01.01.2010 | Orthopaedic Surgery

Quadriceps force in relation of intrinsic anteroposterior stability of TKA design

verfasst von: Thomas Jan Heyse, Christoph Becher, Nadine Kron, Sven Ostermeier, Christof Hurschler, Markus D. Schofer, Susanne Fuchs-Winkelmann, Carsten O. Tibesku

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 1/2010

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Decreased quadriceps strength and fatigue is suspected to be one of the contributing factors for anterior knee pain and malfunction after total knee arthroplasty (TKA). The purpose of this in vitro study was to investigate the amount of quadriceps force required to extend the knee isokinetically after TKA in dependence of different prosthesis designs and the state of the posterior cruciate ligament (PCL).

Materials and methods

Eight fresh frozen human knee specimens underwent testing in a kinematic device simulating an isokinetic knee extension cycle from 120° of flexion to full extension. The quadriceps force was measured after implantation of a cruciate retaining (CR) TKA (Genesis II, Smith&Nephew, Memphis, TN, USA) applying a conventional CR (11 mm) and a highly conforming (deep dished, DD) polyethylene (PE) inlay consecutively before and after resection of the PCL. Finally, tests were repeated with a posterior-stabilized (PS) design.

Results

Simulating a physiological knee extension, no significant differences in the average quadriceps force were detected between the cruciate preserving inlays (CR 1,146.57 ± 88.04 N, DD 1,150.19 ± 97.54 N, P = 0.86) as long as the PCL was intact. After resection of the PCL, the required quadriceps force increased significantly for both designs (CR 1,203.17 ± 91.51 N, P < 0.01 and DD 1,191.88 ± 80.07 N, P < 0.03). After implantation of the posterior stabilized femoral component quad force decreased to its initial levels with forces significantly lower compared to the PCL deficient knees provided with a CR or DD (PS 1,130.91 ± 107.88 N, P < 0.01) inlay. With a deficient PCL there were no statistical differences for the DD design in comparison with CR in mean quad forces (CR 1,203.17 ± 91.51 N vs. DD 1,191.88 ± 80.07 N, P = 0.50) nor in peak forces (CR 1,729.44 ± 161.86 N, DD 1,688.66 ± 123.18 N, P = 0.17).

Discussion

At intact PCL peak quad forces and mean forces beyond 70° of flexion could be shown to be significantly lower with a PS TKA design in comparison with cruciate preserving designs such as CR and DD. In the PCL deficient knee quad forces with a highly conforming implant (DD) and CR were significantly higher than with a PS TKA. The use of PS implants in all PCL deficient knees seems to be advisable.
Literatur
1.
Zurück zum Zitat Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg Am 64(9):1328–1335PubMed Andriacchi TP, Galante JO, Fermier RW (1982) The influence of total knee-replacement design on walking and stair-climbing. J Bone Joint Surg Am 64(9):1328–1335PubMed
2.
Zurück zum Zitat Rand JA (1994) The patellofemoral joint in total knee arthroplasty. J Bone Joint Surg Am 76(4):612–620PubMed Rand JA (1994) The patellofemoral joint in total knee arthroplasty. J Bone Joint Surg Am 76(4):612–620PubMed
4.
Zurück zum Zitat Boyd AD Jr, Ewald FC, Thomas WH, Poss R, Sledge CB (1993) Long-term complications after total knee arthroplasty with or without resurfacing of the patella. J Bone Joint Surg Am 75(5):674–681PubMed Boyd AD Jr, Ewald FC, Thomas WH, Poss R, Sledge CB (1993) Long-term complications after total knee arthroplasty with or without resurfacing of the patella. J Bone Joint Surg Am 75(5):674–681PubMed
6.
Zurück zum Zitat Kaufer H (1971) Mechanical function of the patella. J Bone Joint Surg Am 53(8):1551–1560PubMed Kaufer H (1971) Mechanical function of the patella. J Bone Joint Surg Am 53(8):1551–1560PubMed
7.
Zurück zum Zitat Dennis DA, Komistek RD, Mahfouz MR, Haas BD, Stiehl JB (2003) Multicenter determination of in vivo kinematics after total knee arthroplasty. Clin Orthop Relat Res Nov(416):37–57 Dennis DA, Komistek RD, Mahfouz MR, Haas BD, Stiehl JB (2003) Multicenter determination of in vivo kinematics after total knee arthroplasty. Clin Orthop Relat Res Nov(416):37–57
8.
Zurück zum Zitat Dennis DA, Komistek RD, Colwell CE, Jr., Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop Relat Res Nov(356):47–57 Dennis DA, Komistek RD, Colwell CE, Jr., Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop Relat Res Nov(356):47–57
9.
Zurück zum Zitat Berman AT, Bosacco SJ, Israelite C (1991) Evaluation of total knee arthroplasty using isokinetic testing. Clin Orthop Relat Res Oct(271):106–113 Berman AT, Bosacco SJ, Israelite C (1991) Evaluation of total knee arthroplasty using isokinetic testing. Clin Orthop Relat Res Oct(271):106–113
10.
Zurück zum Zitat Ostermeier S, Hurschler C, Stukenborg-Colsman C (2004) Quadriceps function after TKA—an in vitro study in a knee kinematic simulator. Clin Biomech (Bristol, Avon) 19(3):270–276CrossRef Ostermeier S, Hurschler C, Stukenborg-Colsman C (2004) Quadriceps function after TKA—an in vitro study in a knee kinematic simulator. Clin Biomech (Bristol, Avon) 19(3):270–276CrossRef
11.
Zurück zum Zitat Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty: comparison of fixed and mobile-bearing inlay designs. Acta Orthop Scand 73(6):638–646CrossRefPubMed Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty: comparison of fixed and mobile-bearing inlay designs. Acta Orthop Scand 73(6):638–646CrossRefPubMed
12.
Zurück zum Zitat Stukenborg-Colsman C, Ostermeier S, Wenger KH, Wirth CJ (2002) Relative motion of a mobile bearing inlay after total knee arthroplasty—dynamic in vitro study. Clin Biomech (Bristol, Avon) 17(1):49–55CrossRef Stukenborg-Colsman C, Ostermeier S, Wenger KH, Wirth CJ (2002) Relative motion of a mobile bearing inlay after total knee arthroplasty—dynamic in vitro study. Clin Biomech (Bristol, Avon) 17(1):49–55CrossRef
13.
Zurück zum Zitat Laskin RS, Maruyama Y, Villaneuva M, Bourne R (2000) Deep-dish congruent tibial component use in total knee arthroplasty: a randomized prospective study. Clin Orthop Relat Res Nov(380):36–44 Laskin RS, Maruyama Y, Villaneuva M, Bourne R (2000) Deep-dish congruent tibial component use in total knee arthroplasty: a randomized prospective study. Clin Orthop Relat Res Nov(380):36–44
14.
Zurück zum Zitat Fuchs S, Tibesku CO, Jerosch J (1998) Isokinetic measurements in total knee arthroplasty patients compared to healthy volunteers. In: 8th Congress of the European Society of Sports Traumatology, Knee Surgery and Arthroscopy, 1998, Nice, France Fuchs S, Tibesku CO, Jerosch J (1998) Isokinetic measurements in total knee arthroplasty patients compared to healthy volunteers. In: 8th Congress of the European Society of Sports Traumatology, Knee Surgery and Arthroscopy, 1998, Nice, France
16.
Zurück zum Zitat Ostermeier S, Friesecke C, Fricke S, Hurschler C, Stukenborg-Colsman C (2008) Quadriceps force during knee extension after non-hinged and hinged TKA: An in vitro study. Acta Orthop 79(1):34–38CrossRefPubMed Ostermeier S, Friesecke C, Fricke S, Hurschler C, Stukenborg-Colsman C (2008) Quadriceps force during knee extension after non-hinged and hinged TKA: An in vitro study. Acta Orthop 79(1):34–38CrossRefPubMed
17.
Zurück zum Zitat Ostermeier S, Hurschler C, Windhagen H, Stukenborg-Colsman C (2006) In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 14(10):934–939CrossRefPubMed Ostermeier S, Hurschler C, Windhagen H, Stukenborg-Colsman C (2006) In vitro investigation of the influence of tibial slope on quadriceps extension force after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 14(10):934–939CrossRefPubMed
18.
Zurück zum Zitat Nisell R (1985) Mechanics of the knee. A study of joint and muscle load with clinical applications. Acta Orthop Scand Suppl 216:1–42PubMed Nisell R (1985) Mechanics of the knee. A study of joint and muscle load with clinical applications. Acta Orthop Scand Suppl 216:1–42PubMed
19.
Zurück zum Zitat Hollister AM, Jatana S, Singh AK, Sullivan WW, Lupichuk AG (1993) The axes of rotation of the knee. Clin Orthop Relat Res May(290):259–268 Hollister AM, Jatana S, Singh AK, Sullivan WW, Lupichuk AG (1993) The axes of rotation of the knee. Clin Orthop Relat Res May(290):259–268
20.
Zurück zum Zitat Komistek RD, Mahfouz MR, Bertin KC, Rosenberg A, Kennedy W (2008) In vivo determination of total knee arthroplasty kinematics: a multicenter analysis of an asymmetrical posterior cruciate retaining total knee arthroplasty. J Arthroplasty 23(1):41–50CrossRefPubMed Komistek RD, Mahfouz MR, Bertin KC, Rosenberg A, Kennedy W (2008) In vivo determination of total knee arthroplasty kinematics: a multicenter analysis of an asymmetrical posterior cruciate retaining total knee arthroplasty. J Arthroplasty 23(1):41–50CrossRefPubMed
21.
Zurück zum Zitat Insall JN, Lachiewicz PF, Burstein AH (1982) The posterior stabilized condylar prosthesis: a modification of the total condylar design. Two to four-year clinical experience. J Bone Joint Surg Am 64(9):1317–1323PubMed Insall JN, Lachiewicz PF, Burstein AH (1982) The posterior stabilized condylar prosthesis: a modification of the total condylar design. Two to four-year clinical experience. J Bone Joint Surg Am 64(9):1317–1323PubMed
22.
Zurück zum Zitat Dennis DA, Komistek RD, Mahfouz MR (2003) In vivo fluoroscopic analysis of fixed-bearing total knee replacements. Clin Orthop Relat Res May(410):114–130 Dennis DA, Komistek RD, Mahfouz MR (2003) In vivo fluoroscopic analysis of fixed-bearing total knee replacements. Clin Orthop Relat Res May(410):114–130
23.
Zurück zum Zitat Banks SA, Hodge WA (2004) Implant design affects knee arthroplasty kinematics during stair-stepping. Clin Orthop Relat Res Sep(426):187–193 Banks SA, Hodge WA (2004) Implant design affects knee arthroplasty kinematics during stair-stepping. Clin Orthop Relat Res Sep(426):187–193
24.
Zurück zum Zitat Scott RD, Thornhill TS (1994) Posterior cruciate supplementing total knee replacement using conforming inserts and cruciate recession. Effect on range of motion and radiolucent lines. Clin Orthop Relat Res Dec(309):146–149 Scott RD, Thornhill TS (1994) Posterior cruciate supplementing total knee replacement using conforming inserts and cruciate recession. Effect on range of motion and radiolucent lines. Clin Orthop Relat Res Dec(309):146–149
Metadaten
Titel
Quadriceps force in relation of intrinsic anteroposterior stability of TKA design
verfasst von
Thomas Jan Heyse
Christoph Becher
Nadine Kron
Sven Ostermeier
Christof Hurschler
Markus D. Schofer
Susanne Fuchs-Winkelmann
Carsten O. Tibesku
Publikationsdatum
01.01.2010
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 1/2010
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-009-0927-4

Weitere Artikel der Ausgabe 1/2010

Archives of Orthopaedic and Trauma Surgery 1/2010 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.