LW, Cohnen M, Brachten S, Ewen K, Mödder U (2002) Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation (“ECG pulsing”): phantom measurements. Rofo 174:1500–1505PubMed Poll LW, Cohnen M, Brachten S, Ewen K, Mödder U (2002) Dose reduction in multi-slice CT of the heart by use of ECG-controlled tube current modulation (“ECG pulsing”): phantom measurements. Rofo 174:1500–1505PubMed
19.
Zurück zum Zitat Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ (2007) Radiation dose to patients from cardiac diagnostic imaging. Circulation 116:1290–1305PubMedCrossRef Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ (2007) Radiation dose to patients from cardiac diagnostic imaging. Circulation 116:1290–1305PubMedCrossRef
20.
Zurück zum Zitat Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting international commission on radiological protection publication 103 or dual-energy scanning. Am J Roentgenol 194:881–889CrossRef Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting international commission on radiological protection publication 103 or dual-energy scanning. Am J Roentgenol 194:881–889CrossRef
21.
Zurück zum Zitat Leung KC, Martin CJ (1996) Effective doses for coronary angiography. Br J Radiol 69:426–431PubMedCrossRef Leung KC, Martin CJ (1996) Effective doses for coronary angiography. Br J Radiol 69:426–431PubMedCrossRef
22.
Zurück zum Zitat Budoff MJ, Achenbach S, Blumenthal RS et al (2006) Assessment of coronary artery disease by cardiac computed tomography. A scientific statement from the American Heart Association committee on cardiovascular imaging and intervention, council on cardiovascular radiology and intervention, and committee on cardiac imaging, council on clinical cardiology. Circulation 114:1761–1791PubMedCrossRef Budoff MJ, Achenbach S, Blumenthal RS et al (2006) Assessment of coronary artery disease by cardiac computed tomography. A scientific statement from the American Heart Association committee on cardiovascular imaging and intervention, council on cardiovascular radiology and intervention, and committee on cardiac imaging, council on clinical cardiology. Circulation 114:1761–1791PubMedCrossRef
23.
Zurück zum Zitat Marcassa C, Bax JJ, Bengel F et al (2008) Clinical value, cost-effectiveness, and safety of myocardial perfusion scintigraphy: a position statement. Eur Heart J 29:557–563PubMedCrossRef Marcassa C, Bax JJ, Bengel F et al (2008) Clinical value, cost-effectiveness, and safety of myocardial perfusion scintigraphy: a position statement. Eur Heart J 29:557–563PubMedCrossRef
24.
Zurück zum Zitat Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323PubMedCrossRef Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323PubMedCrossRef
25.
Zurück zum Zitat Min JK, Swaminathan RV, Vass M, Gallagher S, Weinsaft JW (2009) High-definition multidetector computed tomography for evaluation of coronary artery stents: comparison to standard-definition 64-detector row computed tomography. J Cardiovasc Comput Tomogr 3:246–251PubMedCrossRef Min JK, Swaminathan RV, Vass M, Gallagher S, Weinsaft JW (2009) High-definition multidetector computed tomography for evaluation of coronary artery stents: comparison to standard-definition 64-detector row computed tomography. J Cardiovasc Comput Tomogr 3:246–251PubMedCrossRef
26.
Zurück zum Zitat Pugliese F, Weustink AC, Van Mieghem C et al (2008) Dual-source coronary computed tomography angiography for detecting in-stent restenosis. Heart 94:848–854PubMedCrossRef Pugliese F, Weustink AC, Van Mieghem C et al (2008) Dual-source coronary computed tomography angiography for detecting in-stent restenosis. Heart 94:848–854PubMedCrossRef
27.
Zurück zum Zitat Abada HT, Larchez C, Daoud B, Sigal-Cinqualbre A, Paul JF (2006) MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol 186:387–390CrossRef Abada HT, Larchez C, Daoud B, Sigal-Cinqualbre A, Paul JF (2006) MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol 186:387–390CrossRef
28.
Zurück zum Zitat Husmann L, Schepis T, Scheffel H et al (2008) Comparison of diagnostic accuracy of 64-slice computed tomography coronary angiography in patients with low, intermediate, and high cardiovascular risk. Acad Radiol 15:452–461PubMedCrossRef Husmann L, Schepis T, Scheffel H et al (2008) Comparison of diagnostic accuracy of 64-slice computed tomography coronary angiography in patients with low, intermediate, and high cardiovascular risk. Acad Radiol 15:452–461PubMedCrossRef
Metadaten
Titel
High diagnostic accuracy of prospective ECG-gating 64-slice computed tomography coronary angiography for the detection of in-stent restenosis
In-stent restenosis assessment by low-dose MDCT
verfasst von
Daniele Andreini
Gianluca Pontone
Antonio L. Bartorelli
Saima Mushtaq
Daniela Trabattoni
Erika Bertella
Sarah Cortinovis
Andrea Annoni
Alberto Formenti
Giovanni Ballerini
Piergiuseppe Agostoni
Cesare Fiorentini
Mauro Pepi
Publikationsdatum
01.07.2011
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 7/2011
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2085-7

Weitere Artikel der Ausgabe 7/2011

European Radiology 7/2011 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.