Acetabular component orientation in intra- and postoperative positions in total hip arthroplasty

verfasst von: Keiko Hayakawa, Yukihide Minoda, Masaharu Aihara, Akira Sakawa, Kenji Ohzono, Koichi Tada

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 9/2009

Einloggen, um Zugang zu erhalten

Abstract

Introduction

In total hip arthroplasty (THA), acetabular component orientation has critically important effects on dislocation, range of motion, polyethylene wear, pelvic osteolysis, and component migration. The differences in the pelvic orientation in the intraoperative lateral position for insertion of acetabular component during operation and that in the postoperative supine position for evaluation of acetabular component orientation will be one of the factors, which make outliers in acetabular component orientation. We compared acetabular component orientation between intraoperative lateral position and postoperative supine position in 100 consecutive primary THAs.

Materials and methods

A total of 100 consecutive primary THAs (between October 2004 and December 2005) in 100 patients performed by a single surgical team were investigated. Intraoperative anteroposterior radiographs of pelvis in the lateral position and postoperative anteroposterior radiographs of pelvis in the supine position were taken. Acetabular component orientation (vertical tilt and anteversion) were measured using computer software.

Results

The absolute values of difference between measurements in the two positions were 5.3° ± 4.5° (mean ± SD) for vertical tilt and 5.1° ± 3.7° for anteversion. The difference in the vertical tilt between the two positions was significant (P < 0.0001).

Conclusion

The difference in the acetabular component orientation between the two positions, which might be caused by the difference between intra- and postoperative pelvic orientation, should be considered during THA.
Literatur
1.
Zurück zum Zitat Amiot LP, Poulin F (2004) Computed tomography-based navigation for hip, knee, and spine surgery. Clin Orthop Relat Res 421:77–86PubMedCrossRef Amiot LP, Poulin F (2004) Computed tomography-based navigation for hip, knee, and spine surgery. Clin Orthop Relat Res 421:77–86PubMedCrossRef
2.
Zurück zum Zitat Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87-B:762–769CrossRef Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stockl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87-B:762–769CrossRef
3.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed
4.
Zurück zum Zitat Digioia AM III, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, Levison TJ, Picard F (2002) Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty 17:359–364PubMedCrossRef Digioia AM III, Jaramaz B, Plakseychuk AY, Moody JE Jr, Nikou C, Labarca RS, Levison TJ, Picard F (2002) Comparison of a mechanical acetabular alignment guide with computer placement of the socket. J Arthroplasty 17:359–364PubMedCrossRef
5.
Zurück zum Zitat DiGioia AM, Jaramaz B, Blackwell M, Simon DA, Morgan F, Moody JE, Nikou C, Colgan BD, Aston CA, Labarca RS, Kischell E, Kanade T (1998) Image guided navigation system to measure intraoperatively acetabular implant alignment. Clin Orthop Relat Res 355:8–22PubMedCrossRef DiGioia AM, Jaramaz B, Blackwell M, Simon DA, Morgan F, Moody JE, Nikou C, Colgan BD, Aston CA, Labarca RS, Kischell E, Kanade T (1998) Image guided navigation system to measure intraoperatively acetabular implant alignment. Clin Orthop Relat Res 355:8–22PubMedCrossRef
6.
Zurück zum Zitat DiGioia AM, Hafez MA, Jaramaz B, Levison TJ, Moody JE (2006) Functional pelvic orientation measured from lateral standing and sitting radiographs. Clin Orthop Relat Res 453:272–276PubMedCrossRef DiGioia AM, Hafez MA, Jaramaz B, Levison TJ, Moody JE (2006) Functional pelvic orientation measured from lateral standing and sitting radiographs. Clin Orthop Relat Res 453:272–276PubMedCrossRef
7.
Zurück zum Zitat Hassan DM, Johnston GH, Dust WN, Watson G, Dolovich AT (1998) Accuracy of intraoperative assessment of acetabular prosthesis placement. J Arthroplasty 13:80–84PubMedCrossRef Hassan DM, Johnston GH, Dust WN, Watson G, Dolovich AT (1998) Accuracy of intraoperative assessment of acetabular prosthesis placement. J Arthroplasty 13:80–84PubMedCrossRef
8.
Zurück zum Zitat Hube R, Birke A, Hein W, Klima S (2003) CT-based and fluoroscopy-based navigation for cup implantation in total hip arthroplasty (THA). Surg Technol Int 11:275–280PubMed Hube R, Birke A, Hein W, Klima S (2003) CT-based and fluoroscopy-based navigation for cup implantation in total hip arthroplasty (THA). Surg Technol Int 11:275–280PubMed
9.
Zurück zum Zitat Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17:282–288PubMedCrossRef Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17:282–288PubMedCrossRef
10.
Zurück zum Zitat Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty 13:530–534PubMedCrossRef Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty 13:530–534PubMedCrossRef
11.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60-A:217–220 Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60-A:217–220
12.
Zurück zum Zitat Minoda Y, Kadowaki T, Kim M (2006) Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res 445:186–191PubMed Minoda Y, Kadowaki T, Kim M (2006) Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res 445:186–191PubMed
13.
Zurück zum Zitat Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg Br 75-B:228–232 Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg Br 75-B:228–232
14.
Zurück zum Zitat Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H (2003) Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res 411:140–151PubMedCrossRef Nishihara S, Sugano N, Nishii T, Ohzono K, Yoshikawa H (2003) Measurements of pelvic flexion angle using three-dimensional computed tomography. Clin Orthop Relat Res 411:140–151PubMedCrossRef
15.
Zurück zum Zitat Sarmiento A, Latta LL (2006) A radiographic review of 135 total hip Charnley arthroplasties followed between 15 and 35 years. Acta Chir Orthop Traumatol Cech 73:145–150PubMed Sarmiento A, Latta LL (2006) A radiographic review of 135 total hip Charnley arthroplasties followed between 15 and 35 years. Acta Chir Orthop Traumatol Cech 73:145–150PubMed
16.
Zurück zum Zitat Schmalzried TP, Guttmann D, Grecula M, Amstutz HC (1994) The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J Bone Joint Surg Am 76:668–677 Schmalzried TP, Guttmann D, Grecula M, Amstutz HC (1994) The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J Bone Joint Surg Am 76:668–677
17.
Zurück zum Zitat Wentzensen A, Zheng G, Vock B, Langlotz U, Korber J, Nolte LP, Grutzner PA (2003) Image-based hip navigation. Int Orthop 27(Suppl):S43–S46PubMed Wentzensen A, Zheng G, Vock B, Langlotz U, Korber J, Nolte LP, Grutzner PA (2003) Image-based hip navigation. Int Orthop 27(Suppl):S43–S46PubMed
Metadaten
Titel
Acetabular component orientation in intra- and postoperative positions in total hip arthroplasty
verfasst von
Keiko Hayakawa
Yukihide Minoda
Masaharu Aihara
Akira Sakawa
Kenji Ohzono
Koichi Tada
Publikationsdatum
01.09.2009
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2009
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-008-0638-2

Weitere Artikel der Ausgabe 9/2009

Archives of Orthopaedic and Trauma Surgery 9/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.