Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar “Bovie” instrument in comparison with traditional laparoscopy

verfasst von: Nicole T. Townsend, Edward L. Jones, Doug Overbey, Bruce Dunne, Jennifer McHenry, Thomas N. Robinson

Erschienen in: Surgical Endoscopy | Ausgabe 8/2017

Einloggen, um Zugang zu erhalten

Abstract

Background

Single-incision laparoscopic surgery (SILS) places multiple instruments in close, parallel proximity, an orientation that may have implications in the production of stray current from the monopolar “Bovie” instrument. The purpose of this study was to compare the energy transferred during SILS compared to traditional four-port laparoscopic surgery (TRD).

Method

In a laparoscopic simulator, instruments were inserted via SILS or TRD setup. The monopolar generator delivered energy to a laparoscopic L-hook instrument for 5-s activations on 30-Watts coag mode. The primary outcome (stray current) was quantified by measuring the heat of liver tissue held adjacent to the non-electrically active 10-mm telescope tip and Maryland grasper in both the SILS and TRD setups. To control for the potential confounder of stray energy coupling via wires outside the surgical field, the camera cord and active electrode wires were oriented parallel or completely separated.

Results

SILS and TRD setups create similar amounts of stray current as measured by increased tissue temperature at the non-electrically active telescope tip (41 ± 12 vs. 39 ± 10 °C; p = 0.71). Stray current was greater in SILS compared to TRD at the tip of the non-electrically active Maryland forceps (38 ± 9 vs. 20 ± 10 °C; p < 0.01). Separation of the active electrode and camera cords did not change the amount of stray energy in the SILS orientation for either telescope (39 ± 10 °C bundled vs. 36 ± 10 °C separated; p = 0.40) or grasper (38 ± 9 °C bundled vs. 34 ± 11 °C separated; p = 0.19) but did in the TRD orientation (41 ± 12 bundled vs. 24 ± 10 separated; p < 0.01). When SILS was compared to TRD with the cords separated, SILS increased stray energy at both the telescope tip and grasper tip (36 ± 10 vs. 24 ± 10 °C; p < 0.01 and 34 ± 11 vs. 17 ± 8 °C; p < 0.01).

Conclusion

SILS increases stray energy transfer nearly twice as much as TRD with the use of the monopolar instrument. Strategies to mitigate the amount of stray energy in the TRD setup such as separation of the active electrode and camera cords are not effective in the SILS setup. These practical findings should enhance surgeons using the SILS approach of increased stray energy that could result in injury.
Literatur
1.
Zurück zum Zitat Nduka CC et al (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179(2):161–170PubMed Nduka CC et al (1994) Cause and prevention of electrosurgical injuries in laparoscopy. J Am Coll Surg 179(2):161–170PubMed
2.
Zurück zum Zitat Hulka JF et al (1997) Laparoscopic-assisted vaginal hysterectomy: American Association of Gynecologic Laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4(2):167–171CrossRefPubMed Hulka JF et al (1997) Laparoscopic-assisted vaginal hysterectomy: American Association of Gynecologic Laparoscopists’ 1995 membership survey. J Am Assoc Gynecol Laparosc 4(2):167–171CrossRefPubMed
3.
Zurück zum Zitat Feder BJ (2006) Surgical device poses a rare but serious peril. New York Times, March 17, New York Feder BJ (2006) Surgical device poses a rare but serious peril. New York Times, March 17, New York
4.
Zurück zum Zitat Jones EL et al (2012) Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications. Surg Endosc 26(11):3053–3057CrossRefPubMed Jones EL et al (2012) Radiofrequency energy antenna coupling to common laparoscopic instruments: practical implications. Surg Endosc 26(11):3053–3057CrossRefPubMed
5.
Zurück zum Zitat Robinson TN et al (2012) Antenna coupling—a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg 256(2):213–218CrossRefPubMed Robinson TN et al (2012) Antenna coupling—a novel mechanism of radiofrequency electrosurgery complication: practical implications. Ann Surg 256(2):213–218CrossRefPubMed
6.
Zurück zum Zitat Tucker RD, Voyles CR, Silvis SE (1992) Capacitive coupled stray currents during laparoscopic and endoscopic electrosurgical procedures. Biomed Instrum Technol 26(4):303–311PubMed Tucker RD, Voyles CR, Silvis SE (1992) Capacitive coupled stray currents during laparoscopic and endoscopic electrosurgical procedures. Biomed Instrum Technol 26(4):303–311PubMed
7.
Zurück zum Zitat Robinson TN et al (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20(5):317–320CrossRefPubMed Robinson TN et al (2010) Surgeon-controlled factors that reduce monopolar electrosurgery capacitive coupling during laparoscopy. Surg Laparosc Endosc Percutan Tech 20(5):317–320CrossRefPubMed
8.
Zurück zum Zitat Robinson TN, Jones EL, Dunn CL, Johnson E, Townsend NT, Paniccia A, Stiegmann GV (2015) Separating the laparoscopic camera cord from the monopolar “Bovie” cord reduces unintended thermal injury from antenna coupling: a randomized controlled trial. Ann Surg 261(6):1056–1060CrossRefPubMed Robinson TN, Jones EL, Dunn CL, Johnson E, Townsend NT, Paniccia A, Stiegmann GV (2015) Separating the laparoscopic camera cord from the monopolar “Bovie” cord reduces unintended thermal injury from antenna coupling: a randomized controlled trial. Ann Surg 261(6):1056–1060CrossRefPubMed
9.
Zurück zum Zitat Phillips MS et al (2012) Intermediate results of a prospective randomized controlled trial of traditional four-port laparoscopic cholecystectomy versus single-incision laparoscopic cholecystectomy. Surg Endosc 26(5):1296–1303CrossRefPubMed Phillips MS et al (2012) Intermediate results of a prospective randomized controlled trial of traditional four-port laparoscopic cholecystectomy versus single-incision laparoscopic cholecystectomy. Surg Endosc 26(5):1296–1303CrossRefPubMed
11.
Zurück zum Zitat Abu-Rafea B et al (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18(6):734–740CrossRefPubMed Abu-Rafea B et al (2011) Monopolar electrosurgery through single-port laparoscopy: a potential hidden hazard for bowel burns. J Minim Invasive Gynecol 18(6):734–740CrossRefPubMed
12.
Zurück zum Zitat Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164(1):57–62CrossRefPubMed Voyles CR, Tucker RD (1992) Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgery. Am J Surg 164(1):57–62CrossRefPubMed
13.
Zurück zum Zitat Montero PN et al (2010) Insulation failure in laparoscopic instruments. Surg Endosc 24(2):462–465CrossRefPubMed Montero PN et al (2010) Insulation failure in laparoscopic instruments. Surg Endosc 24(2):462–465CrossRefPubMed
14.
Zurück zum Zitat Govekar HR et al (2011) Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue? Surg Endosc 25(11):3499–3502CrossRefPubMed Govekar HR et al (2011) Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue? Surg Endosc 25(11):3499–3502CrossRefPubMed
15.
Zurück zum Zitat Wu MP et al (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179(1):67–73CrossRefPubMed Wu MP et al (2000) Complications and recommended practices for electrosurgery in laparoscopy. Am J Surg 179(1):67–73CrossRefPubMed
16.
Zurück zum Zitat Overbey DM, Townsend NT, Chapman BC, Bennett DT, Foley LS, Rau AS, Yi JA, Jones LE, Steigmann GV, Robinson TN (2015) Surgical energy-based device injuries and fatalities reported to the food and drug administration. J Am Coll Surg 221(1):197–205CrossRefPubMed Overbey DM, Townsend NT, Chapman BC, Bennett DT, Foley LS, Rau AS, Yi JA, Jones LE, Steigmann GV, Robinson TN (2015) Surgical energy-based device injuries and fatalities reported to the food and drug administration. J Am Coll Surg 221(1):197–205CrossRefPubMed
17.
Zurück zum Zitat Kamangar F (2012) Effect modification in epidemiology and medicine. Arch Iran Med 15(9):575–582PubMed Kamangar F (2012) Effect modification in epidemiology and medicine. Arch Iran Med 15(9):575–582PubMed
Metadaten
Titel
Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar “Bovie” instrument in comparison with traditional laparoscopy
verfasst von
Nicole T. Townsend
Edward L. Jones
Doug Overbey
Bruce Dunne
Jennifer McHenry
Thomas N. Robinson
Publikationsdatum
18.11.2016
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 8/2017
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-016-5339-2

Weitere Artikel der Ausgabe 8/2017

Surgical Endoscopy 8/2017 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.