In vitro study of accuracy of cervical pedicle screw insertion using an electronic conductivity device (ATPS part III)

verfasst von: Heiko Koller, Wolfgang Hitzl, Frank Acosta, Mark Tauber, Juliane Zenner, Herbert Resch, Yasutsugu Yukawa, Oliver Meier, Rene Schmidt, Michael Mayer

Erschienen in: European Spine Journal | Ausgabe 9/2009

Einloggen, um Zugang zu erhalten

Abstract

Reconstruction of the highly unstable, anteriorly decompressed cervical spine poses biomechanical challenges to current stabilization strategies, including circumferential instrumented fusion, to prevent failure. To avoid secondary posterior surgery, particularly in the elderly population, while increasing primary construct rigidity of anterior-only reconstructions, the authors introduced the concept of anterior transpedicular screw (ATPS) fixation and plating. We demonstrated its morphological feasibility, its superior biomechanical pull-out characteristics compared with vertebral body screws and the accuracy of inserting ATPS using a manual fluoroscopically assisted technique. Although accuracy was high, showing non-critical breaches in the axial and sagittal plane in 78 and 96%, further research was indicated refining technique and increasing accuracy. In light of first clinical case series, the authors analyzed the impact of using an electronic conductivity device (ECD, PediGuard) on the accuracy of ATPS insertion. As there exist only experiences in thoracolumbar surgery the versatility of the ECD was also assessed for posterior cervical pedicle screw fixation (pCPS). 30 ATPS and 30 pCPS were inserted alternately into the C3–T1 vertebra of five fresh-frozen specimen. Fluoroscopic assistance was only used for the entry point selection, pedicle tract preparation was done using the ECD. Preoperative CT scans were assessed for sclerosis at the pedicle entrance or core, and vertebrae with dense pedicles were excluded. Pre- and postoperative reconstructed CT scans were analyzed for pedicle screw positions according to a previously established grading system. Statistical analysis revealed an astonishingly high accuracy for the ATPS group with no critical screw position (0%) in axial or sagittal plane. In the pCPS group, 88.9% of screws inserted showed non-critical screw position, while 11.1% showed critical pedicle perforations. The usage of an ECD for posterior and anterior pedicle screw tract preparation with the exclusion of dense cortical pedicles was shown to be a successful and clinically sound concept with high-accuracy rates for ATPS and pCPS. In concert with fluoroscopic guidance and pedicle axis views, application of an ECD and exclusion of dense cortical pedicles might increase comfort and safety with the clinical use of pCPS. In addition, we presented a reasonable laboratory setting for the clinical introduction of an ATPS-plate system.
Literatur
4.
Zurück zum Zitat Acosta FL Jr, Aryan HE, Chou D, Ames CP (2008) Long-term biomechanical stability and clinical improvement after extended multilevel corpectomy and circumferential reconstruction of the cervical spine using titanium mesh cages. J Spinal Disord Tech 21:165–174. doi:10.1097/BSD.0b013e3180654205 PubMedCrossRef Acosta FL Jr, Aryan HE, Chou D, Ames CP (2008) Long-term biomechanical stability and clinical improvement after extended multilevel corpectomy and circumferential reconstruction of the cervical spine using titanium mesh cages. J Spinal Disord Tech 21:165–174. doi:10.​1097/​BSD.​0b013e3180654205​ PubMedCrossRef
5.
Zurück zum Zitat Acosta AH Jr, FL Ames CP (2006) Successful outcome of six-level cervicothoracic corpectomy and circumferential reconstruction: case report and review of literature on multilevel cervicothoracic corpectomy. Eur Spine J 15(Suppl 5):670–674. doi:10.1007/s00586-006-0203-z PubMedCrossRef Acosta AH Jr, FL Ames CP (2006) Successful outcome of six-level cervicothoracic corpectomy and circumferential reconstruction: case report and review of literature on multilevel cervicothoracic corpectomy. Eur Spine J 15(Suppl 5):670–674. doi:10.​1007/​s00586-006-0203-z PubMedCrossRef
6.
Zurück zum Zitat Aramomi M, Masaki Y, Koshizuka Kadota R, Okawa A, Koda M, Yamazaki M (2008) Anterior pedicle screw fixation for multilevel cervical corpectomy and spinal fusion. Acta Neurochir (Wien) 150:575–582. doi:10.1007/s00701-008-1574-1 CrossRef Aramomi M, Masaki Y, Koshizuka Kadota R, Okawa A, Koda M, Yamazaki M (2008) Anterior pedicle screw fixation for multilevel cervical corpectomy and spinal fusion. Acta Neurochir (Wien) 150:575–582. doi:10.​1007/​s00701-008-1574-1 CrossRef
7.
8.
Zurück zum Zitat Barrey C, Mertens P, Rumelhart C, Cotton F, Jund J, Perrin G (2004) Biomechanical evaluation of cervical lateral mass fixation: a comparison of the Roy-Camille and Magerl screw techniques. J Neurosurg (Spine 3) 100:268–276CrossRef Barrey C, Mertens P, Rumelhart C, Cotton F, Jund J, Perrin G (2004) Biomechanical evaluation of cervical lateral mass fixation: a comparison of the Roy-Camille and Magerl screw techniques. J Neurosurg (Spine 3) 100:268–276CrossRef
9.
Zurück zum Zitat Betz R (2008) Clinical application of the pediguard. Free Paper Presentation and Work-shop, Geneva/Switzerland Betz R (2008) Clinical application of the pediguard. Free Paper Presentation and Work-shop, Geneva/Switzerland
10.
Zurück zum Zitat Bolger C, Kelleher MO, McEvoy L, Brayda-Bruno Kaelin A, Lazennec J-Y, Le Huec J-C, Logroscino C, Mata P, Moreta P, Saillant G, Zeller R (2007) Electrical conductivity measurement: a new technique to detect iatrogenic initial pedicle perforation. Eur Spine J 16:1919–1924. doi:10.1007/s00586-007-0409-8 PubMedCrossRef Bolger C, Kelleher MO, McEvoy L, Brayda-Bruno Kaelin A, Lazennec J-Y, Le Huec J-C, Logroscino C, Mata P, Moreta P, Saillant G, Zeller R (2007) Electrical conductivity measurement: a new technique to detect iatrogenic initial pedicle perforation. Eur Spine J 16:1919–1924. doi:10.​1007/​s00586-007-0409-8 PubMedCrossRef
13.
Zurück zum Zitat Cha SH, Kim C, Choi BK, Kim HJ, Baek SY (2007) C-arm assessment of cervical pedicle screw: screw coaxial fluoroscopy and oblique view. Eur Spine J 32:1721–1727 Cha SH, Kim C, Choi BK, Kim HJ, Baek SY (2007) C-arm assessment of cervical pedicle screw: screw coaxial fluoroscopy and oblique view. Eur Spine J 32:1721–1727
15.
Zurück zum Zitat Choi WG, Tsan Y-S, Lee C-J, Lee S-H (2008) Microscopically assisted subaxial cervical pedicle screw fixation. Spineweek, Geneva Paper Choi WG, Tsan Y-S, Lee C-J, Lee S-H (2008) Microscopically assisted subaxial cervical pedicle screw fixation. Spineweek, Geneva Paper
18.
Zurück zum Zitat Deen HG, Nottmeier EW, Reimer R (2006) Early complications of posterior rod-screw fixation of the cervical and upper thoracic spine. Neurosurg 59:1062–1067 Deen HG, Nottmeier EW, Reimer R (2006) Early complications of posterior rod-screw fixation of the cervical and upper thoracic spine. Neurosurg 59:1062–1067
20.
Zurück zum Zitat Epstein NE (2008) What can we learn from a review of medicolegal malpractice suits involving cervical spine surgery. Annual meeting of the CSRS-A, Austin, TX, USA, Paper Epstein NE (2008) What can we learn from a review of medicolegal malpractice suits involving cervical spine surgery. Annual meeting of the CSRS-A, Austin, TX, USA, Paper
21.
Zurück zum Zitat Garces-Ambrossi GL, McGirt MJ, Parker SL, Mehta VA, Sciubba DM, Witham TF, Wolinsky JP, Ziya GK (2008) Trans-laminar versus pedicle screw fixation of C2: comparison of surgical morbidity and accuracy of 313 consecutive screws. Annual meeting of the CSRS-A, Austin, Texas, USA: Paper Garces-Ambrossi GL, McGirt MJ, Parker SL, Mehta VA, Sciubba DM, Witham TF, Wolinsky JP, Ziya GK (2008) Trans-laminar versus pedicle screw fixation of C2: comparison of surgical morbidity and accuracy of 313 consecutive screws. Annual meeting of the CSRS-A, Austin, Texas, USA: Paper
24.
Zurück zum Zitat Hasegawa K, Hirano T, Shimoda H, Homma T, Morita O (2008) Indications for cervical pedicle screw instrumentation in nontraumatic lesions. Spine 33:2284–2289PubMedCrossRef Hasegawa K, Hirano T, Shimoda H, Homma T, Morita O (2008) Indications for cervical pedicle screw instrumentation in nontraumatic lesions. Spine 33:2284–2289PubMedCrossRef
25.
Zurück zum Zitat Hee HT, Majd ME, Holt RT, Whitecloud TS 3rd, Pemkowski D (2003) Complications of multilevel cervical corpectomies and reconstruction with titanium cages and anterior plating. J Spinal Disord Tech 16:1–9PubMed Hee HT, Majd ME, Holt RT, Whitecloud TS 3rd, Pemkowski D (2003) Complications of multilevel cervical corpectomies and reconstruction with titanium cages and anterior plating. J Spinal Disord Tech 16:1–9PubMed
26.
Zurück zum Zitat Heller SD, Sutterlin CE (1995) Complications of posterior cervical plating. Spine 20:2442–2448 Heller SD, Sutterlin CE (1995) Complications of posterior cervical plating. Spine 20:2442–2448
31.
33.
Zurück zum Zitat Jeanneret B, Gebhard JS, Magerl F (1994) Transpedicular screw fixation of articular mass fracture-separation: results of an anatomical study and operative technique. Spine 19:2529–2539CrossRef Jeanneret B, Gebhard JS, Magerl F (1994) Transpedicular screw fixation of articular mass fracture-separation: results of an anatomical study and operative technique. Spine 19:2529–2539CrossRef
36.
Zurück zum Zitat Koller H, Acosta F, Tauber M, Fox M, Martin H, Forstner R, Augat P, Penzkofer R, Pirich C, Kässmann H, Resch H, Hitzl W (2008) Cervical anterior transpedicular screw fixation (ATPS): Part II. Accuracy of manual insertion and pull-out strenght of ATPS. Eur Spine J 17:539–555. doi:10.1007/s00586-007-0573-x PubMedCrossRef Koller H, Acosta F, Tauber M, Fox M, Martin H, Forstner R, Augat P, Penzkofer R, Pirich C, Kässmann H, Resch H, Hitzl W (2008) Cervical anterior transpedicular screw fixation (ATPS): Part II. Accuracy of manual insertion and pull-out strenght of ATPS. Eur Spine J 17:539–555. doi:10.​1007/​s00586-007-0573-x PubMedCrossRef
37.
Zurück zum Zitat Koller H, Hempfing A, Acosta F, Fox M, Scheiter A, Tauber M, Holz U, Resch H, Hitzl W (2008) Cervical anterior transpedicular screw fixation. Part I: study on morphological feasibility, indications, and technical prerequisites. Eur Spine J 17:523–538. doi:10.1007/s00586-007-0572-y PubMedCrossRef Koller H, Hempfing A, Acosta F, Fox M, Scheiter A, Tauber M, Holz U, Resch H, Hitzl W (2008) Cervical anterior transpedicular screw fixation. Part I: study on morphological feasibility, indications, and technical prerequisites. Eur Spine J 17:523–538. doi:10.​1007/​s00586-007-0572-y PubMedCrossRef
38.
Zurück zum Zitat Koller H, Hempfing A, Ferraris L, Meier O, Metz-Stavenhagen P (2006) 4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results. Eur Spine J 16:2055–2071. doi:10.1007/s00586-007-0398-7 CrossRef Koller H, Hempfing A, Ferraris L, Meier O, Metz-Stavenhagen P (2006) 4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results. Eur Spine J 16:2055–2071. doi:10.​1007/​s00586-007-0398-7 CrossRef
39.
Zurück zum Zitat Koller H, Zenner J, Forstner R, Hempfing A, Maislinger I, Kolb K, Tauber M, Resch H, Mayer M, Hitzl W (2009) Mid- to long-term outcome of instrumented anterior cervical fusion for subaxial injuries. Eur Spine J 18(5):630–653 Koller H, Zenner J, Forstner R, Hempfing A, Maislinger I, Kolb K, Tauber M, Resch H, Mayer M, Hitzl W (2009) Mid- to long-term outcome of instrumented anterior cervical fusion for subaxial injuries. Eur Spine J 18(5):630–653
40.
Zurück zum Zitat Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-asisted cervical pedicle screw insertion. J Neurosurg (Spine 3) 99:257–263CrossRef Kotani Y, Abumi K, Ito M, Minami A (2003) Improved accuracy of computer-asisted cervical pedicle screw insertion. J Neurosurg (Spine 3) 99:257–263CrossRef
43.
Zurück zum Zitat Meier O, Hempfing A, Koller H (2008) Answer to the Letter to the Editor concerning ′4- and 5-level anterior fusions of the cervical spine: “natura non-facit saltus” ‘by Franco JL. Eur Spine J 17:1121–1122. doi:10.1007/s00586-008-0697-7 CrossRef Meier O, Hempfing A, Koller H (2008) Answer to the Letter to the Editor concerning ′4- and 5-level anterior fusions of the cervical spine: “natura non-facit saltus” ‘by Franco JL. Eur Spine J 17:1121–1122. doi:10.​1007/​s00586-008-0697-7 CrossRef
44.
Zurück zum Zitat Melcher RP, Harms J, Boyaci B, Schmelzle R (2008) Resection and reconstruction of destructive lesions of the second cervical vertebra. Eur Spine J 17:1565 Melcher RP, Harms J, Boyaci B, Schmelzle R (2008) Resection and reconstruction of destructive lesions of the second cervical vertebra. Eur Spine J 17:1565
47.
Zurück zum Zitat Murakami H, Yamazaki K, Yoshida S, Kazuhiro S, Endo H, Shimamura T (2008) Posterior occipitocervical reconstruction using cervical pedicle screw system for rheumatoid arthritis: comparison with segmental spinal instrumentation. Spineweek, Geneva Murakami H, Yamazaki K, Yoshida S, Kazuhiro S, Endo H, Shimamura T (2008) Posterior occipitocervical reconstruction using cervical pedicle screw system for rheumatoid arthritis: comparison with segmental spinal instrumentation. Spineweek, Geneva
48.
Zurück zum Zitat Namikawa T, Taneichi H, Inami S, Daisaku T, Nakayuki K, Nohara Y (2008) Morphological analysis of the pedicles with risk of screw malplacement in degenerative cervical spine. Spineweek, Geneva Namikawa T, Taneichi H, Inami S, Daisaku T, Nakayuki K, Nohara Y (2008) Morphological analysis of the pedicles with risk of screw malplacement in degenerative cervical spine. Spineweek, Geneva
49.
Zurück zum Zitat Narotam PK, Pauley SM, McGinn GJ (2003) Titanium mesh cages for cervical spine stabilization after corpectomy: a clinical and radiological study. J Neurosurg Spine 99:172–180CrossRef Narotam PK, Pauley SM, McGinn GJ (2003) Titanium mesh cages for cervical spine stabilization after corpectomy: a clinical and radiological study. J Neurosurg Spine 99:172–180CrossRef
53.
Zurück zum Zitat Pitzen T, Barbier D, Tintinger F, Steudel WI, Strowitzki M (2002) Screw fixation to the posterior cortical shell does not influence peak torque and pullout in anterior cervical plating. Eur Spine J 11:494–499. doi:10.1007/s00586-002-0447-1 PubMedCrossRef Pitzen T, Barbier D, Tintinger F, Steudel WI, Strowitzki M (2002) Screw fixation to the posterior cortical shell does not influence peak torque and pullout in anterior cervical plating. Eur Spine J 11:494–499. doi:10.​1007/​s00586-002-0447-1 PubMedCrossRef
54.
Zurück zum Zitat Pitzen T, Drumm J, Steudel WI, Allam Y, Branea I, Sova L (2008) The complication rate in multisegmental cervical spine surgery is at least doubled when compared to routine cervical spine procedures. Eur Spine J 17:1621 Pitzen T, Drumm J, Steudel WI, Allam Y, Branea I, Sova L (2008) The complication rate in multisegmental cervical spine surgery is at least doubled when compared to routine cervical spine procedures. Eur Spine J 17:1621
55.
Zurück zum Zitat Rath SA, Moszko S, Schäffner PM, Cantone G, Braun V, Richter HP, Antoniadis G (2008) Accuracy of pedicle screw insertion in the cervical spine for internal fixation using frameless stereotactic guidance. J Neurosurg Spine 8:237–245. doi:10.3171/SPI/2008/8/3/237 PubMedCrossRef Rath SA, Moszko S, Schäffner PM, Cantone G, Braun V, Richter HP, Antoniadis G (2008) Accuracy of pedicle screw insertion in the cervical spine for internal fixation using frameless stereotactic guidance. J Neurosurg Spine 8:237–245. doi:10.​3171/​SPI/​2008/​8/​3/​237 PubMedCrossRef
56.
58.
Zurück zum Zitat Roh JS, Teng AL, Rice JA, Huang RC, Davis JA (2004) Accurate cervical pedicle screw placement using laser-guided fluoroscopy: the ‘perfect pedicle’ technique. 32nd Annual Meeting, Boston Roh JS, Teng AL, Rice JA, Huang RC, Davis JA (2004) Accurate cervical pedicle screw placement using laser-guided fluoroscopy: the ‘perfect pedicle’ technique. 32nd Annual Meeting, Boston
60.
Zurück zum Zitat Russo A, Nguyen Q, Bellabarba C, Brnasford R, Chapman JR (2008) Posterior C2 instrumentation: complications and risk. Annual meeting of the CSRS-A, Austin, TX, USA Russo A, Nguyen Q, Bellabarba C, Brnasford R, Chapman JR (2008) Posterior C2 instrumentation: complications and risk. Annual meeting of the CSRS-A, Austin, TX, USA
62.
Zurück zum Zitat Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2005) Effect of constrained posterior screw and rod systems for primary stability: biomechanical in vitro comparison of various instrumentations in a single-level corporectomy model. Eur Spine J 14:372–380. doi:10.1007/s00586-004-0763-8 PubMedCrossRef Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2005) Effect of constrained posterior screw and rod systems for primary stability: biomechanical in vitro comparison of various instrumentations in a single-level corporectomy model. Eur Spine J 14:372–380. doi:10.​1007/​s00586-004-0763-8 PubMedCrossRef
63.
Zurück zum Zitat Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corporectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instruments. Spine 16:1821–1828. doi:10.1097/01.BRS.0000083287.23521.48 CrossRef Schmidt R, Wilke HJ, Claes L, Puhl W, Richter M (2003) Pedicle screws enhance primary stability in multilevel cervical corporectomies: biomechanical in vitro comparison of different implants including constrained and nonconstrained posterior instruments. Spine 16:1821–1828. doi:10.​1097/​01.​BRS.​0000083287.​23521.​48 CrossRef
64.
Zurück zum Zitat Shamji MF, Cook C, Pietrobon R, Tackett S, Brown C, Isaacs RE (2008) Impact of surgical approach on complications and resource utilization of cervical spine fusion: a nationwide perspective to the surgical treatment of diffuse cervical spondylosis. Spine J 9:10–16 Shamji MF, Cook C, Pietrobon R, Tackett S, Brown C, Isaacs RE (2008) Impact of surgical approach on complications and resource utilization of cervical spine fusion: a nationwide perspective to the surgical treatment of diffuse cervical spondylosis. Spine J 9:10–16
65.
Zurück zum Zitat Shen HX, Buchowski JM, Yeom JS, Liu G, Riew KD (2008) Pseudoarthrosis in multi-level anterior cervical fusion with rhBMP-2 and allograft: analysis of 127 cases with minimum 2-year follow-up. IMAST, Hongong/China, Paper Shen HX, Buchowski JM, Yeom JS, Liu G, Riew KD (2008) Pseudoarthrosis in multi-level anterior cervical fusion with rhBMP-2 and allograft: analysis of 127 cases with minimum 2-year follow-up. IMAST, Hongong/China, Paper
66.
Zurück zum Zitat Smajic S, Krauspe R, Muhr G, Schinkel C (2008) Zervikale Wirbelfrakturen bei Patienten mit ankylosierender Spondylitis. Annual Meeting of the German Societies for Trauma and Orthopedic Surgery, Berlin, FRG, Paper Smajic S, Krauspe R, Muhr G, Schinkel C (2008) Zervikale Wirbelfrakturen bei Patienten mit ankylosierender Spondylitis. Annual Meeting of the German Societies for Trauma and Orthopedic Surgery, Berlin, FRG, Paper
68.
Zurück zum Zitat Steinmetz MP, Stewart TJ, Kager CD, Benzel EC, Vaccaro AR (2007) Cervical deformity correction. Neurosurg 60(Suppl):S90–S97CrossRef Steinmetz MP, Stewart TJ, Kager CD, Benzel EC, Vaccaro AR (2007) Cervical deformity correction. Neurosurg 60(Suppl):S90–S97CrossRef
71.
Zurück zum Zitat Suda K, Kajino T, Moridaira H, Limoto S, Taneichi H (2008) How to avoid fatal vascular complications caused by pedicle screws: surgical strategy for safe screw placement. Spineweek, Geneva Suda K, Kajino T, Moridaira H, Limoto S, Taneichi H (2008) How to avoid fatal vascular complications caused by pedicle screws: surgical strategy for safe screw placement. Spineweek, Geneva
72.
74.
Zurück zum Zitat Tay B, Campos Daziano C, Geula D, Chi J, Deviren V (2008) Circumferential osteotomy for fixed cervical kyphosis: a novel surgical technique. Spineweek, Geneva Tay B, Campos Daziano C, Geula D, Chi J, Deviren V (2008) Circumferential osteotomy for fixed cervical kyphosis: a novel surgical technique. Spineweek, Geneva
75.
Zurück zum Zitat Tomasino A, Parikh K, Koller H, Zink W, Tsiouris AJ, Steinberger J, Härtl R (2009) The vertebral artery and the cervical pedicle: morphometrical analysis of a critical neighborhood. J Neurosurg (in review process personal communication) Tomasino A, Parikh K, Koller H, Zink W, Tsiouris AJ, Steinberger J, Härtl R (2009) The vertebral artery and the cervical pedicle: morphometrical analysis of a critical neighborhood. J Neurosurg (in review process personal communication)
77.
79.
Zurück zum Zitat Westerveld LA, Verlaan JJ, Oner FC (2008) Spinal fractures in patients with ankylosing spinal disorders: a systematic review of the literature on treatment, neurological status and complications. Eur Spine J Eur Spine J: 14:5–156 Westerveld LA, Verlaan JJ, Oner FC (2008) Spinal fractures in patients with ankylosing spinal disorders: a systematic review of the literature on treatment, neurological status and complications. Eur Spine J Eur Spine J: 14:5–156
82.
Zurück zum Zitat Yukawa Y, Kato F, Ito K, Horie Y, Nakashima H, Ito Z-y, Wakao N (2008) Placement and complications of cervical pedicle screw in 144 cervical trauma patients using pedicle axis view techniques. Spineweek, Geneva Yukawa Y, Kato F, Ito K, Horie Y, Nakashima H, Ito Z-y, Wakao N (2008) Placement and complications of cervical pedicle screw in 144 cervical trauma patients using pedicle axis view techniques. Spineweek, Geneva
83.
Zurück zum Zitat Yukawa Y, Kato F, Yoshirara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg 5:488–493 Yukawa Y, Kato F, Yoshirara H, Yanase M, Ito K (2006) Cervical pedicle screw fixation in 100 cases of unstable cervical injuries: pedicle axis views obtained using fluoroscopy. J Neurosurg 5:488–493
Metadaten
Titel
In vitro study of accuracy of cervical pedicle screw insertion using an electronic conductivity device (ATPS part III)
verfasst von
Heiko Koller
Wolfgang Hitzl
Frank Acosta
Mark Tauber
Juliane Zenner
Herbert Resch
Yasutsugu Yukawa
Oliver Meier
Rene Schmidt
Michael Mayer
Publikationsdatum
01.09.2009
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe 9/2009
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-009-1054-1

Weitere Artikel der Ausgabe 9/2009

European Spine Journal 9/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.