Does intraoperative fluoroscopic 3D imaging provide extra information for fracture surgery?

verfasst von: B. Carelsen, R. Haverlag, D. Th. Ubbink, J. S. K. Luitse, J. C. Goslings

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 12/2008

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Fracture surgery of the extremities using 2D fluoroscopy frequently fails to detect the suboptimal positioning of implants and joint incongruities. The use of intraoperative 3D-rotational X-ray (3D-RX) imaging with a new X-ray device potentially reveals these failures. We compared 50 intraoperative (2D) results of surgery and certainty about the effectiveness of different aspects of fracture reduction as interpreted from conventional (2D) methods versus intraoperative 3D-RX in 42 distal extremity fractures by means of a surgery questionnaire. In addition, we investigated the need for revision surgery based on postoperative radiological findings in 81 patients. After fracture reduction, just before a 3D-RX scan, the surgeon preoperatively assessed the result of surgery. Three months after surgery, the 3D-RX scan was judged by three experienced surgeons independently. Intraoperative 3D-RX showed significantly more information as to screw positioning and rotation of the fracture reduction than the conventional method (p < 0.005). None of the 81 patients in whom 3D-RX was performed needed surgical revision based on postoperative radiological examinations. Intraoperative 3D-RX with this new device scanning offers additional information about extremity fracture reduction as compared to conventional intraoperative 2D imaging, and may reduce the need for revision surgery. The value of 3D-RX on functional outcomes still needs to be assessed.
Hinweise
The Medical Ethics commission of the Academic Medical Center, University of Amsterdam has approved “Evaluation of the clinical merit and image quality of intraoperative 3D-RX imaging for fracture surgery of the wrist, elbow and ankle.”

Introduction

With the use of conventional C-arm fluoroscopy in fracture surgery of the extremities, suboptimal positioning of implants (viz., screws, plates) and joint incongruities frequently remain unrevealed in clinical practice [2, 4, 5, 14, 16, 17]. These imperfections are often only recognized on postoperative computed tomography (CT) scans [2, 4, 12, 14, 21]. AO guidelines recommend an anatomical reduction of the fracture as a basis for optimal outcome. The reported malreduction, redisplacement and complication rate for fracture surgery of the extremities based on radiological result can be up to 26% [3, 6, 9, 11, 19]. To obtain more intraoperative information, one possible solution could be surgical navigation based on preoperative CT. However, registration of small joints and multiple fragments as encountered in fracture surgery of the distal extremities is laborious and inaccurate [18]. Another solution to avoid unexpected and/or suboptimal postoperative radiological results is better intraoperative imaging.
Intraoperative use of 3-dimensional rotational X-ray (3D-RX) for fracture surgery of the extremities has been reported regarding the Arcadis Orbic 3D in literature to lower revision in 21–39% of the cases [8, 12, 14, 15, 20]. Intraoperative revision of reduction, as indicated by the 3D scan, occurred for 11–19%, and for implant position 11–26% of the operations in these studies [8, 14]. These authors did not register the pre-3D-scan status based on conventional examination and 2D fluoroscopy, but considered this to be optimal. However, intraoperative imaging is often used to achieve an optimal result in a stepwise and minimally invasive manner. Therefore, instant optimal results and confidence regarding these results are not self-evident, that is, intraoperative 3D-RX scanning guides the different stages of an operation. The threshold for revision surgery is much higher when the patient has left the OR. Small imperfections that were revealed with aid of intraoperative 3D-RX could have been revised directly, but less likely afterwards. Further investigation of revisions avoided is desired as well as of revision rates during follow-up.
All above referenced publications used the Arcadis Orbic 3D (Siemens Healthcare, Erlangen, Germany), we used a recently new introduced mobile C-arm with 3D-RX capabilities, the BV Pulsera with 3D-RX (Philips Healthcare, Best, The Netherlands).
Our research objective was to compare performance and certainty about the procedure result as interpreted from conventional methods versus additional intraoperative 3D-RX for fracture reduction of the extremities with this new device. We hypothesized that 3D imaging would reveal a different, less satisfying surgical performance, but with more certainty about the procedure result. Subsequently, when acted upon these findings during the same surgical procedure, we expect a low revision rate because suboptimal reduction and fixation are detected intraoperatively and directly adjusted.

Materials and methods

All consecutive patients treated for intra-articular fractures of the foot, wrist, elbow or shoulder, between August 2005 and December 2006 in the Academic Medical Center (a tertiary referral center) at the University of Amsterdam were eligible for inclusion. Final inclusion with additional application of the 3D-RX system depended on the availability of the system, practical accessibility of the scanning area by the C-arm and a fracture type according to an AO classification of B2 or higher. Approval of the medical ethics board for this study was obtained, and all patients signed an informed consent.
Surgical procedures were performed according to standard protocols. For ankle and foot surgery, a standard table (AlphaMaquet®, Maquet, Baambrugge, NL) was used. As long as the foot is not in line with the table side rails no metal artifacts will appear (Fig. 1, right). For hand and wrist surgery, we utilized a RADIUS table (Liftac AG, Zurich, Switzerland), which is a unilateral suspended-arm table with a carbon fiber tabletop without ground support to allow rotational imaging (Fig. 1, left panel). Before positioning the C-arm, a sterile bag (such as normally used to cover the image intensifier or an instrument table) was placed over the operation zone by the surgeon and instrumenting nurse. Thus, the extremity, table and drapes are inside the bag to keep the operation zone sterile. An anteroposterior and lateral low-dose X-ray image was taken to confirm the alignment of the system to the patient’s anatomy.
During the operation, the surgeon decided whether and when to perform a subsequent 3D-RX scan. If no scans were made during the operation, at least one was made at the end of the operation just before wound closure. The surgeon was free to check in any way the current operation state by means of 2D fluoroscopy or examination. After the scan was made, but before the 3D image was shown, the surgeon answered the questions used for intraoperative assessment of the operation result as shown in “Appendix A”, Table 4. The first 50 scans were included in this assessment. No questionnaire was taken immediately after the 3D to avoid bias.
For 3D-RX imaging, the BV Pulsera (Philips Healthcare, Best, The Netherlands) with 3D-RX system was used. This is a new mobile C-arm, modified for motorized movement and integrated with a Philips 3D-RA workstation. For this study, a series of 225 projection images are taken during a 200° rotation of the C-arm in 30 s, which is used to reconstruct a 3D-RX scan. The dose of each image in this scanning run is dynamically adjusted to obtain low dose and optimal image quality. The maximum effective dose for a hand/wrist 3D-RX scan is 6 μSv, for a foot 10 μSv [12]. For comparison, for CT imaging of the lower extremities a typical dose is 120 μSv and the dose of plain hand and foot radiograph are 0.17 μSv and 0.7 μSv, respectively [7, 13]. Depending on the size of the anatomy of interest, we adjusted the magnification to either a spherical shape with a diameter of 12.5 or 18 cm, with slices of 0.5 or 0.7 mm thick in all directions, respectively. After reconstruction, the 3D-RX data set could be viewed on a workstation in the operating room. Both volume rendering and coronal, sagittal and axial slices were available. After this, the operation could continue, while the images were being processed (4 min). Sterile coverage, aligning and scanning takes approximately 3 min, in total 7 min are needed to obtain a 3D image. Next, the surgeon examined the 3D-RX images and continued or finished the operation.
Four months after obtaining the first 50 scans, three experienced trauma surgeons, who also operated the majority of the patients included, reviewed the 3D-RX images. The reviewers were offered the images in a blinded and random fashion and were asked to answer the questions regarding the intraoperative assessment of the operation result. Primary outcome was the result of pre-3D scan surgery as judged by the surgeon, based on conventional examination and 2D fluoroscopy, in comparison with the judgment based on the 3D-RX scan by means of the questionnaire. Secondary outcome was the number of revision surgeries based on postoperative radiological findings like radiographs or CT imaging, for patients treated with the help of 3D-RX. Record was kept for the number of revision surgery procedures within 3 months of follow-up, based on the radiological outcomes.
The judgment results from the three reviewers were averaged and are presented as median values. A Mann–Whitney U test was used to analyze differences between the surgeon’s answer and the median value of the three reviewers for every question (“Appendix A”, Table 4). A two-tailed significance level of p < 0.05 was considered significant.
Based on the literature findings, a 20% difference in revision rates can be expected when 3D-RX imaging is used [8, 14]. This difference was used to calculate a minimum number of 14 patients required to achieve statistical significance: if for three out of 14 patients (21.4%) a revision is avoided, this outcome has a 95% confidence interval of 5–51%, that is., the revision rate is significantly above 0. We included a larger number of patients to be able to increase the precision of the point estimates also regarding other outcomes.
To analyze the agreement among the three reviewers regarding their judgment of the 3D-RX scans, we determined the percentage of agreement (i.e., the accuracy) among the surgeons within one point on the 5-point scale. This was also used to assess the intra-observer agreement. For this purpose, three patients (a hind foot, an ankle and a wrist fracture) were reviewed a second time by all reviewers six weeks later.

Results

A total of 81 operations concerning 81 patients were operated with aid of 3D-RX and included in this study. In total, 97 3D-RX scans were performed, 65 patients were scanned once, 16 patients twice, during different stages of the operation (on average 1.2 scans per patient). The average operation time was 126 min. The mean age of the group was 44 years (ranging from 12 to 86) and comprised 45 males and 36 females. Patient data are shown in Table 1. Excluded 3D-RX scans were those with a poor image quality due to wrong aligning (3), too much metal (3), or inexplicable failure (2).
Table 1
Description of the study population. Numbers between parenthesis indicate the number of patients included in the questionnaire
Extremity
Patients
Scans
Classification
Scans
Foot
52 (24)
66 (31)
Hind foot
26 (12)
Ankle
30 (16)
Forefoot
5 (2)
Arthrodesis
4 (1)
Perioperative imaginga
1a
Wrist
25 (16)
27 (17)
Distal radius
23 (14)
Osteosynthesis carpal/metacarpal fracture
4 (3)
Elbow
3 (1)
3 (1)
Radial head
3 (1)
Shoulder
1 (1)
1 (1)
Scapula
1(1)
Total
81(42)
97(50)
  
aIn one case after removal of osteosynthesis material a 3D-RX scan was made and on the basis of the image no further intervention was done
The comparison between conventional methods and intraoperative 3D-RX scanning for fracture surgery of the extremities showed a significantly lower performance as to screw positioning and rotation of the fracture reduction after 3D-RX (p < 0.002, Table 2). The same trend could be observed for the other aspects interrogated. Certainty about the rotation of the fracture reduction performance was significantly larger (p < 0.05) if based on 3D-RX images (Table 2).
Table 2
Questionnaire results of the 50 scans investigated
 
N
S < R
S > R
Significance (p)
Performance
Positioning K-wires?
14
1
6
0.12
Positioning screws?
34
4
20
0.0004*
Reduction fracture: alignment?
47
11
16
0.09
Reduction fracture: rotation?
36
4
18
0.002*
Joint surface?
48
12
22
0.1
Certainty
Positioning K-wires?
14
6
4
0.6
Positioning screws?
34
14
7
0.09
Reduction fracture: alignment?
47
13
15
0.9
Reduction fracture: rotation?
36
17
10
0.05*
Joint surface?
48
17
19
0.9
N is the total number of times the question was answered or applicable. S is the intraoperative surgeons score, R the median of the reviewers. S > R concerning the performance means that the reviewers are less satisfied with the performance of the surgeon before the 3D-RX scan than her/himself. S > R concerning the certainty question means that reviewers are less certain about the diagnosis than the surgeon before the 3D-RX scan about his/her performance
Values in * represents a significant difference
Inter-observer accuracy was good (85 ± 2%) for a difference of up to one point on the 5-point scale, while 40 ± 4% was scored unanimously. Intra-observer accuracy was excellent (94 ± 3%) for a difference of up to one point between the first and second judgment, while 62% ± 3% was scored unanimously.
None of the 81 patients included in this study needed revision based on the postoperative radiological examinations. An example of 3D-RX imaging versus 2D fluoroscopy is shown in Fig. 2.

Discussion

Fracture surgery of the extremities with conventional examination methods frequently misses out suboptimal positioning of implants and joint incongruities during the operation. These problems remain unrevealed until postoperative radiography or CT imaging is performed [2, 4, 1416, 21].
This study was performed with a new device for intraoperative 3D-RX imaging. The main differences between a longer known X-ray device, the Arcadis Orbic 3D (Siemens Healthcare, Erlangen, Germany) are shown in Table 3. The 3D movement of the Arcadis is able to reach more parts of the body. According to technical specifications, the Pulsera should at least provide the comparable image quality.
Table 3
Overview of relevant properties which clearly are different for both systems
Property
BV Pulsera with 3D-RX
Arcadis Orbic 3D
3D movement
Head or extremity
Lateral
Duration of scan
30 s
60 s or 120 s
No. of images/scan
90–225–450
50–100
Kind of X-ray images
Pulsed (10 ms)
Fluoroscopy
II magnifications
6″–9″–12″
6″–9″
X-ray generator power
15 kW
2.3 kW
Limitations of this study include patient selection. Although we prospectively included all patients, the scan movement was not predefined. However, a protocol to scan at predefined moment is cumbersome in patients suffering from a great variety of fractures qualifying for 3D-RX scanning. In addition, 3D-RX scanning may be inconclusive, which occurred in 8%, of the scans in our study. However, these occurred in the first 58 consecutive patients (14%). Improved experience of the radiology technicians and anticipating metal artifacts avoided problems in the remaining scans.
Furthermore, the golden standard as to the effectiveness of fracture surgery is long-term functional outcome. However, examination of CT scans or postoperative 2D radiographs is considered an acceptably predictive intermediate-term outcome measure [3, 6, 9, 11, 19]. Currently, there is no golden standard available during surgery, which hampers clinical evaluation of this technique which is a weakness of our study. Evaluation of the radiological diagnostic value of 3D-RX has been done previously and was considered adequate and similar to CT and radiography for imaging bony structures [1, 8, 10, 12, 14, 15, 20]. We compared 3D-RX with the conventional intraoperative practice, which is known to have shortcomings [2, 4, 12, 14, 21]. Thus, the particular shortcomings could be exposed. In this manner, we proved that intraoperative 3D imaging provides additional information, an import condition to further investigate long-term functional patient outcome.
Since no general classification system exists to rate radiological outcomes of fracture surgery, we designed our own, based on the standard way surgeons and radiologists examine radiographs and/or CT scans. The questionnaire gives a comprehensive judgment about the results of the surgical procedure. The intra- and inter-observer accuracy of this study was good. Nevertheless, the questionnaire should be validated to allow comparison of the results with those from other centers.
This study showed that intraoperative 3D-RX provides extra information to the surgeon, especially regarding the assessment of rotation of fracture reduction and screw placement. These are the two important performance parameters that are hard to appreciate by conventional examination and fluoroscopic inspection. Especially with fluoroscopy, the angle of imaging and over-projection can lead to unreliable results. With intraoperative 3D-RX, images can be viewed from every possible angle without extra fluoroscopy time, although metal artifacts may distort the images. Our results showed that 18% of the screw positionings at least two points less satisfying were rated on the 3D-RX images than expected intraoperatively. If we consider this condition a revision it is between the results of Richter et al. (26%) and Hufner et al. (10.3%) [8, 14]. Also, if we consider inclusion time and number for this study a revision rate of 16% or higher is required in order to be cost-beneficial based on Hufner et al. [8]. Moreover, during the course of this study we experienced that the frequency of use increased and the device is shared with other disciplines as well. Nevertheless, it should be stressed that most intraoperative failures are revised more frequently than when revealed postoperatively.
In this study, the rotation of the reduction was diagnosed with more certainty by the reviewers than the surgeon without the 3D-RX scan. Except for the diagnosis of the alignment of the reduction and the joint surface the reviewers were more certain about their diagnosis than the surgeon. This (un)certainty, however, does not only appreciate the result of surgery, but also the character of the surgeons involved. Moreover, the ‘noisy’ 3D-RX images do not provide such confidence as obtained by CT and radiography.
In this study no revisions were indicated on the basis of postoperative radiological results, which can be attributed to the fact that 3D-RX imaging was already in use. The reported radiology-based failure rate, if treated conventionally, is higher and ranges from 10% in distal radius fractures to over 26% in calcaneus or malleolus tertius fractures [6, 9]. These failures would be, at least in most cases, a strong indication for revision surgery.

Conclusion

Intraoperative 3D-RX clearly provides more information about the result of fracture reduction procedures than conventional 2D-scanning. The reduction in indications for revision surgery after a relatively short follow-up period is promising, but not equivalent to a good functional outcome. This patient-relevant parameter still needs to be investigated after a longer follow-up period which is meaningful as proven by this study. In our hospital, the use of intraoperative 3D-RX imaging has become a standard procedure during surgical treatment of complex joint fractures of the extremities.

Acknowledgments

The authors wish to thank K. J. Ponsen and G. J. Streekstra of the AMC, Amsterdam, and P. van Kemenade of Philips Healthcare, Best, for their contribution to this study.

Conflict of interest statement

The PhD project of B.C. is funded by Philips Healthcare and the Dutch ministry of economical affairs.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Anhänge

Appendix A

Table 4
Questions used for intraoperative assessment of the operation state and review
Question
Range performance
Range certainty
Positioning K-wires?
1. Redo
1. Not
5. Satisfied
5. Sure
Positioning screws?
1. Redo
1. Not
5. Satisfied
5. Sure
Reduction fracture: alignment?
1. Redo
1. Not
5. Satisfied
5. Sure
Reduction fracture: rotation?
1. Redo
1. Not
5. Satisfied
5. Sure
Joint surface?
1. Redo
1. Not
5. Satisfied
5. Sure
Literatur
1.
Zurück zum Zitat Carelsen B, Bakker NH, Boon SN, Fokkens WJ, Freling NJ, Noordhoek N (2004) Mobile 3D rotational X-ray: comparison with CT in sinus surgery. Medica Mundi 48:4–9 Carelsen B, Bakker NH, Boon SN, Fokkens WJ, Freling NJ, Noordhoek N (2004) Mobile 3D rotational X-ray: comparison with CT in sinus surgery. Medica Mundi 48:4–9
2.
Zurück zum Zitat Catalano LW III, Barron OA, Glickel SZ (2004) Assessment of articular displacement of distal radius fractures. Clin Orthop Relat Res 423:79–84PubMedCrossRef Catalano LW III, Barron OA, Glickel SZ (2004) Assessment of articular displacement of distal radius fractures. Clin Orthop Relat Res 423:79–84PubMedCrossRef
3.
Zurück zum Zitat Chen SH, Wu PH, Lee YS (2007) Long-term results of pilon fractures. Arch Orthop Trauma Surg 127:55–60PubMedCrossRef Chen SH, Wu PH, Lee YS (2007) Long-term results of pilon fractures. Arch Orthop Trauma Surg 127:55–60PubMedCrossRef
4.
Zurück zum Zitat Edwards CC II, Haraszti CJ, McGillivary GR, Gutow AP (2001) Intra-articular distal radius fractures: arthroscopic assessment of radiographically assisted reduction. J Hand Surg 26:1036–1041CrossRef Edwards CC II, Haraszti CJ, McGillivary GR, Gutow AP (2001) Intra-articular distal radius fractures: arthroscopic assessment of radiographically assisted reduction. J Hand Surg 26:1036–1041CrossRef
5.
Zurück zum Zitat Euler E, Wirth S, Linsenmaier U, Mutschler W, Pfeifer KJ, Hebecker A (2001) Comparative study of the quality of C-arm based 3D imaging of the talus. Unfallchirurg 104:839–846PubMedCrossRef Euler E, Wirth S, Linsenmaier U, Mutschler W, Pfeifer KJ, Hebecker A (2001) Comparative study of the quality of C-arm based 3D imaging of the talus. Unfallchirurg 104:839–846PubMedCrossRef
6.
Zurück zum Zitat Handoll HH, Madhok R (2003) Surgical interventions for treating distal radial fractures in adults. Cochrane database of systematic reviews (Online).CD003209 Handoll HH, Madhok R (2003) Surgical interventions for treating distal radial fractures in adults. Cochrane database of systematic reviews (Online).CD003209
7.
Zurück zum Zitat Huda W, Gkanatsios NA (1998) Radiation dosimetry for extremity radiographs. Health Phys 75:492–499PubMed Huda W, Gkanatsios NA (1998) Radiation dosimetry for extremity radiographs. Health Phys 75:492–499PubMed
8.
Zurück zum Zitat Hufner T, Stubig T, Gosling T, Kendoff D, Geerling J, Krettek C (2007) Cost–benefit analysis of intraoperative 3D imaging. Unfallchirurg 110:114–121CrossRef Hufner T, Stubig T, Gosling T, Kendoff D, Geerling J, Krettek C (2007) Cost–benefit analysis of intraoperative 3D imaging. Unfallchirurg 110:114–121CrossRef
9.
Zurück zum Zitat Kapoor H, Agarwal A, Dhaon BK (2000) Displaced intra-articular fractures of distal radius: a comparative evaluation of results following closed reduction, external fixation and open reduction with internal fixation. Injury 31:75–79PubMedCrossRef Kapoor H, Agarwal A, Dhaon BK (2000) Displaced intra-articular fractures of distal radius: a comparative evaluation of results following closed reduction, external fixation and open reduction with internal fixation. Injury 31:75–79PubMedCrossRef
10.
Zurück zum Zitat Kotsianos D, Wirth S, Fischer T, Euler E, Rock C, Linsenmaier U, Pfeifer KJ, Reiser M (2004) 3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT. Eur Radiol 14:1590–1595PubMedCrossRef Kotsianos D, Wirth S, Fischer T, Euler E, Rock C, Linsenmaier U, Pfeifer KJ, Reiser M (2004) 3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT. Eur Radiol 14:1590–1595PubMedCrossRef
11.
Zurück zum Zitat Langenhuijsen JF, Heetveld MJ, Ultee JM, Steller EP, Butzelaar RM (2002) Results of ankle fractures with involvement of the posterior tibial margin. J Trauma 53:55–60PubMedCrossRef Langenhuijsen JF, Heetveld MJ, Ultee JM, Steller EP, Butzelaar RM (2002) Results of ankle fractures with involvement of the posterior tibial margin. J Trauma 53:55–60PubMedCrossRef
12.
Zurück zum Zitat Meier R, Kfuri M Jr, Geerling J, Hufner T, Krimmer H, Krettek C (2005) Intraoperative three-dimensional imaging with an isocentric mobile C-arm at the wrist. Handchir Mikrochir Plast Chir 37:256–259PubMedCrossRef Meier R, Kfuri M Jr, Geerling J, Hufner T, Krimmer H, Krettek C (2005) Intraoperative three-dimensional imaging with an isocentric mobile C-arm at the wrist. Handchir Mikrochir Plast Chir 37:256–259PubMedCrossRef
14.
Zurück zum Zitat Richter M, Geerling J, Zech S, Goesling T, Krettek C (2005) Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C–3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma 19:259–266PubMedCrossRef Richter M, Geerling J, Zech S, Goesling T, Krettek C (2005) Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C–3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma 19:259–266PubMedCrossRef
15.
Zurück zum Zitat Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg 109:112–118PubMedCrossRef Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg 109:112–118PubMedCrossRef
16.
Zurück zum Zitat Ruch DS, Vallee J, Poehling GG, Smith BP, Kuzma GR (2004) Arthroscopic reduction versus fluoroscopic reduction in the management of intra-articular distal radius fractures. Arthroscopy 20:225–230PubMedCrossRef Ruch DS, Vallee J, Poehling GG, Smith BP, Kuzma GR (2004) Arthroscopic reduction versus fluoroscopic reduction in the management of intra-articular distal radius fractures. Arthroscopy 20:225–230PubMedCrossRef
17.
Zurück zum Zitat Sanders R, Fortin P, DiPasquale T, Walling A (1993) Operative treatment in 120 displaced intraarticular calcaneal fractures. Results using a prognostic computed tomography scan classification. Clinical Orthop Relat Res 290:87–95 Sanders R, Fortin P, DiPasquale T, Walling A (1993) Operative treatment in 120 displaced intraarticular calcaneal fractures. Results using a prognostic computed tomography scan classification. Clinical Orthop Relat Res 290:87–95
18.
Zurück zum Zitat Schep NW, van Vugt AB (2006) Navigation surgery and fracture treatment. Nederlands tijdschrift voor geneeskunde 150:2301–2306PubMed Schep NW, van Vugt AB (2006) Navigation surgery and fracture treatment. Nederlands tijdschrift voor geneeskunde 150:2301–2306PubMed
19.
Zurück zum Zitat Stulik J, Stehlik J, Rysavy M, Wozniak A (2006) Minimally-invasive treatment of intra-articular fractures of the calcaneum. J Bone Joint Surg Br 88:1634–1641PubMedCrossRef Stulik J, Stehlik J, Rysavy M, Wozniak A (2006) Minimally-invasive treatment of intra-articular fractures of the calcaneum. J Bone Joint Surg Br 88:1634–1641PubMedCrossRef
20.
Zurück zum Zitat Wich M, Spranger N, Ekkernkamp A (2004) Intraoperative imaging with the ISO C(3D). Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen 75:982–987PubMed Wich M, Spranger N, Ekkernkamp A (2004) Intraoperative imaging with the ISO C(3D). Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen 75:982–987PubMed
21.
Zurück zum Zitat Zwipp H, Dahlen C, Randt T, Gavlik JM (1997) Complex trauma of the foot. Der Orthopade 26:1046–1056PubMed Zwipp H, Dahlen C, Randt T, Gavlik JM (1997) Complex trauma of the foot. Der Orthopade 26:1046–1056PubMed
Metadaten
Titel
Does intraoperative fluoroscopic 3D imaging provide extra information for fracture surgery?
verfasst von
B. Carelsen
R. Haverlag
D. Th. Ubbink
J. S. K. Luitse
J. C. Goslings
Publikationsdatum
01.12.2008
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 12/2008
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-008-0740-5

Weitere Artikel der Ausgabe 12/2008

Archives of Orthopaedic and Trauma Surgery 12/2008 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.