Accuracy of minimally invasive navigated acetabular and iliosacral fracture stabilization using a targeting and noninvasive registration device

verfasst von: Ralf E. Rosenberger, B. Dolati, R. Larndorfer, M. Blauth, D. Krappinger, Reto J. Bale

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 2/2010

Einloggen, um Zugang zu erhalten

Abstract

Background

To assess the feasibility and accuracy of guide pin (GP) placement using a combined noninvasive patient immobilization and stereotactic targeting system in computer-assisted percutaneous pelvic fracture stabilization.

Methods

A total of 12 patients with negligible dislocated unstable pelvic fractures were enrolled in this study, performed between February 2002 and October 2005. Our original plans included 13 GP placements in the iliosacral area (SF) and 8 in the acetabular (AF) area. Patients were bedded on a noninvasive dual-vacuum immobilization device. Interventions were planned on a navigation system using intraoperatively acquired CT data. Radiodense markers glued to the skin and the immobilization device provided synchronization between virtual data set and real anatomical situation. A stereotactic targeting device was used for stabilization of GP tracking. GP positions were verified intraoperatively by CT, followed by fracture stabilization with cannulated screws.

Results

Mean GP placement accuracy according to plan: (1) SF-cohort: 2.8 mm (SD 2.0 mm, range 0.5–9.0 mm) at the bony entry point and 3.8 mm (SD 2.3 mm, range 0.6–9.5 mm) at the target point. (2) AF-cohort: 3.0 mm (SD 0.9 mm, range 1.6–4.9 mm) at the bony entry point and 3.9 mm (SD 1.9 mm, range 1.6–7.5 mm) at the target point. GP placement succeeded optimally in 11 out of 13 cases in the SF-cohort, and 6 out of 8 cases in the AF-cohort. The individual average dose–length product (DLP) per successful finished procedure was 1,576 mGy × cm (SD 812 mGy × cm, range 561–2,739 mGy × cm).

Conclusion

Our findings substantiate application of the noninvasive patient immobilization and stereotactic targeting system as effective in computer-assited percutaneous stabilization of sacral bone fractures/SI joint disruptions and coronally oriented acetabular dome fractures. We recommend according to the ALARA (as low as reasonable achievable) principle: first, the kV and mAs values have to be reduced. Second, the scanned volume has to be strictly limited to the area of interest. Third, the number of control CTs have to be minimized. Also, the IsoC might be a better choice for implant tracking below 12 cm to reduce the radiation dose to the minimum. We believe that for all high-precise GP placements in the acetabular column area, further improvements in GP guidance (inhibiting pin tip slipping and detecting intraosseous GP deflection) are necessary.
Literatur
20.
Zurück zum Zitat Bale RJ et al (2001) Osteochondral lesions of the talus: computer-assisted retrograde drilling: feasibility and accuracy in initial experiences. Radiology 218(1):278–282PubMed Bale RJ et al (2001) Osteochondral lesions of the talus: computer-assisted retrograde drilling: feasibility and accuracy in initial experiences. Radiology 218(1):278–282PubMed
21.
Zurück zum Zitat Adams L, van den Brug WP, Vogele M, Bale R. Alignment device (EasyTaxis) European Patent PHN 16.013 Adams L, van den Brug WP, Vogele M, Bale R. Alignment device (EasyTaxis) European Patent PHN 16.013
22.
Zurück zum Zitat Bale RJ et al (2002) A novel vacuum device for extremity immobilisation during digital angiography: preliminary clinical experiences. Eur Radiol 12(12):2890–2894PubMed Bale RJ et al (2002) A novel vacuum device for extremity immobilisation during digital angiography: preliminary clinical experiences. Eur Radiol 12(12):2890–2894PubMed
23.
Zurück zum Zitat Fuss M et al (2004) Repositioning accuracy of a commercially available double-vacuum whole body immobilization system for stereotactic body radiation therapy. Technol Cancer Res Treat 3(1):59–67PubMed Fuss M et al (2004) Repositioning accuracy of a commercially available double-vacuum whole body immobilization system for stereotactic body radiation therapy. Technol Cancer Res Treat 3(1):59–67PubMed
26.
Zurück zum Zitat Day CS et al (2000) Transsacral versus modified pelvic landmarks for percutaneous iliosacral screw placement: a computed tomographic analysis and cadaveric study. Am J Orthop 29(Suppl 9):16–21 Day CS et al (2000) Transsacral versus modified pelvic landmarks for percutaneous iliosacral screw placement: a computed tomographic analysis and cadaveric study. Am J Orthop 29(Suppl 9):16–21
31.
Zurück zum Zitat Wieners G et al (2005) Comparison of radiation dose and image quality of Siremobil-IsoC(3D) with a 16-slice spiral CT for diagnosis and intervention in the human pelvic bone. Rofo 177(2):258–264PubMed Wieners G et al (2005) Comparison of radiation dose and image quality of Siremobil-IsoC(3D) with a 16-slice spiral CT for diagnosis and intervention in the human pelvic bone. Rofo 177(2):258–264PubMed
Metadaten
Titel
Accuracy of minimally invasive navigated acetabular and iliosacral fracture stabilization using a targeting and noninvasive registration device
verfasst von
Ralf E. Rosenberger
B. Dolati
R. Larndorfer
M. Blauth
D. Krappinger
Reto J. Bale
Publikationsdatum
01.02.2010
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 2/2010
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-009-0932-7

Weitere Artikel der Ausgabe 2/2010

Archives of Orthopaedic and Trauma Surgery 2/2010 Zur Ausgabe

Orthopaedic Surgery

Total femur replacement

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.