PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging

verfasst von: Tim Hilgenfeld, Marcel Prager, Alexander Heil, Franz Sebastian Schwindling, Mathias Nittka, David Grodzki, Peter Rammelsberg, Martin Bendszus, Sabine Heiland

Erschienen in: European Radiology | Ausgabe 12/2017

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Dental MRI is often impaired by artefacts due to metallic dental materials. Several sequences were developed to reduce susceptibility artefacts. Here, we evaluated a set of sequences for artefact reduction for dental MRI for the first time.

Methods

Artefact volume, signal-to-noise ratio (SNR) and image quality were assessed on a 3-T MRI for pointwise encoding time reduction with radial acquisition (PETRA), multiple-slab acquisition with view angle tilting gradient, based on a sampling perfection with application-optimised contrasts using different flip angle evolution (SPACE) sequence (MSVAT-SPACE), slice-encoding for metal-artefact correction (SEMAC) and compared to a standard SPACE and a standard turbo-spin-echo (TSE) sequence. Field-of-view and acquisition times were chosen to enable in vivo application. Two implant-supported prostheses were tested (porcelain fused to metal non-precious alloy and monolithic zirconia).

Results

Smallest artefact was measured for TSE sequences with no difference between the standard TSE and the SEMAC. MSVAT-SPACE reduced artefacts about 56% compared to the standard SPACE. Effect of the PETRA was dependent on sample used. Image quality and SNR were comparable for all sequences except PETRA, which yielded poor results.

Conclusion

There is no benefit in terms of artefact reduction for SEMAC compared to standard TSE. Usage of MSVAT-SPACE is advantageous since artefacts are reduced and higher resolution is achieved.

Key Points

SEMAC is not superior to TSE in terms of artefact reduction.
MSVAT-SPACE reduces susceptibility artefacts while maintaining comparable image quality.
PETRA reduces susceptibility artefacts depending on material but offers poor image quality
Literatur
1.
Zurück zum Zitat Vandenberghe B, Jacobs R, Bosmans H (2010) Modern dental imaging: a review of the current technology and clinical applications in dental practice. Eur Radiol 20:2637–2655CrossRefPubMed Vandenberghe B, Jacobs R, Bosmans H (2010) Modern dental imaging: a review of the current technology and clinical applications in dental practice. Eur Radiol 20:2637–2655CrossRefPubMed
2.
Zurück zum Zitat Patel S, Dawood A, Ford TP, Whaites E (2007) The potential applications of cone beam computed tomography in the management of endodontic problems. Int Endod J 40:818–830CrossRefPubMed Patel S, Dawood A, Ford TP, Whaites E (2007) The potential applications of cone beam computed tomography in the management of endodontic problems. Int Endod J 40:818–830CrossRefPubMed
3.
Zurück zum Zitat Gahleitner A, Watzek G, Imhof H (2003) Dental CT: imaging technique, anatomy, and pathologic conditions of the jaws. Eur Radiol 13:366–376PubMed Gahleitner A, Watzek G, Imhof H (2003) Dental CT: imaging technique, anatomy, and pathologic conditions of the jaws. Eur Radiol 13:366–376PubMed
5.
Zurück zum Zitat Monsour P, Dudhia R (2008) Implant radiography and radiology. Aust Dent J 53(Suppl 1):25 Monsour P, Dudhia R (2008) Implant radiography and radiology. Aust Dent J 53(Suppl 1):25
6.
7.
Zurück zum Zitat Liang X, Lambrichts I, Sun Y et al (2010) A comparative evaluation of cone beam computed tomography (CBCT) and multi-slice CT (MSCT). Part II: On 3D model accuracy. Eur J Radiol 75:270–274CrossRefPubMed Liang X, Lambrichts I, Sun Y et al (2010) A comparative evaluation of cone beam computed tomography (CBCT) and multi-slice CT (MSCT). Part II: On 3D model accuracy. Eur J Radiol 75:270–274CrossRefPubMed
8.
Zurück zum Zitat Eggers G, Rieker M, Kress B, Fiebach J, Dickhaus H, Hassfeld S (2005) Artefacts in magnetic resonance imaging caused by dental material. MAGMA 18:103–111CrossRefPubMed Eggers G, Rieker M, Kress B, Fiebach J, Dickhaus H, Hassfeld S (2005) Artefacts in magnetic resonance imaging caused by dental material. MAGMA 18:103–111CrossRefPubMed
9.
Zurück zum Zitat Ai T, Padua A, Goerner F et al (2012) SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Investig Radiol 47:267–276CrossRef Ai T, Padua A, Goerner F et al (2012) SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Investig Radiol 47:267–276CrossRef
10.
Zurück zum Zitat Lee YH, Lim D, Kim E, Kim S, Song HT, Suh JS (2013) Usefulness of slice encoding for metal artifact correction (SEMAC) for reducing metallic artifacts in 3-T MRI. Magn Reson Imaging 31:703–706CrossRefPubMed Lee YH, Lim D, Kim E, Kim S, Song HT, Suh JS (2013) Usefulness of slice encoding for metal artifact correction (SEMAC) for reducing metallic artifacts in 3-T MRI. Magn Reson Imaging 31:703–706CrossRefPubMed
11.
Zurück zum Zitat Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW (2012) Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology 265:204–214CrossRefPubMed Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW (2012) Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology 265:204–214CrossRefPubMed
12.
Zurück zum Zitat Cho ZH, Kim DJ, Kim YK (1988) Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys 15:7–11CrossRefPubMed Cho ZH, Kim DJ, Kim YK (1988) Total inhomogeneity correction including chemical shifts and susceptibility by view angle tilting. Med Phys 15:7–11CrossRefPubMed
13.
Zurück zum Zitat Zho S-Y, Kim M-O, Lee K-W, Kim D-H (2013) Artifact reduction from metallic dental materials in T1-weighted spin-echo imaging at 3.0 tesla. J Magn Reson Imaging 37:471–478CrossRefPubMed Zho S-Y, Kim M-O, Lee K-W, Kim D-H (2013) Artifact reduction from metallic dental materials in T1-weighted spin-echo imaging at 3.0 tesla. J Magn Reson Imaging 37:471–478CrossRefPubMed
14.
Zurück zum Zitat Penarrocha Diago M, Boronat Lopez A, Lamas Pelayo J (2006) Update in dental implant periapical surgery. Med Oral Patol Oral Cir Bucal 11:E429–E432PubMed Penarrocha Diago M, Boronat Lopez A, Lamas Pelayo J (2006) Update in dental implant periapical surgery. Med Oral Patol Oral Cir Bucal 11:E429–E432PubMed
15.
Zurück zum Zitat Li G, Nittka M, Paul D, Lauer L (2011) MSVAT-SPACE for fast metal implants imaging. Proc Intl Soc Mag Reson Med 19:3171 Li G, Nittka M, Paul D, Lauer L (2011) MSVAT-SPACE for fast metal implants imaging. Proc Intl Soc Mag Reson Med 19:3171
16.
Zurück zum Zitat Grodzki D, Jakob P, Heismann B (2012) Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA). Magn Reson Med 67:510–518CrossRefPubMed Grodzki D, Jakob P, Heismann B (2012) Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA). Magn Reson Med 67:510–518CrossRefPubMed
17.
Zurück zum Zitat Klinke T, Daboul A, Maron J et al (2012) Artifacts in magnetic resonance imaging and computed tomography caused by dental materials. PLoS One 7:e31766CrossRefPubMedPubMedCentral Klinke T, Daboul A, Maron J et al (2012) Artifacts in magnetic resonance imaging and computed tomography caused by dental materials. PLoS One 7:e31766CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Imai H, Tanaka Y, Nomura N et al (2013) Three-dimensional quantification of susceptibility artifacts from various metals in magnetic resonance images. Acta Biomater 9:8433–8439CrossRefPubMed Imai H, Tanaka Y, Nomura N et al (2013) Three-dimensional quantification of susceptibility artifacts from various metals in magnetic resonance images. Acta Biomater 9:8433–8439CrossRefPubMed
19.
Zurück zum Zitat Dietrich O, Raya JGG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385CrossRefPubMed Dietrich O, Raya JGG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385CrossRefPubMed
20.
Zurück zum Zitat Friedman L, Glover GH (2006) Report on a multicenter fMRI quality assurance protocol. J Magn Reson Imaging 23:827–839CrossRefPubMed Friedman L, Glover GH (2006) Report on a multicenter fMRI quality assurance protocol. J Magn Reson Imaging 23:827–839CrossRefPubMed
21.
Zurück zum Zitat Landis J, Koch G (1977) An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 33:363–374CrossRefPubMed Landis J, Koch G (1977) An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 33:363–374CrossRefPubMed
22.
Zurück zum Zitat Eley KA, Watt-Smith SR, Golding SJ (2013) "Black Bone" MRI: a potential non-ionizing method for three-dimensional cephalometric analysis–a preliminary feasibility study. Dentomaxillofac Radiol 42:20130236CrossRefPubMedPubMedCentral Eley KA, Watt-Smith SR, Golding SJ (2013) "Black Bone" MRI: a potential non-ionizing method for three-dimensional cephalometric analysis–a preliminary feasibility study. Dentomaxillofac Radiol 42:20130236CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Tymofiyeva O, Rottner K, Jakob PM, Richter EJJ, Proff P (2010) Three-dimensional localization of impacted teeth using magnetic resonance imaging. Clin Oral Investig 14:169–176CrossRefPubMed Tymofiyeva O, Rottner K, Jakob PM, Richter EJJ, Proff P (2010) Three-dimensional localization of impacted teeth using magnetic resonance imaging. Clin Oral Investig 14:169–176CrossRefPubMed
24.
Zurück zum Zitat Kress B, Buhl Y, Hähnel S, Eggers G, Sartor K, Schmitter M (2007) Age- and tooth-related pulp cavity signal intensity changes in healthy teeth: a comparative magnetic resonance imaging analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:134–137CrossRefPubMed Kress B, Buhl Y, Hähnel S, Eggers G, Sartor K, Schmitter M (2007) Age- and tooth-related pulp cavity signal intensity changes in healthy teeth: a comparative magnetic resonance imaging analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:134–137CrossRefPubMed
25.
Zurück zum Zitat Tymofiyeva O, Rottner K, Gareis D et al (2008) In vivo MRI-based dental impression using an intraoral RF receiver coil. Concepts Magn Reson 33B:244–251CrossRef Tymofiyeva O, Rottner K, Gareis D et al (2008) In vivo MRI-based dental impression using an intraoral RF receiver coil. Concepts Magn Reson 33B:244–251CrossRef
26.
Zurück zum Zitat Schara R, Sersa I, Skaleric U (2009) T1 relaxation time and magnetic resonance imaging of inflamed gingival tissue. Dentomaxillofac Radiol 38:216–223CrossRefPubMed Schara R, Sersa I, Skaleric U (2009) T1 relaxation time and magnetic resonance imaging of inflamed gingival tissue. Dentomaxillofac Radiol 38:216–223CrossRefPubMed
27.
Zurück zum Zitat Tartaglino LM, Flanders AE, Vinitski S, Friedman DP (1994) Metallic artifacts on MR images of the postoperative spine: reduction with fast spin-echo techniques. Radiology 190:565–569CrossRefPubMed Tartaglino LM, Flanders AE, Vinitski S, Friedman DP (1994) Metallic artifacts on MR images of the postoperative spine: reduction with fast spin-echo techniques. Radiology 190:565–569CrossRefPubMed
29.
Zurück zum Zitat Viano A, Gronemeyer S, Haliloglu M, Hoffer F (2000) Improved MR imaging for patients with metallic implants. Magn Reson Imaging 18:287–295CrossRefPubMed Viano A, Gronemeyer S, Haliloglu M, Hoffer F (2000) Improved MR imaging for patients with metallic implants. Magn Reson Imaging 18:287–295CrossRefPubMed
30.
Zurück zum Zitat Cox RJ, Kau CH, Rasche V (2012) Three-dimensional ultrashort echo magnetic resonance imaging of orthodontic appliances in the natural dentition. Am J Orthod Dentofacial Orthop 142:552–561CrossRefPubMed Cox RJ, Kau CH, Rasche V (2012) Three-dimensional ultrashort echo magnetic resonance imaging of orthodontic appliances in the natural dentition. Am J Orthod Dentofacial Orthop 142:552–561CrossRefPubMed
31.
Zurück zum Zitat Carl M, Koch K, Du J (2012) MR imaging near metal with undersampled 3D radial UTE-MAVRIC sequences. Magn Reson Med 69:27–36CrossRefPubMed Carl M, Koch K, Du J (2012) MR imaging near metal with undersampled 3D radial UTE-MAVRIC sequences. Magn Reson Med 69:27–36CrossRefPubMed
Metadaten
Titel
PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging
verfasst von
Tim Hilgenfeld
Marcel Prager
Alexander Heil
Franz Sebastian Schwindling
Mathias Nittka
David Grodzki
Peter Rammelsberg
Martin Bendszus
Sabine Heiland
Publikationsdatum
11.07.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2017
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4901-1

Weitere Artikel der Ausgabe 12/2017

European Radiology 12/2017 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.