Diagnostic accuracy of coronary CT angiography using 3rd-generation dual-source CT and automated tube voltage selection: Clinical application in a non-obese and obese patient population
verfasst von:
Stefanie Mangold, Julian L. Wichmann, U. Joseph Schoepf, Damiano Caruso, Christian Tesche, Daniel H. Steinberg, Akos Varga-Szemes, Andrew C. Stubenrauch, Richard R. Bayer II, Matthew Biancalana, Konstantin Nikolaou, Carlo N. De Cecco
Erschienen in:
European Radiology
|
Ausgabe 6/2017
Einloggen, um Zugang zu erhalten
Abstract
Purpose
To investigate diagnostic accuracy of 3rd-generation dual-source CT (DSCT) coronary angiography in obese and non-obese patients.
Methods
We retrospectively analyzed 76 patients who underwent coronary CT angiography (CCTA) and invasive coronary angiography. Prospectively ECG-triggered acquisition was performed with automated tube voltage selection (ATVS). Patients were dichotomized based on body mass index in groups A (<30 kg/m2, n = 37) and B (≥30 kg/m2, n = 39) and based on tube voltage in groups C (<120 kV, n = 46) and D (120 kV, n = 30). Coronary arteries were assessed for significant stenoses (≥50 % luminal narrowing) and diagnostic accuracy was calculated.
Results
Per-patient overall sensitivity, specificity, positive predictive value, negative predictive value (NPV) and accuracy were 96.9 %, 95.5 %, 93.9 %, 97.7 % and 96.1 %, respectively. Sensitivity and NPV were lower in groups B and D compared to groups A and C, but no statistically significant differences were observed (group A vs. B: sensitivity, 100.0 % vs. 93.3 %, p = 0.9493; NPV, 100 % vs. 95.5 %, p = 0.9812; group C vs. D: sensitivity, 100.0 % vs. 92.3 %, p = 0.8462; NPV, 100.0 % vs. 94.1 %, p = 0.8285).
Conclusion
CCTA using 3rd-generation DSCT and (ATVS) provides high diagnostic accuracy in both non-obese and obese patients.
Key Points
• Coronary CTA provides high diagnostic accuracy in non-obese and obese patients.
• Diagnostic accuracy between obese and non-obese patients showed no significant difference.
• <120 kV studies were performed in 44 % of obese patients.
• Current radiation dose-saving approaches can be applied independent of body habitus.