3D MR Sequence Capable of Simultaneous Image Acquisitions with and without Blood Vessel Suppression: Utility in Diagnosing Brain Metastases

verfasst von: Kazufumi Kikuchi, Akio Hiwatashi, Osamu Togao, Koji Yamashita, Masami Yoneyama, Makoto Obara, Junji Kishimoto, Takashi Yoshiura, Hiroshi Honda

Erschienen in: European Radiology | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Objective

Volume isotropic simultaneous interleaved bright- and black-blood examination (VISIBLE) is a recently developed 3D MR sequence that provides simultaneous acquisitions of images with blood vessel suppression (Black) and images without it (Bright). Our purpose was to evaluate the usefulness of VISIBLE in detecting brain metastases.

Methods

This prospective study included patients with suspected brain metastasis imaged with both VISIBLE and MPRAGE. From a data set, we compared the number of visualized blood vessels and the lesion-to-normal contrast-to-noise ratio (CNR) in 60 patients. We also performed an observer test to compare their diagnostic performance with VISIBLE, MPRAGE and only Black in 34 patients. Diagnostic performance was evaluated using a figure of merit (FOM), sensitivity, false-positive results per case (FPs/case) and reading time.

Results

The number of vessels was significantly fewer in Black compared to MPRAGE and Bright (P < 0.0001). CNR was significantly higher with both Black and Bright than with MPRAGE (P < 0.005). In the observer test, significantly higher sensitivity (P < 0.0001) and FOM (P < 0.0001), significantly shorter reading time (P = 0.0001) and similar FPs/case were achieved with VISIBLE compared to MPRAGE. Compared to only Black, VISIBLE resulted in comparable sensitivity, but significantly fewer FPs/case (P = 0.0008).

Conclusion

VISIBLE can improve radiologists’ diagnostic performance for brain metastasis.

Key Points

VISIBLE can achieve higher sensitivity and shorter reading time than MPRAGE.
VISIBLE can achieve lower false-positive rates than blood vessel suppressed images.
Compared to MPRAGE, VISIBLE can improve diagnostic performance for brain metastasis.
Literatur
1.
Zurück zum Zitat Posner JB, Chernik NL (1978) Intracranial metastases from systemic cancer. Adv Neurol 19:579–592PubMed Posner JB, Chernik NL (1978) Intracranial metastases from systemic cancer. Adv Neurol 19:579–592PubMed
2.
Zurück zum Zitat Rosner D, Nemoto T, Pickren J, Lane W (1983) Management of brain metastases from breast cancer by combination chemotherapy. J Neurooncol 1:131–137CrossRefPubMed Rosner D, Nemoto T, Pickren J, Lane W (1983) Management of brain metastases from breast cancer by combination chemotherapy. J Neurooncol 1:131–137CrossRefPubMed
3.
Zurück zum Zitat Linskey ME, Andrews DW, Asher AL et al (2010) The role of stereotactic radiosurgery in the management of patients with newly diagnosed brain metastases: a systematic review and evidence-based clinical practice guideline. J Neurooncol 96:45–68CrossRefPubMedCentralPubMed Linskey ME, Andrews DW, Asher AL et al (2010) The role of stereotactic radiosurgery in the management of patients with newly diagnosed brain metastases: a systematic review and evidence-based clinical practice guideline. J Neurooncol 96:45–68CrossRefPubMedCentralPubMed
4.
Zurück zum Zitat Lippitz B, Lindquist C, Paddick I, Peterson D, O’Neill K, Beaney R (2014) Stereotactic radiosurgery in the treatment of brain metastases: the current evidence. Cancer Treat Rev 40:48–59CrossRefPubMed Lippitz B, Lindquist C, Paddick I, Peterson D, O’Neill K, Beaney R (2014) Stereotactic radiosurgery in the treatment of brain metastases: the current evidence. Cancer Treat Rev 40:48–59CrossRefPubMed
5.
Zurück zum Zitat Monaco EA 3rd, Faraji AH, Berkowitz O et al (2013) Leukoencephalopathy after whole-brain radiation therapy plus radiosurgery versus radiosurgery alone for metastatic lung cancer. Cancer 119:226–232CrossRefPubMed Monaco EA 3rd, Faraji AH, Berkowitz O et al (2013) Leukoencephalopathy after whole-brain radiation therapy plus radiosurgery versus radiosurgery alone for metastatic lung cancer. Cancer 119:226–232CrossRefPubMed
6.
Zurück zum Zitat Serizawa T, Hirai T, Nagano O et al (2010) Gamma knife surgery for 1–10 brain metastases without prophylactic whole-brain radiation therapy: analysis of cases meeting the Japanese prospective multi-institute study (JLGK0901) inclusion criteria. J Neurooncol 98:163–167CrossRefPubMed Serizawa T, Hirai T, Nagano O et al (2010) Gamma knife surgery for 1–10 brain metastases without prophylactic whole-brain radiation therapy: analysis of cases meeting the Japanese prospective multi-institute study (JLGK0901) inclusion criteria. J Neurooncol 98:163–167CrossRefPubMed
7.
Zurück zum Zitat Hunter GK, Suh JH, Reuther AM et al (2012) Treatment of five or more brain metastases with stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 83:1394–1398CrossRefPubMed Hunter GK, Suh JH, Reuther AM et al (2012) Treatment of five or more brain metastases with stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 83:1394–1398CrossRefPubMed
8.
Zurück zum Zitat Schellinger PD, Meinck HM, Thron A (1999) Diagnostic accuracy of MRI compared to CCT in patients with brain metastases. J Neurooncol 44:275–281CrossRefPubMed Schellinger PD, Meinck HM, Thron A (1999) Diagnostic accuracy of MRI compared to CCT in patients with brain metastases. J Neurooncol 44:275–281CrossRefPubMed
9.
Zurück zum Zitat Kakeda S, Korogi Y, Hirai Y et al (2007) Detection of brain metastasis at 3T: comparison among SE, IR-FSE and 3D-GRE sequences. Eur Radiol 17:2345–2351CrossRefPubMed Kakeda S, Korogi Y, Hirai Y et al (2007) Detection of brain metastasis at 3T: comparison among SE, IR-FSE and 3D-GRE sequences. Eur Radiol 17:2345–2351CrossRefPubMed
10.
Zurück zum Zitat Takeda T, Takeda A, Nagaoka T et al (2008) Gadolinium-enhanced three-dimensional magnetization-prepared rapid gradient-echo (3D MP-RAGE) imaging is superior to spin-echo imaging in delineating brain metastases. Acta Radiol 49:1167–1173CrossRefPubMed Takeda T, Takeda A, Nagaoka T et al (2008) Gadolinium-enhanced three-dimensional magnetization-prepared rapid gradient-echo (3D MP-RAGE) imaging is superior to spin-echo imaging in delineating brain metastases. Acta Radiol 49:1167–1173CrossRefPubMed
11.
Zurück zum Zitat Komada T, Naganawa S, Ogawa H et al (2008) Contrast-enhanced MR imaging of metastatic brain tumor at 3 tesla: utility of T(1)-weighted SPACE compared with 2D spin echo and 3D gradient echo sequence. Magn Reson Med Sci 7:13–21CrossRefPubMed Komada T, Naganawa S, Ogawa H et al (2008) Contrast-enhanced MR imaging of metastatic brain tumor at 3 tesla: utility of T(1)-weighted SPACE compared with 2D spin echo and 3D gradient echo sequence. Magn Reson Med Sci 7:13–21CrossRefPubMed
12.
Zurück zum Zitat Kato Y, Higano S, Tamura H et al (2009) Usefulness of contrast-enhanced T1-weighted sampling perfection with application-optimized contrasts by using different flip angle evolutions in detection of small brain metastasis at 3T MR imaging: comparison with magnetization-prepared rapid acquisition of gradient echo imaging. AJNR Am J Neuroradiol 30:923–929CrossRefPubMed Kato Y, Higano S, Tamura H et al (2009) Usefulness of contrast-enhanced T1-weighted sampling perfection with application-optimized contrasts by using different flip angle evolutions in detection of small brain metastasis at 3T MR imaging: comparison with magnetization-prepared rapid acquisition of gradient echo imaging. AJNR Am J Neuroradiol 30:923–929CrossRefPubMed
13.
Zurück zum Zitat Nagao E, Yoshiura T, Hiwatashi A et al (2011) 3D turbo spin-echo sequence with motion-sensitized driven-equilibrium preparation for detection of brain metastases on 3T MR imaging. AJNR Am J Neuroradiol 32:664–670CrossRefPubMed Nagao E, Yoshiura T, Hiwatashi A et al (2011) 3D turbo spin-echo sequence with motion-sensitized driven-equilibrium preparation for detection of brain metastases on 3T MR imaging. AJNR Am J Neuroradiol 32:664–670CrossRefPubMed
14.
Zurück zum Zitat Park J, Kim EY (2010) Contrast-enhanced, three-dimensional, whole-brain, black-blood imaging: application to small brain metastases. Magn Reson Med 63:553–561CrossRefPubMed Park J, Kim EY (2010) Contrast-enhanced, three-dimensional, whole-brain, black-blood imaging: application to small brain metastases. Magn Reson Med 63:553–561CrossRefPubMed
15.
Zurück zum Zitat Yoneyama M, Nakamura M, Tabuchi T et al (2013) Whole-brain black-blood imaging with magnetization-transfer prepared spin echo-like contrast: a novel sequence for contrast-enhanced brain metastasis screening at 3T. Radiol Phys Technol 6:431–436CrossRefPubMed Yoneyama M, Nakamura M, Tabuchi T et al (2013) Whole-brain black-blood imaging with magnetization-transfer prepared spin echo-like contrast: a novel sequence for contrast-enhanced brain metastasis screening at 3T. Radiol Phys Technol 6:431–436CrossRefPubMed
16.
Zurück zum Zitat Wang J, Yarnykh VL, Hatsukami T, Chu B, Balu N, Yuan C (2007) Improved suppression of plaque-mimicking artifacts in black-blood carotid atherosclerosis imaging using a multislice motion-sensitized driven-equilibrium (MSDE) turbo spin-echo (TSE) sequence. Magn Reson Med 58:973–981CrossRefPubMed Wang J, Yarnykh VL, Hatsukami T, Chu B, Balu N, Yuan C (2007) Improved suppression of plaque-mimicking artifacts in black-blood carotid atherosclerosis imaging using a multislice motion-sensitized driven-equilibrium (MSDE) turbo spin-echo (TSE) sequence. Magn Reson Med 58:973–981CrossRefPubMed
17.
Zurück zum Zitat Wang J, Yarnykh VL, Yuan C (2010) Enhanced image quality in black-blood MRI using the improved motion-sensitized driven-equilibrium (iMSDE) sequence. J Magn Reson Imaging 31:1256–1263CrossRefPubMedCentralPubMed Wang J, Yarnykh VL, Yuan C (2010) Enhanced image quality in black-blood MRI using the improved motion-sensitized driven-equilibrium (iMSDE) sequence. J Magn Reson Imaging 31:1256–1263CrossRefPubMedCentralPubMed
18.
Zurück zum Zitat Dong L, Wang J, Yarnykh VL et al (2010) Efficient flow suppressed MRI improves interscan reproducibility of carotid atherosclerosis plaque burden measurements. J Magn Reson Imaging 32:452–458CrossRefPubMedCentralPubMed Dong L, Wang J, Yarnykh VL et al (2010) Efficient flow suppressed MRI improves interscan reproducibility of carotid atherosclerosis plaque burden measurements. J Magn Reson Imaging 32:452–458CrossRefPubMedCentralPubMed
20.
Zurück zum Zitat Obara M, Kuroda K, Wang J et al (2013) Comparison between two types of improved motion-sensitized driven-equilibrium (iMSDE) for intracranial black-blood imaging at 3.0 tesla. J Magn Reson Imaging. doi:10.1002/jmri.24430 Obara M, Kuroda K, Wang J et al (2013) Comparison between two types of improved motion-sensitized driven-equilibrium (iMSDE) for intracranial black-blood imaging at 3.0 tesla. J Magn Reson Imaging. doi:10.​1002/​jmri.​24430
21.
Zurück zum Zitat Yoneyama M, Obara M, Takahara T et al (2014) Volume isotropic simultaneous interleaved black-blood and bright-blood imaging: a novel sequence for contrast-enhanced brain metastasis screening. Magn Reson Med Sci. doi:10.2463/mrms. 2013-0065 Yoneyama M, Obara M, Takahara T et al (2014) Volume isotropic simultaneous interleaved black-blood and bright-blood imaging: a novel sequence for contrast-enhanced brain metastasis screening. Magn Reson Med Sci. doi:10.​2463/​mrms.​ 2013-0065
22.
Zurück zum Zitat Chakraborty DP, Berbaum KS (2004) Observer studies involving detection and localization: modeling, analysis, and validation. Med Phys 31:2313–2330CrossRefPubMed Chakraborty DP, Berbaum KS (2004) Observer studies involving detection and localization: modeling, analysis, and validation. Med Phys 31:2313–2330CrossRefPubMed
23.
Zurück zum Zitat Chakraborty DP (2006) Analysis of location specific observer performance data: validated extensions of the jackknife free-response (JAFROC) method. Acad Radiol 13:1187–1193CrossRefPubMed Chakraborty DP (2006) Analysis of location specific observer performance data: validated extensions of the jackknife free-response (JAFROC) method. Acad Radiol 13:1187–1193CrossRefPubMed
Metadaten
Titel
3D MR Sequence Capable of Simultaneous Image Acquisitions with and without Blood Vessel Suppression: Utility in Diagnosing Brain Metastases
verfasst von
Kazufumi Kikuchi
Akio Hiwatashi
Osamu Togao
Koji Yamashita
Masami Yoneyama
Makoto Obara
Junji Kishimoto
Takashi Yoshiura
Hiroshi Honda
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 4/2015
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3496-z

Weitere Artikel der Ausgabe 4/2015

European Radiology 4/2015 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.