Cardiac memory in humans: vectocardiographic quantification in cardiac resynchronization therapy

verfasst von: Luigi Padeletti, Chiara Fantappiè, Laura Perrotta, Giuseppe Ricciardi, Paolo Pieragnoli, Marco Chiostri, Sergio Valsecchi, Maria Cristina Porciani, Antonio Michelucci, Fabio Fantini

Erschienen in: Clinical Research in Cardiology | Ausgabe 1/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

“Cardiac memory” (CM) refers to a change in repolarization induced by an altered pathway of activation. The effects of biventricular pacing on CM induction have not been investigated.

Objective

To investigate the development of CM during cardiac resynchronization therapy (CRT) through vectorcardiography (VCG).

Methods

Eleven patients undergoing CRT were enrolled. VCG was acquired during spontaneous ventricular activation at baseline and during AAI and DDD pacing immediately after and 7, 14, 21 and 60 days after the implantation.

Results

At 1-week follow-up, during AAI pacing T vector angles significantly changed (azimuth 23 ± 19°; p = 0.002; elevation 23 ± 27°; p = 0.019) and magnitude significantly increased (baseline 1.13 ± 0.69 mV; 7 days: 1.77 ± 1.27 mV; p = 0.026). T angle changes remained stable throughout the follow-up period while a further significant increase in magnitude was observed at 60 days (2.21 ± 1.50 mV; p = 0.01 vs. baseline and p = 0.04 vs. 7 days). Paced T vector magnitude at implant (2.24 ± 1.25 mV) decreased significantly at 7 days (1.64 ± 1.26 mV; p = 0.030) with a further significant decrease at 60 days (1.40 ± 1.18 mV; p = 0.003 vs. baseline; p = 0.02 vs. 7 days).

Conclusion

CRT induces a significant change in T vector magnitude, azimuth, and elevation after resumption of spontaneous ventricular activation after 7 days from implantation. While further changes in T vector angle were not observed, after 2 months of CRT a significant decrease of paced T vector magnitude and a significant increase of spontaneous T vector magnitude were observed.
Literatur
1.
Zurück zum Zitat Chatterjee K, Harris A, Davies G, Leatham A (1969) Electrocardiographic changes subsequent to artificial ventricular depolarization. Br Heart J 31:770–779CrossRefPubMed Chatterjee K, Harris A, Davies G, Leatham A (1969) Electrocardiographic changes subsequent to artificial ventricular depolarization. Br Heart J 31:770–779CrossRefPubMed
2.
Zurück zum Zitat Chatterjee K, Harris AM, Davies JG, Leatham A (1969) T-wave changes after artificial pacing. Lancet 2(7629):1061–1063CrossRefPubMed Chatterjee K, Harris AM, Davies JG, Leatham A (1969) T-wave changes after artificial pacing. Lancet 2(7629):1061–1063CrossRefPubMed
3.
Zurück zum Zitat Rosenbaum MB, Blanco HH, Elizari MV, Lazzari JO, Davidenko JM (1982) Electrotonic modulation of the T wave and cardiac memory. Am J Cardiol 50:213–222CrossRefPubMed Rosenbaum MB, Blanco HH, Elizari MV, Lazzari JO, Davidenko JM (1982) Electrotonic modulation of the T wave and cardiac memory. Am J Cardiol 50:213–222CrossRefPubMed
4.
Zurück zum Zitat Wecke L, Gadler F, Linde C, Lundahl G, Rosen MR, Bergfeldt L (2005) Temporal characteristics of cardiac memory in humans: vectorcardiographic quantification in a model of cardiac pacing. Heart Rhythm 2:28–34CrossRefPubMed Wecke L, Gadler F, Linde C, Lundahl G, Rosen MR, Bergfeldt L (2005) Temporal characteristics of cardiac memory in humans: vectorcardiographic quantification in a model of cardiac pacing. Heart Rhythm 2:28–34CrossRefPubMed
5.
Zurück zum Zitat Shvilkin A, Bojovic B, Vajdic B, Gussak I, Zimetbaum P, Josephson ME (2009) Vectorcardiographic determinants of cardiac memory during normal ventricular activation and continuous ventricular pacing. Heart Rhythm 6:943–948CrossRefPubMed Shvilkin A, Bojovic B, Vajdic B, Gussak I, Zimetbaum P, Josephson ME (2009) Vectorcardiographic determinants of cardiac memory during normal ventricular activation and continuous ventricular pacing. Heart Rhythm 6:943–948CrossRefPubMed
6.
Zurück zum Zitat Abraham WT, Fisher WG, Smith AL et al (2002) Cardiac resynchronization in chronic heart failure. N Engl J Med 346:1845–1853CrossRefPubMed Abraham WT, Fisher WG, Smith AL et al (2002) Cardiac resynchronization in chronic heart failure. N Engl J Med 346:1845–1853CrossRefPubMed
7.
Zurück zum Zitat Bristow MR, Saxon LA, Boehmer J et al (2004) Cardiac resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 350:2140–2150CrossRefPubMed Bristow MR, Saxon LA, Boehmer J et al (2004) Cardiac resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 350:2140–2150CrossRefPubMed
8.
Zurück zum Zitat Cleland JG, Daubert JC, Erdmann E et al (2005) The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 352:1539–1549CrossRefPubMed Cleland JG, Daubert JC, Erdmann E et al (2005) The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 352:1539–1549CrossRefPubMed
9.
Zurück zum Zitat Ricci R, Pignalberi C, Ansalone G et al (2002) Early and late QRS morphology and width in biventricular pacing: relationship to lead site and electrical remodeling. J Interv Card Electrophysiol 6:279–285CrossRefPubMed Ricci R, Pignalberi C, Ansalone G et al (2002) Early and late QRS morphology and width in biventricular pacing: relationship to lead site and electrical remodeling. J Interv Card Electrophysiol 6:279–285CrossRefPubMed
10.
Zurück zum Zitat Barold SS, Herweg B, Giudici M (2005) Electrocardiographic follow up of biventricular pacemakers. ANE 10(2):231–255PubMed Barold SS, Herweg B, Giudici M (2005) Electrocardiographic follow up of biventricular pacemakers. ANE 10(2):231–255PubMed
11.
Zurück zum Zitat Barold SS, Giudici CM, Herweg B (2006) Importance of the surface electrocardiogram for the assessment of cardiac resynchronization. In: Yu CM, Hayes DL, Auricchio A (eds) Cardiac resynchronization therapy, 2nd edn. Blackwell Futura, Malden, pp 309–323 Barold SS, Giudici CM, Herweg B (2006) Importance of the surface electrocardiogram for the assessment of cardiac resynchronization. In: Yu CM, Hayes DL, Auricchio A (eds) Cardiac resynchronization therapy, 2nd edn. Blackwell Futura, Malden, pp 309–323
12.
Zurück zum Zitat Kenny T (2007) The nuts and the bolts of cardiac resynchronization therapy. Blackwell Futura (St. Jude Medical), Malden pp 93–102; 135–143 Kenny T (2007) The nuts and the bolts of cardiac resynchronization therapy. Blackwell Futura (St. Jude Medical), Malden pp 93–102; 135–143
13.
Zurück zum Zitat ACC/AHA/HRS (2008) Guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American Heart Association task force on practice guidelines (writing committee to revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. JACC 51(21):e1–62 ACC/AHA/HRS (2008) Guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American Heart Association task force on practice guidelines (writing committee to revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. JACC 51(21):e1–62
14.
Zurück zum Zitat Ziegler R, Bloomfield DK (1970) A study of the normal QRS-T angle in the frontal plane. J Electrocardiol 3:161–167CrossRefPubMed Ziegler R, Bloomfield DK (1970) A study of the normal QRS-T angle in the frontal plane. J Electrocardiol 3:161–167CrossRefPubMed
15.
Zurück zum Zitat Alessandrini RS, McPherson DD, Kadish AH, Kane BJ, Goldberger JJ (1997) Cardiac memory: a mechanical and electrical phenomenon. Am J Physiol 272:H1952–H1959PubMed Alessandrini RS, McPherson DD, Kadish AH, Kane BJ, Goldberger JJ (1997) Cardiac memory: a mechanical and electrical phenomenon. Am J Physiol 272:H1952–H1959PubMed
16.
Zurück zum Zitat Rosen MR, Cohen IS, Danilo P Jr et al (1998) The heart remembers. Cardiovasc Res 40:469–482CrossRefPubMed Rosen MR, Cohen IS, Danilo P Jr et al (1998) The heart remembers. Cardiovasc Res 40:469–482CrossRefPubMed
17.
Zurück zum Zitat Shvilkin A, Danilo P Jr, Wang J, Burkhoff D, Anyukhovsky EP, Sosunov EA, Hara M, Rosen MR (1998) Evolution and resolution of long-term cardiac memory. Circulation 97:1810–1817PubMed Shvilkin A, Danilo P Jr, Wang J, Burkhoff D, Anyukhovsky EP, Sosunov EA, Hara M, Rosen MR (1998) Evolution and resolution of long-term cardiac memory. Circulation 97:1810–1817PubMed
18.
Zurück zum Zitat Fantoni C, Auricchio A (2006) Electrical activation sequence. In: Yu CM, Hayes DL, Auricchio A (eds) Cardiac resynchronization therapy, 2nd edn. Blackwell Futura, Malden, pp 35–54 Fantoni C, Auricchio A (2006) Electrical activation sequence. In: Yu CM, Hayes DL, Auricchio A (eds) Cardiac resynchronization therapy, 2nd edn. Blackwell Futura, Malden, pp 35–54
Metadaten
Titel
Cardiac memory in humans: vectocardiographic quantification in cardiac resynchronization therapy
verfasst von
Luigi Padeletti
Chiara Fantappiè
Laura Perrotta
Giuseppe Ricciardi
Paolo Pieragnoli
Marco Chiostri
Sergio Valsecchi
Maria Cristina Porciani
Antonio Michelucci
Fabio Fantini
Publikationsdatum
01.01.2011
Verlag
Springer-Verlag
Erschienen in
Clinical Research in Cardiology / Ausgabe 1/2011
Print ISSN: 1861-0684
Elektronische ISSN: 1861-0692
DOI
https://doi.org/10.1007/s00392-010-0209-8

Weitere Artikel der Ausgabe 1/2011

Clinical Research in Cardiology 1/2011 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.