01.05.2005 | Original Article

Measurement of meniscofemoral contact pressure after repair of bucket-handle tears with biodegradable implants

verfasst von: Roland Becker, Dieter Wirz, Cornelius Wolf, Beat Göpfert, Wolfgang Nebelung, Niklaus Friederich

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 4/2005

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Biodegradable implants are frequently used for meniscus repair. Articular cartilage damage has been reported recently after meniscus repair with biodegradable implants. The aim of the study was to investigate the meniscofemoral contact pressure at the posterior horn of the medial and lateral meniscus after repair of bucket-handle lacerations.

Materials and methods

Specimens were mounted in a materials testing machine (Bionix 858, MTS) which was equipped with a load cell. The quadriceps tendon was attached to a hydraulic cylinder, and knee motion was controlled via tension of the quadriceps tendon. A piezo-resistive system (Tekscan, Boston, MA, USA) measured the meniscofemoral contact pressure. Five different types of biodegradable implants (Arrow, Dart, Fastener, Stinger and Meniscal Screw) and horizontal suture (no. 2 Ethibond) were tested. The knee was extended from 90° of flexion to 0° under a constant load of 350 N due to adjustment of the tension force of the quadriceps tendon. The femorotibial pressure and contact area were recorded at 0°, 30°, 60° and 90° of flexion.

Results

The meniscofemoral pressure did not increase after meniscus repair with biodegradable implants or sutures. The meniscofemoral peak pressure at the posterior horn was 1.46±1.54 MPa in the medial compartment and 1.08±1.17 MPa in the lateral compartment at full knee extension. The meniscofemoral pressure increased significantly in both compartments with knee flexion from 0° to 90°.

Conclusion

Biodegradable implants for meniscus repair do not affect the meniscofemoral pressure. However, there remains a risk of damage to the cartilage when barbed implants are used. If the implant is not entirely advanced into the meniscus, the sharp head or some of the barbs at the column of the implant may come into direct contact with the articular cartilage of the femoral condyle or tibial plateau. The authors presume that incorrect positioning of the implant seems to be the major reason for cartilage damage.
Literatur
1.
Zurück zum Zitat Aagaard H, Verdonk R (1999) Function of the normal meniscus and consequences of meniscal resection. Scand J Med Sci Sports 9:134–140PubMed Aagaard H, Verdonk R (1999) Function of the normal meniscus and consequences of meniscal resection. Scand J Med Sci Sports 9:134–140PubMed
2.
Zurück zum Zitat Ahmed AM (1983) A pressure distribution transducer for in-vitro static measurements in synovial joints. J Biomech Eng 105:309–314PubMed Ahmed AM (1983) A pressure distribution transducer for in-vitro static measurements in synovial joints. J Biomech Eng 105:309–314PubMed
3.
Zurück zum Zitat Allen CR, Wong EK, Livesay GA, Sakane M, Fu FH, Woo SL (2000) Importance of the medial meniscus in the anterior cruciate ligament-deficient knee. J Orthop Res 18:109–115PubMed Allen CR, Wong EK, Livesay GA, Sakane M, Fu FH, Woo SL (2000) Importance of the medial meniscus in the anterior cruciate ligament-deficient knee. J Orthop Res 18:109–115PubMed
4.
Zurück zum Zitat Anderson K, Marx RG, Hannafin J, Warren RF (2000) Chondral injury following meniscal repair with a biodegradable implant. Arthroscopy 16:749–753PubMed Anderson K, Marx RG, Hannafin J, Warren RF (2000) Chondral injury following meniscal repair with a biodegradable implant. Arthroscopy 16:749–753PubMed
5.
Zurück zum Zitat Aspden RM, Yarker YE, Hukins DW (1985) Collagen orientations in the meniscus of the knee joint. J Anat 140:371–380PubMed Aspden RM, Yarker YE, Hukins DW (1985) Collagen orientations in the meniscus of the knee joint. J Anat 140:371–380PubMed
7.
Zurück zum Zitat Bargar WL, Moreland JR, Markolf KL, Shoemaker SC, Amstutz HC, Grant TT (1980) In vivo stability testing of post-meniscectomy knees. Clin Orthop 150:247–252PubMed Bargar WL, Moreland JR, Markolf KL, Shoemaker SC, Amstutz HC, Grant TT (1980) In vivo stability testing of post-meniscectomy knees. Clin Orthop 150:247–252PubMed
8.
Zurück zum Zitat Becker R, Schroder M, Starke C, Urbach D, Nebelung W (2001) Biomechanical investigations of different meniscal repair implants in comparison with horizontal sutures on human meniscus. Arthroscopy 17:439–444CrossRefPubMed Becker R, Schroder M, Starke C, Urbach D, Nebelung W (2001) Biomechanical investigations of different meniscal repair implants in comparison with horizontal sutures on human meniscus. Arthroscopy 17:439–444CrossRefPubMed
9.
Zurück zum Zitat Becker R, Starke C, Heymann M, Nebelung W (2002) Biomechanical properties under cyclic loading of seven meniscus repair techniques. Clin Orthop 400:236–245PubMed Becker R, Starke C, Heymann M, Nebelung W (2002) Biomechanical properties under cyclic loading of seven meniscus repair techniques. Clin Orthop 400:236–245PubMed
10.
Zurück zum Zitat Becker R, Voigt D, Starke C, Heymann M, Wilson GA, Nebelung W (2001) Biomechanical properties of quadruple tendon and patellar tendon femoral fixation techniques. Knee Surg Sports Traumatol Arthrosc 9:337–342PubMed Becker R, Voigt D, Starke C, Heymann M, Wilson GA, Nebelung W (2001) Biomechanical properties of quadruple tendon and patellar tendon femoral fixation techniques. Knee Surg Sports Traumatol Arthrosc 9:337–342PubMed
11.
Zurück zum Zitat Favenesi J, Shaffer J, Mow VC (1983) Biphasic mechanical properties of knee meniscus. Orthop Trans 8:174 Favenesi J, Shaffer J, Mow VC (1983) Biphasic mechanical properties of knee meniscus. Orthop Trans 8:174
12.
Zurück zum Zitat Felson DT, Naimark A, Anderson J, Kazis L, Castelli W, Meenan RF (1987) The prevalence of knee osteoarthritis in the elderly. The Framingham Osteoarthritis Study. Arthritis Rheum 30:914–918PubMed Felson DT, Naimark A, Anderson J, Kazis L, Castelli W, Meenan RF (1987) The prevalence of knee osteoarthritis in the elderly. The Framingham Osteoarthritis Study. Arthritis Rheum 30:914–918PubMed
13.
Zurück zum Zitat Fithian DC, Kelly MA, Mow VC (1990) Material properties and structure-function relationships in the menisci. Clin Orthop 252:19–31PubMed Fithian DC, Kelly MA, Mow VC (1990) Material properties and structure-function relationships in the menisci. Clin Orthop 252:19–31PubMed
14.
Zurück zum Zitat Fukubayashi T, Kurosawa H (1980) The contact area and pressure distribution pattern of the knee. A study of normal and osteoarthrotic knee joints. Acta Orthop Scand 51:871–879PubMed Fukubayashi T, Kurosawa H (1980) The contact area and pressure distribution pattern of the knee. A study of normal and osteoarthrotic knee joints. Acta Orthop Scand 51:871–879PubMed
15.
Zurück zum Zitat Goertzen DJ, Budney DR, Cinats JG (1997) Methodology and apparatus to determine material properties of the knee joint meniscus. Med Eng Phys 19:412–419CrossRefPubMed Goertzen DJ, Budney DR, Cinats JG (1997) Methodology and apparatus to determine material properties of the knee joint meniscus. Med Eng Phys 19:412–419CrossRefPubMed
16.
Zurück zum Zitat Harris ML, Morberg P, Bruce WJ, Walsh WR (1999) An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji film. J Biomech 32:951–958CrossRefPubMed Harris ML, Morberg P, Bruce WJ, Walsh WR (1999) An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji film. J Biomech 32:951–958CrossRefPubMed
17.
Zurück zum Zitat Higuchi H, Kimura M, Shirakura K, Terauchi M, Takagishi K (2000) Factors affecting long-term results after arthroscopic partial meniscectomy. Clin Orthop 377:161–168CrossRefPubMed Higuchi H, Kimura M, Shirakura K, Terauchi M, Takagishi K (2000) Factors affecting long-term results after arthroscopic partial meniscectomy. Clin Orthop 377:161–168CrossRefPubMed
18.
Zurück zum Zitat Hsieh HH, Walker PS (1976) Stabilizing mechanisms of the loaded and unloaded knee joint. J Bone Joint Surg Am 58:87–93PubMed Hsieh HH, Walker PS (1976) Stabilizing mechanisms of the loaded and unloaded knee joint. J Bone Joint Surg Am 58:87–93PubMed
19.
Zurück zum Zitat Kohn D, Mussack T, Plitz W (1993) Material properties of meniscus tissue and the effect of the menisci on the compression behavior of the knee joint. Z Orthop 131:397–403PubMed Kohn D, Mussack T, Plitz W (1993) Material properties of meniscus tissue and the effect of the menisci on the compression behavior of the knee joint. Z Orthop 131:397–403PubMed
20.
Zurück zum Zitat Kohn D, Siebert W (1989) Meniscus suture techniques: a comparative biomechanical cadaver study. Arthroscopy 5:324–327CrossRefPubMed Kohn D, Siebert W (1989) Meniscus suture techniques: a comparative biomechanical cadaver study. Arthroscopy 5:324–327CrossRefPubMed
21.
Zurück zum Zitat Lanzer WL, Komenda G (1990) Changes in articular cartilage after meniscectomy. Clin Orthop 252:41–48PubMed Lanzer WL, Komenda G (1990) Changes in articular cartilage after meniscectomy. Clin Orthop 252:41–48PubMed
22.
Zurück zum Zitat Lechner K, Hull ML, Howell SM (2000) Is the circumferential tensile modulus within a human medial meniscus affected by the test sample location and cross-sectional area? J Orthop Res 18:945–951PubMed Lechner K, Hull ML, Howell SM (2000) Is the circumferential tensile modulus within a human medial meniscus affected by the test sample location and cross-sectional area? J Orthop Res 18:945–951PubMed
23.
Zurück zum Zitat Levy IM, Torzilli PA, Warren RF (1982) The effect of medial meniscectomy on anterior-posterior motion of the knee. J Bone Joint Surg Am 64:883–888PubMed Levy IM, Torzilli PA, Warren RF (1982) The effect of medial meniscectomy on anterior-posterior motion of the knee. J Bone Joint Surg Am 64:883–888PubMed
24.
Zurück zum Zitat Messner K, Gao J (1998) The menisci of the knee joint. Anatomical and functional characteristics, and a rationale for clinical treatment. J Anat 193:161–178CrossRefPubMed Messner K, Gao J (1998) The menisci of the knee joint. Anatomical and functional characteristics, and a rationale for clinical treatment. J Anat 193:161–178CrossRefPubMed
26.
Zurück zum Zitat Mow V, Ratcliffe A, Chern K, Kelly M (1992) Structure and function relationships of the menisci of the knee. In: Mow V, Arnoczky SP, Jackson DW (eds) Knee meniscus: basic and clinical foundations. Raven Press, New York, pp 37–53 Mow V, Ratcliffe A, Chern K, Kelly M (1992) Structure and function relationships of the menisci of the knee. In: Mow V, Arnoczky SP, Jackson DW (eds) Knee meniscus: basic and clinical foundations. Raven Press, New York, pp 37–53
27.
Zurück zum Zitat Neyret P, Donell ST, DeJour D, DeJour H (1993) Partial meniscectomy and anterior cruciate ligament rupture in soccer players. A study with a minimum 20-year followup. Am J Sports Med 21:455–460PubMed Neyret P, Donell ST, DeJour D, DeJour H (1993) Partial meniscectomy and anterior cruciate ligament rupture in soccer players. A study with a minimum 20-year followup. Am J Sports Med 21:455–460PubMed
28.
Zurück zum Zitat Paletta GAJ, Manning T, Snell E, Parker R, Bergfeld J (1997) The effect of allograft meniscal replacement on intraarticular contact area and pressures in the human knee. A biomechanical study. Am J Sports Med 25:692–698PubMed Paletta GAJ, Manning T, Snell E, Parker R, Bergfeld J (1997) The effect of allograft meniscal replacement on intraarticular contact area and pressures in the human knee. A biomechanical study. Am J Sports Med 25:692–698PubMed
29.
Zurück zum Zitat Petersen W, Tillmann B (1999) Structure and vascularization of the knee joint menisci. Z Orthop 137:31–37PubMed Petersen W, Tillmann B (1999) Structure and vascularization of the knee joint menisci. Z Orthop 137:31–37PubMed
30.
Zurück zum Zitat Proctor CS, Schmidt MB, Whipple RR, Kelly MA, Mow VC (1989) Material properties of the normal medial bovine meniscus. J Orthop Res 7:771–782PubMed Proctor CS, Schmidt MB, Whipple RR, Kelly MA, Mow VC (1989) Material properties of the normal medial bovine meniscus. J Orthop Res 7:771–782PubMed
31.
Zurück zum Zitat Rockborn P, Gillquist J (1996) Long-term results after arthroscopic meniscectomy. The role of preexisting cartilage fibrillation in a 13 year follow-up of 60 patients. Int J Sports Med 17:608–613PubMed Rockborn P, Gillquist J (1996) Long-term results after arthroscopic meniscectomy. The role of preexisting cartilage fibrillation in a 13 year follow-up of 60 patients. Int J Sports Med 17:608–613PubMed
32.
Zurück zum Zitat Ross G, Grabill J, McDevitt E (2000) Chondral injury after meniscal repair with bioabsorbable arrows. Arthroscopy 16:754–756PubMed Ross G, Grabill J, McDevitt E (2000) Chondral injury after meniscal repair with bioabsorbable arrows. Arthroscopy 16:754–756PubMed
33.
Zurück zum Zitat Schatzmann L, Brunner P, Staubli HU (1998) Effect of cyclic preconditioning on the tensile properties of human quadriceps tendons and patellar ligaments. Knee Surg Sports Traumatol Arthrosc 6 [Suppl 1]:S56–61 Schatzmann L, Brunner P, Staubli HU (1998) Effect of cyclic preconditioning on the tensile properties of human quadriceps tendons and patellar ligaments. Knee Surg Sports Traumatol Arthrosc 6 [Suppl 1]:S56–61
34.
Zurück zum Zitat Seil R, Rupp S, Mai C, Kohn D (2001) The footprint of meniscus fixation devices on the femoral surface of the medial meniscus: a biomechanical cadaver study. 3rd ISAKOS Congress, Montreaux, Switzerland, May 14–18, poster no. 400 Seil R, Rupp S, Mai C, Kohn D (2001) The footprint of meniscus fixation devices on the femoral surface of the medial meniscus: a biomechanical cadaver study. 3rd ISAKOS Congress, Montreaux, Switzerland, May 14–18, poster no. 400
35.
Zurück zum Zitat Seil R, Rupp S, Dienst M, Mueller B, Bonkhoff H, Kohn DM (2000) Chondral lesions after arthroscopic meniscus repair using meniscus arrows. Arthroscopy 16:E17PubMed Seil R, Rupp S, Dienst M, Mueller B, Bonkhoff H, Kohn DM (2000) Chondral lesions after arthroscopic meniscus repair using meniscus arrows. Arthroscopy 16:E17PubMed
36.
Zurück zum Zitat Shoemaker SC, Markolf KL (1982) In vivo rotatory knee stability. Ligamentous and muscular contributions. J Bone Joint Surg Am 64:208–216PubMed Shoemaker SC, Markolf KL (1982) In vivo rotatory knee stability. Ligamentous and muscular contributions. J Bone Joint Surg Am 64:208–216PubMed
37.
Zurück zum Zitat Skaggs DL, Warden WH, Mow VC (1994) Radial tie fibers influence the tensile properties of the bovine medial meniscus. J Orthop Res 12:176–185PubMed Skaggs DL, Warden WH, Mow VC (1994) Radial tie fibers influence the tensile properties of the bovine medial meniscus. J Orthop Res 12:176–185PubMed
38.
Zurück zum Zitat Spilker RL, Donzelli PS, Mow VC (1992) A transversely isotropic biphasic finite element model of the meniscus. J Biomech 25:1027–1045PubMed Spilker RL, Donzelli PS, Mow VC (1992) A transversely isotropic biphasic finite element model of the meniscus. J Biomech 25:1027–1045PubMed
39.
Zurück zum Zitat Tekscan (2001) Tekscan product information—medical product overview. Tekscan, Boston Tekscan (2001) Tekscan product information—medical product overview. Tekscan, Boston
40.
Zurück zum Zitat Tissakht M, Ahmed AM (1995) Tensile stress-strain characteristics of the human meniscal material. J Biomech 28:411–422CrossRefPubMed Tissakht M, Ahmed AM (1995) Tensile stress-strain characteristics of the human meniscal material. J Biomech 28:411–422CrossRefPubMed
41.
Zurück zum Zitat Walker PS, Erkman MJ (1975) The role of the menisci in force transmission across the knee. Clin Orthop 109:184–192PubMed Walker PS, Erkman MJ (1975) The role of the menisci in force transmission across the knee. Clin Orthop 109:184–192PubMed
42.
Zurück zum Zitat Wirz D, Becker R, Li SF, Friederich NF, Muller W (2002) Validation of the Tekscan system for statistic and dynamic pressure measurements of the human femorotibial joint. Biomed Tech (Berl) 47:195–201 Wirz D, Becker R, Li SF, Friederich NF, Muller W (2002) Validation of the Tekscan system for statistic and dynamic pressure measurements of the human femorotibial joint. Biomed Tech (Berl) 47:195–201
Metadaten
Titel
Measurement of meniscofemoral contact pressure after repair of bucket-handle tears with biodegradable implants
verfasst von
Roland Becker
Dieter Wirz
Cornelius Wolf
Beat Göpfert
Wolfgang Nebelung
Niklaus Friederich
Publikationsdatum
01.05.2005
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 4/2005
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-004-0739-5

Weitere Artikel der Ausgabe 4/2005

Archives of Orthopaedic and Trauma Surgery 4/2005 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.