Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology 6/2008

01.06.2008 | Glaucoma

Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range

verfasst von: S. Munkwitz, A. Elkarmouty, E. M. Hoffmann, N. Pfeiffer, H. Thieme

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 6/2008

Einloggen, um Zugang zu erhalten

Abstract

Background

The aim of this study was to compare the intraocular pressure (IOP) results measured by the iCare rebound tonometer with those obtained by the Goldmann applanation tonometer (GAT) over a wide range of IOP values. Furthermore, the comfort level of the iCare measurement was evaluated.

Method

The study included 75 eyes of 75 patients. The patients were divided into three groups (7–15 mmHg n = 25, 16–22 mmHg n = 25, 23–60 mmHg n = 25). The measurements were taken by two independent observers in a masked fashion. All patients were asked about discomfort during the iCare measurement. To establish the agreement between the two devices, a Bland-Altman analysis was performed.

Results

Overall, the 95% confidence interval of the differences between the two devices was −8.67 to 10.25 mmHg and in 62.7%, the iCare measurement was within ±3 mmHg of the GAT measurements. The distribution of the differences in IOP was similar, from 7–22 mmHg. In the higher IOP range (23–60 mmHg), however, the deviation was almost twice as large. The measurement with the iCare tonometer was well tolerated; 100% of the patients denied any discomfort.

Conclusions

The iCare tonometer is a mobile alternative to GAT in a low to moderate IOP range, but our findings show a greater deviation than previously reported. In high IOP values, measurements with the iCare tonometer do not correlate well with GAT.
Literatur
1.
Zurück zum Zitat The Advanced Glaucoma Intervention Study (AGIS) : 7 (2000) The relationship between control of intraocular pressure and visual field deterioration. The AGIS Investigators. Am J Ophthalmol 130(4):429–440CrossRef The Advanced Glaucoma Intervention Study (AGIS) : 7 (2000) The relationship between control of intraocular pressure and visual field deterioration. The AGIS Investigators. Am J Ophthalmol 130(4):429–440CrossRef
2.
Zurück zum Zitat Internationaler Standard für Augentonometer ISO 8612 (2001) Beuth-Verlag GmbH, Berlin Internationaler Standard für Augentonometer ISO 8612 (2001) Beuth-Verlag GmbH, Berlin
3.
Zurück zum Zitat Bechrakis E (1966) On spontaneous decrease of pressure in applanation tonometry. Ophthalmologica 151(5):604–614PubMedCrossRef Bechrakis E (1966) On spontaneous decrease of pressure in applanation tonometry. Ophthalmologica 151(5):604–614PubMedCrossRef
4.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed
5.
Zurück zum Zitat Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15(3):213–217PubMedCrossRef Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15(3):213–217PubMedCrossRef
6.
Zurück zum Zitat Chapman CR, Casey KL, Dubner R, Foley KM, Gracely RH, Reading AE (1985) Pain measurement: an overview. Pain 22(1):1–31PubMedCrossRef Chapman CR, Casey KL, Dubner R, Foley KM, Gracely RH, Reading AE (1985) Pain measurement: an overview. Pain 22(1):1–31PubMedCrossRef
7.
Zurück zum Zitat Davies LN, Bartlett H, Mallen EA, Wolffsohn JS (2006) Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 84(2):206–209PubMedCrossRef Davies LN, Bartlett H, Mallen EA, Wolffsohn JS (2006) Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 84(2):206–209PubMedCrossRef
8.
Zurück zum Zitat Dekking HM, Coster HD (1967) Dynamic tonometry. Ophthalmologica 154(1):59–74PubMed Dekking HM, Coster HD (1967) Dynamic tonometry. Ophthalmologica 154(1):59–74PubMed
9.
Zurück zum Zitat Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25(5):436–440PubMedCrossRef Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25(5):436–440PubMedCrossRef
10.
Zurück zum Zitat Goldmann H, Schmidt T (1957) About applanation tonometry. Ophthalmologica 134(4):221–242PubMed Goldmann H, Schmidt T (1957) About applanation tonometry. Ophthalmologica 134(4):221–242PubMed
11.
Zurück zum Zitat Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol 120(10):1268–1279PubMed Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol 120(10):1268–1279PubMed
12.
Zurück zum Zitat Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedCrossRef Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedCrossRef
13.
Zurück zum Zitat Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124(4):471–476PubMedCrossRef Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124(4):471–476PubMedCrossRef
14.
Zurück zum Zitat Kontiola A, Puska P (2004) Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol 242(1):3–7PubMedCrossRef Kontiola A, Puska P (2004) Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol 242(1):3–7PubMedCrossRef
15.
Zurück zum Zitat Kontiola AI (2000) A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand 78(2):142–145PubMedCrossRef Kontiola AI (2000) A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand 78(2):142–145PubMedCrossRef
16.
Zurück zum Zitat Kontiola AI (2003) Developing impact tonometers for clinical use and glaucoma research. Finland: Department of Ophthalmology, University of Helsinki Kontiola AI (2003) Developing impact tonometers for clinical use and glaucoma research. Finland: Department of Ophthalmology, University of Helsinki
17.
Zurück zum Zitat Kontiola AI, Goldblum D, Mittag T, Danias J (2001) The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 73(6):781–785PubMedCrossRef Kontiola AI, Goldblum D, Mittag T, Danias J (2001) The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 73(6):781–785PubMedCrossRef
18.
Zurück zum Zitat Krakau CE, Wilke K (1971) On repeated tonometry. Acta Ophthalmol (Copenh) 49(4):611–614CrossRef Krakau CE, Wilke K (1971) On repeated tonometry. Acta Ophthalmol (Copenh) 49(4):611–614CrossRef
19.
Zurück zum Zitat Obbink J (1931) Onderzoek naar het verband tusschen inwendigen oogdruk en ballistische reacties. Thesis, Utrecht, The Netherlands Obbink J (1931) Onderzoek naar het verband tusschen inwendigen oogdruk en ballistische reacties. Thesis, Utrecht, The Netherlands
20.
Zurück zum Zitat Parker VA, Herrtage J, Sarkies NJ (2001) Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry. Br J Ophthalmol 85(11):1303–1304PubMedCrossRef Parker VA, Herrtage J, Sarkies NJ (2001) Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry. Br J Ophthalmol 85(11):1303–1304PubMedCrossRef
21.
Zurück zum Zitat Phelps CD, Phelps GK (1976) Measurement of intraocular pressure: a study of its reproducibility. Albrecht Von Graefes Arch Klin Exp Ophthalmol 198(1):39–43PubMedCrossRef Phelps CD, Phelps GK (1976) Measurement of intraocular pressure: a study of its reproducibility. Albrecht Von Graefes Arch Klin Exp Ophthalmol 198(1):39–43PubMedCrossRef
22.
Zurück zum Zitat Sandhu SS, Chattopadhyay S, Birch MK, Ray-Chaudhuri N (2005) Frequency of goldmann applanation tonometer calibration error checks. J Glaucoma 14(3):215–218PubMedCrossRef Sandhu SS, Chattopadhyay S, Birch MK, Ray-Chaudhuri N (2005) Frequency of goldmann applanation tonometer calibration error checks. J Glaucoma 14(3):215–218PubMedCrossRef
23.
Zurück zum Zitat Schreiber W, Vorwerk CK, Langenbucher A, Behrens-Baumann W, Viestenz A (2007) A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry. Ophthalmologe 104:299–304PubMedCrossRef Schreiber W, Vorwerk CK, Langenbucher A, Behrens-Baumann W, Viestenz A (2007) A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry. Ophthalmologe 104:299–304PubMedCrossRef
24.
Zurück zum Zitat van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*. Ophthalmic Physiol Opt 25(5):429–435PubMedCrossRef van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*. Ophthalmic Physiol Opt 25(5):429–435PubMedCrossRef
25.
Zurück zum Zitat Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–1712PubMed Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–1712PubMed
Metadaten
Titel
Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range
verfasst von
S. Munkwitz
A. Elkarmouty
E. M. Hoffmann
N. Pfeiffer
H. Thieme
Publikationsdatum
01.06.2008
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 6/2008
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-007-0758-3

Weitere Artikel der Ausgabe 6/2008

Graefe's Archive for Clinical and Experimental Ophthalmology 6/2008 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.