Effect of a new heated and humidified breathing circuit with a fluid-warming device on intraoperative core temperature: a prospective randomized study

verfasst von: Eugene Kim, Sue-Young Lee, Young-Jin Lim, Jung-Yoon Choi, Young-Tae Jeon, Jung-Won Hwang, Hee-Pyoung Park

Erschienen in: Journal of Anesthesia | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

The effect of the Mega Acer kit@, a new heated and humidified breathing circuit (HHBC) containing a fluid-warming device, was investigated on intraoperative core temperature (T c).

Methods

A total of 102 patients undergoing elective craniotomies were randomly divided into three groups based on the breathing circuit used: a conventional breathing circuit (group C, n = 34), a Fisher & Paykel HHBC (group F, n = 34), and the Mega (group M, n = 34). From baseline to the end of the surgery, T c and infusion fluid temperature (T f) were recorded at 15-min intervals. If T c became lower than 35.5 °C, a forced-air warmer was used.

Results

Baseline temperatures were 36.7 ± 0.3, 36.6 ± 0.2, and 36.6 ± 0.2 °C in groups C, M, and F, respectively. T c at the end of surgery dropped from baseline values by 1.0 ± 0.4, 0.5 ± 0.5, and 0.8 ± 0.5 °C in groups C, M, and F, respectively. From 60 min of post-induction to the end of surgery, T c was higher in group M than group C (p < 0.05). From 105 min of post-induction to the end of surgery, T c was higher in group M than group F (p < 0.05). The number of patients receiving forced-air warmer and total forced-air warmer using time were significantly lower in group M than groups C and F (p < 0.05). T f was higher in group M than groups C and F throughout the study period (31.0 ± 1.0 vs. 23.5 ± 0.5 and 24.0 ± 0.4 °C; p < 0.01).

Conclusions

The Mega significantly reduced the drop in intraoperative T c by delivering warm fluids, compared with the other breathing circuits tested.

Trial Registration

Clinicaltrials.gov identifier: NCT01831843.
Literatur
1.
Zurück zum Zitat Rohrer MJ, Natale AM. Effect of hypothermia on the coagulation cascade. Crit Care Med. 1992;20:1402–5.CrossRefPubMed Rohrer MJ, Natale AM. Effect of hypothermia on the coagulation cascade. Crit Care Med. 1992;20:1402–5.CrossRefPubMed
2.
Zurück zum Zitat Rajagopalan S, Mascha E, Na J, Sessler DI. The effects of mild perioperative hypothermia on blood loss and transfusion requirement. Anesthesiology. 2008;108:71–7.CrossRefPubMed Rajagopalan S, Mascha E, Na J, Sessler DI. The effects of mild perioperative hypothermia on blood loss and transfusion requirement. Anesthesiology. 2008;108:71–7.CrossRefPubMed
3.
Zurück zum Zitat Lenhardt R, Marker E, Goll V, Tschernich H, Kurz A, Sessler DI, Narzt E, Lackner F. Mild intraoperative hypothermia prolongs postanesthetic recovery. Anesthesiology. 1997;87:1318–23.CrossRefPubMed Lenhardt R, Marker E, Goll V, Tschernich H, Kurz A, Sessler DI, Narzt E, Lackner F. Mild intraoperative hypothermia prolongs postanesthetic recovery. Anesthesiology. 1997;87:1318–23.CrossRefPubMed
4.
Zurück zum Zitat Polderman KH. Mechanisms of action, physiological effects, and complications of hypothermia. Crit Care Med. 2009;37:S186–202.CrossRefPubMed Polderman KH. Mechanisms of action, physiological effects, and complications of hypothermia. Crit Care Med. 2009;37:S186–202.CrossRefPubMed
5.
Zurück zum Zitat Camus Y, Delva E, Cohen S, Lienhart A. The effects of warming intravenous fluids on intraoperative hypothermia and postoperative shivering during prolonged abdominal surgery. Acta Anaesthesiol Scand. 1996;40:779–82.CrossRefPubMed Camus Y, Delva E, Cohen S, Lienhart A. The effects of warming intravenous fluids on intraoperative hypothermia and postoperative shivering during prolonged abdominal surgery. Acta Anaesthesiol Scand. 1996;40:779–82.CrossRefPubMed
6.
Zurück zum Zitat Ihn CH, Joo JD, Chung HS, Choi JW, Kim DW, Jeon YS, Kim YS, Choi WY. Comparison of three warming devices for the prevention of core hypothermia and post-anaesthesia shivering. J Int Med Res. 2008;36:923–31.CrossRefPubMed Ihn CH, Joo JD, Chung HS, Choi JW, Kim DW, Jeon YS, Kim YS, Choi WY. Comparison of three warming devices for the prevention of core hypothermia and post-anaesthesia shivering. J Int Med Res. 2008;36:923–31.CrossRefPubMed
7.
Zurück zum Zitat Smith CE, Gerdes E, Sweda S, Myles C, Punjabi A, Pinchak AC, Hagen JF. Warming intravenous fluids reduces perioperative hypothermia in women undergoing ambulatory gynecological surgery. Anesth Analg. 1998;87:37–41.PubMed Smith CE, Gerdes E, Sweda S, Myles C, Punjabi A, Pinchak AC, Hagen JF. Warming intravenous fluids reduces perioperative hypothermia in women undergoing ambulatory gynecological surgery. Anesth Analg. 1998;87:37–41.PubMed
8.
Zurück zum Zitat Yoo HS, Park SW, Yi JW, Kwon MI, Rhee YG. The effect of forced-air warming during arthroscopic shoulder surgery with general anesthesia. Arthroscopy. 2009;25:510–4.CrossRefPubMed Yoo HS, Park SW, Yi JW, Kwon MI, Rhee YG. The effect of forced-air warming during arthroscopic shoulder surgery with general anesthesia. Arthroscopy. 2009;25:510–4.CrossRefPubMed
9.
Zurück zum Zitat Matsukawa T, Sessler DI, Sessler AM, Schroeder M, Ozaki M, Kurz A, Cheng C. Heat flow and distribution during induction of general anesthesia. Anesthesiology. 1995;82:662–73.CrossRefPubMed Matsukawa T, Sessler DI, Sessler AM, Schroeder M, Ozaki M, Kurz A, Cheng C. Heat flow and distribution during induction of general anesthesia. Anesthesiology. 1995;82:662–73.CrossRefPubMed
10.
Zurück zum Zitat Nakagawa NK, Macchione M, Petrolino HM, Guimaraes ET, King M, Saldiva PH, Lorenzi-Filho G. Effects of a heat and moisture exchanger and a heated humidifier on respiratory mucus in patients undergoing mechanical ventilation. Crit Care Med. 2000;28:312–7.CrossRefPubMed Nakagawa NK, Macchione M, Petrolino HM, Guimaraes ET, King M, Saldiva PH, Lorenzi-Filho G. Effects of a heat and moisture exchanger and a heated humidifier on respiratory mucus in patients undergoing mechanical ventilation. Crit Care Med. 2000;28:312–7.CrossRefPubMed
11.
Zurück zum Zitat Hynson JM, Sessler DI. Intraoperative warming therapies: a comparison of three devices. J Clin Anesth. 1992;4:194–9.CrossRefPubMed Hynson JM, Sessler DI. Intraoperative warming therapies: a comparison of three devices. J Clin Anesth. 1992;4:194–9.CrossRefPubMed
12.
Zurück zum Zitat Lee HK, Jang YH, Choi KW, Lee JH. The effect of electrically heated humidifier on the body temperature and blood loss in spinal surgery under general anesthesia. Korean J Anesthesiol. 2011;61:112–6.PubMedCentralCrossRefPubMed Lee HK, Jang YH, Choi KW, Lee JH. The effect of electrically heated humidifier on the body temperature and blood loss in spinal surgery under general anesthesia. Korean J Anesthesiol. 2011;61:112–6.PubMedCentralCrossRefPubMed
13.
Zurück zum Zitat Stone DR, Downs JB, Paul WL, Perkins HM. Adult body temperature and heated humidification of anesthetic gases during general anesthesia. Anesth Analg. 1981;60:736–41.CrossRefPubMed Stone DR, Downs JB, Paul WL, Perkins HM. Adult body temperature and heated humidification of anesthetic gases during general anesthesia. Anesth Analg. 1981;60:736–41.CrossRefPubMed
14.
Zurück zum Zitat Goldberg ME, Epstein R, Rosenblum F, Larijani GE, Marr A, Lessin J, Torjman M, Seltzer J. Do heated humidifiers and heat and moisture exchangers prevent temperature drop during lower abdominal surgery? J Clin Anesth. 1992;4:16–20.CrossRefPubMed Goldberg ME, Epstein R, Rosenblum F, Larijani GE, Marr A, Lessin J, Torjman M, Seltzer J. Do heated humidifiers and heat and moisture exchangers prevent temperature drop during lower abdominal surgery? J Clin Anesth. 1992;4:16–20.CrossRefPubMed
15.
Zurück zum Zitat Jo YY, Kim HS, Chang YJ, Yun SY, Kwak HJ. The effect of warmed inspired gases on body temperature during arthroscopic shoulder surgery under general anesthesia. Korean J Anesthesiol. 2013;65:14–8.PubMedCentralCrossRefPubMed Jo YY, Kim HS, Chang YJ, Yun SY, Kwak HJ. The effect of warmed inspired gases on body temperature during arthroscopic shoulder surgery under general anesthesia. Korean J Anesthesiol. 2013;65:14–8.PubMedCentralCrossRefPubMed
16.
Zurück zum Zitat Ralley FE, Ramsay JG, Wynands JE, Townsend GE, Whalley DG, DelliColli P. Effect of heated humidified gases on temperature drop after cardiopulmonary bypass. Anesth Analg. 1984;63:1106–10.CrossRefPubMed Ralley FE, Ramsay JG, Wynands JE, Townsend GE, Whalley DG, DelliColli P. Effect of heated humidified gases on temperature drop after cardiopulmonary bypass. Anesth Analg. 1984;63:1106–10.CrossRefPubMed
17.
Zurück zum Zitat Kwak IS, Choi do Y, Lee TH, Bae JY, Lim TW, Kim KM. The effect of heated breathing circuit on body temperature and humidity of anesthetic gas in major burns. Korean J Anesthesiol. 2013;64:6–11.PubMedCentralCrossRefPubMed Kwak IS, Choi do Y, Lee TH, Bae JY, Lim TW, Kim KM. The effect of heated breathing circuit on body temperature and humidity of anesthetic gas in major burns. Korean J Anesthesiol. 2013;64:6–11.PubMedCentralCrossRefPubMed
18.
Zurück zum Zitat Kaufman RD. Relationship between esophageal temperature gradient and heart and lung sounds heard by esophageal stethoscope. Anesth Analg. 1987;66:1046–8.CrossRefPubMed Kaufman RD. Relationship between esophageal temperature gradient and heart and lung sounds heard by esophageal stethoscope. Anesth Analg. 1987;66:1046–8.CrossRefPubMed
20.
Zurück zum Zitat Leslie K, Sessler DI. Reduction in the shivering threshold is proportional to spinal block height. Anesthesiology. 1996;84:1327–31.CrossRefPubMed Leslie K, Sessler DI. Reduction in the shivering threshold is proportional to spinal block height. Anesthesiology. 1996;84:1327–31.CrossRefPubMed
21.
Zurück zum Zitat Turner M, Hodzovic I, Mapleson WW. Simulated clinical evaluation of four fluid warming devices. Anaesthesia. 2006;61:571–5.CrossRefPubMed Turner M, Hodzovic I, Mapleson WW. Simulated clinical evaluation of four fluid warming devices. Anaesthesia. 2006;61:571–5.CrossRefPubMed
22.
Zurück zum Zitat Presson RG Jr, Bezruczko AP, Hillier SC, McNiece WL. Evaluation of a new fluid warmer effective at low to moderate flow rates. Anesthesiology. 1993;78:974–80.CrossRefPubMed Presson RG Jr, Bezruczko AP, Hillier SC, McNiece WL. Evaluation of a new fluid warmer effective at low to moderate flow rates. Anesthesiology. 1993;78:974–80.CrossRefPubMed
23.
Zurück zum Zitat Horowitz PE, Delagarza MA, Pulaski JJ, Smith RA. Flow rates and warming efficacy with Hotline and Ranger blood/fluid warmers. Anesth Analg. 2004;99:788–92.CrossRefPubMed Horowitz PE, Delagarza MA, Pulaski JJ, Smith RA. Flow rates and warming efficacy with Hotline and Ranger blood/fluid warmers. Anesth Analg. 2004;99:788–92.CrossRefPubMed
24.
Zurück zum Zitat de Castro J Jr, Bolfi F, Carvalho LR, Braz JR. The temperature and humidity in a low-flow anesthesia workstation with and without a heat and moisture exchanger. Anesth Analg. 2011;113:534–8.PubMed de Castro J Jr, Bolfi F, Carvalho LR, Braz JR. The temperature and humidity in a low-flow anesthesia workstation with and without a heat and moisture exchanger. Anesth Analg. 2011;113:534–8.PubMed
25.
Zurück zum Zitat Chalon J, Loew DA, Malebranche J. Effects of dry anesthetic gases on tracheobronchial ciliated epithelium. Anesthesiology. 1972;37:338–43.CrossRefPubMed Chalon J, Loew DA, Malebranche J. Effects of dry anesthetic gases on tracheobronchial ciliated epithelium. Anesthesiology. 1972;37:338–43.CrossRefPubMed
26.
Zurück zum Zitat Pollett HF, Reid WD. Prevention of obstruction of nasopharyngeal CPAP tubes by adequate humidification of inspired gases. Can Anaesth Soc J. 1977;24:615–7.CrossRefPubMed Pollett HF, Reid WD. Prevention of obstruction of nasopharyngeal CPAP tubes by adequate humidification of inspired gases. Can Anaesth Soc J. 1977;24:615–7.CrossRefPubMed
27.
Zurück zum Zitat Jung TH, Rho JH, Hwang JH, Lee JH, Cha SC, Woo SC. The effect of the humidifier on sore throat and cough after thyroidectomy. Korean J Anesthesiol. 2011;61:470–4.PubMedCentralCrossRefPubMed Jung TH, Rho JH, Hwang JH, Lee JH, Cha SC, Woo SC. The effect of the humidifier on sore throat and cough after thyroidectomy. Korean J Anesthesiol. 2011;61:470–4.PubMedCentralCrossRefPubMed
28.
Zurück zum Zitat Patel N, Smith CE, Pinchak AC, Hagen JF. Comparison of esophageal, tympanic, and forehead skin temperatures in adult patients. J Clin Anesth. 1996;8:462–8.CrossRefPubMed Patel N, Smith CE, Pinchak AC, Hagen JF. Comparison of esophageal, tympanic, and forehead skin temperatures in adult patients. J Clin Anesth. 1996;8:462–8.CrossRefPubMed
Metadaten
Titel
Effect of a new heated and humidified breathing circuit with a fluid-warming device on intraoperative core temperature: a prospective randomized study
verfasst von
Eugene Kim
Sue-Young Lee
Young-Jin Lim
Jung-Yoon Choi
Young-Tae Jeon
Jung-Won Hwang
Hee-Pyoung Park
Publikationsdatum
01.08.2015
Verlag
Springer Japan
Erschienen in
Journal of Anesthesia / Ausgabe 4/2015
Print ISSN: 0913-8668
Elektronische ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-015-1994-z

Weitere Artikel der Ausgabe 4/2015

Journal of Anesthesia 4/2015 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.