Magnetic resonance imaging of the time course of hyperpolarized 129Xe gas exchange in the human lungs and heart

verfasst von: Ozkan Doganay, Mitchell Chen, Tahreema Matin, Marzia Rigolli, Julie-Ann Phillips, Anthony McIntyre, Fergus V. Gleeson

Erschienen in: European Radiology | Ausgabe 5/2019

Abstract

Purpose

To perform magnetic resonance imaging (MRI), human lung imaging, and quantification of the gas-transfer dynamics of hyperpolarized xenon-129 (HPX) from the alveoli into the blood plasma.

Materials and methods

HPX MRI with iterative decomposition of water and fat with echo asymmetry and least-square estimation (IDEAL) approach were used with multi-interleaved spiral k-space sampling to obtain HPX gas and dissolved phase images. IDEAL time-series images were then obtained from ten subjects including six normal subjects and four patients with pulmonary emphysema to test the feasibility of the proposed technique for capturing xenon-129 gas-transfer dynamics (XGTD). The dynamics of xenon gas diffusion over the entire lung was also investigated by measuring the signal intensity variations between three regions of interest, including the left and right lungs and the heart using Welch’s t test.

Results

The technique enabled the acquisition of HPX gas and dissolved phase compartment images in a single breath-hold interval of 8 s. The y-intersect of the XGTD curves were also found to be statistically lower in the patients with lung emphysema than in the healthy group (p < 0.05).

Conclusion

This time-series IDEAL technique enables the visualization and quantification of inhaled xenon from the alveoli to the left ventricle with a clinical gradient strength magnet during a single breath-hold, in healthy and diseased lungs.

Key Points

The proposed hyperpolarized xenon-129 gas and dissolved magnetic resonance imaging technique can provide regional and temporal measurements of xenon-129 gas-transfer dynamics.
• Quantitative measurement of xenon-129 gas-transfer dynamics from the alveolar to the heart was demonstrated in normal subjects and pulmonary emphysema.
• Comparison of gas-transfer dynamics in normal subjects and pulmonary emphysema showed that the proposed technique appears sensitive to changes affecting the alveoli, pulmonary interstitium, and capillaries.
Begleitmaterial
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00330-018-5853-9) contains supplementary material, which is available to authorized users.
Abkürzungen
Gas-transfer time points
initial
Initial gas-transfer time point for spoiling the initial dissolved phase signal
late
Late gas-transfer time points
short
Short gas-transfer time points
129Xe
Xenon-129
COPD
Chronic obstructive pulmonary disease
HP
Hyperpolarized
HPX
Hyperpolarized xenon-129
IDEAL
Iterative decomposition of water and fat with echo asymmetry and least-square estimation
MR
Magnetic resonance
MRI
Magnetic resonance imaging
PTPB
Pulmonary tissue and plasma blood
RBC
Red blood cell
ROI
Region of interest
S 1
The linear increase of XGTD curve
SGas
Signal gas
SNR
Signal-to-noise ratio
S o
The y-intersect of XGTD curve
SPTBP
Signal-PTBP compartment
SRBC
Signal-RBC compartment
T 1
The longitudinal decay time
T 2 *
Transverse signal decay time
TE-I, TE-II, and TE-III
The first, second, and third echo images
TE1, TE2, TE3
The first, second, and third echo times
TR
Pulse reputation times
TRmin
The minimum TR
XGTD
Xenon-129 gas-transfer dynamics
XTC
Xenon polarization transfer constant
αGas
Flip angle on gas phase
τ1
The gas-transfer time constant

Introduction

Hyperpolarized xenon-129 (HPX) magnetic resonance imaging (MRI) may be used to evaluate pulmonary ventilation, in patients with respiratory diseases such as chronic obstructive pulmonary disease (COPD) and asthma [18]. Due to the solubility of xenon, HPX MRI may also provide additional functional information about the gas exchange efficiency in the alveolar epithelium, the interstitium, the pulmonary capillary endothelium, and the blood plasma of the lungs [9]. At clinical MR magnetic field strengths (1.5–3 T), the dissolved xenon gas splits into two chemical shifts corresponding to the pulmonary tissue and blood plasma (PTBP) and red blood cell (RBC) compartments of the lung. The vast majority of the previous HP dissolved xenon studies focused on the measurement of diffusion uptake (i.e., gas-transfer dynamics) from xenon in the alveoli into the PTBP and RBC compartments of the entire thoracic cavity using MR spectroscopy techniques [1012]. These techniques allow measurement of lung function when used in corroboration with numerical modeling [1317]. Recent HPX MRI techniques have enabled the spatial distribution of regional diseases, addressing the challenges arising from low signal-to-noise ratios (SNR) and short transverse signal decay time (T2* ≈ 2 ms) of PTBP and RBC peaks [1823]. Concurrent imaging of the dissolved phases of xenon and the measurement of gas-transfer dynamics would enable quantitative information about the efficiency or otherwise of gas exchange pulmonary function.
An early dissolved phase imaging technique, the xenon polarization transfer constant (XTC) has been shown to be sensitive to measurement of gas-transfer dynamics combining with the spatial information [24, 25]. In contrast to MR spectroscopy, XTC enabled the indirect measurement of the dissolved phase signal from the estimation of depolarization of the reservoir gas-phase magnetization with respect to various delay times [26, 27]. Unlike other dissolved phase imaging techniques, XTC did not provide imaging of PTBP and RBC compartments of the lungs separately.
MR imaging with iterative decomposition of water and fat with echo asymmetry and least-square estimation (IDEAL) Cartesian k-space sampling has been reported to be an accurate technique for imaging the chemical shift of hydrogen protons in muscle and fat compartments [28]. The IDEAL technique has been used in the metabolic imaging of hyperpolarized (HP) 13C using a single-shot spiral k-space sampling, taking advantage of the relatively longer T2* (T2* ≈ 30 ms) [29]. This IDEAL with spiral k-space sampling technique was subsequently applied to HPX MR for imaging gas, PTBP, and RBC compartments in rat lungs [30]. The short T2* of 129Xe in the PTBP and RBC compartments was compensated by using an insertable high-performance gradient system (maximum gradient amplitude, 300 mT/m; slew rate, 2500 T/m/s) [31, 32]. While this strategy provided quantitative measurement of the physiological parameters and dynamics of HPX gas diffusion from alveoli air sacs to the heart in a rat model [31, 32], it was difficult to obtain sub-centimeter pixel resolution in humans using this single-shot spiral sampling technique due to the limited gradient performance (maximum gradient amplitude, 50 mT/m; slew rate, 200 T/m/s) of clinical MR scanners.
In this study, HPX MRI single-shot k-space sampling IDEAL technique has now been extended to human lung imaging using multi-interleave spiral k-space sampling in a standard clinical gradient MR scanner. By taking advantage of rapid k-space sampling, the feasibility of capturing the time-series IDEAL gas, PTBP, and RBC compartment images in a single breath-hold across eight different time points has been investigated in six healthy volunteers and four patients with pulmonary emphysema.

Materials and methods

Six healthy participants (aged 25–45 years) with no known cardiopulmonary disease, no evidence of emphysema or other pulmonary disease on CT, and with normal lung function were scanned using a 1.5 T MR scanner (Signa HDx Twin Speed, GE Healthcare) and a proton-blocked vest-shaped RF coil (Clinical MR solutions). Enriched xenon gas (87% 129Xe) was polarized by a commercially available polarizer (Model 9300, Polarean with 10–15% polarization). Subjects were instructed to inhale 1.0 L of pure xenon gas following a protocol approved by the National Research Ethics Service Committee.
Conventional proton MR 3-plane localizer images were acquired using the standard fast gradient echo sequence and double inversion-recovery (DIR) black-blood imaging sequence to acquire a diagnostic display of the lung, cardiac, and mediastinal anatomy as previously explained [6]. Following the conventional proton scan, IDEAL HPX gas, PTBP, and RBC compartment images were obtained using a multi-interleave spiral k-space with a three-point IDEAL approach. Briefly, three-point IDEAL approach acquisition produced three different echo images (TE-I, TE-II, and TE-III) with three echo times (TE1 = 0 μs, TE2 = 160 μs, and TE3 = 320 μs) which were calculated using the number of signal averaging approach [30]. Each of the echo images were acquired using 12 RF pulses and spiral interleaves for a particular echo time as shown in Fig. 1a using a standard clinical MR gradient. Imaging parameters were field of view. 34 cm; number of pixels, 32; bandwidth, 62.5 kHz; and readout time duration per interleaf, 4 ms. The scan time required for a single echo image was 12 × TRmin = 240 ms by setting TRmin to the minimum value of 20 ms as shown in Fig. 1a. The TRmin consisted of the time required for the slice selection, RF pulse (a hard pulse), the echo times, readout time (~ 4 ms), rewinders, and spoiler gradients per interleaves.
To obtain time-series IDEAL images, the imaging pulse sequence in Fig. 1a was repeated with additional delay times as shown in Fig. 1b. This allowed the acquisition of time-series IDEAL images with various gas-transfer time points, ∆. To investigate the optimum selection of ∆, time-series IDEAL images were acquired from a healthy subject including late ∆ points late: initial, 20, 25, 50, 100, 500, 1000, and 2000 ms leading to a long scan time of ~ 17 s. ∆initial was the spoiling acquisition used to destroy initial buildup of dissolved phase signal which accumulated during the time between the inhalation of HP 129Xe gas and the scan starting. To investigate the feasibility of obtaining XGTD, the time-series IDEAL images of four healthy subjects were collected with the following gas-transfer time points: short: initial, 300, 200, 100, 50, 30, 25, and 20 ms leading to a shorter total scan time of ~ 8 s.
IDEAL images were reconstructed onto a 128 × 128 matrix using a similar algorithm to those described in Wiesinger et al [33] and Schulte et al [29]. Image intensity variations in the PTBP and RBC compartments were analyzed selecting three ROIs, the left and right lungs with a matrix size of 80 × 40 and the heart with a matrix size of 30 × 30 (as shown in Supplementary Fig. 1). ROI-left with matrix size 80 × 40 included signal from the left lung and heart. Similarly, ROI-heart with matrix size 30 × 30 included signal from the left lung and heart. The dissolved phase 129Xe heart signal in ROI-left and the dissolved phase 129Xe lung signal in ROI-heart were separated by solving two variable equations as explained in appendix.
The normalized dissolved phase signals from ROI-left, ROI-right, and ROI-heart in the time-series PTBP and RBC images were fitted to the simple exponential function as described by Mansson et al [13, 21] as explained in Supplementary Fig. 1. Collectively, the normalized data points and the exponential fit were described as XGTD curves. The gas-transfer time constant, τ1, which corresponded to the rapid exponential increase, S1, the linear increase, and So, the y- intersect of XGTD curve were measured in all subjects as reported in Table 1. Additionally, the gas compartment time-series images were analyzed by fitting the signal loss equation to the gas signal at right and left ROIs (as explained in Supplementary Fig. 1) as demonstrated in the appendix.
Table 1
Summary of patient with lung emphysema FEVi -> FEV1
Subject number
Sex (male/female)
Age (years)
CT emphysema (% left/right)
FEV1(% predicted)
FEV1/FVC (% predicted)
1
Male
89
6/20
78
73
2
Female
74
6/7
76
85
3
Male
69
8/5
43
48
4
Female
74
2/3
92
81
FEV1 forced expiratory volume in 1 s, FVC forced vital capacity
The Welch’s t test in GraphPad Prism (GraphPad Prism version 7.00, GraphPad Software) was used to investigate any potential difference in the fitting parameters (i.e,. S1, So, and τ1) between ROIs within the subjects with a 95% confidence interval. For proof of concept in clinical practice, fours patients currently involved in another HPX study with CT proven pulmonary emphysema and abnormal lung function were then scanned using the same technique. The CT diagnosis of emphysema was made using CT density maps, with a threshold of – 940 Hounsfield units. After optimization of gas-transfer time points from the images were obtained from a healthy subject as explained previously, XGTD curves were compared between the normal (n = 5) and pulmonary emphysema (n = 4) subjects as shown in Table 1.

Results

The averages of all time points over the late in time-series IDEAL gas compartment images including ROI-Left and ROI-Right are shown in Fig. 2a. The corresponding gas signal from each time point in the time-series IDEAL gas compartment images and the corresponding fit from Eq. (3) for the ROI-left and ROI-right are shown in Fig. 2d. As expected, the Sgas decreases continuously during the acquisition of time-series IDEAL images as a function of T1 = 35 ± 10 s and number of RF pulses (36 per IDEAL image) with αGas = 3.6 ± 0.4°. During the five time points between initial and ∆ = 0.1 s which correspond to the first 8 s of the entire scan time (17 s), the reservoir gas signal drops monotonically. However, for the later three time points between ∆ = 0.5 s and ∆ = 2 s, corresponding to the scan time interval between 8 s and 17 s, the reservoir gas signal drops more rapidly due to T1. As the scan time increases, the effect of T1 becomes more noticeable in the reservoir gas signal as also shown in Supplementary Fig. 2.
The average of PTBP and RBC compartment images of all time points over the late in time-series IDEAL is shown in Fig. 2b, c. The time-series PTBP and RBC compartment imaging of the lungs are also shown for each time points in Supplementary Fig. 2. A strong signal is visible from ROI-heart (Fig. 2b, c) resulting in a different image texture compared to the gas compartment image (Fig. 2a). The temporal PTBP and RBC signal from each time point over late from ROI-left, ROI-right, and ROI-heart is shown in Fig. 2e, f, respectively. The SPTBP(ROI-left) and SPTBP(ROI-right) in Fig. 2e and SRBC(ROI-left) and SRBC(ROI-right) in Fig. 2f follow a different trend than Sgas in Fig. 2d corresponding to the replenishment of dissolved phase signal as a function of ∆. In particular, the SPTBP(ROI-heart) and SRBC(ROI-heart) increase significantly at time points of ∆ = 0.5 s:2 s but remain low at the early time point of ∆ = 20 ms:100 ms in Fig. 2e. f. For example, SPTBP(ROI-heart) = 0.2 × 104 is only 5% of the SPTBP(ROI-right) = 4.4 × 104 at ∆ = 25 ms in Fig. 2e. However, Fig. 2e demonstrates SPTBP(ROI-heart) = 1 × 104 is 42% of the SPTBP(ROI-right) = 2.4 × 104 at ∆ = 2 s showing that the arrival of xenon from the alveoli to the heart is longer than the arrival of xenon to the pulmonary tissue and corresponding pulmonary capillary network. This difference in the arrival time of xenon in ROI-heart and ROI-right Fig. 2e, f also confirms that the separation of heart signal was effective (as explained in Appendix: Separation of Heart Signal). After correction of the heart signal, ROI-left and ROI-right were identical in the healthy subject as expected. The delayed arrival of xenon to the heart (ROI-heart) is also shown in Supplementary Fig. 2 in axial and coronal views confirming the delayed arrival of dissolved phase HPX to the heart for the late gas-transfer time points (> 200 ms).
The time-series IDEAL images from another healthy subject are shown in Fig. 3a in which the effect of T1 signal loss was minimized by capturing more short gas-transfer time points (short: initial, 300, 200, 100, 50, 30, 25, and 20 ms) resulting in a shorter scan time of ~ 8 s. The signal is strongest in the center of the PTBP and RBC compartment images at the initial acquisition, ∆initial, in Fig. 3a as shown with arrows. Three coronal slices of 1H MRI of the thorax from the same subjects are also shown in Fig. 3b–d, including the 3D render of six coronal slices in Fig. 3e. The left ventricle of the heart and the corresponding arterial vasculature vessels are shown with arrows in Fig. 3b and e, respectively. In comparison to the 1H MR images, the strong dissolved phase signal in the center (Fig. 3a) at ∆initial is likely to have originated from the dissolved phase 129Xe signal in the left ventricle, where the dissolved 129Xe accumulated during the time between inhalation of the 129Xe gas and the scan starting.
The mean of two XGTD curves from the same healthy subject including the standard deviation are shown in Fig. 4a, b demonstrating the separation of PTBP and RBC compartments. The average standard deviation of the repeated measurements was less than 5.2% suggesting that the repeated measurements are reasonably consistent. There was no statistical difference between the ROI-left and ROI-right (p value > 0.175 and p value > 0.128) of five healthy participants. The τ1 from ROI-heart (τ1 = 31-37 ms) demonstrated a statistical difference compared to ROI-left (τ1 = 9.3–13 ms) and ROI-right (τ1 = 9.4–13 ms) (p value < 0.001) and approximately three times greater than the lung signal due to the delayed arrival of dissolved phase xenon in the heart. XGTD curves from a patient with pulmonary emphysema are also shown in Fig. 4c, d. The amplitude of XGTD curves from the emphysema subjects was substantially smaller than the healthy subjects. The actual time-series IDEAL images of the healthy and the patient with emphysema are also shown in Supplementary Fig. 3.
The XGTD curve fitting parameters were measured with a very high fitting accuracy (R-squared > 0.98) and summarized in Table 2 for the normal and abnormal lungs. For a statistical comparison, So in Table 2 was compared between normal and abnormal lungs as shown in Fig. 5a–b (including the p values). So values of the patients with emphysema were smaller and statistically different to the normal lungs. The mean of So in the normal lungs was nearly two times greater than the emphysema group in Table 2. Interestingly, the ratio of τ1 between PTBP and RBC compartments was 0.97 ± 0.001 over the healthy subjects suggesting the gas transfer was repeatable in healthy subjects. However, the pulmonary emphysema patients had a greater standard deviation (1.07 ± 0.07) showing a larger variation due emphysema as shown in Fig. 5c.
Table 2
Summary of the XGTD curve fitting parameters in each participant
Subjects
Scans
ROI-Right
ROI-Left
ROI-Heart
S1 (×106)
So (×103)
τ1 (ms)
S1 (×106)
So (×103)
τ1 (ms)
τ1 (ms)
Normal Lungs
1
PTBP
3.28
2.50
10.0
4.18
2.74
12.0
35.9
RBC
3.11
2.34
10.5
3.18
2.33
12.6
37.8
2
PTBP
4.34
2.84
10.7
3.56
2.55
11.3
34.6
RBC
4.30
3.05
11.1
3.29
2.42
10.9
34.2
3
PTBP
3.15
2.26
9.4
3.39
2.48
10.6
35.8
RBC
3.10
2.23
10.6
3.25
2.17
10.5
35.3
4
PTBP
6.14
4.44
9.7
6.89
4.33
11.8
37.5
RBC
5.95
4.28
9.6
5.80
3.97
12.9
37.0
5
PTBP
4.03
2.74
8.7
3.26
2.33
10.0
34.3
RBC
3.36
2.51
8.5
3.12
2.17
10.6
33.7
Pulmonary Emphysema
1
PTBP
2.61
1.77
13.0
2.55
1.75
11.8
31.7
RBC
2.41
1.58
9.6
2.12
1.57
9.3
31.5
2
PTBP
2.76
1.85
11.5
2.63
1.80
12.0
31.2
RBC
2.63
1.78
13.0
2.61
1.77
13.0
31.6
3
PTBP
2.71
1.86
10.8
2.93
1.90
12.3
32.9
RBC
2.70
1.87
12.5
2.73
1.86
11.6
33.3
4
PTBP
2.45
1.73
12.2
2.80
1.90
8.55
34.0
RBC
1.96
1.43
9.4
2.72
1.84
9.17
33.1
PTBP is pulmonary tissue and blood plasma, and RBC is the red blood cell compartment of lungs. S1, So, and τ1 are the the fitting parameters. ROI-Right, ROI-Left, and ROI-Heart are the region of interests.
Particularly, for the subjects in the Supplementary Fig. 3, the ratio of XGTD curve fitting parameters was compared between four ROIs as shown in Sup Fig. 3 a–b, showing the feasibility of regional analysis. The ventilation and corresponding gas compartment signal were comparable between the healthy and pulmonary diseased subjects from the upper and lower right lungs (ratio of Mi = 0.88 and 1.02 in Table 3). The ratio of the XGTD curve fitting parameters (i.e., So and S1 in Table 3) was a factor of 5.69–2.05 higher in the healthy subject compared to pulmonary emphysema subject.
Table 3
Ratio of fitting parameters from four ROIs between a healthy subject and pulmonary emphysema in Supplementary Figure 3
ROIs
Gas
PTBP
RBC
Mi
S1
So
S1
So
Right upper
0.88
5.07
2.70
5.69
3.05
Right lower
1.02
3.11
4.44
3.35
2.05
Left upper
2.69
5.48
5.02
5.93
3.89
Left lower
2.04
5.09
5.36
5.84
2.55
ROIs are shown in Supplementary Fig. 3. Mi is a scaling factor that is dependent on the degree of polarization, concentration of HPX gas. PTBP is pulmonary tissue and blood plasma, and RBC is the red blood cell compartment of lungs. S1 is the linear increase and So is the y- intersect of XGTD curve

Discussion

The multi-interleaved spiral k-space sampling IDEAL technique enabled time-series IDEAL imaging of HP 129Xe in the gas, PTBP, and RBC compartments in human lungs using a standard clinical MR scanner. As a proof of concept study, the feasibility of measuring the XGTD curves has been successfully demonstrated in both participant groups with normal and abnormal lungs. The proposed technique in this study showed a statistical difference between the subjects with abnormal and normal lungs. To our knowledge, this is the first in-man study showing the transfer of 129Xe gas in the alveoli into PTBP and RBC compartments of the lungs, and into the pulmonary vasculature and the left ventricle of the heart. Additionally, the amount of gas transfer from lung alveolar sacs to the PTBP and RBC compartments and thereby XGTD curves were impaired significantly in emphysema subjects compared to the healthy cohort.
The early XTC imaging technique is conceptually most relevant to the time-series IDEAL imaging in this study. As shown in the time-series IDEAL images the dissolved phase (PTBP and RBC), image texture is different to the gas ventilation distribution of HPX gas. Direct imaging of the dissolved phase compartments allows the capture of the diffusion of dissolved phase 129Xe from the lungs to the heart. Therefore, the dissolved phase HPX can be directly and quantitatively measured from the actual PTBP and RBC compartment images by selecting ROIs in the heart and lungs separately. In this respect, the time-series IDEAL imaging technique we have developed may provide insights into the measurement of XGTD similar to those of whole lung MR spectroscopy, with the added advantage of regional measurement of gas-transfer dynamics.
The measured τ1 reported in this study was comparable to those of measured with XTC imaging technique (~ 11 ms) [3436] and MR spectroscopy (26 ± 17 ms) [15, 16]. The τ1 corresponded to ~ 50% of the dissolved phase signal compared to the dissolved phase xenon signal at ∆ = 300 ms in Fig. 4a suggesting that most of the diffusion from alveoli to the pulmonary tissue and capillaries occurs in the order of milliseconds. Additionally, the time-series IDEAL technique in this study allowed measurement of τ1 from the heart. Since the dissolved phase xenon at ROI-heart followed a longer pathway, the τ1 has been found to be approximately three times longer than that of xenon arriving to the ROI-left and ROI-right. Furthermore, the ratio of τ1 between PTBP and RBC compartments showed a lower variance in the heathy group than in patients with pulmonary emphysema suggesting that the emphysema affects not only the gas ventilation in the lung air sacs but also the transfer of xenon between the PTBP and RBC compartments.
The So corresponding to the y-intersect of the XGTC curves and dissolved phase signal amplitude was lower in the patients with lung emphysema than in the subjects with normal lungs. This is in agreement with the findings reported by Qing et al in a cohort of COPD patients [22]. Accordingly, the decrease in So may be attributed to changes caused by emphysema in the abnormal lung group. Physiologically, the decreased So would be attributed to the decreased gas transfer surface area between alveolar airs sacs to pulmonary tissue and capillary network of the lungs as a result of emphysema. The decreased gas transfer surface area is likely to be due to the loss of alveolar walls and the destruction of capillaries which may not significantly affect the gas compartment signal but can result in significant changes in XGTD curves. Therefore, in addition to the gas ventilation, the pattern of change in XGTD curves (i.e., So, S1, and τ1) would allow for additional functional information to be derived and would be dependent on the type of lung disease present. This was demonstrated in the increase of dissolved phase signal associated with the increased interstitial thickening in patients with idiopathic pulmonary fibrosis [18, 37]. Furthermore, in the radiation-induced lung injury model, the XGTD curves from the PTBP and RBC compartments and apparent diffusion coefficient from HPX gas were shown to change independently due to the radiation pneumonitis and inflammation [11, 31, 3841]. In the future, this technique (i.e., measured So, S1, and τ1) may potentially be used to investigate the severity of pulmonary disease, including the quantitative measurement of lung physiological parameters (i.e., tissue thickness, capillary diffusion length, perfusion, and mean transit time) when incorporated into a suitable gas exchange numerical model.
One major limitation of our study is that the XGTD curves were not fitted to the numerical gas exchange models, although the technique was developed for quantitative measurement of lung physiology by fitting to the numerical models. The existing gas exchange models [1315, 17, 42] cannot be directly applied to the time-series IDEAL images with multi-interleave spiral k-space acquisition for calculating the lung gas exchange parameters without a theoretical justification. As the existing models were suitable for fitting the MR spectroscopy of the whole lungs where the spatial variation in the dissolved phase images was not considered, single-shot spiral IDEAL data as the replenishment of dissolved phase magnetization due to the TRmin is negligible. Incorporating the fitting parameters (So, S1, and τ1) reported in Table 2 in a suitable numerical gas exchange model, the lung gas exchange parameters (i.e., tissue thickness, capillary diffusion length, perfusion, and mean transit time) and variation between the same subject scans and the mean of healthy subject scan can be investigated quantitatively. Another limitation was the relatively low polarization (10–13%) limiting the voxel by voxel analysis of the XGTD particularly for the diseased subjects with short gas-transfer times (∆ = 20–50 ms) when the replenishment of dissolved phase magnetization is low. The signal detected should be significantly improved using higher polarization levels and better spatial and frequency selective RF pulses in the future [43].
In conclusion, the multi-interleave spiral k-space sampling IDEAL approach enabled imaging HPX in gas, PTBP, and RBC compartments in human lungs using a standard clinical strength gradient MR scanner within an easily tolerated breath-hold of 8 s. Furthermore, the feasibility of measuring the gas-transfer curves has been successfully demonstrated in five subjects with normal lungs and four subjects with pulmonary emphysema. The proposed XGTD technique suggests that time-series IDEAL imaging can capture the transfer of 129Xe gas in the alveoli into the PTBP and RBC compartments of the lungs and into the pulmonary vasculature and to the left ventricle of the heart. This initial small participant study has also shown that this technique may potentially differentiate between patients with normal and abnormal lungs.

Acknowledgments

The authors are thankful to Claire Trumper, Jennifer Lee, and Kenneth Jacob for assistance with the gas production and delivery. We also acknowledge Nigar Syed, Alexandra Eves, and other MRI radiographers from the Churchill Hospital for their kind help with MR scans.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is Professor Fergus Gleeson (fgleeson@icloud.com).

Conflict of interest

The authors declare that they have no conflict of interest.

Statistics and biometry

One of the authors has significant statistical expertise.
No complex statistical methods were necessary for this paper.
Written informed consent was obtained from all subjects (patients) in this study.

Ethical approval

Institutional Review Board approval was obtained.

Methodology

• Prospective
• Case-control study
• Performed at one institution
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Literatur
1.
Zurück zum Zitat Kirby M, Svenningsen S, Owrangi A et al (2012) Hyperpolarized He-3 and Xe-129 MR imaging in healthy volunteers and patients with chronic obstructive pulmonary disease. Radiology 265:600–610CrossRefPubMed Kirby M, Svenningsen S, Owrangi A et al (2012) Hyperpolarized He-3 and Xe-129 MR imaging in healthy volunteers and patients with chronic obstructive pulmonary disease. Radiology 265:600–610CrossRefPubMed
2.
Zurück zum Zitat Driehuys B, Martinez-Jimenez S, Cleveland ZI et al (2012) Chronic obstructive pulmonary disease: safety and tolerability of hyperpolarized 129Xe MR imaging in healthy volunteers and patients. Radiology 262:279–289CrossRefPubMedPubMedCentral Driehuys B, Martinez-Jimenez S, Cleveland ZI et al (2012) Chronic obstructive pulmonary disease: safety and tolerability of hyperpolarized 129Xe MR imaging in healthy volunteers and patients. Radiology 262:279–289CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Shukla Y, Wheatley A, Kirby M et al (2012) Hyperpolarized 129Xe magnetic resonance imaging: tolerability in healthy volunteers and subjects with pulmonary disease. Acad Radiol 19:941–951CrossRefPubMed Shukla Y, Wheatley A, Kirby M et al (2012) Hyperpolarized 129Xe magnetic resonance imaging: tolerability in healthy volunteers and subjects with pulmonary disease. Acad Radiol 19:941–951CrossRefPubMed
4.
Zurück zum Zitat Svenningsen S, Guo F, Kirby M et al (2014) Pulmonary functional magnetic resonance imaging: asthma temporal-spatial maps. Acad Radiol 21:1402–1410CrossRefPubMed Svenningsen S, Guo F, Kirby M et al (2014) Pulmonary functional magnetic resonance imaging: asthma temporal-spatial maps. Acad Radiol 21:1402–1410CrossRefPubMed
5.
Zurück zum Zitat He M, Driehuys B, Que LG, Huang YT (2016) Using hyperpolarized 129Xe MRI to quantify the pulmonary ventilation distribution. Acad Radiol 23:1521–1531CrossRefPubMedPubMedCentral He M, Driehuys B, Que LG, Huang YT (2016) Using hyperpolarized 129Xe MRI to quantify the pulmonary ventilation distribution. Acad Radiol 23:1521–1531CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Matin TN, Rahman N, Nickol AH et al (2017) Chronic obstructive pulmonary disease: lobar analysis with hyperpolarized 129Xe MR imaging. Radiology 282:857–868CrossRefPubMed Matin TN, Rahman N, Nickol AH et al (2017) Chronic obstructive pulmonary disease: lobar analysis with hyperpolarized 129Xe MR imaging. Radiology 282:857–868CrossRefPubMed
7.
Zurück zum Zitat Ebner L, Kammerman J, Driehuys B, Schiebler ML, Cadman RV, Fain SB (2017) The role of hyperpolarized 129xenon in MR imaging of pulmonary function. Eur J Radiol 86:343–352CrossRefPubMed Ebner L, Kammerman J, Driehuys B, Schiebler ML, Cadman RV, Fain SB (2017) The role of hyperpolarized 129xenon in MR imaging of pulmonary function. Eur J Radiol 86:343–352CrossRefPubMed
8.
Zurück zum Zitat Doganay O, Matin TN, Mcintyre A et al (2018) Fast dynamic ventilation MRI of hyperpolarized (129) Xe using spiral imaging. Magn Reson Med 79:2597–2606 Doganay O, Matin TN, Mcintyre A et al (2018) Fast dynamic ventilation MRI of hyperpolarized (129) Xe using spiral imaging. Magn Reson Med 79:2597–2606
9.
Zurück zum Zitat Mugler JP 3rd, Driehuys B, Brookeman JR et al (1997) MR imaging and spectroscopy using hyperpolarized 129Xe gas: preliminary human results. Magn Reson Med 37:809–815CrossRefPubMed Mugler JP 3rd, Driehuys B, Brookeman JR et al (1997) MR imaging and spectroscopy using hyperpolarized 129Xe gas: preliminary human results. Magn Reson Med 37:809–815CrossRefPubMed
10.
Zurück zum Zitat Stewart NJ, Leung G, Norquay G et al (2014) Experimental validation of the hyperpolarized (129) Xe chemical shift saturation recovery technique in healthy volunteers and subjects with interstitial lung disease. Magn Reson Med. https://doi.org/10.1002/mrm.25400 Stewart NJ, Leung G, Norquay G et al (2014) Experimental validation of the hyperpolarized (129) Xe chemical shift saturation recovery technique in healthy volunteers and subjects with interstitial lung disease. Magn Reson Med. https://​doi.​org/​10.​1002/​mrm.​25400
11.
Zurück zum Zitat Fox MS, Ouriadov A, Thind K et al (2014) Detection of radiation induced lung injury in rats using dynamic hyperpolarized (129)Xe magnetic resonance spectroscopy. Med Phys 41 Fox MS, Ouriadov A, Thind K et al (2014) Detection of radiation induced lung injury in rats using dynamic hyperpolarized (129)Xe magnetic resonance spectroscopy. Med Phys 41
13.
Zurück zum Zitat Månsson S, Wolber J, Driehuys B, Wollmer P, Golman K (2003) Characterization of diffusing capacity and perfusion of the rat lung in a lipopolysaccaride disease model using hyperpolarized 129Xe. Magn Reson Med 50:1170–1179 Månsson S, Wolber J, Driehuys B, Wollmer P, Golman K (2003) Characterization of diffusing capacity and perfusion of the rat lung in a lipopolysaccaride disease model using hyperpolarized 129Xe. Magn Reson Med 50:1170–1179
14.
Zurück zum Zitat Patz S, Muradyan I, Hrovat MI et al (2011) Diffusion of hyperpolarized 129Xe in the lung: a simplified model of 129Xe septal uptake and experimental results. New J Phys 13 Patz S, Muradyan I, Hrovat MI et al (2011) Diffusion of hyperpolarized 129Xe in the lung: a simplified model of 129Xe septal uptake and experimental results. New J Phys 13
15.
Zurück zum Zitat Chang YV (2013) MOXE: a model of gas exchange for hyperpolarized 129Xe magnetic resonance of the lung. Magn Reson Med 69:884–890CrossRefPubMed Chang YV (2013) MOXE: a model of gas exchange for hyperpolarized 129Xe magnetic resonance of the lung. Magn Reson Med 69:884–890CrossRefPubMed
16.
Zurück zum Zitat Chang YV, Quirk JD, Ruset IC, Atkinson JJ, Hersman FW, Woods JC (2014) Quantification of human lung structure and physiology using hyperpolarized 129Xe. Magn Reson Med 71:339–344CrossRefPubMed Chang YV, Quirk JD, Ruset IC, Atkinson JJ, Hersman FW, Woods JC (2014) Quantification of human lung structure and physiology using hyperpolarized 129Xe. Magn Reson Med 71:339–344CrossRefPubMed
17.
Zurück zum Zitat Stewart NJ, Parra-Robles J, Wild JM (2016) Finite element modeling of (129)Xe diffusive gas exchange NMR in the human alveoli. J Magn Reson 271:21–33CrossRefPubMed Stewart NJ, Parra-Robles J, Wild JM (2016) Finite element modeling of (129)Xe diffusive gas exchange NMR in the human alveoli. J Magn Reson 271:21–33CrossRefPubMed
18.
Zurück zum Zitat Wang JM, Robertson SH, Wang Z et al (2018) Using hyperpolarized (129)Xe MRI to quantify regional gas transfer in idiopathic pulmonary fibrosis. Thorax 73:21–28CrossRefPubMed Wang JM, Robertson SH, Wang Z et al (2018) Using hyperpolarized (129)Xe MRI to quantify regional gas transfer in idiopathic pulmonary fibrosis. Thorax 73:21–28CrossRefPubMed
19.
Zurück zum Zitat Kaushik SS, Freeman MS, Cleveland ZI et al (2013) Probing the regional distribution of pulmonary gas exchange through single-breath gas- and dissolved-phase 129Xe MR imaging. J Appl Physiol (1985) 115:850–860CrossRef Kaushik SS, Freeman MS, Cleveland ZI et al (2013) Probing the regional distribution of pulmonary gas exchange through single-breath gas- and dissolved-phase 129Xe MR imaging. J Appl Physiol (1985) 115:850–860CrossRef
21.
Zurück zum Zitat Driehuys B, Cofer GP, Pollaro J, Mackel JB, Hedlund LW, Johnson GA (2006) Imaging alveolar-capillary gas transfer using hyperpolarized 129Xe MRI. Proc Natl Acad Sci U S A 103:18278–18283CrossRefPubMedPubMedCentral Driehuys B, Cofer GP, Pollaro J, Mackel JB, Hedlund LW, Johnson GA (2006) Imaging alveolar-capillary gas transfer using hyperpolarized 129Xe MRI. Proc Natl Acad Sci U S A 103:18278–18283CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Qing K, Mugler JP 3rd, Altes TA et al (2014) Assessment of lung function in asthma and COPD using hyperpolarized 129Xe chemical shift saturation recovery spectroscopy and dissolved-phase MRI. NMR Biomed 27:1490–1501 Qing K, Mugler JP 3rd, Altes TA et al (2014) Assessment of lung function in asthma and COPD using hyperpolarized 129Xe chemical shift saturation recovery spectroscopy and dissolved-phase MRI. NMR Biomed 27:1490–1501
23.
Zurück zum Zitat Qing K, Ruppert K, Jiang Y et al (2014) Regional mapping of gas uptake by blood and tissue in the human lung using hyperpolarized xenon-129 MRI. J Magn Reson Imaging 39:346–359CrossRefPubMed Qing K, Ruppert K, Jiang Y et al (2014) Regional mapping of gas uptake by blood and tissue in the human lung using hyperpolarized xenon-129 MRI. J Magn Reson Imaging 39:346–359CrossRefPubMed
24.
Zurück zum Zitat Ruppert K, Mata JF, Brookeman JR, Hagspiel KD, Mugler JP 3rd (2004) Exploring lung function with hyperpolarized (129)Xe nuclear magnetic resonance. Magn Reson Med 51:676–687CrossRefPubMed Ruppert K, Mata JF, Brookeman JR, Hagspiel KD, Mugler JP 3rd (2004) Exploring lung function with hyperpolarized (129)Xe nuclear magnetic resonance. Magn Reson Med 51:676–687CrossRefPubMed
25.
Zurück zum Zitat Ruppert K, Brookeman JR, Hagspiel KD, Driehuys B, Mugler JP 3rd (2000) NMR of hyperpolarized (129)Xe in the canine chest: spectral dynamics during a breath-hold. NMR Biomed 13:220–228CrossRefPubMed Ruppert K, Brookeman JR, Hagspiel KD, Driehuys B, Mugler JP 3rd (2000) NMR of hyperpolarized (129)Xe in the canine chest: spectral dynamics during a breath-hold. NMR Biomed 13:220–228CrossRefPubMed
26.
Zurück zum Zitat Muradyan I, Butler JP, Dabaghyan M et al (2013) Single-breath xenon polarization transfer contrast (SB-XTC): implementation and initial results in healthy humans. J Magn Reson Imaging 37:457–470CrossRefPubMed Muradyan I, Butler JP, Dabaghyan M et al (2013) Single-breath xenon polarization transfer contrast (SB-XTC): implementation and initial results in healthy humans. J Magn Reson Imaging 37:457–470CrossRefPubMed
27.
Zurück zum Zitat Patz S, Muradian I, Hrovat MI et al (2008) Human pulmonary imaging and spectroscopy with hyperpolarized 129Xe at 0.2T. Acad Radiol 15:713–727 Patz S, Muradian I, Hrovat MI et al (2008) Human pulmonary imaging and spectroscopy with hyperpolarized 129Xe at 0.2T. Acad Radiol 15:713–727
28.
Zurück zum Zitat Reeder SB, Pineda AR, Wen Z et al (2005) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med 54:636–644CrossRef Reeder SB, Pineda AR, Wen Z et al (2005) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med 54:636–644CrossRef
29.
Zurück zum Zitat Schulte RF, Sperl JI, Weidl E et al (2013) Saturation-recovery metabolic-exchange rate imaging with hyperpolarized [1-13C] pyruvate using spectral-spatial excitation. Magn Reson Med 69:1209–1216CrossRefPubMed Schulte RF, Sperl JI, Weidl E et al (2013) Saturation-recovery metabolic-exchange rate imaging with hyperpolarized [1-13C] pyruvate using spectral-spatial excitation. Magn Reson Med 69:1209–1216CrossRefPubMed
30.
Zurück zum Zitat Doganay O, Wade T, Hegarty E, McKenzie C, Schulte RF, Santyr GE (2016) Hyperpolarized (129) Xe imaging of the rat lung using spiral IDEAL. Magn Reson Med 76:566–576CrossRefPubMed Doganay O, Wade T, Hegarty E, McKenzie C, Schulte RF, Santyr GE (2016) Hyperpolarized (129) Xe imaging of the rat lung using spiral IDEAL. Magn Reson Med 76:566–576CrossRefPubMed
31.
Zurück zum Zitat Doganay O, Stirrat E, McKenzie C, Schulte RF, Santyr GE (2016) Quantification of regional early stage gas exchange changes using hyperpolarized (129)Xe MRI in a rat model of radiation-induced lung injury. Med Phys 43:2410CrossRefPubMed Doganay O, Stirrat E, McKenzie C, Schulte RF, Santyr GE (2016) Quantification of regional early stage gas exchange changes using hyperpolarized (129)Xe MRI in a rat model of radiation-induced lung injury. Med Phys 43:2410CrossRefPubMed
32.
Zurück zum Zitat Zanette B, Stirrat E, Jelveh S, Hope A, Santyr G (2017) Physiological gas exchange mapping of hyperpolarized (129) Xe using spiral-IDEAL and MOXE in a model of regional radiation-induced lung injury. Med Phys. https://doi.org/10.1002/mp.12730 Zanette B, Stirrat E, Jelveh S, Hope A, Santyr G (2017) Physiological gas exchange mapping of hyperpolarized (129) Xe using spiral-IDEAL and MOXE in a model of regional radiation-induced lung injury. Med Phys. https://​doi.​org/​10.​1002/​mp.​12730
33.
Zurück zum Zitat Wiesinger F, Weidl E, Menzel MI et al (2012) IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1-13C]pyruvate. Magn Reson Med 68:8–16CrossRefPubMed Wiesinger F, Weidl E, Menzel MI et al (2012) IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1-13C]pyruvate. Magn Reson Med 68:8–16CrossRefPubMed
34.
Zurück zum Zitat Dregely I, Ruset IC, Mata JF et al (2012) Multiple-exchange-time xenon polarization transfer contrast (MXTC) MRI: initial results in animals and healthy volunteers. Magn Reson Med 67:943–953CrossRefPubMed Dregely I, Ruset IC, Mata JF et al (2012) Multiple-exchange-time xenon polarization transfer contrast (MXTC) MRI: initial results in animals and healthy volunteers. Magn Reson Med 67:943–953CrossRefPubMed
35.
Zurück zum Zitat Dregely I, Mugler JP 3rd, Ruset IC et al (2011) Hyperpolarized Xenon-129 gas-exchange imaging of lung microstructure: first case studies in subjects with obstructive lung disease. J Magn Reson Imaging 33:1052–1062CrossRefPubMedPubMedCentral Dregely I, Mugler JP 3rd, Ruset IC et al (2011) Hyperpolarized Xenon-129 gas-exchange imaging of lung microstructure: first case studies in subjects with obstructive lung disease. J Magn Reson Imaging 33:1052–1062CrossRefPubMedPubMedCentral
36.
Zurück zum Zitat Ruppert K, Brookeman JR, Hagspiel KD, Mugler JP 3rd (2000) Probing lung physiology with xenon polarization transfer contrast (XTC). Magn Reson Med 44:349–357CrossRefPubMed Ruppert K, Brookeman JR, Hagspiel KD, Mugler JP 3rd (2000) Probing lung physiology with xenon polarization transfer contrast (XTC). Magn Reson Med 44:349–357CrossRefPubMed
37.
Zurück zum Zitat Kaushik SS, Freeman MS, Yoon SW et al (2014) Measuring diffusion limitation with a perfusion-limited gas--hyperpolarized 129Xe gas-transfer spectroscopy in patients with idiopathic pulmonary fibrosis. J Appl Physiol (1985) 117:577–585CrossRef Kaushik SS, Freeman MS, Yoon SW et al (2014) Measuring diffusion limitation with a perfusion-limited gas--hyperpolarized 129Xe gas-transfer spectroscopy in patients with idiopathic pulmonary fibrosis. J Appl Physiol (1985) 117:577–585CrossRef
38.
Zurück zum Zitat Santyr G, Fox M, Thind K et al (2014) Anatomical, functional and metabolic imaging of radiation-induced lung injury using hyperpolarized MRI. NMR Biomed 27:1515–1524CrossRefPubMed Santyr G, Fox M, Thind K et al (2014) Anatomical, functional and metabolic imaging of radiation-induced lung injury using hyperpolarized MRI. NMR Biomed 27:1515–1524CrossRefPubMed
39.
Zurück zum Zitat Zanette B, Stirrat E, Jelveh S, Hope A, Santyr G (2017) Detection of regional radiation-induced lung injury using hyperpolarized 129 Xe chemical shift imaging in a rat model involving partial lung irradiation: proof-of-concept demonstration. Adv Radiat Oncol 2:475–484CrossRefPubMedPubMedCentral Zanette B, Stirrat E, Jelveh S, Hope A, Santyr G (2017) Detection of regional radiation-induced lung injury using hyperpolarized 129 Xe chemical shift imaging in a rat model involving partial lung irradiation: proof-of-concept demonstration. Adv Radiat Oncol 2:475–484CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Ouriadov A, Fox M, Hegarty E, Parraga G, Wong E, Santyr EG (2016) Early stage radiation-induced lung injury detected using hyperpolarized (129) Xe morphometry: proof-of-concept demonstration in a rat model. Magn Reson Med Sci 75:2421–2431 Ouriadov A, Fox M, Hegarty E, Parraga G, Wong E, Santyr EG (2016) Early stage radiation-induced lung injury detected using hyperpolarized (129) Xe morphometry: proof-of-concept demonstration in a rat model. Magn Reson Med Sci 75:2421–2431
41.
Zurück zum Zitat Li H, Zhang Z, Zhao X, Sun X, Ye C, Zhou X (2016) Quantitative evaluation of radiation-induced lung injury with hyperpolarized xenon magnetic resonance. Magn Reson Med 76:408–416CrossRefPubMed Li H, Zhang Z, Zhao X, Sun X, Ye C, Zhou X (2016) Quantitative evaluation of radiation-induced lung injury with hyperpolarized xenon magnetic resonance. Magn Reson Med 76:408–416CrossRefPubMed
42.
Zurück zum Zitat Doganay O, Fox M, Santyr GE (2014) Measurement of pulmonary perfusion and gas exchange using hyperpolarized 129Xe in a rodent model of radiation-induced lung Injury. Proceedings of the 22th annual meeting of ISMRM (abstract ID 2290), Milan, Italy Doganay O, Fox M, Santyr GE (2014) Measurement of pulmonary perfusion and gas exchange using hyperpolarized 129Xe in a rodent model of radiation-induced lung Injury. Proceedings of the 22th annual meeting of ISMRM (abstract ID 2290), Milan, Italy
43.
Zurück zum Zitat Leung G, Norquay G, Schulte RF, Wild JM (2015) Radiofrequency pulse design for the selective excitation of dissolved 129Xe. Magn Reson Med 73:21–30CrossRefPubMed Leung G, Norquay G, Schulte RF, Wild JM (2015) Radiofrequency pulse design for the selective excitation of dissolved 129Xe. Magn Reson Med 73:21–30CrossRefPubMed
Metadaten
Titel
Magnetic resonance imaging of the time course of hyperpolarized 129Xe gas exchange in the human lungs and heart
verfasst von
Ozkan Doganay
Mitchell Chen
Tahreema Matin
Marzia Rigolli
Julie-Ann Phillips
Anthony McIntyre
Fergus V. Gleeson
Publikationsdatum
05.12.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 5/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5853-9

Weitere Artikel der Ausgabe 5/2019

European Radiology 5/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.