A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair

verfasst von: Corey R. Deeken, Keith M. Faucher, Brent D. Matthews

Erschienen in: Surgical Endoscopy | Ausgabe 2/2012

Einloggen, um Zugang zu erhalten

Abstract

Background

The objective of this review was to provide an overview of the components that comprise each of the eight barrier mesh prostheses commonly utilized for LVHR and to review the current literature related to the characteristics and effectiveness of these materials to guide the general surgeon in selecting the most appropriate material for LVHR.

Methods

Composite prostheses with permanent barriers (Bard™ Composix™ E/X, Bard™ Composix™ L/P, and DUALMESH® Biomaterial) were compared to composite prostheses with absorbable barriers (C-QUR™ Mesh, PROCEED™ Surgical Mesh, Bard™ Sepramesh™ IP Composite, Parietex™ Composite, and PHYSIOMESH™) using scanning electron microscopy and a review of the current preclinical and clinical literature.

Results

Clinical studies and preclinical animal models have attempted to determine the adhesion characteristics and effectiveness of barrier mesh prostheses available for ventral hernia repair applications. However, it is difficult to make any definitive statements about the adhesion characteristics and effectiveness of these materials because all meshes were not included in all studies and likewise not compared under identical conditions. Overall, Parietex™ Composite and DUALMESH® Biomaterial were cited most frequently for improvement of adhesion characteristics, followed closely by Bard™ Sepramesh™ IP Composite and C-QUR™ Mesh. Bard™ Composix™, PROCEED™ Surgical Mesh, and uncoated polypropylene were cited most frequently as having the most tenacious and extensive adhesions.

Conclusions

Differences observed between the various barrier prostheses are likely attributable to the chemical composition of the barrier or the conditions required for resorption and metabolism of the barrier components. It is likely that the components of these barriers incite a wide range of inflammatory responses resulting in the range of adhesion coverage and tenacity observed in the preclinical and clinical studies reviewed. Clinical trials are needed to more appropriately define the clinical effectiveness of these barriers.
Literatur
1.
Zurück zum Zitat Burger JW, Luijendijk RW, Hop WC, Halm JA, Verdaasdonk EG, Jeekel J (2004) Long-term follow-up of a randomized controlled trial of suture versus mesh repair of incisional hernia. Ann Surg 240:578–583PubMed Burger JW, Luijendijk RW, Hop WC, Halm JA, Verdaasdonk EG, Jeekel J (2004) Long-term follow-up of a randomized controlled trial of suture versus mesh repair of incisional hernia. Ann Surg 240:578–583PubMed
2.
Zurück zum Zitat Heniford BT, Park A, Ramshaw BJ, Voeller G (2003) Laparoscopic repair of ventral hernias: nine years’ experience with 850 consecutive hernias. Ann Surg 238:391–399PubMed Heniford BT, Park A, Ramshaw BJ, Voeller G (2003) Laparoscopic repair of ventral hernias: nine years’ experience with 850 consecutive hernias. Ann Surg 238:391–399PubMed
3.
Zurück zum Zitat Pierce RA, Spitler JA, Frisella MM, Matthews BD, Brunt LM (2007) Pooled data analysis of laparoscopic vs. open ventral hernia repair: 14 years of patient data accrual. Surg Endosc 21:378–386PubMedCrossRef Pierce RA, Spitler JA, Frisella MM, Matthews BD, Brunt LM (2007) Pooled data analysis of laparoscopic vs. open ventral hernia repair: 14 years of patient data accrual. Surg Endosc 21:378–386PubMedCrossRef
4.
Zurück zum Zitat Ray NF, Denton WG, Thamer M, Henderson SC, Perry S (1998) Abdominal adhesiolysis: inpatient care and expenditures in the United States in 1994. J Am Coll Surg 186:1–9PubMedCrossRef Ray NF, Denton WG, Thamer M, Henderson SC, Perry S (1998) Abdominal adhesiolysis: inpatient care and expenditures in the United States in 1994. J Am Coll Surg 186:1–9PubMedCrossRef
5.
Zurück zum Zitat Gray SH, Vick CC, Graham LA, Finan KR, Neumayer LA, Hawn MT (2008) Risk of complications from enterotomy or unplanned bowel resection during elective hernia repair. Arch Surg 143:582–586PubMedCrossRef Gray SH, Vick CC, Graham LA, Finan KR, Neumayer LA, Hawn MT (2008) Risk of complications from enterotomy or unplanned bowel resection during elective hernia repair. Arch Surg 143:582–586PubMedCrossRef
6.
Zurück zum Zitat Halm JA, de Wall LL, Steyerberg EW, Jeekel J, Lange JF (2007) Intraperitoneal polypropylene mesh hernia repair complicates subsequent abdominal surgery. World J Surg 31:423–429PubMedCrossRef Halm JA, de Wall LL, Steyerberg EW, Jeekel J, Lange JF (2007) Intraperitoneal polypropylene mesh hernia repair complicates subsequent abdominal surgery. World J Surg 31:423–429PubMedCrossRef
7.
Zurück zum Zitat Usher FC, Ochsner J, Tuttle LJ (1958) Use of Marlex mesh in the repair of incisional hernias. Am Surg 24:969–974PubMed Usher FC, Ochsner J, Tuttle LJ (1958) Use of Marlex mesh in the repair of incisional hernias. Am Surg 24:969–974PubMed
8.
Zurück zum Zitat (2007) PROCEED Surgical mesh instructions for use (2007) PROCEED Surgical mesh instructions for use
9.
Zurück zum Zitat (2005) Sepramesh IP composite instructions for use (2005) Sepramesh IP composite instructions for use
10.
Zurück zum Zitat Pierce RA, Perrone J, Nimeri A, Sexton J, Walcutt J, Frisella M, Matthews B (2009) 120-day comparative analysis of adhesion grade and quantity, mesh contraction, and tissue response to a novel omega-3 fatty acid bioabsorbable barrier macroporous mesh after intraperitoneal placement. Surg Innov 16:46–54PubMedCrossRef Pierce RA, Perrone J, Nimeri A, Sexton J, Walcutt J, Frisella M, Matthews B (2009) 120-day comparative analysis of adhesion grade and quantity, mesh contraction, and tissue response to a novel omega-3 fatty acid bioabsorbable barrier macroporous mesh after intraperitoneal placement. Surg Innov 16:46–54PubMedCrossRef
11.
Zurück zum Zitat (2011) PHYSIOMESH instructions for use. Ethicon, Inc (2011) PHYSIOMESH instructions for use. Ethicon, Inc
12.
Zurück zum Zitat Tingstedt B, Isaksson K, Andersson E, Andersson R (2007) Prevention of abdominal adhesions—present state and what’s beyond the horizon? Eur Surg Res 39:259–268PubMedCrossRef Tingstedt B, Isaksson K, Andersson E, Andersson R (2007) Prevention of abdominal adhesions—present state and what’s beyond the horizon? Eur Surg Res 39:259–268PubMedCrossRef
13.
Zurück zum Zitat Schreinemacher MH, Emans PJ, Gijbels MJ, Greve JW, Beets GL, Bouvy ND (2009) Degradation of mesh coatings and intraperitoneal adhesion formation in an experimental model. Br J Surg 96:305–313PubMedCrossRef Schreinemacher MH, Emans PJ, Gijbels MJ, Greve JW, Beets GL, Bouvy ND (2009) Degradation of mesh coatings and intraperitoneal adhesion formation in an experimental model. Br J Surg 96:305–313PubMedCrossRef
14.
Zurück zum Zitat (2005) Bard* Composix E/X mesh instructions for use (2005) Bard* Composix E/X mesh instructions for use
15.
Zurück zum Zitat (2006) Composix L/P instructions for use. CR Bard (2006) Composix L/P instructions for use. CR Bard
16.
Zurück zum Zitat Deeken CR, Abdo MS, Frisella MM, Matthews BD (2011) Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair. Surg Endosc 25:1541–1552PubMedCrossRef Deeken CR, Abdo MS, Frisella MM, Matthews BD (2011) Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair. Surg Endosc 25:1541–1552PubMedCrossRef
17.
Zurück zum Zitat (2007) DualMesh biomaterial instructions for use (2007) DualMesh biomaterial instructions for use
18.
Zurück zum Zitat (2008) Gore DUALMESH PLUS biomaterial instructions for use (2008) Gore DUALMESH PLUS biomaterial instructions for use
19.
Zurück zum Zitat Fox C et al (1991) Infection-resistant compositions, medical devices and surfaces and methods for preparing the same. U.S. Patent #5019096 Fox C et al (1991) Infection-resistant compositions, medical devices and surfaces and methods for preparing the same. U.S. Patent #5019096
20.
Zurück zum Zitat Ratner BD, Hoffman AS, Schoen FJ, Lemons JE (2004) Biomaterials science. Academic Press, San Diego Ratner BD, Hoffman AS, Schoen FJ, Lemons JE (2004) Biomaterials science. Academic Press, San Diego
22.
Zurück zum Zitat Turaev A (1995) Dependence of the biodegradability of carboxymethylcellulose on its supermolecular structure and molecular parameters. Chem Nat Compds 31:254–259CrossRef Turaev A (1995) Dependence of the biodegradability of carboxymethylcellulose on its supermolecular structure and molecular parameters. Chem Nat Compds 31:254–259CrossRef
23.
Zurück zum Zitat Yamaoka T, Tabata Y, Ikada Y (1995) Fate of water-soluble polymers administered via different routes. J Pharm Sci 84:349–354PubMedCrossRef Yamaoka T, Tabata Y, Ikada Y (1995) Fate of water-soluble polymers administered via different routes. J Pharm Sci 84:349–354PubMedCrossRef
24.
Zurück zum Zitat Earle DB, Mark LA (2008) Prosthetic material in inguinal hernia repair: how do I choose? Surg Clin North Am 88:179–201PubMedCrossRef Earle DB, Mark LA (2008) Prosthetic material in inguinal hernia repair: how do I choose? Surg Clin North Am 88:179–201PubMedCrossRef
25.
Zurück zum Zitat (2009) C-QUR mesh instructions for use (2009) C-QUR mesh instructions for use
27.
Zurück zum Zitat Kabara J, Swieczkowski D, Conley AJ, Truant J (1972) Fatty acids and derivatives as antimicrobial agents. Antimicrob Agents Chemother 2:23–28PubMed Kabara J, Swieczkowski D, Conley AJ, Truant J (1972) Fatty acids and derivatives as antimicrobial agents. Antimicrob Agents Chemother 2:23–28PubMed
28.
Zurück zum Zitat Thormar H, Isaacs CE, Brown HR, Barshatzky MR, Pessolano T (1987) Inactivation of enveloped viruses and killing of cells by fatty acids and monoglycerides. Antimicrob Agents Chemother 31:27–31PubMed Thormar H, Isaacs CE, Brown HR, Barshatzky MR, Pessolano T (1987) Inactivation of enveloped viruses and killing of cells by fatty acids and monoglycerides. Antimicrob Agents Chemother 31:27–31PubMed
29.
Zurück zum Zitat Ferrucci L, Cherubini A, Bandinelli S, Bartali B, Corsi A, Lauretani F, Martin A, Andres-Lacueva C, Senin U, Guralnik JM (2006) Relationship of plasma polyunsaturated fatty acids to circulating inflammatory markers. J Clin Endocrinol Metab 91:439–446PubMedCrossRef Ferrucci L, Cherubini A, Bandinelli S, Bartali B, Corsi A, Lauretani F, Martin A, Andres-Lacueva C, Senin U, Guralnik JM (2006) Relationship of plasma polyunsaturated fatty acids to circulating inflammatory markers. J Clin Endocrinol Metab 91:439–446PubMedCrossRef
30.
Zurück zum Zitat Arita M, Yoshida M, Hong S, Tjonahen E, Glickman JN, Petasis NA, Blumberg RS, Serhan CN (2005) Resolvin E1, an endogenous lipid mediator derived from omega-3 eicosapentaenoic acid, protects against 2,4,6-trinitrobenzene sulfonic acid-induced colitis. Proc Natl Acad Sci USA 102:7671–7676PubMedCrossRef Arita M, Yoshida M, Hong S, Tjonahen E, Glickman JN, Petasis NA, Blumberg RS, Serhan CN (2005) Resolvin E1, an endogenous lipid mediator derived from omega-3 eicosapentaenoic acid, protects against 2,4,6-trinitrobenzene sulfonic acid-induced colitis. Proc Natl Acad Sci USA 102:7671–7676PubMedCrossRef
31.
Zurück zum Zitat (2010) Parietex composite instructions for use (2010) Parietex composite instructions for use
32.
Zurück zum Zitat Ory F et al (2002) Composite prosthesis for preventing post-surgical adhesions and method for obtaining same. U.S. Patent #09463313 Ory F et al (2002) Composite prosthesis for preventing post-surgical adhesions and method for obtaining same. U.S. Patent #09463313
33.
Zurück zum Zitat Tayot J et al (2002) Collagenic material useful in particular for preventing post-operative adhesions. U.S. Patent #6391939 Tayot J et al (2002) Collagenic material useful in particular for preventing post-operative adhesions. U.S. Patent #6391939
34.
Zurück zum Zitat van Amerongen MJ, Harmsen MC, Petersen AH, Kors G, van Luyn MJ (2006) The enzymatic degradation of scaffolds and their replacement by vascularized extracellular matrix in the murine myocardium. Biomaterials 27:2247–2257PubMedCrossRef van Amerongen MJ, Harmsen MC, Petersen AH, Kors G, van Luyn MJ (2006) The enzymatic degradation of scaffolds and their replacement by vascularized extracellular matrix in the murine myocardium. Biomaterials 27:2247–2257PubMedCrossRef
35.
Zurück zum Zitat Campbell M (1995) Biochemistry. Saunders, New York Campbell M (1995) Biochemistry. Saunders, New York
36.
Zurück zum Zitat Boland ED, Coleman BD, Barnes CP, Simpson DG, Wnek GE, Bowlin GL (2005) Electrospinning polydioxanone for biomedical applications. Acta Biomater 1:115–123PubMedCrossRef Boland ED, Coleman BD, Barnes CP, Simpson DG, Wnek GE, Bowlin GL (2005) Electrospinning polydioxanone for biomedical applications. Acta Biomater 1:115–123PubMedCrossRef
37.
Zurück zum Zitat Pillai CK, Sharma CP (2010) Review paper: absorbable polymeric surgical sutures: chemistry, production, properties, biodegradability, and performance. J Biomater Appl 25:291–366PubMedCrossRef Pillai CK, Sharma CP (2010) Review paper: absorbable polymeric surgical sutures: chemistry, production, properties, biodegradability, and performance. J Biomater Appl 25:291–366PubMedCrossRef
38.
Zurück zum Zitat Ashton W et al (1968) Oxidized cellulose product and method for preparing the same. U.S. Patent #3364200 Ashton W et al (1968) Oxidized cellulose product and method for preparing the same. U.S. Patent #3364200
39.
Zurück zum Zitat Boardman F et al (1993) Cellulose oxidation by a perfluoronated hydrocarbon solution of nitrogen dioxide. U.S. Patent #5180398 Boardman F et al (1993) Cellulose oxidation by a perfluoronated hydrocarbon solution of nitrogen dioxide. U.S. Patent #5180398
40.
Zurück zum Zitat Wiseman D et al (2003) Bioabsorbable medical devices from oxidized polysaccharides. U.S. Patent #20030073663 Wiseman D et al (2003) Bioabsorbable medical devices from oxidized polysaccharides. U.S. Patent #20030073663
41.
Zurück zum Zitat Dimitrijevich SD, Tatarko M, Gracy RW, Linsky CB, Olsen C (1990) Biodegradation of oxidized regenerated cellulose. Carbohydr Res 195:247–256PubMedCrossRef Dimitrijevich SD, Tatarko M, Gracy RW, Linsky CB, Olsen C (1990) Biodegradation of oxidized regenerated cellulose. Carbohydr Res 195:247–256PubMedCrossRef
42.
Zurück zum Zitat Dimitrijevich SD, Tatarko M, Gracy RW, Wise GE, Oakford LX, Linsky CB, Kamp L (1990) In vivo degradation of oxidized, regenerated cellulose. Carbohydr Res 198:331–341PubMedCrossRef Dimitrijevich SD, Tatarko M, Gracy RW, Wise GE, Oakford LX, Linsky CB, Kamp L (1990) In vivo degradation of oxidized, regenerated cellulose. Carbohydr Res 198:331–341PubMedCrossRef
43.
Zurück zum Zitat Nary FH, Matsumoto MA, Batista AC, Lopes LC, de Goes FC, Consolaro A (2002) Comparative study of tissue response to polyglecaprone 25, polyglactin 910 and polytetrafluoroethylene suture materials in rats. Braz Dent J 13:86–91 Nary FH, Matsumoto MA, Batista AC, Lopes LC, de Goes FC, Consolaro A (2002) Comparative study of tissue response to polyglecaprone 25, polyglactin 910 and polytetrafluoroethylene suture materials in rats. Braz Dent J 13:86–91
44.
Zurück zum Zitat Arnaud JP, Hennekinne-Mucci S, Pessaux P, Tuech JJ, Aube C (2003) Ultrasound detection of visceral adhesion after intraperitoneal ventral hernia treatment: a comparative study of protected versus unprotected meshes. Hernia 7:85–88PubMedCrossRef Arnaud JP, Hennekinne-Mucci S, Pessaux P, Tuech JJ, Aube C (2003) Ultrasound detection of visceral adhesion after intraperitoneal ventral hernia treatment: a comparative study of protected versus unprotected meshes. Hernia 7:85–88PubMedCrossRef
45.
Zurück zum Zitat Cobb WS, Harris JB, Lokey JS, McGill ES, Klove KL (2003) Incisional herniorrhaphy with intraperitoneal composite mesh: a report of 95 cases. Am Surg 69:784–787PubMed Cobb WS, Harris JB, Lokey JS, McGill ES, Klove KL (2003) Incisional herniorrhaphy with intraperitoneal composite mesh: a report of 95 cases. Am Surg 69:784–787PubMed
46.
Zurück zum Zitat Koehler RH, Begos D, Berger D, Carey S, LeBlanc K, Park A, Ramshaw B, Smoot R, Voeller G (2003) Minimal adhesions to ePTFE mesh after laparoscopic ventral incisional hernia repair: reoperative findings in 65 cases. JSLS 7:335–340PubMed Koehler RH, Begos D, Berger D, Carey S, LeBlanc K, Park A, Ramshaw B, Smoot R, Voeller G (2003) Minimal adhesions to ePTFE mesh after laparoscopic ventral incisional hernia repair: reoperative findings in 65 cases. JSLS 7:335–340PubMed
47.
Zurück zum Zitat Leber GE, Garb JL, Alexander AI, Reed WP (1998) Long-term complications associated with prosthetic repair of incisional hernias. Arch Surg 133:378–382PubMedCrossRef Leber GE, Garb JL, Alexander AI, Reed WP (1998) Long-term complications associated with prosthetic repair of incisional hernias. Arch Surg 133:378–382PubMedCrossRef
48.
Zurück zum Zitat Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM, Matthews BD (2010) Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007PubMedCrossRef Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM, Matthews BD (2010) Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007PubMedCrossRef
49.
Zurück zum Zitat Chelala E, Debardemaeker Y, Elias B, Charara F, Dessily M, Alle JL (2010) Eighty-five redo surgeries after 733 laparoscopic treatments for ventral and incisional hernia: adhesion and recurrence analysis. Hernia 14:123–129PubMedCrossRef Chelala E, Debardemaeker Y, Elias B, Charara F, Dessily M, Alle JL (2010) Eighty-five redo surgeries after 733 laparoscopic treatments for ventral and incisional hernia: adhesion and recurrence analysis. Hernia 14:123–129PubMedCrossRef
50.
Zurück zum Zitat Wassenaar EB, Schoenmaeckers EJ, Raymakers JT, Rakic S (2010) Subsequent abdominal surgery after laparoscopic ventral and incisional hernia repair with an expanded polytetrafluoroethylene mesh: a single institution experience with 72 reoperations. Hernia 14:137–142PubMedCrossRef Wassenaar EB, Schoenmaeckers EJ, Raymakers JT, Rakic S (2010) Subsequent abdominal surgery after laparoscopic ventral and incisional hernia repair with an expanded polytetrafluoroethylene mesh: a single institution experience with 72 reoperations. Hernia 14:137–142PubMedCrossRef
51.
Zurück zum Zitat Duffy AJ, Hogle NJ, LaPerle KM, Fowler DL (2004) Comparison of two composite meshes using two fixation devices in a porcine laparoscopic ventral hernia repair model. Hernia 8:358–364PubMedCrossRef Duffy AJ, Hogle NJ, LaPerle KM, Fowler DL (2004) Comparison of two composite meshes using two fixation devices in a porcine laparoscopic ventral hernia repair model. Hernia 8:358–364PubMedCrossRef
52.
Zurück zum Zitat McGinty JJ, Hogle NJ, McCarthy H, Fowler DL (2005) A comparative study of adhesion formation and abdominal wall ingrowth after laparoscopic ventral hernia repair in a porcine model using multiple types of mesh. Surg Endosc 19:786–790PubMedCrossRef McGinty JJ, Hogle NJ, McCarthy H, Fowler DL (2005) A comparative study of adhesion formation and abdominal wall ingrowth after laparoscopic ventral hernia repair in a porcine model using multiple types of mesh. Surg Endosc 19:786–790PubMedCrossRef
53.
Zurück zum Zitat Jacob BP, Hogle NJ, Durak E, Kim T, Fowler DL (2007) Tissue ingrowth and bowel adhesion formation in an animal comparative study: polypropylene versus proceed versus parietex composite. Surg Endosc 21:629–633PubMedCrossRef Jacob BP, Hogle NJ, Durak E, Kim T, Fowler DL (2007) Tissue ingrowth and bowel adhesion formation in an animal comparative study: polypropylene versus proceed versus parietex composite. Surg Endosc 21:629–633PubMedCrossRef
54.
Zurück zum Zitat Gonzalez R, Rodeheaver GT, Moody DL, Foresman PA, Ramshaw BJ (2004) Resistance to adhesion formation: a comparative study of treated and untreated mesh products placed in the abdominal cavity. Hernia 8:213–219PubMedCrossRef Gonzalez R, Rodeheaver GT, Moody DL, Foresman PA, Ramshaw BJ (2004) Resistance to adhesion formation: a comparative study of treated and untreated mesh products placed in the abdominal cavity. Hernia 8:213–219PubMedCrossRef
55.
Zurück zum Zitat Gaertner WB, Bonsack ME, Delaney JP (2010) Visceral adhesions to hernia prostheses. Hernia 14:375–381PubMedCrossRef Gaertner WB, Bonsack ME, Delaney JP (2010) Visceral adhesions to hernia prostheses. Hernia 14:375–381PubMedCrossRef
56.
Zurück zum Zitat Kayaoglu HA, Ozkan N, Hazinedaroglu SM, Ersoy OF, Erkek AB, Koseoglu RD (2005) Comparison of adhesive properties of five different prosthetic materials used in hernioplasty. J Invest Surg 18:89–95PubMedCrossRef Kayaoglu HA, Ozkan N, Hazinedaroglu SM, Ersoy OF, Erkek AB, Koseoglu RD (2005) Comparison of adhesive properties of five different prosthetic materials used in hernioplasty. J Invest Surg 18:89–95PubMedCrossRef
57.
Zurück zum Zitat Matthews BD, Mostafa G, Carbonell AM, Joels CS, Kercher KW, Austin C, Norton HJ, Heniford BT (2005) Evaluation of adhesion formation and host tissue response to intra-abdominal polytetrafluoroethylene mesh and composite prosthetic mesh. J Surg Res 123:227–234PubMedCrossRef Matthews BD, Mostafa G, Carbonell AM, Joels CS, Kercher KW, Austin C, Norton HJ, Heniford BT (2005) Evaluation of adhesion formation and host tissue response to intra-abdominal polytetrafluoroethylene mesh and composite prosthetic mesh. J Surg Res 123:227–234PubMedCrossRef
58.
Zurück zum Zitat Matthews BD, Pratt BL, Pollinger HS, Backus CL, Kercher KW, Sing RF, Heniford BT (2003) Assessment of adhesion formation to intra-abdominal polypropylene mesh and polytetrafluoroethylene mesh. J Surg Res 114:126–132PubMedCrossRef Matthews BD, Pratt BL, Pollinger HS, Backus CL, Kercher KW, Sing RF, Heniford BT (2003) Assessment of adhesion formation to intra-abdominal polypropylene mesh and polytetrafluoroethylene mesh. J Surg Res 114:126–132PubMedCrossRef
59.
Zurück zum Zitat Novitsky YW, Harrell AG, Cristiano JA, Paton BL, Norton HJ, Peindl RD, Kercher KW, Heniford BT (2007) Comparative evaluation of adhesion formation, strength of ingrowth, and textile properties of prosthetic meshes after long-term intra-abdominal implantation in a rabbit. J Surg Res 140:6–11PubMedCrossRef Novitsky YW, Harrell AG, Cristiano JA, Paton BL, Norton HJ, Peindl RD, Kercher KW, Heniford BT (2007) Comparative evaluation of adhesion formation, strength of ingrowth, and textile properties of prosthetic meshes after long-term intra-abdominal implantation in a rabbit. J Surg Res 140:6–11PubMedCrossRef
60.
Zurück zum Zitat Greenawalt KE, Butler TJ, Rowe EA, Finneral AC, Garlick DS, Burns JW (2000) Evaluation of Sepramesh biosurgical composite in a rabbit hernia repair model. J Surg Res 94:92–98PubMedCrossRef Greenawalt KE, Butler TJ, Rowe EA, Finneral AC, Garlick DS, Burns JW (2000) Evaluation of Sepramesh biosurgical composite in a rabbit hernia repair model. J Surg Res 94:92–98PubMedCrossRef
Metadaten
Titel
A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair
verfasst von
Corey R. Deeken
Keith M. Faucher
Brent D. Matthews
Publikationsdatum
01.02.2012
Verlag
Springer-Verlag
Erschienen in
Surgical Endoscopy / Ausgabe 2/2012
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-011-1899-3

Weitere Artikel der Ausgabe 2/2012

Surgical Endoscopy 2/2012 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.