Comparison of intraocular tonometry using three different non-invasive tonometers in children

verfasst von: Julia Lasseck, Thomas Jehle, Nicolas Feltgen, Wolf Alexander Lagrèze

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 10/2008

Einloggen, um Zugang zu erhalten

Abstract

Background

In childhood glaucoma, the correct determination of intraocular pressure (IOP) is crucial in clinical decision-making. We therefore investigated how intraocular tonometry readings correlate with readings from commonly used tonometers.

Methods

IOP was measured unilaterally in 20 children suffering from congenital (n = 7) or secondary glaucoma (n = 13), 10 minutes after the induction of general anaesthesia. The children were aged from 1 month to 17 years (mean age 4.3 years, median age 1.3 years). Non-invasive applanation (Tono-Pen XL, Perkins tonometer) and indentation tonometry (Schiötz tonometer) were performed in random order prior to intraocular tonometry with a 26-gauge needle connected to a pressure sensor. Linear regression analysis and the coefficients of variance (CV) were used to compare the data obtained from the various tonometers.

Results

Compared with intraocular pressure, the CV was 10% for the Tono-Pen XL, 17% for the Schiötz, and 19% for the Perkins tonometer. The coefficient of determination (r2) was 0.74 for Tono-Pen XL, 0.60 for Schiötz and 0.78 for Perkins tonometry. The IOP values obtained with the Tono-Pen XL scattered homogeneously around the intraocular IOP, while the Perkins and Schiötz tonometers underestimated intraocular measured IOP.

Conclusion

Of the three tonometers evaluated in this series of paediatric glaucoma patients, the Tono-Pen XL most closely reflected true IOP.
Literatur
1.
Zurück zum Zitat Amaral WO, Teixeira RM, Alencar LM, Cronemberger S, Calixto N (2006) Central and peripheral corneal thickness: influence on the iop measurement by Tonopen. Arq Bras Oftalmol 69:41–45PubMedCrossRef Amaral WO, Teixeira RM, Alencar LM, Cronemberger S, Calixto N (2006) Central and peripheral corneal thickness: influence on the iop measurement by Tonopen. Arq Bras Oftalmol 69:41–45PubMedCrossRef
2.
Zurück zum Zitat Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS (2002) Effect of Corneal Thickness on Intraocular Pressure Measurements with the Pneumotonometer, Goldmann Applanation Tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 43:1389–1392PubMed Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS (2002) Effect of Corneal Thickness on Intraocular Pressure Measurements with the Pneumotonometer, Goldmann Applanation Tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 43:1389–1392PubMed
3.
Zurück zum Zitat Blumberg D, Congdon N, Jampel H, Gilbert D, Elliott R, Rivers R, Munoz B, Quigley H (2007) The effects of sevoflurane and ketamine on intraocular pressure in children during examination under anesthesia. Am J Ophthalmol 143:494–499PubMedCrossRef Blumberg D, Congdon N, Jampel H, Gilbert D, Elliott R, Rivers R, Munoz B, Quigley H (2007) The effects of sevoflurane and ketamine on intraocular pressure in children during examination under anesthesia. Am J Ophthalmol 143:494–499PubMedCrossRef
4.
Zurück zum Zitat Bordon AF, Katsumi O, Hirose T (1995) Tonometry in pediatric patients: a comparative study among Tono-pen, Perkins, and Schiötz tonometers. J Pediatr Ophthalmol Strabismus 32:373–377PubMed Bordon AF, Katsumi O, Hirose T (1995) Tonometry in pediatric patients: a comparative study among Tono-pen, Perkins, and Schiötz tonometers. J Pediatr Ophthalmol Strabismus 32:373–377PubMed
5.
Zurück zum Zitat Deuter CM, Schlote T, Hahn GA, Bende T, Derse M (2002) Messung des Augeninnendrucks mit dem Tono-Pen im Vergleich zum Applanations-tonometer nach Goldmann - eine klinische Studie an 100 Augen. Klin Monatsbl Augenheilkd 219:138–142PubMedCrossRef Deuter CM, Schlote T, Hahn GA, Bende T, Derse M (2002) Messung des Augeninnendrucks mit dem Tono-Pen im Vergleich zum Applanations-tonometer nach Goldmann - eine klinische Studie an 100 Augen. Klin Monatsbl Augenheilkd 219:138–142PubMedCrossRef
6.
Zurück zum Zitat Dietlein TS, Jacobi PC, Krieglstein GK (1996) Klinische Diskrepanz zwischen Schiötz- und Perkins-Tonometrie beim kindlichen Glaukom. Klin Monatsbl Augenheilkd 209:299–303PubMed Dietlein TS, Jacobi PC, Krieglstein GK (1996) Klinische Diskrepanz zwischen Schiötz- und Perkins-Tonometrie beim kindlichen Glaukom. Klin Monatsbl Augenheilkd 209:299–303PubMed
7.
Zurück zum Zitat Dohadwala AA, Munger R, Damji KF (1998) Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 105:1849–1854PubMedCrossRef Dohadwala AA, Munger R, Damji KF (1998) Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 105:1849–1854PubMedCrossRef
8.
Zurück zum Zitat Eisenberg DL, Sherman BG, McKeown CA, Schuman JS (1998) Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105:1173–1181PubMedCrossRef Eisenberg DL, Sherman BG, McKeown CA, Schuman JS (1998) Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105:1173–1181PubMedCrossRef
9.
Zurück zum Zitat Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87PubMedCrossRef Feltgen N, Leifert D, Funk J (2001) Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 85:85–87PubMedCrossRef
10.
Zurück zum Zitat Friedenwald JS (1937) Contribution to the theory and practice of tonometry. Am J Ophthalmol 20:985–1024 Friedenwald JS (1937) Contribution to the theory and practice of tonometry. Am J Ophthalmol 20:985–1024
11.
12.
Zurück zum Zitat Horowitz GS, Byles J, Lee J, D’Este C (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Experiment Ophthalmol 32:584–589PubMedCrossRef Horowitz GS, Byles J, Lee J, D’Este C (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin Experiment Ophthalmol 32:584–589PubMedCrossRef
13.
Zurück zum Zitat Kniestedt C, Nee M, Stamper RL (2005) Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243:359–366PubMedCrossRef Kniestedt C, Nee M, Stamper RL (2005) Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243:359–366PubMedCrossRef
14.
Zurück zum Zitat Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef
15.
Zurück zum Zitat Krieglstein G, Langham ME (1975) Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica 171:132–145PubMed Krieglstein G, Langham ME (1975) Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica 171:132–145PubMed
16.
Zurück zum Zitat Lam DS, Leung DY, Chiu TY, Fan DS, Cheung EY, Wong TY, Lai JS, Tham CC (2004) Pressure phosphene self-tonometry: a comparison with Goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci 45:3131–3136PubMedCrossRef Lam DS, Leung DY, Chiu TY, Fan DS, Cheung EY, Wong TY, Lai JS, Tham CC (2004) Pressure phosphene self-tonometry: a comparison with Goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci 45:3131–3136PubMedCrossRef
17.
Zurück zum Zitat Lee GA, Khaw PT, Ficker LA, Shah P (2002) The corneal thickness and intraocular pressure story: where are we now? Clin Experiment Ophthalmol 30:334–337PubMedCrossRef Lee GA, Khaw PT, Ficker LA, Shah P (2002) The corneal thickness and intraocular pressure story: where are we now? Clin Experiment Ophthalmol 30:334–337PubMedCrossRef
18.
Zurück zum Zitat Levy J, Lifshitz T, Rosen S, Tessler Z, Biedner BZ (2005) Is the tono-pen accurate for measuring intraocular pressure in young children with congenital glaucoma? J AAPOS 9:321–325PubMedCrossRef Levy J, Lifshitz T, Rosen S, Tessler Z, Biedner BZ (2005) Is the tono-pen accurate for measuring intraocular pressure in young children with congenital glaucoma? J AAPOS 9:321–325PubMedCrossRef
19.
Zurück zum Zitat Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef Liu J, Roberts CJ (2005) Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg 31:146–155PubMedCrossRef
20.
Zurück zum Zitat Madjlessi F, Marx W, Reinhard T, Althaus C, Sundmacher R (2000) Impressions- und Applanationstonometrie bei pathologischen Hornhäuten im Vergleich mit der intraokularen Nadeldruckmessung. Ophthalmologe 97:478–481PubMedCrossRef Madjlessi F, Marx W, Reinhard T, Althaus C, Sundmacher R (2000) Impressions- und Applanationstonometrie bei pathologischen Hornhäuten im Vergleich mit der intraokularen Nadeldruckmessung. Ophthalmologe 97:478–481PubMedCrossRef
21.
Zurück zum Zitat Marx W, Madjlessi F, Reinhard T, Althaus C, Sundmacher R (1999) Mehr als vier Jahre Erfahrung mit der elektronischen, intraokularen Nadel-Druckmessung bei irregulären Hornhäuten. Ophthalmologe 96:498–502PubMedCrossRef Marx W, Madjlessi F, Reinhard T, Althaus C, Sundmacher R (1999) Mehr als vier Jahre Erfahrung mit der elektronischen, intraokularen Nadel-Druckmessung bei irregulären Hornhäuten. Ophthalmologe 96:498–502PubMedCrossRef
22.
Zurück zum Zitat Mirakhur RK, Elliott P, Shepherd WF, McGalliard JN (1990) Comparison of the effects of isoflurane and halothane on intraocular pressure. Acta Anaesthesiol Scand 34:282–285PubMed Mirakhur RK, Elliott P, Shepherd WF, McGalliard JN (1990) Comparison of the effects of isoflurane and halothane on intraocular pressure. Acta Anaesthesiol Scand 34:282–285PubMed
23.
Zurück zum Zitat Mirakhur RK, Shepherd WF, Darrah WC (1987) Propofol or thiopentone: effects on intraocular pressure associated with induction of anaesthesia and tracheal intubation (facilitated with suxamethonium). Br J Anaesth 59:431–436PubMedCrossRef Mirakhur RK, Shepherd WF, Darrah WC (1987) Propofol or thiopentone: effects on intraocular pressure associated with induction of anaesthesia and tracheal intubation (facilitated with suxamethonium). Br J Anaesth 59:431–436PubMedCrossRef
24.
Zurück zum Zitat Nagdeve NG, Yaddanapudi S, Pandav SS (2006) The effect of different doses of ketamine on intraocular pressure in anesthetized children. J Pediatr Ophthalmol Strabismus 43:219–223PubMed Nagdeve NG, Yaddanapudi S, Pandav SS (2006) The effect of different doses of ketamine on intraocular pressure in anesthetized children. J Pediatr Ophthalmol Strabismus 43:219–223PubMed
25.
Zurück zum Zitat Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR (1988) Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol 106:1697–1700PubMed Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR (1988) Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol 106:1697–1700PubMed
26.
Zurück zum Zitat Schiötz H (1905) Ein neuer Tonometer. Arch Augenheilkunde 52:410–424 Schiötz H (1905) Ein neuer Tonometer. Arch Augenheilkunde 52:410–424
Metadaten
Titel
Comparison of intraocular tonometry using three different non-invasive tonometers in children
verfasst von
Julia Lasseck
Thomas Jehle
Nicolas Feltgen
Wolf Alexander Lagrèze
Publikationsdatum
01.10.2008
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 10/2008
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-008-0863-y

Weitere Artikel der Ausgabe 10/2008

Graefe's Archive for Clinical and Experimental Ophthalmology 10/2008 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.