14.08.2015 | Computed Tomography

Reduction of metallic coil artefacts in computed tomography body imaging: effects of a new single-energy metal artefact reduction algorithm

verfasst von: Masafumi Kidoh, Daisuke Utsunomiya, Osamu Ikeda, Yoshitaka Tamura, Seitaro Oda, Yoshinori Funama, Hideaki Yuki, Takeshi Nakaura, Takayuki Kawano, Toshinori Hirai, Yasuyuki Yamashita

Erschienen in: European Radiology | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

Objectives

We evaluated the effect of a single-energy metal artefact reduction (SEMAR) algorithm for metallic coil artefact reduction in body imaging.

Methods

Computed tomography angiography (CTA) was performed in 30 patients with metallic coils (10 men, 20 women; mean age, 67.9 ± 11 years). Non-SEMAR images were reconstructed with iterative reconstruction alone, and SEMAR images were reconstructed with the iterative reconstruction plus SEMAR algorithms. We compared image noise around metallic coils and the maximum diameters of artefacts from coils between the non-SEMAR and SEMAR images. Two radiologists visually evaluated the metallic coil artefacts utilizing a four-point scale: 1 = extensive; 2 = strong; 3 = mild; 4 = minimal artefacts.

Results

The image noise and maximum diameters of the artefacts of the SEMAR images were significantly lower than those of the non-SEMAR images (65.1 ± 33.0 HU vs. 29.7 ± 10.3 HU; 163.9 ± 54.8 mm vs. 10.3 ± 19.0 mm, respectively; P < 0.001). Better visual scores were obtained with the SEMAR technique (3.4 ± 0.6 vs. 1.0 ± 0.0, P < 0.001).

Conclusions

The SEMAR algorithm significantly reduced artefacts caused by metallic coils compared with the non-SEMAR algorithm. This technique can potentially increase CT performance for the evaluation of post-coil embolization complications.

Key Points

The new algorithm involves a raw data- and image-based reconstruction technique.
The new algorithm mitigates artefacts from metallic coils on body CT images.
The new algorithm significantly reduced artefacts caused by metallic coils.
The metal artefact reduction algorithm improves CT image quality after coil embolization.
Literatur
1.
Zurück zum Zitat Ikeda O, Tamura Y, Nakasone Y, Iryou Y, Yamashita Y (2008) Nonoperative management of unruptured visceral artery aneurysms: treatment by transcatheter coil embolization. J Vasc Surg 47:1212–1219CrossRefPubMed Ikeda O, Tamura Y, Nakasone Y, Iryou Y, Yamashita Y (2008) Nonoperative management of unruptured visceral artery aneurysms: treatment by transcatheter coil embolization. J Vasc Surg 47:1212–1219CrossRefPubMed
2.
Zurück zum Zitat Ikeda O, Nakasone Y, Tamura Y, Yamashita Y (2010) Endovascular management of visceral artery pseudoaneurysms: transcatheter coil embolization using the isolation technique. Cardiovasc Intervent Radiol 33:1128–1134CrossRefPubMed Ikeda O, Nakasone Y, Tamura Y, Yamashita Y (2010) Endovascular management of visceral artery pseudoaneurysms: transcatheter coil embolization using the isolation technique. Cardiovasc Intervent Radiol 33:1128–1134CrossRefPubMed
3.
Zurück zum Zitat Han YM, Song SK, Hwang HP et al (2013) Images in vascular medicine. Large congenital renal arteriovenous malformation treated with interlock coil embolization. Vasc Med 18:237–238CrossRefPubMed Han YM, Song SK, Hwang HP et al (2013) Images in vascular medicine. Large congenital renal arteriovenous malformation treated with interlock coil embolization. Vasc Med 18:237–238CrossRefPubMed
4.
Zurück zum Zitat Johnston SC, Zhao S, Dudley RA, Berman MF, Gress DR (2001) Treatment of unruptured cerebral aneurysms in California. Stroke 32:597–605CrossRefPubMed Johnston SC, Zhao S, Dudley RA, Berman MF, Gress DR (2001) Treatment of unruptured cerebral aneurysms in California. Stroke 32:597–605CrossRefPubMed
5.
Zurück zum Zitat Etezadi V, Gandhi RT, Benenati JF et al (2011) Endovascular treatment of visceral and renal artery aneurysms. J Vasc Interv Radiol 22:1246–1253CrossRefPubMed Etezadi V, Gandhi RT, Benenati JF et al (2011) Endovascular treatment of visceral and renal artery aneurysms. J Vasc Interv Radiol 22:1246–1253CrossRefPubMed
6.
Zurück zum Zitat Yasumoto T, Osuga K, Yamamoto H et al (2013) Long-term outcomes of coil packing for visceral aneurysms: correlation between packing density and incidence of coil compaction or recanalization. J Vasc Interv Radiol 24:1798–1807CrossRefPubMed Yasumoto T, Osuga K, Yamamoto H et al (2013) Long-term outcomes of coil packing for visceral aneurysms: correlation between packing density and incidence of coil compaction or recanalization. J Vasc Interv Radiol 24:1798–1807CrossRefPubMed
7.
Zurück zum Zitat Gallas S, Januel AC, Pasco A et al (2009) Long-term follow-up of 1036 cerebral aneurysms treated by bare coils: a multicentric cohort treated between 1998 and 2003. AJNR Am J Neuroradiol 30:1986–1992CrossRefPubMed Gallas S, Januel AC, Pasco A et al (2009) Long-term follow-up of 1036 cerebral aneurysms treated by bare coils: a multicentric cohort treated between 1998 and 2003. AJNR Am J Neuroradiol 30:1986–1992CrossRefPubMed
8.
Zurück zum Zitat Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRefPubMed Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691CrossRefPubMed
9.
Zurück zum Zitat Lell MM, Meyer E, Schmid M et al (2013) Frequency split metal artefact reduction in pelvic computed tomography. Eur Radiol 23:2137–2145CrossRefPubMed Lell MM, Meyer E, Schmid M et al (2013) Frequency split metal artefact reduction in pelvic computed tomography. Eur Radiol 23:2137–2145CrossRefPubMed
10.
Zurück zum Zitat Morsbach F, Wurnig M, Kunz DM et al (2013) Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions. Eur Radiol 23:2687–2694CrossRefPubMed Morsbach F, Wurnig M, Kunz DM et al (2013) Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions. Eur Radiol 23:2687–2694CrossRefPubMed
11.
Zurück zum Zitat Funama Y, Taguchi K, Utsunomiya D et al (2015) A newly-developed metal artifact reduction algorithm improves the visibility of oral cavity lesions on 320-MDCT volume scans. Phys Med 31:66–71CrossRefPubMed Funama Y, Taguchi K, Utsunomiya D et al (2015) A newly-developed metal artifact reduction algorithm improves the visibility of oral cavity lesions on 320-MDCT volume scans. Phys Med 31:66–71CrossRefPubMed
12.
Zurück zum Zitat Gondim Teixeira PA, Meyer JB, Baumann C et al (2014) Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skelet Radiol 43:1237–1246CrossRef Gondim Teixeira PA, Meyer JB, Baumann C et al (2014) Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skelet Radiol 43:1237–1246CrossRef
13.
Zurück zum Zitat Lemmens C, Faul D, Nuyts J (2009) Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans Med Imaging 28:250–260CrossRefPubMed Lemmens C, Faul D, Nuyts J (2009) Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans Med Imaging 28:250–260CrossRefPubMed
14.
Zurück zum Zitat Geisel D, Gebauer B, Malinowski M, Stockmann M, Denecke T (2014) Comparison of CT and MRI artefacts from coils and vascular plugs used for portal vein embolization. Eur J Radiol 83:692–695CrossRefPubMed Geisel D, Gebauer B, Malinowski M, Stockmann M, Denecke T (2014) Comparison of CT and MRI artefacts from coils and vascular plugs used for portal vein embolization. Eur J Radiol 83:692–695CrossRefPubMed
15.
Zurück zum Zitat Kidoh M, Nakaura T, Nakamura S et al (2014) Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR). Clin Radiol 69:e11–e16CrossRefPubMed Kidoh M, Nakaura T, Nakamura S et al (2014) Reduction of dental metallic artefacts in CT: value of a newly developed algorithm for metal artefact reduction (O-MAR). Clin Radiol 69:e11–e16CrossRefPubMed
16.
Zurück zum Zitat Patel A, Weintraub JL, Nowakowski FS et al (2012) Single-center experience with elective transcatheter coil embolization of splenic artery aneurysms: technique and midterm follow-up. J Vasc Interv Radiol 23:893–899CrossRefPubMed Patel A, Weintraub JL, Nowakowski FS et al (2012) Single-center experience with elective transcatheter coil embolization of splenic artery aneurysms: technique and midterm follow-up. J Vasc Interv Radiol 23:893–899CrossRefPubMed
17.
Zurück zum Zitat Tekola BD, Arner DM, Behm BW (2013) Coil migration after transarterial coil embolization of a splenic artery pseudoaneurysm. Case Rep Gastroenterol 7:487–491CrossRefPubMedPubMedCentral Tekola BD, Arner DM, Behm BW (2013) Coil migration after transarterial coil embolization of a splenic artery pseudoaneurysm. Case Rep Gastroenterol 7:487–491CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Xu C, Verhaegen F, Laurendeau D, Enger SA, Beaulieu L (2011) An algorithm for efficient metal artifact reductions in permanent seed. Med Phys 38:47–56CrossRefPubMed Xu C, Verhaegen F, Laurendeau D, Enger SA, Beaulieu L (2011) An algorithm for efficient metal artifact reductions in permanent seed. Med Phys 38:47–56CrossRefPubMed
20.
Zurück zum Zitat Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37:5482–5493CrossRefPubMed Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M (2010) Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys 37:5482–5493CrossRefPubMed
21.
Zurück zum Zitat Zhao S, Robertson DD, Wang G, Whiting B, Bae KT (2000) X-ray CT metal artifact reduction using wavelets: an application for imaging total hip prostheses. IEEE Trans Med Imaging 19:1238–1247CrossRefPubMed Zhao S, Robertson DD, Wang G, Whiting B, Bae KT (2000) X-ray CT metal artifact reduction using wavelets: an application for imaging total hip prostheses. IEEE Trans Med Imaging 19:1238–1247CrossRefPubMed
22.
Zurück zum Zitat Jeong S, Kim SH, Hwang EJ, Shin CI, Han JK, Choi BI (2015) Usefulness of a metal artifact reduction algorithm for orthopedic implants in abdominal CT: phantom and clinical study results. AJR Am J Roentgenol 204:307–317CrossRefPubMed Jeong S, Kim SH, Hwang EJ, Shin CI, Han JK, Choi BI (2015) Usefulness of a metal artifact reduction algorithm for orthopedic implants in abdominal CT: phantom and clinical study results. AJR Am J Roentgenol 204:307–317CrossRefPubMed
23.
Zurück zum Zitat Hilgers G, Nuver T, Minken A (2014) The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants. J Appl Clin Med Phys 15:4597PubMed Hilgers G, Nuver T, Minken A (2014) The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants. J Appl Clin Med Phys 15:4597PubMed
24.
Zurück zum Zitat Cheng PM, Romero M, Duddalwar VA (2014) Pulmonary pseudoemboli: a new artifact arising from a commercial metal artifact reduction algorithm for computed tomographic image reconstruction. J Comput Assist Tomogr 38:159–162CrossRefPubMed Cheng PM, Romero M, Duddalwar VA (2014) Pulmonary pseudoemboli: a new artifact arising from a commercial metal artifact reduction algorithm for computed tomographic image reconstruction. J Comput Assist Tomogr 38:159–162CrossRefPubMed
25.
Zurück zum Zitat Li H, Noel C, Chen H et al (2012) Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys 39:7507–7517CrossRefPubMedPubMedCentral Li H, Noel C, Chen H et al (2012) Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys 39:7507–7517CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Guggenberger R, Winklhofer S, Osterhoff G et al (2012) Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur Radiol 22:2357–2364CrossRefPubMed Guggenberger R, Winklhofer S, Osterhoff G et al (2012) Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur Radiol 22:2357–2364CrossRefPubMed
27.
Zurück zum Zitat Pessis E, Campagna R, Sverzut JM et al (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed Pessis E, Campagna R, Sverzut JM et al (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed
28.
Zurück zum Zitat Kuchenbecker S, Faby S, Sawall S, Lell M, Kachelriess M (2015) Dual energy CT: how well can pseudo-monochromatic imaging reduce metal artifacts? Med Phys 42:1023–1036CrossRefPubMed Kuchenbecker S, Faby S, Sawall S, Lell M, Kachelriess M (2015) Dual energy CT: how well can pseudo-monochromatic imaging reduce metal artifacts? Med Phys 42:1023–1036CrossRefPubMed
29.
Zurück zum Zitat Mamourian AC, Erkmen K, Pluta DJ (2008) Nonhelical acquisition CT angiogram after aneurysmal clipping: in vitro testing shows diminished artifact. AJNR Am J Neuroradiol 29:660–662CrossRefPubMed Mamourian AC, Erkmen K, Pluta DJ (2008) Nonhelical acquisition CT angiogram after aneurysmal clipping: in vitro testing shows diminished artifact. AJNR Am J Neuroradiol 29:660–662CrossRefPubMed
Metadaten
Titel
Reduction of metallic coil artefacts in computed tomography body imaging: effects of a new single-energy metal artefact reduction algorithm
verfasst von
Masafumi Kidoh
Daisuke Utsunomiya
Osamu Ikeda
Yoshitaka Tamura
Seitaro Oda
Yoshinori Funama
Hideaki Yuki
Takeshi Nakaura
Takayuki Kawano
Toshinori Hirai
Yasuyuki Yamashita
Publikationsdatum
14.08.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 5/2016
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-015-3950-6

Weitere Artikel der Ausgabe 5/2016

European Radiology 5/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.