01.12.2013 | Gastrointestinal

Clinical impact of an adaptive statistical iterative reconstruction algorithm for detection of hypervascular liver tumours using a low tube voltage, high tube current MDCT technique

verfasst von: Daniele Marin, Kingshuk Roy Choudhury, Rajan T. Gupta, Lisa M. Ho, Brian C. Allen, Sebastian T. Schindera, James G. Colsher, Ehsan Samei, Rendon C. Nelson

Erschienen in: European Radiology | Ausgabe 12/2013

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To investigate the impact of an adaptive statistical iterative reconstruction (ASiR) algorithm on diagnostic accuracy and confidence for the diagnosis of hypervascular liver tumours, as well as the reader’s perception of image quality, using a low tube voltage (80 kVp), high tube current computed tomography (CT) technique.

Methods

Forty patients (29 men, 11 women) with 65 hypervascular liver tumours underwent dual energy CT. The 80 kV set of the dual energy acquisition was reconstructed with standard filtered backprojection (FBP) and ASiR at different blending levels. Lesion contrast-to-noise ratio (CNR), reader’s confidence for lesion detection and characterisation, and reader’s evaluation of image quality were recorded.

Results

ASiR yielded significantly higher CNR values compared with FBP (P < 0.0001 for all comparisons). Reader’s perception of lesion conspicuity and confidence in the diagnosis of malignancy were also higher with 60 % and 80 % ASiR, compared with FBP (P = 0.01 and < 0.001, respectively). Compared with FBP, ASiR yielded nearly significantly lower specificity for lesion detection and a substantial decrease in the reader’s perception of image quality.

Conclusions

Compared with the standard FBP algorithm, ASiR significantly improves conspicuity of hypervascular liver lesions. This improvement may come at the cost of decreased specificity and reader’s perception of image quality.

Key Points

• Adaptive statistical iterative reconstruction algorithms (ASiRs) offer increasing potential in multidetector CT.
• An ASiR algorithm significantly improves conspicuity of hypervascular liver lesions at MDCT.
• Improved lesion conspicuity translates into increased reader’s confidence for diagnosis of malignancy.
• False positive findings may increase with ASiR, leading to potentially lower specificity.
Literatur
1.
Zurück zum Zitat Schindera ST, Nelson RC, Mukundan S et al (2008) Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection—phantom study. Radiology 246:125–132PubMedCrossRef Schindera ST, Nelson RC, Mukundan S et al (2008) Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection—phantom study. Radiology 246:125–132PubMedCrossRef
2.
Zurück zum Zitat Marin D, Nelson RC, Samei E et al (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection—initial clinical experience. Radiology 251:771–779PubMedCrossRef Marin D, Nelson RC, Samei E et al (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection—initial clinical experience. Radiology 251:771–779PubMedCrossRef
3.
Zurück zum Zitat Huda W, Scalzetti EM, Levin G (2000) technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435PubMedCrossRef Huda W, Scalzetti EM, Levin G (2000) technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435PubMedCrossRef
4.
Zurück zum Zitat Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526PubMedCrossRef Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526PubMedCrossRef
5.
Zurück zum Zitat Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef
6.
Zurück zum Zitat Singh S, Kalra MK, Hsieh J et al (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257:373–383PubMedCrossRef Singh S, Kalra MK, Hsieh J et al (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257:373–383PubMedCrossRef
7.
Zurück zum Zitat Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q (2010) Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195:713–719PubMedCrossRef Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q (2010) Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195:713–719PubMedCrossRef
8.
Zurück zum Zitat Prakash P, Kalra MK, Kambadakone AK et al (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Investig Radiol 45:202CrossRef Prakash P, Kalra MK, Kambadakone AK et al (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Investig Radiol 45:202CrossRef
9.
Zurück zum Zitat Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2012) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol 22:138–143PubMedCrossRef Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2012) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol 22:138–143PubMedCrossRef
10.
Zurück zum Zitat Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm—initial clinical experience. Radiology 254:145–153PubMedCrossRef Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm—initial clinical experience. Radiology 254:145–153PubMedCrossRef
11.
Zurück zum Zitat Bossuyt PM, Reitsma JB, Bruns DE et al (2003) Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. Clin Chem Lab Med 41:68–73PubMed Bossuyt PM, Reitsma JB, Bruns DE et al (2003) Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. Clin Chem Lab Med 41:68–73PubMed
12.
Zurück zum Zitat Sultana S, Awai K, Nakayama Y et al (2007) Hypervascular hepatocellular carcinomas: bolus tracking with a 40-detector CT scanner to time arterial phase imaging. Radiology 243:140–147PubMedCrossRef Sultana S, Awai K, Nakayama Y et al (2007) Hypervascular hepatocellular carcinomas: bolus tracking with a 40-detector CT scanner to time arterial phase imaging. Radiology 243:140–147PubMedCrossRef
13.
Zurück zum Zitat Miéville FA, Ayestaran P, Argaud C et al (2010) Potential benefit of the CT adaptive statistical iterative reconstruction method for pediatric cardiac diagnosis. Proc SPIE 7622:2D1–2D11 Miéville FA, Ayestaran P, Argaud C et al (2010) Potential benefit of the CT adaptive statistical iterative reconstruction method for pediatric cardiac diagnosis. Proc SPIE 7622:2D1–2D11
14.
Zurück zum Zitat Couinaud C (1957) Le foie: etudes anatomiques et chirurgicales. Masson et Cie, Paris Couinaud C (1957) Le foie: etudes anatomiques et chirurgicales. Masson et Cie, Paris
15.
16.
Zurück zum Zitat Nino-Murcia M, Olcott EW, Jeffrey RB, Lamm RL, Beaulieu CF, Jain KA (2000) Focal liver lesions: pattern-based classification scheme for enhancement at arterial phase CT. Radiology 215:746–751PubMedCrossRef Nino-Murcia M, Olcott EW, Jeffrey RB, Lamm RL, Beaulieu CF, Jain KA (2000) Focal liver lesions: pattern-based classification scheme for enhancement at arterial phase CT. Radiology 215:746–751PubMedCrossRef
17.
Zurück zum Zitat Medical imaging: the assessment of image quality (report no. 54). International Commission on Radiation Units and Measurements. Bethesda, MD: International Commission on Radiation Units and Measurements, 1996 Medical imaging: the assessment of image quality (report no. 54). International Commission on Radiation Units and Measurements. Bethesda, MD: International Commission on Radiation Units and Measurements, 1996
18.
Zurück zum Zitat Jessen K, Panzer W, Shrimpton P et al (2000) EUR 16262: European guidelines on quality criteria for computed tomography. Office for Official Publications of the European Communities, Luxembourg Jessen K, Panzer W, Shrimpton P et al (2000) EUR 16262: European guidelines on quality criteria for computed tomography. Office for Official Publications of the European Communities, Luxembourg
19.
Zurück zum Zitat Obuchowski NA, Rockette HE (1995) Hypothesis testing of the diagnostic accuracy for multiple diagnostic tests: an ANOVA approach with dependent observations. J Stat Comput Simul 24:285–308CrossRef Obuchowski NA, Rockette HE (1995) Hypothesis testing of the diagnostic accuracy for multiple diagnostic tests: an ANOVA approach with dependent observations. J Stat Comput Simul 24:285–308CrossRef
20.
Zurück zum Zitat Pinheiro J, Bates D, DebRoy S, Sarkar D; the R Development Core Team (2012) nlme: linear and nonlinear mixed effects models. R package version 3.1-105 Pinheiro J, Bates D, DebRoy S, Sarkar D; the R Development Core Team (2012) nlme: linear and nonlinear mixed effects models. R package version 3.1-105
21.
Zurück zum Zitat Kulkarni NM, Uppot RN, Eisner BH, Sahani DV (2012) radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: how low can we go? Radiology 265:158–166PubMedCrossRef Kulkarni NM, Uppot RN, Eisner BH, Sahani DV (2012) radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: how low can we go? Radiology 265:158–166PubMedCrossRef
22.
Zurück zum Zitat Brancatelli G, Baron RL, Peterson MS, Marsh W (2003) Helical CT screening for hepatocellular carcinoma in patients with cirrhosis: frequency and causes of false-positive interpretation. AJR Am J Roentgenol 180:1007–1014PubMedCrossRef Brancatelli G, Baron RL, Peterson MS, Marsh W (2003) Helical CT screening for hepatocellular carcinoma in patients with cirrhosis: frequency and causes of false-positive interpretation. AJR Am J Roentgenol 180:1007–1014PubMedCrossRef
23.
Zurück zum Zitat Hayashi PH, Trotter JF, Forman L et al (2004) Impact of pretransplant diagnosis of hepatocellular carcinoma on cadaveric liver allocation in the era of MELD. Liver Transpl 10:42–48PubMedCrossRef Hayashi PH, Trotter JF, Forman L et al (2004) Impact of pretransplant diagnosis of hepatocellular carcinoma on cadaveric liver allocation in the era of MELD. Liver Transpl 10:42–48PubMedCrossRef
24.
Zurück zum Zitat Freeman RB, Mithoefer A, Ruthazer R et al (2006) Optimizing staging for hepatocellular carcinoma before liver transplantation: a retrospective analysis of the UNOS/OPTN database. Liver Transpl 12:1504–1511PubMedCrossRef Freeman RB, Mithoefer A, Ruthazer R et al (2006) Optimizing staging for hepatocellular carcinoma before liver transplantation: a retrospective analysis of the UNOS/OPTN database. Liver Transpl 12:1504–1511PubMedCrossRef
25.
Zurück zum Zitat Richard S, Husarik DB, Yadava G, Murphy SN, Samei E (2012) Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys 39:4115PubMedCrossRef Richard S, Husarik DB, Yadava G, Murphy SN, Samei E (2012) Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys 39:4115PubMedCrossRef
26.
Zurück zum Zitat Guimarães LS, Fletcher JG, Harmsen WS et al (2010) Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. Radiology 257:732–742PubMedCrossRef Guimarães LS, Fletcher JG, Harmsen WS et al (2010) Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. Radiology 257:732–742PubMedCrossRef
Metadaten
Titel
Clinical impact of an adaptive statistical iterative reconstruction algorithm for detection of hypervascular liver tumours using a low tube voltage, high tube current MDCT technique
verfasst von
Daniele Marin
Kingshuk Roy Choudhury
Rajan T. Gupta
Lisa M. Ho
Brian C. Allen
Sebastian T. Schindera
James G. Colsher
Ehsan Samei
Rendon C. Nelson
Publikationsdatum
01.12.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2013
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-013-2964-1

Weitere Artikel der Ausgabe 12/2013

European Radiology 12/2013 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.