Cobalt serum levels differ in well functioning Birmingham resurfacing and Birmingham modular THA

verfasst von: Lisa Renner, Martin Faschingbauer, Tom Schmidt-Braekling, Friedrich Boettner

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Metal-on-metal (MoM) bearings are known to release metal ions secondary to wear and corrosion. This may cause local reactions (adverse soft tissue reactions and osteolysis) and systemic effects. Little is known about the exact pattern and the differences between large head MoM total hip replacements (THA) and resurfacings (HR).

Questions

(1) Is there a difference in metal ion concentrations between HR and MoM-THR using the same bearing design (Birmingham Hip Resurfacing System, Smith & Nephew, Inc. Memphis, TN, USA)? (2) Are metal ion levels changing over time in MoM-THA or HR? (3) Do acetabular inclination angle and femoral component size influence cobalt and chromium levels? Is there a correlation between clinical outcome and metal ion levels?

Materials and methods

A retrospective analysis was conducted in 77 well functioning unilateral Birmingham HR and 42 well functioning unilateral modular Birmingham MoM-THA (Smith & Nephew, Inc. Memphis, TN, USA) operated on between 2007 and 2012. Blood samples were taken at a minimum of 13 months and subsequent during annual follow-ups.

Results

(1) Cobalt levels were significantly higher in MoM-THA compared to HR (p < 0.001). There was no significant difference in chromium levels (p = 0.313). (2) Cobalt is increasing over time in MoM-THA (p = 0.030) whereas metal ions remain stable in HR. (3) Metal ion levels were not affected by acetabular inclination angle and femoral component size in MoM-THA. Chromium levels correlate with the femoral component size (r = −0.240; p = 0.037), the UCLA activity score (r = −0.344; p = 0.003) and the VAS (r = 0.263; p = 0.38) in HR.

Conclusion

Considering that HR and MoM-THA used the same MoM bearing design, increased cobalt levels may be related to trunnion wear or corrosion. Elevated cobalt levels should raise concern for corrosion related failure in MoM-THA.
Literatur
1.
Zurück zum Zitat Sadoghi P et al (2013) Revision surgery after total joint arthroplasty: a complication-based analysis using worldwide arthroplasty registers. J Arthroplasty 28(8):1329–1332CrossRefPubMed Sadoghi P et al (2013) Revision surgery after total joint arthroplasty: a complication-based analysis using worldwide arthroplasty registers. J Arthroplasty 28(8):1329–1332CrossRefPubMed
2.
Zurück zum Zitat Langton DJ et al (2011) Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br 93(2):164–171CrossRefPubMed Langton DJ et al (2011) Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br 93(2):164–171CrossRefPubMed
3.
Zurück zum Zitat Carr AM, DeSteiger R (2008) Osteolysis in patients with a metal-on-metal hip arthroplasty. ANZ J Surg 78(3):144–147CrossRefPubMed Carr AM, DeSteiger R (2008) Osteolysis in patients with a metal-on-metal hip arthroplasty. ANZ J Surg 78(3):144–147CrossRefPubMed
4.
Zurück zum Zitat Bisschop R et al (2013) High prevalence of pseudotumors in patients with a Birmingham hip resurfacing prosthesis: a prospective cohort study of one hundred and twenty-nine patients. J Bone Joint Surg Am 95(17):1554–1560CrossRefPubMed Bisschop R et al (2013) High prevalence of pseudotumors in patients with a Birmingham hip resurfacing prosthesis: a prospective cohort study of one hundred and twenty-nine patients. J Bone Joint Surg Am 95(17):1554–1560CrossRefPubMed
5.
Zurück zum Zitat Bosker BH et al (2012) High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br 94(6):755–761CrossRefPubMed Bosker BH et al (2012) High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement: a prospective cohort study. J Bone Joint Surg Br 94(6):755–761CrossRefPubMed
6.
Zurück zum Zitat Huk OL et al (2004) Induction of apoptosis and necrosis by metal ions in vitro. J Arthroplasty 19(8 Suppl 3):84–87CrossRefPubMed Huk OL et al (2004) Induction of apoptosis and necrosis by metal ions in vitro. J Arthroplasty 19(8 Suppl 3):84–87CrossRefPubMed
7.
Zurück zum Zitat Ladon D et al (2004) Changes in metal levels and chromosome aberrations in the peripheral blood of patients after metal-on-metal hip arthroplasty. J Arthroplasty 19(8 Suppl 3):78–83CrossRefPubMed Ladon D et al (2004) Changes in metal levels and chromosome aberrations in the peripheral blood of patients after metal-on-metal hip arthroplasty. J Arthroplasty 19(8 Suppl 3):78–83CrossRefPubMed
8.
Zurück zum Zitat Hannemann F et al (2013) European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 99(3):263–271CrossRefPubMed Hannemann F et al (2013) European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 99(3):263–271CrossRefPubMed
9.
Zurück zum Zitat Lombardi AV et al (2012) The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br 94(11 Suppl A):14–18CrossRefPubMed Lombardi AV et al (2012) The Hip Society: algorithmic approach to diagnosis and management of metal-on-metal arthroplasty. J Bone Joint Surg Br 94(11 Suppl A):14–18CrossRefPubMed
10.
Zurück zum Zitat Hart AJ et al (2011) Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br 93(10):1308–1313CrossRefPubMed Hart AJ et al (2011) Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br 93(10):1308–1313CrossRefPubMed
11.
Zurück zum Zitat Van Der Straeten C et al (2013) The 2012 Otto Aufranc Award: the interpretation of metal ion levels in unilateral and bilateral hip resurfacing. Clin Orthop Relat Res 471(2):377–385CrossRef Van Der Straeten C et al (2013) The 2012 Otto Aufranc Award: the interpretation of metal ion levels in unilateral and bilateral hip resurfacing. Clin Orthop Relat Res 471(2):377–385CrossRef
12.
Zurück zum Zitat Heisel C et al (2008) Characterization of the running-in period in total hip resurfacing arthroplasty: an in vivo and in vitro metal ion analysis. J Bone Joint Surg Am 90(Suppl 3):125–133CrossRefPubMed Heisel C et al (2008) Characterization of the running-in period in total hip resurfacing arthroplasty: an in vivo and in vitro metal ion analysis. J Bone Joint Surg Am 90(Suppl 3):125–133CrossRefPubMed
13.
14.
Zurück zum Zitat Hart AJ et al (2014) Surveillance of patients with metal-on-metal hip resurfacing and total hip prostheses: a prospective cohort study to investigate the relationship between blood metal ion levels and implant failure. J Bone Joint Surg Am 96(13):1091–1099CrossRefPubMed Hart AJ et al (2014) Surveillance of patients with metal-on-metal hip resurfacing and total hip prostheses: a prospective cohort study to investigate the relationship between blood metal ion levels and implant failure. J Bone Joint Surg Am 96(13):1091–1099CrossRefPubMed
15.
16.
Zurück zum Zitat Lee R, Essner A, Wang A (2008) Tribological considerations in primary and revision metal-on-metal arthroplasty. J Bone Joint Surg Am 90(Suppl 3):118–124CrossRefPubMed Lee R, Essner A, Wang A (2008) Tribological considerations in primary and revision metal-on-metal arthroplasty. J Bone Joint Surg Am 90(Suppl 3):118–124CrossRefPubMed
17.
Zurück zum Zitat Matharu GS et al (2015) Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients. Int Orthop 39(9):1803–1811CrossRefPubMed Matharu GS et al (2015) Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients. Int Orthop 39(9):1803–1811CrossRefPubMed
18.
Zurück zum Zitat Coulter G et al (2012) Birmingham hip resurfacing at a mean of ten years: results from an independent centre. J Bone Joint Surg Br 94(3):315–321CrossRefPubMed Coulter G et al (2012) Birmingham hip resurfacing at a mean of ten years: results from an independent centre. J Bone Joint Surg Br 94(3):315–321CrossRefPubMed
19.
Zurück zum Zitat Daniel J et al (2014) Results of Birmingham hip resurfacing at 12 to 15 years: a single-surgeon series. Bone Joint J 96-B(10):1298–1306CrossRefPubMed Daniel J et al (2014) Results of Birmingham hip resurfacing at 12 to 15 years: a single-surgeon series. Bone Joint J 96-B(10):1298–1306CrossRefPubMed
20.
Zurück zum Zitat Hosny HA et al (2013) Midterm results with Birmingham hip resurfacing/synergy stem modular metal-on-metal total hip arthroplasty. Acta Orthop Belg 79(4):386–391PubMed Hosny HA et al (2013) Midterm results with Birmingham hip resurfacing/synergy stem modular metal-on-metal total hip arthroplasty. Acta Orthop Belg 79(4):386–391PubMed
21.
Zurück zum Zitat Liddle AD et al (2013) Revision of metal-on-metal hip arthroplasty in a tertiary center: a prospective study of 39 hips with between 1 and 4 years of follow-up. Acta Orthop 84(3):237–245CrossRefPubMedPubMedCentral Liddle AD et al (2013) Revision of metal-on-metal hip arthroplasty in a tertiary center: a prospective study of 39 hips with between 1 and 4 years of follow-up. Acta Orthop 84(3):237–245CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Savarino L et al (2013) Do ion levels in metal-on-metal hip resurfacing differ from those in metal-on-metal THA at long-term followup? Clin Orthop Relat Res 471(9):2964–2971CrossRefPubMedPubMedCentral Savarino L et al (2013) Do ion levels in metal-on-metal hip resurfacing differ from those in metal-on-metal THA at long-term followup? Clin Orthop Relat Res 471(9):2964–2971CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Jantzen C et al (2013) Chromium and cobalt ion concentrations in blood and serum following various types of metal-on-metal hip arthroplasties: a literature overview. Acta Orthop 84(3):229–236CrossRefPubMedPubMedCentral Jantzen C et al (2013) Chromium and cobalt ion concentrations in blood and serum following various types of metal-on-metal hip arthroplasties: a literature overview. Acta Orthop 84(3):229–236CrossRefPubMedPubMedCentral
25.
Zurück zum Zitat Kuzyk PR et al (2011) Hip resurfacing versus metal-on-metal total hip arthroplasty—are metal ion levels different? Bull NYU Hosp Jt Dis 69(Suppl 1):S5–S11PubMed Kuzyk PR et al (2011) Hip resurfacing versus metal-on-metal total hip arthroplasty—are metal ion levels different? Bull NYU Hosp Jt Dis 69(Suppl 1):S5–S11PubMed
26.
Zurück zum Zitat Garbuz DS et al (2010) The John Charnley Award: metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 468(2):318–325CrossRefPubMedPubMedCentral Garbuz DS et al (2010) The John Charnley Award: metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 468(2):318–325CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Langton DJ et al (2011) Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br 93(8):1011–1016CrossRefPubMed Langton DJ et al (2011) Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br 93(8):1011–1016CrossRefPubMed
30.
Zurück zum Zitat Cooper HJ et al (2012) Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 94(18):1655–1661PubMed Cooper HJ et al (2012) Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 94(18):1655–1661PubMed
31.
Zurück zum Zitat Kop AM, Swarts E (2009) Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty 24(7):1019–1023CrossRefPubMed Kop AM, Swarts E (2009) Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty 24(7):1019–1023CrossRefPubMed
32.
Zurück zum Zitat Jacobs JJ et al (1998) Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg Am 80(10):1447–1458PubMed Jacobs JJ et al (1998) Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg Am 80(10):1447–1458PubMed
33.
Zurück zum Zitat Matthies AK et al (2013) Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res 31(11):1677–1685PubMed Matthies AK et al (2013) Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res 31(11):1677–1685PubMed
34.
Zurück zum Zitat Lohmann CH et al (2013) Periprosthetic tissue metal content but not serum metal content predicts the type of tissue response in failed small-diameter metal-on-metal total hip arthroplasties. J Bone Joint Surg Am 95(17):1561–1568CrossRefPubMed Lohmann CH et al (2013) Periprosthetic tissue metal content but not serum metal content predicts the type of tissue response in failed small-diameter metal-on-metal total hip arthroplasties. J Bone Joint Surg Am 95(17):1561–1568CrossRefPubMed
35.
Zurück zum Zitat De Haan R et al (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br 90(10):1291–1297CrossRefPubMed De Haan R et al (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br 90(10):1291–1297CrossRefPubMed
36.
Zurück zum Zitat Langton DJ et al (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46CrossRefPubMed Langton DJ et al (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br 92(1):38–46CrossRefPubMed
37.
Zurück zum Zitat Van Der Straeten C et al (2013) Metal ion levels from well-functioning Birmingham hip resurfacings decline significantly at ten years. Bone Joint J 95-B(10):1332–1338CrossRef Van Der Straeten C et al (2013) Metal ion levels from well-functioning Birmingham hip resurfacings decline significantly at ten years. Bone Joint J 95-B(10):1332–1338CrossRef
38.
Zurück zum Zitat Lavigne M et al (2011) Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve. J Bone Joint Surg Am 93(Suppl 2):128–136CrossRefPubMed Lavigne M et al (2011) Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve. J Bone Joint Surg Am 93(Suppl 2):128–136CrossRefPubMed
39.
Zurück zum Zitat Chang EY et al (2013) Relationship of plasma metal ions and clinical and imaging findings in patients with ASR XL metal-on-metal total hip replacements. J Bone Joint Surg Am 95(22):2015–2020CrossRefPubMed Chang EY et al (2013) Relationship of plasma metal ions and clinical and imaging findings in patients with ASR XL metal-on-metal total hip replacements. J Bone Joint Surg Am 95(22):2015–2020CrossRefPubMed
40.
Zurück zum Zitat Emmanuel AR et al (2014) The effect of acetabular inclination on metal ion levels following metal-on-metal hip arthroplasty. J Arthroplasty 29(1):186–191CrossRefPubMed Emmanuel AR et al (2014) The effect of acetabular inclination on metal ion levels following metal-on-metal hip arthroplasty. J Arthroplasty 29(1):186–191CrossRefPubMed
41.
Zurück zum Zitat Vendittoli PA et al (2011) Metal Ion release with large-diameter metal-on-metal hip arthroplasty. J Arthroplasty 26(2):282–288CrossRefPubMed Vendittoli PA et al (2011) Metal Ion release with large-diameter metal-on-metal hip arthroplasty. J Arthroplasty 26(2):282–288CrossRefPubMed
42.
Zurück zum Zitat Azam MQ et al (2016) Survivorship and clinical outcome of Birmingham hip resurfacing: a minimum ten years’ follow-up. Int Orthop 40(1):1–7CrossRefPubMed Azam MQ et al (2016) Survivorship and clinical outcome of Birmingham hip resurfacing: a minimum ten years’ follow-up. Int Orthop 40(1):1–7CrossRefPubMed
Metadaten
Titel
Cobalt serum levels differ in well functioning Birmingham resurfacing and Birmingham modular THA
verfasst von
Lisa Renner
Martin Faschingbauer
Tom Schmidt-Braekling
Friedrich Boettner
Publikationsdatum
16.03.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 5/2016
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-016-2439-3

Weitere Artikel der Ausgabe 5/2016

Archives of Orthopaedic and Trauma Surgery 5/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.