Skip to main content
Erschienen in: European Spine Journal 12/2013

01.12.2013 | Original Article

Use of an ultrasonic osteotome device in spine surgery: experience from the first 128 patients

verfasst von: Xiaobang Hu, Donna D. Ohnmeiss, Isador H. Lieberman

Erschienen in: European Spine Journal | Ausgabe 12/2013

Einloggen, um Zugang zu erhalten

Abstract

Introduction

The ultrasonic BoneScalpel is a tissue-specific device that allows the surgeon to make precise osteotomies while protecting collateral or adjacent soft tissue structures. The device is comprised of a blunt ultrasonic blade that oscillates at over 22,500 cycles/s with an imperceptible microscopic amplitude. The recurring impacts pulverize the noncompliant crystalline structure resulting in a precise cut. The more compliant adjacent soft tissue is not affected by the ultrasonic oscillation. The purpose of this study is to report the experience and safety of using this ultrasonic osteotome device in a variety of spine surgeries.

Methods

Data were retrospectively collected from medical charts and surgical reports for each surgery in which the ultrasonic scalpel was used to perform any type of osteotomy (facetectomy, laminotomy, laminectomy, en bloc resection, Smith Petersen osteotomy, pedicle subtraction osteotomy, etc.). The majority of patients had spinal stenosis, degenerative or adolescent scoliosis, pseudoarthrosis, adjacent segment degeneration, and spondylolisthesis et al. Intra-operative complications were also recorded.

Results

A total of 128 consecutive patients (73 female, 55 male) beginning with our first case experience were included in this study. The mean age of the patients was 58 years (range 12–85 years). Eighty patients (62.5 %) had previous spine surgery and/or spinal deformity. The ultrasonic scalpel was used at all levels of the spine and the average levels operated on each patient were 5. The mean operation time (skin to skin) was 4.3 h and the mean blood loss was 425.4 ml. In all cases, the ultrasonic scalpel was used to create the needed osteotomies to facilitate the surgical procedure without any percussion on the spinal column or injury to the underlying nerves. There was a noticeable absence of bleeding from the cut end of the bone consistent with the ultrasonic application. There were 11 instances of dural injuries (8.6 %) and two of which were directly associated with the use of ultrasonic device. In no procedure was the use of the ultrasonic scalpel abandoned for use of another instrument due to difficulty in using the device or failure to achieve the desired osteotomy.

Conclusions

Overall, the ultrasonic scalpel was safe and performed as desired when used as a bone cutting device to facilitate osteotomies in a variety of spine surgeries. However, caution should be taken to avoid potential thermal injury and dural tear. If used properly, this device may decrease the risk of soft tissue injury associated with the use of high speed burrs and oscillating saws during spine surgery.
Literatur
1.
Zurück zum Zitat Hosono N, Miwa T, Mukai Y, Takenaka S, Makino T, Fuji T (2009) Potential risk of thermal damage to cervical nerve roots by a high-speed drill. J Bone Joint Surg Br 91:1541–1544PubMed Hosono N, Miwa T, Mukai Y, Takenaka S, Makino T, Fuji T (2009) Potential risk of thermal damage to cervical nerve roots by a high-speed drill. J Bone Joint Surg Br 91:1541–1544PubMed
2.
Zurück zum Zitat Hara M, Takayasu M, Takagi T, Yoshida J (2001) En bloc laminoplasty performed with threadwire saw. Neurosurg 48:235–239 Hara M, Takayasu M, Takagi T, Yoshida J (2001) En bloc laminoplasty performed with threadwire saw. Neurosurg 48:235–239
3.
Zurück zum Zitat Dammous S, Darche V, R. G (2010) Use of ultrasonic bone scalpel in orthognatic surgery. Proceedings of the XX Congress of the European Association of Cranio-Maxillo-Facial Surgery, Bruges, Belgium, Medimond International Proceedings Dammous S, Darche V, R. G (2010) Use of ultrasonic bone scalpel in orthognatic surgery. Proceedings of the XX Congress of the European Association of Cranio-Maxillo-Facial Surgery, Bruges, Belgium, Medimond International Proceedings
4.
Zurück zum Zitat Sanborn MR, Balzer J, Gerszten PC, Karausky P, Cheng BC, Welch WC (2011) Safety and efficacy of a novel ultrasonic osteotome device in an ovine model. J Clin Neurosci 18:1528–1533PubMedCrossRef Sanborn MR, Balzer J, Gerszten PC, Karausky P, Cheng BC, Welch WC (2011) Safety and efficacy of a novel ultrasonic osteotome device in an ovine model. J Clin Neurosci 18:1528–1533PubMedCrossRef
5.
Zurück zum Zitat Satomi K, Ogawa J, Ishii Y, Hirabayashi K (2001) Short-term complications and long-term results of expansive open-door laminoplasty for cervical stenotic myelopathy. Spine J 1:26–30PubMedCrossRef Satomi K, Ogawa J, Ishii Y, Hirabayashi K (2001) Short-term complications and long-term results of expansive open-door laminoplasty for cervical stenotic myelopathy. Spine J 1:26–30PubMedCrossRef
6.
Zurück zum Zitat Horton JE, Tarpley TM Jr, Jacoway JR (1981) Clinical applications of ultrasonic instrumentation in the surgical removal of bone. Oral Surg Oral Med Oral Pathol 51:236–242PubMedCrossRef Horton JE, Tarpley TM Jr, Jacoway JR (1981) Clinical applications of ultrasonic instrumentation in the surgical removal of bone. Oral Surg Oral Med Oral Pathol 51:236–242PubMedCrossRef
7.
Zurück zum Zitat Sherman JA, Davies HT (2000) Ultracision: the harmonic scalpel and its possible uses in maxillofacial surgery. Br J Oral Maxillofac Surg 38:530–532PubMedCrossRef Sherman JA, Davies HT (2000) Ultracision: the harmonic scalpel and its possible uses in maxillofacial surgery. Br J Oral Maxillofac Surg 38:530–532PubMedCrossRef
8.
Zurück zum Zitat Stubinger S, Kuttenberger J, Filippi A, Sader R, Zeilhofer HF (2005) Intraoral piezosurgery: preliminary results of a new technique. J Oral Maxillofac Surg 63:1283–1287PubMedCrossRef Stubinger S, Kuttenberger J, Filippi A, Sader R, Zeilhofer HF (2005) Intraoral piezosurgery: preliminary results of a new technique. J Oral Maxillofac Surg 63:1283–1287PubMedCrossRef
9.
Zurück zum Zitat Vercellotti T, Pollack AS (2006) A new bone surgery device: sinus grafting and periodontal surgery. Compend Contin Educ Dent 27:319–325PubMed Vercellotti T, Pollack AS (2006) A new bone surgery device: sinus grafting and periodontal surgery. Compend Contin Educ Dent 27:319–325PubMed
10.
Zurück zum Zitat McMahon P, Dididze M, Levi AD (2012) Incidental durotomy after spinal surgery: a prospective study in an academic institution. J Neurosurg 17:30–36 McMahon P, Dididze M, Levi AD (2012) Incidental durotomy after spinal surgery: a prospective study in an academic institution. J Neurosurg 17:30–36
Metadaten
Titel
Use of an ultrasonic osteotome device in spine surgery: experience from the first 128 patients
verfasst von
Xiaobang Hu
Donna D. Ohnmeiss
Isador H. Lieberman
Publikationsdatum
01.12.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe 12/2013
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-013-2780-y

Weitere Artikel der Ausgabe 12/2013

European Spine Journal 12/2013 Zur Ausgabe

Open Operating Theatre (OOT)

Correction of kyphoscoliosis

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.