Interspinous spacers in the treatment of degenerative lumbar spinal disease: our experience with DIAM and Aperius devices

verfasst von: Antonio P. Fabrizi, Raffaella Maina, Luigi Schiabello

Erschienen in: European Spine Journal | Sonderheft 1/2011

Einloggen, um Zugang zu erhalten

Abstract

Insertion of an interspinous devices has became a common procedure for the treatment of different clinical picture of degenerative spinal disease. We present our experience in 1,575 patients with the use of two different interspinous spacers: Device for Intervertebral Assisted Motion (DIAM) and Aperius PercLID system. From 2000 through 2008, 1,315 consecutive patients underwent DIAM implantation and 260 had an Aperius PercLID procedure. The main surgical indications included: degenerative disc disease (478 patients), canal and/or foraminal stenosis (347 patients), disc herniation (283 patients), black disc and facet syndrome (143) and topping-off (64 patients). 1,100 patients underwent a single level implant and 475 had a multiple level implant. Mean operating time was 35 min for DIAM and 7 min for Aperius. Complications were detected in 20 patients (10 cases of infections, 10 fractures of the posterior spinous processes). 40 patients were subsequently treated with posterior arthrodesis (n = 30) or total disc replacement (n = 10). Patient’s postoperative clinical status was rated according to the modified Macnab criteria: symptoms resolution or improvement was achieved in 1,505 patients; and unchanged or unsatisfactory results in 70. Both techniques are safe, simple and less technically demanding. These approaches appear to be an effective alternative in selected cases, although conventional posterior lumbar decompression and fusion still may be required.
Literatur
1.
Zurück zum Zitat Lindsey DP, Swanson KE, Fuchs P, Hsu KY, Zucherman JF, Yerby SA (2003) The effects of an interspinous implant on the kinematics of the instrumented and adjacent levels in the lumbar spine. Spine 28(19):2192–2197PubMedCrossRef Lindsey DP, Swanson KE, Fuchs P, Hsu KY, Zucherman JF, Yerby SA (2003) The effects of an interspinous implant on the kinematics of the instrumented and adjacent levels in the lumbar spine. Spine 28(19):2192–2197PubMedCrossRef
2.
Zurück zum Zitat Lee J, Hida K, Seki T, Iwasaki Y, Minoru A (2004) An interspinous process distractor (X STOP) for lumbar spinal stenosis in elderly patients: preliminary experiences in 10 consecutive cases. J Spinal Disord Tech 17(1):72–77PubMedCrossRef Lee J, Hida K, Seki T, Iwasaki Y, Minoru A (2004) An interspinous process distractor (X STOP) for lumbar spinal stenosis in elderly patients: preliminary experiences in 10 consecutive cases. J Spinal Disord Tech 17(1):72–77PubMedCrossRef
3.
Zurück zum Zitat Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Influence of X stop on neural foramina and spinal canal area in spinal stenosis. Spine 31(25):2958–2962PubMedCrossRef Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Influence of X stop on neural foramina and spinal canal area in spinal stenosis. Spine 31(25):2958–2962PubMedCrossRef
4.
Zurück zum Zitat Siddiqui M, Nicol M, Karadimas E, Smith F, Wardlaw D (2005) The positional magnetic resonance imaging changes in the lumbar spine following insertion of a novel interspinous process distraction device. Spine 30(23):2677–2682PubMedCrossRef Siddiqui M, Nicol M, Karadimas E, Smith F, Wardlaw D (2005) The positional magnetic resonance imaging changes in the lumbar spine following insertion of a novel interspinous process distraction device. Spine 30(23):2677–2682PubMedCrossRef
5.
Zurück zum Zitat Richards JC, Majumdar S, Lindsey DP, Beaupre GS, Yerby SA (2005) The treatment mechanism of an interspinous process implant for lumbar neurogenic intermittent claudication. Spine 30(7):744–749PubMedCrossRef Richards JC, Majumdar S, Lindsey DP, Beaupre GS, Yerby SA (2005) The treatment mechanism of an interspinous process implant for lumbar neurogenic intermittent claudication. Spine 30(7):744–749PubMedCrossRef
6.
Zurück zum Zitat Bono CM, Vaccaro AR (2007) Interspinous process devices in the lumbar spine. J Spinal Disord Tech 20(3):255–261PubMedCrossRef Bono CM, Vaccaro AR (2007) Interspinous process devices in the lumbar spine. J Spinal Disord Tech 20(3):255–261PubMedCrossRef
7.
Zurück zum Zitat Wilke HJ, Drumm J, Haussler K, Mack C, Steudel WI, Kettler A (2008) Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure. Eur Spine J 17(8):1049–1056PubMedCrossRef Wilke HJ, Drumm J, Haussler K, Mack C, Steudel WI, Kettler A (2008) Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure. Eur Spine J 17(8):1049–1056PubMedCrossRef
8.
Zurück zum Zitat Gunzburg R, Szpalski M, Callary SA, Colloca CJ, Kosmopoulos V, Harrison D, Moore RJ (2009) Effect of a novel interspinous implant on lumbar spinal range of motion. Eur Spine J 18(5):696–703PubMedCrossRef Gunzburg R, Szpalski M, Callary SA, Colloca CJ, Kosmopoulos V, Harrison D, Moore RJ (2009) Effect of a novel interspinous implant on lumbar spinal range of motion. Eur Spine J 18(5):696–703PubMedCrossRef
9.
Zurück zum Zitat Swanson KE, Lindsey DP, Hsu KY, Zucherman JF, Yerby SA (2003) The effects of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976) 28(1):26–32 Swanson KE, Lindsey DP, Hsu KY, Zucherman JF, Yerby SA (2003) The effects of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976) 28(1):26–32
10.
Zurück zum Zitat Senegas J (2002) Mechanical supplementation by non-rigid fixation in degenerative intervertebral lumbar segments: the Wallis system. Eur Spine J 11(Suppl 2):S164–S169PubMed Senegas J (2002) Mechanical supplementation by non-rigid fixation in degenerative intervertebral lumbar segments: the Wallis system. Eur Spine J 11(Suppl 2):S164–S169PubMed
11.
Zurück zum Zitat Caserta S, La Maida GA, Misaggi B, Peroni D, Pietrabissa R, Raimondi MT, Redaelli A (2002) Elastic stabilization alone or combined with rigid fusion in spinal surgery: a biomechanical study and clinical experience based on 82 cases. Eur Spine J 11(Suppl 2):S192–S197PubMed Caserta S, La Maida GA, Misaggi B, Peroni D, Pietrabissa R, Raimondi MT, Redaelli A (2002) Elastic stabilization alone or combined with rigid fusion in spinal surgery: a biomechanical study and clinical experience based on 82 cases. Eur Spine J 11(Suppl 2):S192–S197PubMed
12.
Zurück zum Zitat Whitesides TE Jr (2003) The effect of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976) 28(16):1906–1907 Whitesides TE Jr (2003) The effect of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976) 28(16):1906–1907
13.
Zurück zum Zitat Bellini CM, Galbusera F, Raimondi MT, Mineo GV, Brayda-Bruno M (2007) Biomechanics of the lumbar spine after dynamic stabilization. J Spinal Disord Tech 20(6):423–429PubMedCrossRef Bellini CM, Galbusera F, Raimondi MT, Mineo GV, Brayda-Bruno M (2007) Biomechanics of the lumbar spine after dynamic stabilization. J Spinal Disord Tech 20(6):423–429PubMedCrossRef
14.
Zurück zum Zitat Mariottini A, Pieri S, Giachi S, Carangelo B, Zalaffi A, Muzii FV, Palma L (2005) Preliminary results of a soft novel lumbar intervertebral prothesis (DIAM) in the degenerative spinal pathology. Acta Neurochir Suppl 92:129–131PubMedCrossRef Mariottini A, Pieri S, Giachi S, Carangelo B, Zalaffi A, Muzii FV, Palma L (2005) Preliminary results of a soft novel lumbar intervertebral prothesis (DIAM) in the degenerative spinal pathology. Acta Neurochir Suppl 92:129–131PubMedCrossRef
15.
Zurück zum Zitat Taylor J, Pupin P, Delajoux S, Palmer S (2007) Device for intervertebral assisted motion: technique and initial results. Neurosurg Focus 22(1):E6PubMedCrossRef Taylor J, Pupin P, Delajoux S, Palmer S (2007) Device for intervertebral assisted motion: technique and initial results. Neurosurg Focus 22(1):E6PubMedCrossRef
16.
Zurück zum Zitat Lauryssen C (2007) Appropriate selection of patients with lumbar spinal stenosis for interspinous process decompression with the X STOP device. Neurosurg Focus 22(1):E5PubMedCrossRef Lauryssen C (2007) Appropriate selection of patients with lumbar spinal stenosis for interspinous process decompression with the X STOP device. Neurosurg Focus 22(1):E5PubMedCrossRef
17.
Zurück zum Zitat Siddiqui M, Smith FW, Wardlaw D (2007) One-year results of X Stop interspinous implant for the treatment of lumbar spinal stenosis. Spine 32(12):1345–1348PubMedCrossRef Siddiqui M, Smith FW, Wardlaw D (2007) One-year results of X Stop interspinous implant for the treatment of lumbar spinal stenosis. Spine 32(12):1345–1348PubMedCrossRef
18.
Zurück zum Zitat Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Effects of X-STOP device on sagittal lumbar spine kinematics in spinal stenosis. J Spinal Disord Tech 19(5):328–333PubMedCrossRef Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Effects of X-STOP device on sagittal lumbar spine kinematics in spinal stenosis. J Spinal Disord Tech 19(5):328–333PubMedCrossRef
19.
Zurück zum Zitat Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA (2005) The effect of an interspinous process implant on facet loading during extension. Spine (Phila Pa 1976) 30(8):903–907 Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA (2005) The effect of an interspinous process implant on facet loading during extension. Spine (Phila Pa 1976) 30(8):903–907
20.
Zurück zum Zitat Brussee P, Hauth J, Donk RD, Verbeek AL, Bartels RH (2008) Self-rated evaluation of outcome of the implantation of interspinous process distraction (X-Stop) for neurogenic claudication. Eur Spine J 17(2):200–203PubMedCrossRef Brussee P, Hauth J, Donk RD, Verbeek AL, Bartels RH (2008) Self-rated evaluation of outcome of the implantation of interspinous process distraction (X-Stop) for neurogenic claudication. Eur Spine J 17(2):200–203PubMedCrossRef
21.
Zurück zum Zitat Kuchta J, Sobottke R, Eysel P, Simons P (2009) Two-year results of interspinous spacer (X-Stop) implantation in 175 patients with neurologic intermittent claudication due to lumbar spinal stenosis. Eur Spine J 18(6):823–829PubMedCrossRef Kuchta J, Sobottke R, Eysel P, Simons P (2009) Two-year results of interspinous spacer (X-Stop) implantation in 175 patients with neurologic intermittent claudication due to lumbar spinal stenosis. Eur Spine J 18(6):823–829PubMedCrossRef
22.
Zurück zum Zitat Kondrashov DG, Hannibal M, Hsu KY, Zucherman JF (2006) Interspinous process decompression with the X-STOP device for lumbar spinal stenosis: a 4-year follow-up study. J Spinal Disord Tech 19(5):323–327PubMedCrossRef Kondrashov DG, Hannibal M, Hsu KY, Zucherman JF (2006) Interspinous process decompression with the X-STOP device for lumbar spinal stenosis: a 4-year follow-up study. J Spinal Disord Tech 19(5):323–327PubMedCrossRef
23.
Zurück zum Zitat Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio S, Cauthen JC, Ozuna RM (2004) A prospective randomized multi-center study for the treatment of lumbar spinal stenosis with the X STOP interspinous implant: 1-year results. Eur Spine J 13(1):22–31PubMedCrossRef Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio S, Cauthen JC, Ozuna RM (2004) A prospective randomized multi-center study for the treatment of lumbar spinal stenosis with the X STOP interspinous implant: 1-year results. Eur Spine J 13(1):22–31PubMedCrossRef
24.
Zurück zum Zitat Anderson PA, Tribus CB, Kitchel SH (2006) Treatment of neurogenic claudication by interspinous decompression: application of the X STOP device in patients with lumbar degenerative spondylolisthesis. J Neurosurg Spine 4(6):463–471PubMedCrossRef Anderson PA, Tribus CB, Kitchel SH (2006) Treatment of neurogenic claudication by interspinous decompression: application of the X STOP device in patients with lumbar degenerative spondylolisthesis. J Neurosurg Spine 4(6):463–471PubMedCrossRef
25.
Zurück zum Zitat Hsu KY, Zucherman JF, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Cauthen JC, Ozuna RM (2006) Quality of life of lumbar stenosis-treated patients in whom the X STOP interspinous device was implanted. J Neurosurg Spine 5(6):500–507PubMedCrossRef Hsu KY, Zucherman JF, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Cauthen JC, Ozuna RM (2006) Quality of life of lumbar stenosis-treated patients in whom the X STOP interspinous device was implanted. J Neurosurg Spine 5(6):500–507PubMedCrossRef
26.
Zurück zum Zitat Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio ST, Cauthen JC, Ozuna RM (2005) A multicenter, prospective, randomized trial evaluating the X STOP Interspinous process decompression system for the treatment of neurogenic intermittent claudication: two-year follow-up results. Spine 30(12):1351–1358PubMedCrossRef Zucherman JF, Hsu KY, Hartjen CA, Mehalic TF, Implicito DA, Martin MJ, Johnson DR, Skidmore GA, Vessa PP, Dwyer JW, Puccio ST, Cauthen JC, Ozuna RM (2005) A multicenter, prospective, randomized trial evaluating the X STOP Interspinous process decompression system for the treatment of neurogenic intermittent claudication: two-year follow-up results. Spine 30(12):1351–1358PubMedCrossRef
27.
Zurück zum Zitat Vena P, Franzoso G, Gastaldi D, Contro R, Dallolio V (2005) A finite element model of the L4–L5 spinal motion segment: biomechanical compatibility of an interspinous device. Comput Methods Biomech Biomed Eng 8(1):7–16CrossRef Vena P, Franzoso G, Gastaldi D, Contro R, Dallolio V (2005) A finite element model of the L4–L5 spinal motion segment: biomechanical compatibility of an interspinous device. Comput Methods Biomech Biomed Eng 8(1):7–16CrossRef
28.
Zurück zum Zitat Phillips FM, Voronov LI, Gaitanis IN, Carandang G, Havey RM, Patwardhan AG (2006) Biomechanics of posterior dynamic stabilizing device (DIAM) after facetectomy and discectomy. Spine J 6(6):714–722PubMedCrossRef Phillips FM, Voronov LI, Gaitanis IN, Carandang G, Havey RM, Patwardhan AG (2006) Biomechanics of posterior dynamic stabilizing device (DIAM) after facetectomy and discectomy. Spine J 6(6):714–722PubMedCrossRef
29.
Zurück zum Zitat Kim KA, McDonald M, Pik JH, Khoueir P, Wang MY (2007) Dynamic intraspinous spacer technology for posterior stabilization: case–control study on the safety, sagittal angulation, and pain outcome at 1-year follow-up evaluation. Neurosurg Focus 22(1):E7PubMed Kim KA, McDonald M, Pik JH, Khoueir P, Wang MY (2007) Dynamic intraspinous spacer technology for posterior stabilization: case–control study on the safety, sagittal angulation, and pain outcome at 1-year follow-up evaluation. Neurosurg Focus 22(1):E7PubMed
30.
Zurück zum Zitat Verhoof OJ, Bron JL, Wapstra FH, van Royen BJ (2008) High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J 17(2):188–192PubMedCrossRef Verhoof OJ, Bron JL, Wapstra FH, van Royen BJ (2008) High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J 17(2):188–192PubMedCrossRef
31.
Zurück zum Zitat Schnake KJ, Putzier M, Haas NP, Kandziora F (2006) Mechanical concepts for disc regeneration. Eur Spine J 15(Suppl 3):S354–S360PubMedCrossRef Schnake KJ, Putzier M, Haas NP, Kandziora F (2006) Mechanical concepts for disc regeneration. Eur Spine J 15(Suppl 3):S354–S360PubMedCrossRef
Metadaten
Titel
Interspinous spacers in the treatment of degenerative lumbar spinal disease: our experience with DIAM and Aperius devices
verfasst von
Antonio P. Fabrizi
Raffaella Maina
Luigi Schiabello
Publikationsdatum
01.05.2011
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe Sonderheft 1/2011
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-011-1753-2

Weitere Artikel der Sonderheft 1/2011

European Spine Journal 1/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.