Tidal volume and airway pressure under percutaneous transtracheal ventilation without a jet ventilator: comparison of high-flow oxygen ventilation and manual ventilation in complete and incomplete upper airway obstruction models

verfasst von: Nobuko Sasano, Akemi Tanaka, Ai Muramatsu, Yoshihito Fujita, Shoji Ito, Hiroshi Sasano, Kazuya Sobue

Erschienen in: Journal of Anesthesia | Ausgabe 3/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Percutaneous transtracheal ventilation (PTV) can be life-saving in a cannot ventilate, cannot intubate situation. The aim of this study was to investigate the efficacy of PTV by measuring tidal volumes (VTs) and airway pressure (Paw) in high-flow oxygen ventilation and manual ventilation using a model lung.

Methods

We examined 14G, 16G, 18G, and 20G intravenous catheters and minitracheotomy catheters. In high-flow oxygen ventilation, the flow was set to 10 L/min, while the inspiratory:expiratory phases (I:E) were 1 s:4 s in the complete upper airway obstruction model and 1 s:1 s in the incomplete obstruction model. In manual ventilation, I:E were 2 s:4 s in the complete obstruction model and 2 s:3 s in the incomplete obstruction model. We ventilated through each catheter for 2 min and measured VT and Paw.

Results

In high-flow ventilation, the average VTs were approximately 150 ml and <100 ml with 14G catheters in complete and incomplete upper airway obstruction, respectively. The VTs obtained were reduced when the bore size was decreased. In manual ventilation, the average VTs were over 300 ml and approximately 260 ml with 14G catheters in complete and incomplete upper airway obstruction, respectively. In high-flow ventilation, the airway pressure tended to be higher. The minitracheotomy catheters produced over 800 ml of VT and created almost no positive end-expiratory pressure.

Conclusions

High-flow ventilation tends to result in higher airway pressure despite a smaller VT, which is probably due to a PEEP effect caused by high flow.
Literatur
1.
Zurück zum Zitat American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Practice guidelines for management of the difficult airway: an updated report. Anesthesiology. 2003; 98:1269–77. American Society of Anesthesiologists Task Force on Management of the Difficult Airway. Practice guidelines for management of the difficult airway: an updated report. Anesthesiology. 2003; 98:1269–77.
2.
Zurück zum Zitat Schaefer R, Hueter L, Preussler NP, Schreiber T, Schwarzkopf K. Percutaneous transtracheal emergency ventilation with a self-made device in an animal model. Paediatr Anaesth. 2007;17:972–6.PubMedCrossRef Schaefer R, Hueter L, Preussler NP, Schreiber T, Schwarzkopf K. Percutaneous transtracheal emergency ventilation with a self-made device in an animal model. Paediatr Anaesth. 2007;17:972–6.PubMedCrossRef
3.
Zurück zum Zitat Gaughan SD, Ozaki GT, Benumof JL. A comparison in a lung model of low- and high-flow regulators for transtracheal jet ventilation. Anesthesiology. 1992;77:189–99.PubMedCrossRef Gaughan SD, Ozaki GT, Benumof JL. A comparison in a lung model of low- and high-flow regulators for transtracheal jet ventilation. Anesthesiology. 1992;77:189–99.PubMedCrossRef
4.
Zurück zum Zitat Benumof JL. Transtracheal jet ventilation via percutaneous catheter and high-pressure source. In: Benumof JL, editor. Airway management: principles and practice. St. Louis: Mosby; 1996. p. 455–74. Benumof JL. Transtracheal jet ventilation via percutaneous catheter and high-pressure source. In: Benumof JL, editor. Airway management: principles and practice. St. Louis: Mosby; 1996. p. 455–74.
5.
Zurück zum Zitat American College of Surgeons Committee on Trauma, editor. Airway and ventilatory management. In: Advanced Trauma Life Support® for Doctors ATLS®: student course manual. 8th ed. Chicago: ACSC; 2008. p. 25–53. American College of Surgeons Committee on Trauma, editor. Airway and ventilatory management. In: Advanced Trauma Life Support® for Doctors ATLS®: student course manual. 8th ed. Chicago: ACSC; 2008. p. 25–53.
7.
Zurück zum Zitat Lenfant F, Pean D, Brisard L, Freysz M, Lejus C. Oxygen delivery during transtracheal oxygenation: a comparion of two manual devices. Anesth Analg. 2010;111:922–4.PubMed Lenfant F, Pean D, Brisard L, Freysz M, Lejus C. Oxygen delivery during transtracheal oxygenation: a comparion of two manual devices. Anesth Analg. 2010;111:922–4.PubMed
8.
Zurück zum Zitat Hamaekers A, Borg P, Enk D. The importance of flow and pressure release in emergency jet ventilation devices. Paediatr Anaesth. 2008;19:452–7.CrossRef Hamaekers A, Borg P, Enk D. The importance of flow and pressure release in emergency jet ventilation devices. Paediatr Anaesth. 2008;19:452–7.CrossRef
9.
Zurück zum Zitat Reed JP, Kemph JP, Hamelberg W, Hitchcock FA, Jacoby J. Studies with transtracheal artificial respiration. Anesthesiology. 1954;15:28–41.PubMedCrossRef Reed JP, Kemph JP, Hamelberg W, Hitchcock FA, Jacoby J. Studies with transtracheal artificial respiration. Anesthesiology. 1954;15:28–41.PubMedCrossRef
10.
Zurück zum Zitat Henderson J. Airway management in the adult. In: Miller RD, editor. Miller’s anesthesia. 7th ed. Philadelphia: Churchill Livingstone; 2010. p. 1573–610.CrossRef Henderson J. Airway management in the adult. In: Miller RD, editor. Miller’s anesthesia. 7th ed. Philadelphia: Churchill Livingstone; 2010. p. 1573–610.CrossRef
11.
Zurück zum Zitat Gal TJ. Airway management. In: Miller RD, editor. Miller’s anesthesia. 6th ed. Philadelphia: Churchill Livingstone; 2005. p. 1617–52. Gal TJ. Airway management. In: Miller RD, editor. Miller’s anesthesia. 6th ed. Philadelphia: Churchill Livingstone; 2005. p. 1617–52.
12.
Zurück zum Zitat Preussler NP, Schreiber T, Huter L, Gottschall R, Schubert H, Rek H, Karzai W, Schwarzkopf K. Percutaneous transtracheal ventilation: effects of a new oxygen flow modulator on oxygenation and ventilation in pigs compared with a hand triggered emergency jet injector. Resuscitation. 2003;56:329–33.PubMedCrossRef Preussler NP, Schreiber T, Huter L, Gottschall R, Schubert H, Rek H, Karzai W, Schwarzkopf K. Percutaneous transtracheal ventilation: effects of a new oxygen flow modulator on oxygenation and ventilation in pigs compared with a hand triggered emergency jet injector. Resuscitation. 2003;56:329–33.PubMedCrossRef
Metadaten
Titel
Tidal volume and airway pressure under percutaneous transtracheal ventilation without a jet ventilator: comparison of high-flow oxygen ventilation and manual ventilation in complete and incomplete upper airway obstruction models
verfasst von
Nobuko Sasano
Akemi Tanaka
Ai Muramatsu
Yoshihito Fujita
Shoji Ito
Hiroshi Sasano
Kazuya Sobue
Publikationsdatum
01.06.2014
Verlag
Springer Japan
Erschienen in
Journal of Anesthesia / Ausgabe 3/2014
Print ISSN: 0913-8668
Elektronische ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-013-1733-2

Weitere Artikel der Ausgabe 3/2014

Journal of Anesthesia 3/2014 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.