Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study

verfasst von: Hyun Joo Shin, Myung-Joon Kim, Ha Yan Kim, Yun Ho Roh, Mi-Jung Lee

Erschienen in: European Radiology | Ausgabe 10/2016

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To investigate consistency in shear wave velocities (SWVs) on ultrasound elastography using different machines, transducers and acquisition depths.

Methods

The SWVs were measured using an elasticity phantom with a Young’s modulus of 16.9 kPa, with three recently introduced ultrasound elastography machines (A, B and C from different vendors) and two transducers (low and high frequencies) at four depths (2, 3, 4 and 5 cm). Mean SWVs from 15 measurements and coefficient of variations (CVs) were compared between three machines, two transducers and four acquisition depths.

Results

The SWVs using the high frequency transducer were not acquired at 5 cm depth in machine B, and a high frequency transducer was not available in machine C. The mean SWVs in the three machines were different (p ≤ 0.002). The CVs were 0–0.09 in three machines. The mean SWVs between the two transducers were different (p < 0.001) except at 4 and 5 cm depths in machine A. The SWVs were affected by the acquisition depths in all conditions (p < 0.001).

Conclusion

There is considerable difference in SWVs on ultrasound elastography depending on different machines, transducers and acquisition depths. Caution is needed when using the cutoff values of SWVs in different conditions.

Key Points

The shear wave velocities (SWVs) are different between different ultrasound elastography machines
The SWVs are also different between different transducers and acquisition depths
Caution is needed when using the cutoff SWVs measured under different conditions
Literatur
1.
Zurück zum Zitat Ophir J, Cespedes I, Ponnekanti H, Yazdi Y, Li X (1991) Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging 13:111–134CrossRefPubMed Ophir J, Cespedes I, Ponnekanti H, Yazdi Y, Li X (1991) Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging 13:111–134CrossRefPubMed
2.
Zurück zum Zitat Gennisson JL, Deffieux T, Fink M, Tanter M (2013) Ultrasound elastography: principles and techniques. Diagn Interv Imaging 94:487–495CrossRefPubMed Gennisson JL, Deffieux T, Fink M, Tanter M (2013) Ultrasound elastography: principles and techniques. Diagn Interv Imaging 94:487–495CrossRefPubMed
3.
Zurück zum Zitat Jeong WK, Lim HK, Lee HK, Jo JM, Kim Y (2014) Principles and clinical application of ultrasound elastography for diffuse liver disease. Ultrasonography 33:149–160CrossRefPubMedPubMedCentral Jeong WK, Lim HK, Lee HK, Jo JM, Kim Y (2014) Principles and clinical application of ultrasound elastography for diffuse liver disease. Ultrasonography 33:149–160CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Park MK (2014) Usefulness of acoustic radiation force impulse elastography in the differential diagnosis of benign and malignant solid pancreatic lesions. Ultrasonography 33:26–33CrossRefPubMed Park MK (2014) Usefulness of acoustic radiation force impulse elastography in the differential diagnosis of benign and malignant solid pancreatic lesions. Ultrasonography 33:26–33CrossRefPubMed
5.
Zurück zum Zitat Carlsen JF, Ewertsen C, Saftoiu A, Lonn L, Nielsen MB (2014) Accuracy of visual scoring and semi-quantification of ultrasound strain elastography–a phantom study. PLoS One 9, e88699CrossRefPubMedPubMedCentral Carlsen JF, Ewertsen C, Saftoiu A, Lonn L, Nielsen MB (2014) Accuracy of visual scoring and semi-quantification of ultrasound strain elastography–a phantom study. PLoS One 9, e88699CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Havre RF, Elde E, Gilja OH, Odegaard S, Eide GE, Matre K et al (2008) Freehand real-time elastography: impact of scanning parameters on image quality and in vitro intra- and interobserver validations. Ultrasound Med Biol 34:1638–1650CrossRefPubMed Havre RF, Elde E, Gilja OH, Odegaard S, Eide GE, Matre K et al (2008) Freehand real-time elastography: impact of scanning parameters on image quality and in vitro intra- and interobserver validations. Ultrasound Med Biol 34:1638–1650CrossRefPubMed
7.
Zurück zum Zitat Friedrich-Rust M, Wunder K, Kriener S, Sotoudeh F, Richter S, Bojunga J et al (2009) Liver fibrosis in viral hepatitis: noninvasive assessment with acoustic radiation force impulse imaging versus transient elastography. Radiology 252:595–604CrossRefPubMed Friedrich-Rust M, Wunder K, Kriener S, Sotoudeh F, Richter S, Bojunga J et al (2009) Liver fibrosis in viral hepatitis: noninvasive assessment with acoustic radiation force impulse imaging versus transient elastography. Radiology 252:595–604CrossRefPubMed
8.
Zurück zum Zitat Grazhdani H, Cantisani V, Lodise P, Di Rocco G, Proietto MC, Fioravanti E et al (2014) Prospective evaluation of acoustic radiation force impulse technology in the differentiation of thyroid nodules: accuracy and interobserver variability assessment. J Ultrasound 17:13–20CrossRefPubMedPubMedCentral Grazhdani H, Cantisani V, Lodise P, Di Rocco G, Proietto MC, Fioravanti E et al (2014) Prospective evaluation of acoustic radiation force impulse technology in the differentiation of thyroid nodules: accuracy and interobserver variability assessment. J Ultrasound 17:13–20CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Wang CZ, Zheng J, Huang ZP, Xiao Y, Song D, Zeng J et al (2014) Influence of measurement depth on the stiffness assessment of healthy liver with real-time shear wave elastography. Ultrasound Med Biol 40:461–469CrossRefPubMed Wang CZ, Zheng J, Huang ZP, Xiao Y, Song D, Zeng J et al (2014) Influence of measurement depth on the stiffness assessment of healthy liver with real-time shear wave elastography. Ultrasound Med Biol 40:461–469CrossRefPubMed
10.
Zurück zum Zitat Tozaki M, Saito M, Joo C, Yamaguchi M, Isobe S, Ogawa Y et al (2011) Ultrasonographic tissue quantification of the breast using acoustic radiation force impulse technology: phantom study and clinical application. Jpn J Radiol 29:598–603CrossRefPubMed Tozaki M, Saito M, Joo C, Yamaguchi M, Isobe S, Ogawa Y et al (2011) Ultrasonographic tissue quantification of the breast using acoustic radiation force impulse technology: phantom study and clinical application. Jpn J Radiol 29:598–603CrossRefPubMed
11.
Zurück zum Zitat Carlsen JF, Pedersen MR, Ewertsen C, Saftoiu A, Lonn L, Rafaelsen SR et al (2015) A comparative study of strain and shear-wave elastography in an elasticity phantom. AJR Am J Roentgenol 204:W236–W242CrossRefPubMed Carlsen JF, Pedersen MR, Ewertsen C, Saftoiu A, Lonn L, Rafaelsen SR et al (2015) A comparative study of strain and shear-wave elastography in an elasticity phantom. AJR Am J Roentgenol 204:W236–W242CrossRefPubMed
12.
Zurück zum Zitat Hall TJ, Milkowski A, Garra B, Carson P, Palmeri M, Nightingale K, Lynch T, Alturki A, Andre M, Audiere S (2013) RSNA/QIBA: shear wave speed as a biomarker for liver fibrosis staging. Ultrasonics Symposium (IUS), 2013 I.E. International. IEEE, pp 397–400 Hall TJ, Milkowski A, Garra B, Carson P, Palmeri M, Nightingale K, Lynch T, Alturki A, Andre M, Audiere S (2013) RSNA/QIBA: shear wave speed as a biomarker for liver fibrosis staging. Ultrasonics Symposium (IUS), 2013 I.E. International. IEEE, pp 397–400
13.
Zurück zum Zitat Chang S, Kim MJ, Kim J, Lee MJ (2013) Variability of shear wave velocity using different frequencies in acoustic radiation force impulse (ARFI) elastography: a phantom and normal liver study. Ultraschall Med 34:260–265PubMed Chang S, Kim MJ, Kim J, Lee MJ (2013) Variability of shear wave velocity using different frequencies in acoustic radiation force impulse (ARFI) elastography: a phantom and normal liver study. Ultraschall Med 34:260–265PubMed
14.
Zurück zum Zitat Dillman JR, Chen S, Davenport MS, Zhao H, Urban MW, Song P et al (2015) Superficial ultrasound shear wave speed measurements in soft and hard elasticity phantoms: repeatability and reproducibility using two ultrasound systems. Pediatr Radiol 45:376–385CrossRefPubMed Dillman JR, Chen S, Davenport MS, Zhao H, Urban MW, Song P et al (2015) Superficial ultrasound shear wave speed measurements in soft and hard elasticity phantoms: repeatability and reproducibility using two ultrasound systems. Pediatr Radiol 45:376–385CrossRefPubMed
15.
Zurück zum Zitat Yamanaka N, Kaminuma C, Taketomi-Takahashi A, Tsushima Y (2012) Reliable measurement by virtual touch tissue quantification with acoustic radiation force impulse imaging: phantom study. J Ultrasound Med 31:1239–1244PubMed Yamanaka N, Kaminuma C, Taketomi-Takahashi A, Tsushima Y (2012) Reliable measurement by virtual touch tissue quantification with acoustic radiation force impulse imaging: phantom study. J Ultrasound Med 31:1239–1244PubMed
16.
Zurück zum Zitat Zhao H, Song P, Urban MW, Kinnick RR, Yin M, Greenleaf JF et al (2011) Bias observed in time-of-flight shear wave speed measurements using radiation force of a focused ultrasound beam. Ultrasound Med Biol 37:1884–1892CrossRefPubMedPubMedCentral Zhao H, Song P, Urban MW, Kinnick RR, Yin M, Greenleaf JF et al (2011) Bias observed in time-of-flight shear wave speed measurements using radiation force of a focused ultrasound beam. Ultrasound Med Biol 37:1884–1892CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Liu B, Liang J, Zheng Y, Xie X, Huang G, Zhou L et al (2015) Two-dimensional shear wave elastography as promising diagnostic tool for predicting malignant thyroid nodules: a prospective single-centre experience. Eur Radiol 25:624–634CrossRefPubMed Liu B, Liang J, Zheng Y, Xie X, Huang G, Zhou L et al (2015) Two-dimensional shear wave elastography as promising diagnostic tool for predicting malignant thyroid nodules: a prospective single-centre experience. Eur Radiol 25:624–634CrossRefPubMed
18.
Zurück zum Zitat Ko KH, Jung HK, Kim SJ, Kim H, Yoon JH (2014) Potential role of shear-wave ultrasound elastography for the differential diagnosis of breast non-mass lesions: preliminary report. Eur Radiol 24:305–311CrossRefPubMed Ko KH, Jung HK, Kim SJ, Kim H, Yoon JH (2014) Potential role of shear-wave ultrasound elastography for the differential diagnosis of breast non-mass lesions: preliminary report. Eur Radiol 24:305–311CrossRefPubMed
19.
Zurück zum Zitat Vergari C, Dubois G, Vialle R, Gennisson JL, Tanter M, Dubousset J et al (2015) Lumbar annulus fibrosus biomechanical characterization in healthy children by ultrasound shear wave elastography. Eur Radiol. doi:10.1007/s00330-015-3911-0 PubMed Vergari C, Dubois G, Vialle R, Gennisson JL, Tanter M, Dubousset J et al (2015) Lumbar annulus fibrosus biomechanical characterization in healthy children by ultrasound shear wave elastography. Eur Radiol. doi:10.​1007/​s00330-015-3911-0 PubMed
20.
Zurück zum Zitat Vergari C, Rouch P, Dubois G, Bonneau D, Dubousset J, Tanter M et al (2014) Non-invasive biomechanical characterization of intervertebral discs by shear wave ultrasound elastography: a feasibility study. Eur Radiol 24:3210–3216CrossRefPubMed Vergari C, Rouch P, Dubois G, Bonneau D, Dubousset J, Tanter M et al (2014) Non-invasive biomechanical characterization of intervertebral discs by shear wave ultrasound elastography: a feasibility study. Eur Radiol 24:3210–3216CrossRefPubMed
21.
22.
Zurück zum Zitat Sun Z, Xie M, Xiang F, Song Y, Yu C, Zhang Y et al (2015) Utility of real-time shear wave elastography in the assessment of testicular torsion. PLoS One 10, e0138523CrossRefPubMedPubMedCentral Sun Z, Xie M, Xiang F, Song Y, Yu C, Zhang Y et al (2015) Utility of real-time shear wave elastography in the assessment of testicular torsion. PLoS One 10, e0138523CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Fontanilla T, Canas T, Macia A, Alfageme M, Gutierrez Junquera C, Malalana A et al (2014) Normal values of liver shear wave velocity in healthy children assessed by acoustic radiation force impulse imaging using a convex probe and a linear probe. Ultrasound Med Biol 40:470–477CrossRefPubMed Fontanilla T, Canas T, Macia A, Alfageme M, Gutierrez Junquera C, Malalana A et al (2014) Normal values of liver shear wave velocity in healthy children assessed by acoustic radiation force impulse imaging using a convex probe and a linear probe. Ultrasound Med Biol 40:470–477CrossRefPubMed
24.
Zurück zum Zitat Franchi-Abella S, Elie C, Correas JM (2013) Ultrasound elastography: advantages, limitations and artefacts of the different techniques from a study on a phantom. Diagn Interv Imaging 94:497–501CrossRefPubMed Franchi-Abella S, Elie C, Correas JM (2013) Ultrasound elastography: advantages, limitations and artefacts of the different techniques from a study on a phantom. Diagn Interv Imaging 94:497–501CrossRefPubMed
25.
Zurück zum Zitat Sporea I, Bota S, Gradinaru-Tascau O, Sirli R, Popescu A (2014) Comparative study between two point shear wave elastographic techniques: acoustic radiation force impulse (ARFI) elastography and ElastPQ. Med Ultrason 16:309–314PubMed Sporea I, Bota S, Gradinaru-Tascau O, Sirli R, Popescu A (2014) Comparative study between two point shear wave elastographic techniques: acoustic radiation force impulse (ARFI) elastography and ElastPQ. Med Ultrason 16:309–314PubMed
26.
Zurück zum Zitat Huang ZP, Zhang XL, Zeng J, Zheng J, Wang P, Zheng RQ (2014) Study of detection times for liver stiffness evaluation by shear wave elastography. World J Gastroenterol 20:9578–9584PubMedPubMedCentral Huang ZP, Zhang XL, Zeng J, Zheng J, Wang P, Zheng RQ (2014) Study of detection times for liver stiffness evaluation by shear wave elastography. World J Gastroenterol 20:9578–9584PubMedPubMedCentral
27.
Zurück zum Zitat Lee MJ, Kim MJ, Han KH, Yoon CS (2013) Age-related changes in liver, kidney, and spleen stiffness in healthy children measured with acoustic radiation force impulse imaging. Eur J Radiol 82:e290–e294CrossRefPubMed Lee MJ, Kim MJ, Han KH, Yoon CS (2013) Age-related changes in liver, kidney, and spleen stiffness in healthy children measured with acoustic radiation force impulse imaging. Eur J Radiol 82:e290–e294CrossRefPubMed
28.
Zurück zum Zitat Gallotti A, D'Onofrio M, Pozzi Mucelli R (2010) Acoustic radiation force impulse (ARFI) technique in ultrasound with Virtual Touch tissue quantification of the upper abdomen. Radiol Med 115:889–897CrossRefPubMed Gallotti A, D'Onofrio M, Pozzi Mucelli R (2010) Acoustic radiation force impulse (ARFI) technique in ultrasound with Virtual Touch tissue quantification of the upper abdomen. Radiol Med 115:889–897CrossRefPubMed
29.
Zurück zum Zitat Zheng XZ, Ji P, Mao HW, Zhang XY, Xia EH, Xing G et al (2011) A novel approach to assessing changes in prostate stiffness with age using virtual touch tissue quantification. J Ultrasound Med 30:387–390PubMed Zheng XZ, Ji P, Mao HW, Zhang XY, Xia EH, Xing G et al (2011) A novel approach to assessing changes in prostate stiffness with age using virtual touch tissue quantification. J Ultrasound Med 30:387–390PubMed
Metadaten
Titel
Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study
verfasst von
Hyun Joo Shin
Myung-Joon Kim
Ha Yan Kim
Yun Ho Roh
Mi-Jung Lee
Publikationsdatum
26.01.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 10/2016
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4212-y

Weitere Artikel der Ausgabe 10/2016

European Radiology 10/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.