Comparison of the effects of model-based iterative reconstruction and filtered back projection algorithms on software measurements in pulmonary subsolid nodules

verfasst von: Julien G. Cohen, Hyungjin Kim, Su Bin Park, Bram van Ginneken, Gilbert R. Ferretti, Chang Hyun Lee, Jin Mo Goo, Chang Min Park

Erschienen in: European Radiology | Ausgabe 8/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the differences between filtered back projection (FBP) and model-based iterative reconstruction (MBIR) algorithms on semi-automatic measurements in subsolid nodules (SSNs).

Methods

Unenhanced CT scans of 73 SSNs obtained using the same protocol and reconstructed with both FBP and MBIR algorithms were evaluated by two radiologists. Diameter, mean attenuation, mass and volume of whole nodules and their solid components were measured. Intra- and interobserver variability and differences between FBP and MBIR were then evaluated using Bland–Altman method and Wilcoxon tests.

Results

Longest diameter, volume and mass of nodules and those of their solid components were significantly higher using MBIR (p < 0.05) with mean differences of 1.1% (limits of agreement, −6.4 to 8.5%), 3.2% (−20.9 to 27.3%) and 2.9% (−16.9 to 22.7%) and 3.2% (−20.5 to 27%), 6.3% (−51.9 to 64.6%), 6.6% (−50.1 to 63.3%), respectively. The limits of agreement between FBP and MBIR were within the range of intra- and interobserver variability for both algorithms with respect to the diameter, volume and mass of nodules and their solid components. There were no significant differences in intra- or interobserver variability between FBP and MBIR (p > 0.05).

Conclusion

Semi-automatic measurements of SSNs significantly differed between FBP and MBIR; however, the differences were within the range of measurement variability.

Key points

Intra- and interobserver reproducibility of measurements did not differ between FBP and MBIR.
Differences in SSNs’ semi-automatic measurement induced by reconstruction algorithms were not clinically significant.
Semi-automatic measurement may be conducted regardless of reconstruction algorithm.
SSNs’ semi-automated classification agreement (pure vs. part-solid) did not significantly differ between algorithms.
Literatur
1.
Zurück zum Zitat Naidich DP, Bankier AA, MacMahon H et al (2013) Recommendations for the management of subsolid pulmonary nodules detected at CT: a statement from the Fleischner Society. Radiology 266:304–317CrossRefPubMed Naidich DP, Bankier AA, MacMahon H et al (2013) Recommendations for the management of subsolid pulmonary nodules detected at CT: a statement from the Fleischner Society. Radiology 266:304–317CrossRefPubMed
2.
Zurück zum Zitat Lee HY, Choi YL, Lee KS et al (2014) Pure ground-glass opacity neoplastic lung nodules: histopathology, imaging, and management. AJR Am J Roentgenol 202:W224–233CrossRefPubMed Lee HY, Choi YL, Lee KS et al (2014) Pure ground-glass opacity neoplastic lung nodules: histopathology, imaging, and management. AJR Am J Roentgenol 202:W224–233CrossRefPubMed
3.
Zurück zum Zitat Hasegawa M, Sone S, Takashima S et al (2000) Growth rate of small lung cancers detected on mass CT screening. Br J Radiol 73:1252–1259CrossRefPubMed Hasegawa M, Sone S, Takashima S et al (2000) Growth rate of small lung cancers detected on mass CT screening. Br J Radiol 73:1252–1259CrossRefPubMed
4.
Zurück zum Zitat Kakinuma R, Muramatsu Y, Kusumoto M et al (2015) Solitary pure ground-glass nodules 5 mm or smaller: frequency of growth. Radiology 276:873–882CrossRefPubMed Kakinuma R, Muramatsu Y, Kusumoto M et al (2015) Solitary pure ground-glass nodules 5 mm or smaller: frequency of growth. Radiology 276:873–882CrossRefPubMed
5.
Zurück zum Zitat Song YS, Park CM, Park SJ, Lee SM, Jeon YK, Goo JM (2014) Volume and mass doubling times of persistent pulmonary subsolid nodules detected in patients without known malignancy. Radiology 273:276–284CrossRefPubMed Song YS, Park CM, Park SJ, Lee SM, Jeon YK, Goo JM (2014) Volume and mass doubling times of persistent pulmonary subsolid nodules detected in patients without known malignancy. Radiology 273:276–284CrossRefPubMed
6.
Zurück zum Zitat Chang B, Hwang JH, Choi YH et al (2013) Natural history of pure ground-glass opacity lung nodules detected by low-dose CT scan. Chest 143:172–178CrossRefPubMed Chang B, Hwang JH, Choi YH et al (2013) Natural history of pure ground-glass opacity lung nodules detected by low-dose CT scan. Chest 143:172–178CrossRefPubMed
7.
Zurück zum Zitat Matsuguma H, Mori K, Nakahara R et al (2013) Characteristics of subsolid pulmonary nodules showing growth during follow-up with CT scanning. Chest 143:436–443CrossRefPubMed Matsuguma H, Mori K, Nakahara R et al (2013) Characteristics of subsolid pulmonary nodules showing growth during follow-up with CT scanning. Chest 143:436–443CrossRefPubMed
8.
Zurück zum Zitat Kobayashi Y, Fukui T, Ito S et al (2013) How long should small lung lesions of ground-glass opacity be followed? J Thorac Oncol 8:309–314CrossRefPubMed Kobayashi Y, Fukui T, Ito S et al (2013) How long should small lung lesions of ground-glass opacity be followed? J Thorac Oncol 8:309–314CrossRefPubMed
9.
Zurück zum Zitat Lee SW, Leem CS, Kim TJ et al (2013) The long-term course of ground-glass opacities detected on thin-section computed tomography. Respir Med 107:904–910CrossRefPubMed Lee SW, Leem CS, Kim TJ et al (2013) The long-term course of ground-glass opacities detected on thin-section computed tomography. Respir Med 107:904–910CrossRefPubMed
10.
Zurück zum Zitat Brenner DJ (2004) Radiation risks potentially associated with low-dose CT screening of adult smokers for lung cancer. Radiology 231:440–445CrossRefPubMed Brenner DJ (2004) Radiation risks potentially associated with low-dose CT screening of adult smokers for lung cancer. Radiology 231:440–445CrossRefPubMed
11.
Zurück zum Zitat Vardhanabhuti V, Loader RJ, Mitchell GR, Riordan RD, Roobottom CA (2013) Image quality assessment of standard- and low-dose chest CT using filtered back projection, adaptive statistical iterative reconstruction, and novel model-based iterative reconstruction algorithms. AJR Am J Roentgenol 200:545–552CrossRefPubMed Vardhanabhuti V, Loader RJ, Mitchell GR, Riordan RD, Roobottom CA (2013) Image quality assessment of standard- and low-dose chest CT using filtered back projection, adaptive statistical iterative reconstruction, and novel model-based iterative reconstruction algorithms. AJR Am J Roentgenol 200:545–552CrossRefPubMed
12.
Zurück zum Zitat Padole A, Ali Khawaja RD, Kalra MK, Singh S (2015) CT radiation dose and iterative reconstruction techniques. AJR Am J Roentgenol 204:W384–392CrossRefPubMed Padole A, Ali Khawaja RD, Kalra MK, Singh S (2015) CT radiation dose and iterative reconstruction techniques. AJR Am J Roentgenol 204:W384–392CrossRefPubMed
13.
Zurück zum Zitat Mathieu KB, Ai H, Fox PS et al (2014) Radiation dose reduction for CT lung cancer screening using ASIR and MBIR: a phantom study. J Appl Clin Med Phys 15:4515CrossRefPubMed Mathieu KB, Ai H, Fox PS et al (2014) Radiation dose reduction for CT lung cancer screening using ASIR and MBIR: a phantom study. J Appl Clin Med Phys 15:4515CrossRefPubMed
14.
Zurück zum Zitat Vardhanabhuti V, Ilyas S, Gutteridge C, Freeman SJ, Roobottom CA (2013) Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi. Insights Imaging 4:661–669CrossRefPubMedPubMedCentral Vardhanabhuti V, Ilyas S, Gutteridge C, Freeman SJ, Roobottom CA (2013) Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi. Insights Imaging 4:661–669CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Kim H, Park CM, Kim SH et al (2014) Persistent pulmonary subsolid nodules: model-based iterative reconstruction for nodule classification and measurement variability on low-dose CT. Eur Radiol 24:2700–2708CrossRefPubMed Kim H, Park CM, Kim SH et al (2014) Persistent pulmonary subsolid nodules: model-based iterative reconstruction for nodule classification and measurement variability on low-dose CT. Eur Radiol 24:2700–2708CrossRefPubMed
16.
Zurück zum Zitat Wormanns D, Kohl G, Klotz E et al (2004) Volumetric measurements of pulmonary nodules at multi-row detector CT: in vivo reproducibility. Eur Radiol 14:86–92CrossRefPubMed Wormanns D, Kohl G, Klotz E et al (2004) Volumetric measurements of pulmonary nodules at multi-row detector CT: in vivo reproducibility. Eur Radiol 14:86–92CrossRefPubMed
17.
Zurück zum Zitat Scholten ET, de Jong PA, Jacobs C et al (2015) Interscan variation of semi-automated volumetry of subsolid pulmonary nodules. Eur Radiol 25:1040–1047CrossRefPubMed Scholten ET, de Jong PA, Jacobs C et al (2015) Interscan variation of semi-automated volumetry of subsolid pulmonary nodules. Eur Radiol 25:1040–1047CrossRefPubMed
19.
Zurück zum Zitat Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623CrossRefPubMed Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623CrossRefPubMed
20.
Zurück zum Zitat de Hoop B, Gietema H, van Ginneken B, Zanen P, Groenewegen G, Prokop M (2009) A comparison of six software packages for evaluation of solid lung nodules using semi-automated volumetry: what is the minimum increase in size to detect growth in repeated CT examinations. Eur Radiol 19:800–808CrossRefPubMed de Hoop B, Gietema H, van Ginneken B, Zanen P, Groenewegen G, Prokop M (2009) A comparison of six software packages for evaluation of solid lung nodules using semi-automated volumetry: what is the minimum increase in size to detect growth in repeated CT examinations. Eur Radiol 19:800–808CrossRefPubMed
21.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRefPubMed
22.
Zurück zum Zitat Chen B, Barnhart H, Richard S, Robins M, Colsher J, Samei E (2013) Volumetric quantification of lung nodules in CT with iterative reconstruction (ASiR and MBIR). Med Phys 40:111902CrossRefPubMed Chen B, Barnhart H, Richard S, Robins M, Colsher J, Samei E (2013) Volumetric quantification of lung nodules in CT with iterative reconstruction (ASiR and MBIR). Med Phys 40:111902CrossRefPubMed
23.
Zurück zum Zitat van Riel SJ, Sanchez CI, Bankier AA et al (2015) Observer variability for classification of pulmonary nodules on low-dose CT images and its effect on nodule management. Radiology 277:863–871CrossRefPubMed van Riel SJ, Sanchez CI, Bankier AA et al (2015) Observer variability for classification of pulmonary nodules on low-dose CT images and its effect on nodule management. Radiology 277:863–871CrossRefPubMed
24.
Zurück zum Zitat Penn A, Ma M, Chou BB, Tseng JR, Phan P (2015) Inter-reader variability when applying the 2013 Fleischner guidelines for potential solitary subsolid lung nodules. Acta Radiol 56:1180–1186CrossRefPubMed Penn A, Ma M, Chou BB, Tseng JR, Phan P (2015) Inter-reader variability when applying the 2013 Fleischner guidelines for potential solitary subsolid lung nodules. Acta Radiol 56:1180–1186CrossRefPubMed
Metadaten
Titel
Comparison of the effects of model-based iterative reconstruction and filtered back projection algorithms on software measurements in pulmonary subsolid nodules
verfasst von
Julien G. Cohen
Hyungjin Kim
Su Bin Park
Bram van Ginneken
Gilbert R. Ferretti
Chang Hyun Lee
Jin Mo Goo
Chang Min Park
Publikationsdatum
05.01.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2017
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4716-5

Weitere Artikel der Ausgabe 8/2017

European Radiology 8/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.